K202883 · Osteonic Co., Ltd. · MBI · Nov 25, 2020 · Orthopedic
Device Facts
Record ID
K202883
Device Name
Sterile bone screw (PEEK ACL screw)
Applicant
Osteonic Co., Ltd.
Product Code
MBI · Orthopedic
Decision Date
Nov 25, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sterile bone screw (PEEK ACL screw) is a fixation screw that fixes soft tissue such as ligaments, tendons, and the articular capsules to bone, and is used in orthopedic surgery. Knee • ACL repairs • PCL repairs • Extra-capsular repairs – Medial collateral ligament – Lateral collateral ligament – Posterior oblique ligament • Patellar realignment and tendon repairs – Vastus medialis obliquus advancement • Iliotibial band tenodesis
Device Story
Sterile bone screw (PEEK ACL screw) is a fixation device used in orthopedic surgery to secure soft tissues (ligaments, tendons, articular capsules) to bone. Device is manufactured from PEEK (poly-ether-ether-ketone) and supplied sterile via gamma irradiation in Tyvek pouches. It is intended for single-use. During surgery, the screw is inserted into bone to provide mechanical fixation of soft tissue. Clinical benefit is achieved through stable reattachment of tissues to facilitate healing. Device is operated by orthopedic surgeons in clinical settings.
Clinical Evidence
Bench testing only. Performance verified via ASTM F543-A3 (axial pullout), ASTM F543-A1 (torsional yield), ASTM F543-A2 (driving torque), and fatigue testing (70-220N tensile load, 1000 cycles at 1 Hz). Biocompatibility evaluated per ISO 10993. All tests passed.
Technological Characteristics
Material: PEEK (ASTM F2026). Dimensions: Diameters 7.0-10.0 mm, lengths 25-30 mm. Sterilization: Gamma irradiation. Single-use. Mechanical performance tested per ASTM F543. Packaging validated per ISO 11607 and ASTM F1980/F88/F1929.
Indications for Use
Indicated for fixation of soft tissue (ligaments, tendons, articular capsules) to bone in orthopedic surgery, specifically for knee procedures including ACL/PCL repairs, extra-capsular repairs (MCL, LCL, posterior oblique ligament), patellar realignment, tendon repairs (vastus medialis obliquus advancement), and iliotibial band tenodesis.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
November 25, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that.
Osteonic Co., Ltd. % Sanglok Lee Manager WISE COMPANY Inc. #507-508, 166 Gasan digital 2-ro Geumcheon-gu, Seoul, Republic of Korea 08503
#### Re: K202883
Trade/Device Name: Sterile bone screw (PEEK ACL screw) Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI Dated: September 24, 2020 Received: September 28, 2020
#### Dear Sanglok Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
{3}------------------------------------------------
1206Ho. 38. Digital-ro 29 Tel: +82(2)-6902-8400 Fax: +82(2)-6902-8401 Home page: http://www.os
83 Page 1 of 4
## The assigned 510(k) Number: K202883
## 1. 510(k) Summarv Date: 11. 23. 2020
## 2. Applicant
### Sung-moo Hwang
OSTEONIC Co., Ltd. 303Ho, 405Ho, 505-3Ho, 505-3Ho, 902Ho, 1004Ho, 1201Ho, 1206Ho, 38 Digital-ro 29-gil, Guro-gu, Seoul, Korea, 102Ho, 103Ho, 24, Digital-ro 27-gil, Guro-gu, Seoul, Korea TEL: +82-2-6902-8426 FAX: +82-2-6902-8401 Email: sungmhwang@osteonic.com
## 3. Submission Correspondent
Sang-lok, Lee Wise COMPANY Inc. #303, 142, Gasan digital 1-ro, Geumcheon-gu, Seoul, Korea TEL: +82 70 8812 3619 / +82 2 831 3615 FAX: +82 50 4031 3619 Email: info@wisecompany.org
# 4. Subject Device Identification
- 1) Trade Name: Sterile bone screw (PEEK ACL screw)
- 2) Common Name: Sterile bone screw (PEEK ACL screw)
- 3) Classification Name: fastener, fixation, non-biodegradable, soft tissue
- 4) Product Code: MBI
- 5) Panel: Orthopedic
- 6) Regulation Number: 21 CFR 888.3040
- 7) Device Class: II
## 5. Indication for use
The Sterile bone screw (PEEK ACL screw) is a fixation screw that fixes soft tissue such as ligaments, tendons, and the articular capsules to bone, and is used in orthopedic surgery.
Knee
- · ACL repairs
- · PCL repairs
- Extra-capsular repairs
- Medial collateral ligament
- Lateral collateral ligament
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Osteonic logo. The logo consists of a blue diamond shape with a white line running through it. To the right of the diamond is the word "OSTEONIC" in a dark gray sans-serif font. The logo is simple and modern.
K202883 Page 2 of 4
- Posterior oblique ligament
- Patellar realignment and tendon repairs - Vastus medialis obliquus advancement
- Iliotibial band tenodesis
## 6. Predicate devices
- 1) 510(k) Number: K083635
- 2) Device Name: Biosure PK Interference Screw
- 3) Manufacturer: Smith&Nephew PK Interference Screw
## 7. Device Description
The Sterile bone screw (PEEK ACL screw) is a bone fixation screw that fixes soft tissues such as ligament, tendon, and the articular capsules to bone, and is used in orthopedic surgery.
The implanted screw is made of PEEK (poly-ether-ether-ketone, ASTM F2026).
The Sterile bone screw (PEEK ACL screw) is supplied gamma irradiation sterile state and it is packed in Tyvek Pouch.
## 8. Non-Clinical Test Conclusion
#### 1) General Information
Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:
#### ▪ Material
- ASTM F2026:2017, Standard Specification for Poly-ether-ether-ketone(PEEK)Polymer for Surgical Implant Applications
#### · Mechanical Performance
- ASTM F543:17 Standard Specification and test method for metallic bone screws
- = Sterilization, Shelf-life and Packaging for Sterile Product
- ISO 11138-1:2006, Sterilization of health care products Biological indicators -- Part 1: General requirements
- ISO 11140-1:2014, Sterilization of health care products Chemical indicators Part 1: General requirements
- ISO 11137-1:2006. Sterilization of health care products Radiation Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- ISO 11137-2:2013, Sterilization of health care products Radiation Part 2: Establishing the sterilization dose
- ISO 11137-3:2017, Sterilization of health care products Radiation Part 3: Guidance on dosimetric aspects of development, validation and routine control
- ISO 11737-1:2018 Sterilization of medical devices Microbiological methods- Part 1: Estimation of population of microorganisms on products
- ISO 11737-2:2009 Sterilization of medical devices Microbiological methods- Part 2: Tests of
{5}------------------------------------------------
sterility performed in the validation of a sterilization process
- ISO 11607-1:2006/AMD1:2014 Packaging for terminally sterilized medical devices part 1: requirements for materials, sterile barrier systems and packaging system
- ISO 11607-2:2006/AMD1:2014 Packaging for terminally sterilized medical devices Part 2: Validation requirements for forming, sealing and assembly processes
- ASTM F1980:2016 Standard guide for accelerated aging of sterile barrier systems for medical devices
- ASTM F88/F88M:2015 Standard test method for seal strength of flexible barrier materials.
- ASTM F1929:2015 Standard test method for detecting seal leaks in porous medical packaging by dye penetration
- Bacterial Endotoxin
- USP <85> Bacterial Endotoxin Test
- USP <161> Medical Devices-Bacterial Endotoxin and Pyrogen Tests
#### 2) Summary of Biocompatibility
The Sterile bone screw (PEEK ACL screw) has been evaluated for biocompatibility according to ISO 10993.
#### 3) Summary of Performance
All specimens selected (worst-case) to verify the performance of the screw met the test criteria. Therefore, the performance of the entire model was proven.
| No. | Test Article | Test Method | Result |
|-----|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| 1 | Axial Pullout Strength | ASTM F543-A3.Test Method for Determining the Axial<br>Pullout Strength of Medical Bone Screws. | PASS |
| 2 | Torsional Yield Strength | ASTM F543-A1.Test Method for Determining the<br>Torsional Properties of Metallic Bone Screws. | PASS |
| 3 | Driving Torque<br>Insertion<br>Removal | ASTM F543-A2.Test Method for Driving Torque of Medical<br>Bone Screws. | PASS |
| 4 | Fatigue | After fully inserting the screw with the tendon into the fixed<br>artificial bone (20PCF), a fatigue tensile load of 70 to 220N<br>on the tendon is repeated 1000 cycles at 1 Hz. | PASS |
## 9. Substantially Equivalent Conclusion
#### Table 1: Substantial Equivalence Comparison
| Product<br>Name | SUBJECT Device | PREDICATE Device<br>Biosure PK Interference Screw (K083635) | Equivalence<br>Discussion |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Product<br>code | MBI | MBI | Same |
| Regulatory<br>class | Class II | Class II | Same |
| Regulation<br>Number | 21 CFR 888.3040 | 21 CFR 888.3040 | Same |
| Intended<br>Use | The Sterile bone screw (PEEK<br>ACL screw) is a fixation screw<br>that fixes soft tissue such as<br>ligaments, tendons, and the<br>articular capsules to bone, and is<br>used in orthopedic surgery | The BIOSURE PK Screw is indicated for<br>the reattachment of ligament, tendon, soft<br>tissue, or bone to bone during cruciate<br>ligament reconstruction surgeries of the<br>knee. All screws with a diameter of 9 mm<br>or less and a length of 25 mm or less are also | Same |
| | Knee<br>• ACL repairs<br>• PCL repairs<br>• Extra-capsular repairs<br>– Medial collateral ligament<br>– Lateral collateral ligament<br>– Posterior oblique ligament<br>• Patellar realignment and tendon<br>repairs<br>– Vastus medialis obliquus<br>advancement<br>• Iliotibial band tenodesis | intended for use in the following<br>procedures:<br>Knee<br>• ACL repairs<br>• PCL repairs<br>• Extra-capsular repairs<br>– Medial collateral ligament<br>– Lateral collateral ligament<br>– Posterior oblique ligament<br>• Patellar realignment and tendon repairs<br>– Vastus medialis obliquus advancement<br>• Iliotibial band tenodesis<br>Shoulder<br>• Acromioclavicular separation repairs<br>• Biceps tenodesis<br>Foot and Ankle<br>• Medial or lateral instability repairs/<br>reconstructions<br>• Achilles tendon repairs/reconstructions<br>• Metatarsal ligament/tendon repairs/<br>reconstructions<br>• Flexor hullucis longus (FHL)<br>• Tendon transfers<br>Elbow, Wrist, and Hand<br>• Biceps tendon reattachment<br>• Ulnar or radial collateral ligament<br>reconstructions<br>• Lateral epicondylitis repair<br>• Scapholunate ligament reconstruction<br>• Tendon transfers<br>• Carpomedicarpal joint arthroplasty | |
| Operating<br>Principles | The Sterile bone screw (PEEK<br>ACL screw) is a fixation screw<br>that fixes soft tissue such as<br>ligaments, tendons, and the<br>articular capsules to bone, and is<br>used in orthopedic surgery. | The Biosure PK Interference Screw is a<br>fixation screw that fixes soft tissue such as<br>ligaments, tendons, and the articular<br>capsules to bone, and is used in orthopedic<br>surgery. | Same |
| Material | PEEK 100% (Optima PEEK) | PEEK 100% (Optima PEEK) | Same |
| Product<br>Size | Ø 7.0 x (25,30) mm<br>Ø 8.0 x (25,30) mm<br>Ø 9.0 x (25,30) mm<br>Ø 10.0 x (25,30) mm | Ø 6.0 x (20,25,25(R)) mm<br>Ø 7.0 x (20,25,25(R),30) mm<br>Ø 8.0 x (20,25,25(R),30,35) mm<br>Ø 9.0 x (20,25,30,35) mm<br>Ø 10.0 x (20,25,30,35) mm<br>Ø 11.0 x (25,30,35) mm<br>Ø 12.0 x 35 mm | Similar |
| Sterilization | Sterile (Gamma sterilization) | Sterile (Gamma sterilization) | Same |
| Single Use/<br>Reuse | Single use | Single use | Same |
| Packaging | 1 EA / BOX | 1 EA / BOX | Same |
| Shelf life | 5 Years | 5 Years | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Osteonic logo. The logo consists of a blue geometric shape on the left and the word "OSTEONIC" in gray on the right. The geometric shape is a diamond with a stylized "O" inside.
K202883 Page 4 of 4
Based on above, the subject device, Sterile bone screw (PEEK ACL screw), is determined to be Substantially Equivalent (SE) to the predicate devices, Biosure PK Interference Screw (K083635) in respect of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.