K202864 · Covidien · GDW · Oct 28, 2020 · General, Plastic Surgery
Device Facts
Record ID
K202864
Device Name
Tri-Staple 2.0 Reloads
Applicant
Covidien
Product Code
GDW · General, Plastic Surgery
Decision Date
Oct 28, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Tri-Staple™ 2.0 Reloads have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. It may be used for transection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of pancreas. The Tri-Staple™ 2.0 curved tip reloads can be used to blunt dissect or separate target tissue from other tissue.
Device Story
Tri-Staple 2.0 Reloads are single-use surgical stapling components; place three staggered rows of titanium staples on either side of a cut line while simultaneously dividing tissue. Reloads feature an integrated 1-Wire ID-chip assembly; enables communication with Signia Powered Stapler handles and adapters. Used in OR settings by surgeons; provides feedback to the powered handle system. Benefits include consistent staple placement and tissue division during complex surgical procedures. Subject device represents a line extension adding round-tip anvil configurations to existing intelligent reload technology.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical design verification and reliance on previously cleared data for the 1-Wire ID-chip interface (K160176).
Technological Characteristics
Titanium staples; 1-Wire ID-chip assembly for digital communication; articulating reload form factors (30mm, 45mm, 60mm); compatible with GIA universal staplers, Endo GIA universal staplers, and Covidien powered stapler handles. Round-tip anvil design.
Indications for Use
Indicated for patients undergoing abdominal, gynecologic, pediatric, or thoracic surgery requiring resection, transection, or anastomosis. Includes transection of liver substance, hepatic vasculature, biliary structures, and pancreas. Curved tip reloads indicated for blunt dissection or tissue separation.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
October 28, 2020
Covidien Carolina Cabezas Regulatory Affairs Specialist 60 Middletown Avenue North Haven, Connecticut 06473
Re: K202864
Trade/Device Name: Tri-Staple 2.0 Reloads Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: September 23, 2020 Received: September 28, 2020
Dear Carolina Cabezas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. For Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202864
Device Name Tri-Staple™ 2.0 Reloads
# Indications for Use (Describe)
The Tri-Staple™ 2.0 reloads have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. It may be used for transection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of pancreas.
The Tri-Staple™ 2.0 curved tip reloads can be used to blunt dissect or separate target tissue from other tissue.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="color: black;"> <span style="font-weight: bold;"> </span> </span> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# DATE PREPARED:
September 23, 2020
# SUBMITTER:
Covidien 60 Middletown Avenue North Haven, CT 06473 USA
# CONTACT PERSON:
Carolina Cabezas Regulatory Affairs Specialist Telephone: (203) 492-6054 Fax: (203) 492-5029
# IDENTIFICATION OF DEVICE:
| Proprietary/Trade Name: | Tri-Staple™ 2.0 Reloads |
|-------------------------|---------------------------------------------------------|
| Classification Name: | Staples, Implantable |
| Regulation Number: | 21 CFR 878.4750 |
| Product Code: | GDW |
| Device Class: | Class II |
| Review Panel: | General and Plastic Surgery |
| Common Name: | Surgical Stapler with implantable non-absorbable staple |
#### PREDICATE DEVICE:
| Proprietary/Trade Name: | Signia™ Stapler |
|-------------------------|-----------------------------|
| 510(k) Number: | K160176 (April 26, 2016) |
| Classification Name: | Staples, Implantable |
| Regulation Number: | 21 CFR 878.4750 |
| Product Code: | GDW |
| Device Class: | Class II |
| Review Panel: | General and Plastic Surgery |
| Common Name: | Surgical Stapler |
# DEVICE DESCRIPTION:
The Tri-Staple™ 2.0 Reloads and Tri-Staple™ 2.0 curved tip reloads place staggered rows of titanium staples and simultaneously divides the tissue so that three staggered rows of staples are placed on either side of the cut line. The size of the staples is determined by the selection of the single use reload:
The Tri-Staple™ 2.0 reload is available in articulating 30 mm, 45 mm and 60 mm length, the Tri-Staple™ 2.0 curved tip reload is available in articulating 30 mm, 45 mm lengths and the Tri-Staple™ 2.0 black reload is available in articulating 45 mm and 60 mm lengths.
The Tri-Staple™ 2.0 reloads and Tri-Staple™ 2.0 curved tip reloads can be used with the GIA™ universal stapler, Endo GIA™ universal staplers, Endo GIA™ Ultra universal staplers, and Covidien™ powered stapler handles used with Endo GIA™ adapters.
# INTENDED USE/INDICATIONS FOR USE:
The Tri-Staple™ 2.0 Reloads have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. It may be used for transection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of pancreas.
The Tri-Staple™ 2.0 curved tip reloads can be used to blunt dissect or separate target tissue.
{4}------------------------------------------------
# SUMMARY OF TECHNOLOGICAL CHARACTERISTICS:
The purpose of this submission is a line extension to expand the reorder code offering of Tri-Staple™ 2.0 Reloads that feature the 1-Wire ID-chip assembly. Through this Special 510(k), the five new stapling reload codes will be introduced by adding the 1-wire ID Chip assembly to the Endo GIA™ Tri-Staple Reloads. The 1-wire ID-chip assembly enables communication between Tri-Staple™ 2.0 Reloads and the Signia™ Stapler, and such communication feature was cleared under the predicate K160176. The expanded product code offerings are the round tip "Core" Tri-Staple™ 2.0 Reload reorder codes.
As with current Tri-Staple™ 2.0 Intelligent Reloads and Tri-Staple™ 2.0 Intelligent curved tip reloads, the new five Tri-Staple 2.0 Reloads reorder codes place staggered rows of titanium staples and simultaneously divides the tissue so that three staggered rows of staples are placed on either side of the size of the staple is determined by the selection of the single use reload.
Specifically, the subject "Core" Tri-Staple™ 2.0 Reloads when used with compatible Covidien stapler handles have the same indications for use as the "Specialty" Tri-Staple™ 2.0 Reloads previously reported under K160176. In the same manner as the "Specialty" Tri-Staple™ 2.0 Reloads previously reported under K160176, the "Core" Tri-Staple™ 2.0 Reloads have an encrypted 1-Wire ID-chip, thereby enabling communication between these "Core" reloads and the Signia™ Powered Handle and associated Signia™ Linear Adapters. The main difference between the subject Tri-Staple™ 2.0 Reloads and the predicate device is the subject reloads have a round tip anvil vs. the predicate reloads, which have a curved-tip anvil.
# SUBSTANTIAL EQUIVALENCE:
The subject Tri-Staple™ 2.0 Reloads (new codes) are substantially equivalent to the legally marketed Tri-Staple™ 2.0 Intelligent Reloads and Tri-Staple™ 2.0 Intelligent curved tip reloads (K160176), since the addition of the 1-wire ID Chip assembly to the existing Endo GIA™ Reloads with Tri-Staple™ Technology to extend the reorder code offerings, does not alter the intended use, indications, or user environment of the device.
# SUMMARY OF STUDIES:
No new non-clinical performance data or testing has been performed. The technology and design of the 1-Wire IDchip remains the same in the subject Tri-Staple™ 2.0 Reloads. The 1-Wire communications interface for Covidien stapling reloads was previously reported in 510(k) K160176, Signia™ Stapler. Given the fact that the design inputs and functional requirements have been previously verified, including communication with the Signia™ Stapler via the 1-wire ID-Chip assembly, the existing design verification is transferable to the subject Tri-Staple™ 2.0 Reloads (new codes). The Design Control Activities table enclosed in the submission provides the detailed rationales.
Clinical performance data – No clinical study has been performed. The substantial equivalence has been demonstrated by non-clinical studies.
# CONCLUSION:
Based upon the supporting data summarized above, we concluded that this line extension to expand the reorder code offering of Tri-Staple™ 2.0 Reloads reorder codes is as safe and effective as the legally marketed Tri-Staple™ 2.0 Intelligent Reloads and Tri-Staple™ 2.0 Intelligent curved tip reloads (K160176) and does not raise different questions of safety and effectiveness than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.