Needle Stimulator

K202861 · Wuxi Jiajian Medical Instrument Co., Ltd. · BWK · Aug 27, 2021 · NE

Device Facts

Record IDK202861
Device NameNeedle Stimulator
ApplicantWuxi Jiajian Medical Instrument Co., Ltd.
Product CodeBWK · NE
Decision DateAug 27, 2021
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

Needle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

Device Story

Needle Stimulator is an electro-acupuncture device used by qualified acupuncture practitioners in clinical settings. It consists of a console and 6 channels of electrode cables with alligator connectors. The device is powered by 6 1.5V batteries or AC 100-240V. It does not include needles; practitioners select legally marketed needles. The device delivers electrical stimulation via 3 output modes (continuous, interrupted, dense-disperse) using an asymmetric biphasic square waveform. The console features function knobs/buttons for user control of stimulation parameters and treatment time. It includes safety features like Type BF line current isolation, transformer-based channel isolation, and automatic shut-off. The device is microprocessor-controlled. Practitioners use the device to provide acupuncture therapy, with output affecting clinical decision-making based on the selected stimulation mode and intensity. Benefits include providing controlled electrical stimulation to acupuncture needles to assist in therapy.

Clinical Evidence

Bench testing only. The device underwent performance testing to verify compliance with ANSI AAMI ES60601-1, IEC 60601-1-2, and IEC 60601-2-10. Testing confirmed that electrical output parameters, including maximum current density and power density, remain within safety limits. Biocompatibility was assessed per ISO 10993-5 and ISO 10993-10.

Technological Characteristics

Console-based electro-acupuncture stimulator; 6 output channels; powered by 6x 1.5V DC or 100-240V AC; Type BF patient isolation; transformer-based channel isolation; ABS plastic housing; asymmetric biphasic square waveform; microprocessor-controlled; user-adjustable timer and stimulation intensity; compliant with IEC 60601-1, IEC 60601-2-10, and IEC 60601-1-2.

Indications for Use

Indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services - USA. To the right of the seal is the FDA logo in blue, with the letters "FDA" in a bold, sans-serif font. Underneath the letters "FDA" is the text "U.S. FOOD & DRUG ADMINISTRATION" in a smaller, sans-serif font, also in blue. The logo is simple and professional, and it is easily recognizable. August 27, 2021 Wuxi Jiajian Medical Instrument Co., Ltd Caihong Sun Regulatory Assurance No. 35 Baiqiao Rd., Ehu Town, Xishan District, Wuxi, Jiangsu 214116 China Re: K202861 Trade/Device Name: Needle Stimulator Regulatory Class: Unclassified Product Code: BWK Dated: August 9, 2021 Received: August 13, 2021 Dear Caihong Sun: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Jitendra V. Virani -S For Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K202861 Device Name Needle Stimulator Indications for Use (Describe) Needle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary [As required by 21 CFR 807.92] #### 1. Submission Information: 510(k) Number: K202861 Date: April 1st, 2021 Type of 510(k) Submission: Traditional Modified device Basis for 510(k) Submission: Submitter/Manufacturer: Wuxi Jiajian Medical Instrument Co., Ltd No. 35 Baiqiao Rd., Ehu Town, Xishan District, Wuxi, Jiangsu, China 214116 Contact: Doris Dong [Consultant, from Shanghai CV Technology Co., Ltd.] Add: Room 903, No. 19 Dongbao Road, Songjiang Area, Shanghai, 201613 China E-mail: doris.d(@ceve.org.cn Tel: 86 21-31261348 / Fax: 86 21-57712250 ### 2. Device Description: | Proprietary Name: | Needle Stimulator | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Model: | CMNS6-1, CMNS6-2 | | Common Name: | Electro-Acupuncture | | Product Code: | BWK | | Device Class: | Unclassified | | Review Panel: Device | Neurology | | Description: | Needle Stimulator is an electro-acupuncture device for acupuncture therapy powered by 6 pieces of 1.5V batteries or AC 100-240V. It is composed of a console and 6 channels of electrode cables with alligator type connectors. The console has the operating controls including function knobs or buttons. Needle Stimulator does not equip with acupuncture needles. The practitioners should select legally marketed needles. | | Indications for use: | Needle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states. | #### 3. Substantial Equivalence to Predicate device: Detailed comparison data is included in "Section 10.1 - Substantial Equivalence Discussion" of this 510(k) submission. Table 1- | | Parameters | Modified Device | Predicate Device | Remark | | |----|----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|----------------| | 1 | 510(k) Number: | K202861 | K130768 | -- | | | 2 | Marketing clearance<br>date: | No | August 16, 2013 | -- | | | 3 | Device Name | Needle Stimulator | Jiajian® CMN Stimulator | -- | | | 4 | Model | CMNS6-1 | / | -- | | | 5 | Manufacturer | Wuxi Jiajian Medical Instrument Co.,Ltd | Wuxi Jiajian Medical Instrument Co.,Ltd | Same | | | 6 | Intended use | Needle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states. | Jiajian® CMN Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states. | Same | | | 7 | Type of use | Prescription use | Prescription use | Same | | | | Power Source(s) | DC 1.5Vx6 Type R14 or AC 100-240V | DC 1.5Vx6 Type R14 | Same<br>Note 1 | | | | Method of Line Current Isolation | Type BF | Type BF | Same | | | 8 | Patient Leakage<br>-<br>Current | -- | -- | Same | | | | Normal Condition (μΑ) | 2μΑ | 2μΑ | | | | | Single Fault<br>-<br>Condition (μΑ) | ≤50μΑ | ≤50μΑ | | | | 9 | Average DC current through electrodes when device is on but no pulses are being applied (μΑ) | N/A | N/A | Same | | | 10 | Number of Output Modes | 3 (continuous wave/ interrupted wave/ Dense-disperse wave) | 3 (continuous wave/ interrupted wave/ Dense-disperse wave) | Same | | | | Number of Output channels: | 6 (3 channels at most work together on single patient) | 6 (3 channels at most work together on single patient) | Same | | | 11 | - Synchronous or Alternating? | Synchronous | Synchronous | Same | | | | Method of Channel Isolation | Transformer | Transformer | Same | | | 12 | Regulated Current or Regulated Voltage? | Voltage Control | Voltage Control | Same | | | 13 | Software/Firmware/Microprocessor Control? | Yes | Yes | Same | | | 14 | Automatic Overload Trip? | No | No | Same | | | 15 | Automatic No-Load Trip? | No | No | Same | | | 16 | Automatic Shut Off? | Yes | Yes | Same | | | 17 | User Override | Yes | Yes | Same | | | | Control? | | | | | | 18 | | On/Off<br>Status? | Yes | Yes | Same | | | | Low<br>Battery? | Yes | Yes | Same | | | Indicator<br>Display: | Voltage/Current<br>Level? | Yes | Yes | Same | | | 19 | Timer<br>(minutes) | Range | 0-60min | 0-60min | | 20 | Compliance with<br>Voluntary Standards? | | IEC 60601-1, IEC 60601-2-10,<br>IEC 60601-1-2 | IEC 60601-1, IEC 60601-2-10,<br>IEC 60601-1-2 | Same | | 21 | Compliance with 21<br>CFR 898? | | Yes | Yes | Same | | 22 | Weight (grams) | | 740g | 740g | Same | | 23 | Dimensions [W x H x D] | | 230*155*55mm | 230*155*55mm | Same | | 24 | Housing Materials &<br>Construction | | ABS; Injection molded | ABS; Injection molded | Same | | 25 | Waveform | | Biphasic | Biphasic | Same | | 26 | Shape | | Asymmetric biphasic square<br>wave | Asymmetric biphasic square<br>wave | Same | | 27 | Maximum Output<br>Voltage (volts) | | 27V±10% @500Ω | 27V±10% @500Ω | Same | | | | | 60.4V±10% @2kΩ | 60.4V±10% @2kΩ | Same | | | | | 75V±10% @10kΩ | 75V±10% @10kΩ | Same | | 28 | Maximum Output<br>Current (specify units) | | 54mA±10% @500Ω | 54mA±10% @500Ω | Same | | | | | 30.2mA±10% @2kΩ | 30.2mA±10% @2kΩ | Same | | | | | 7.5mA±10% @10kΩ | 7.5mA±10% @10kΩ | Same | | 29 | Pulse<br>width<br>(µsec) | Positive | 200µs±10% | 200µs±10% | Same | | | | Negative | 1030µs (5.15 x (+Phase) ) | 1030µs (5.15 x (+Phase) ) | Same | | 30 | Pulse | Period (msec) | 10~1000ms | 10~1000ms | Same | | 31 | Max. pulse frequency<br>(Hz) [or Rate (pps)] | | 1~100Hz±10% | 1~100Hz±10% | Same | | 32 | Net Charge (µC per<br>pulse) | | 0µC@500Ω, + and – pulses<br>cancel | 0µC@500Ω, + and – pulses<br>cancel | Same | | 33 | Maximum<br>Charge, (µC) | Phase | 9.4µC @500Ω | 9.4µC @500Ω | Same | | 34 | Maximum Average<br>Current, (mA) | | 1.08mA @500Ω | 1.08mA @500Ω | Same | | 35 | Maximum Current<br>Density, (mA/cm², r.m.s.) | | 9.4mA/cm² @500Ω | 13.3mA/cm² @500Ω | Same<br>Note 2 | | | 36 | Maximum Average | | 0.141W/cm²@500Ω | 0.2W/cm²@500Ω | | | Power<br>(W/cm²) | Density, | | | | | 37 | Biocompatibility | ISO10993-5, ISO 10993-10 | ISO10993-5, ISO 10993-10 | Same | | | 38 | Accessories | Lead wires, Alligator type connectors | Lead wires, Alligator type connectors | Same | | {4}------------------------------------------------ {5}------------------------------------------------ {6}------------------------------------------------ # Note 1: Compared with the predicate device, the proposed device can be connected to AC power. Both of the AC power and DC power of the proposed device have passed IEC 60601-1 and EMC tests, and the description of AC power supply has been added to user manual. Therefore, this difference doesn't raise any safety and effectiveness issues. # Note 2: The maximum current density and maximum power density are different between the proposed device and the predicate device because they are calculated by different needle surface areas. After calculation, both of them are within the safety limit. Therefore, this difference doesn't raise any safety and effectiveness issues. Detailed comparison data is included in "Section 10.2 - Substantial Equivalence Discussion" of this 510(k) submission. Table 2- | | Parameters | Modified Device | Predicate Device | Remark | | | |----|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|-----------------------------------------------------------|----------------| | 1 | 510(k) Number: | K202861 | K130768 | -- | | | | 2 | Marketing clearance date: | No | August 16, 2013 | -- | | | | 3 | Device Name | Needle Stimulator | Jiajian® CMN Stimulator | -- | | | | 4 | Model | CMNS6-2 | / | -- | | | | 5 | Manufacturer | Wuxi Jiajian Medical Instrument Co.,Ltd | Wuxi Jiajian Medical Instrument Co.,Ltd | Same | | | | 6 | Intended use | Needle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states. | Jiajian® CMN Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states. | Same | | | | 7 | Type of use | Prescription use | Prescription use | Same | | | | | Power Source(s) | DC 1.5Vx6 Type R14 or AC 100-240V | DC 1.5Vx6 Type R14 | Same<br>Note 1 | | | | | - Method of Line Current Isolation | Type BF | Type BF | Same | | | | 8 | - Patient Leakage Current | -- | -- | Same | | | | | -- Normal Condition (μΑ) | 2μΑ | 2μΑ | | | | | | -- Single Fault Condition (μΑ) | ≤50μΑ | ≤50μΑ | | | | | 9 | Average DC current | N/A | N/A | Same | | | | | through | electrodes | | | | | | | | when device is on but | | | | | | | | no pulses are being | | | | | | | applied (µA) | | | | | | | 10 | Number of Output Modes | | 3 (continuous wave/<br>interrupted wave/<br>Dense-disperse wave) | 3 (continuous wave/<br>interrupted wave/<br>Dense-disperse wave) | Same | | | | 11 | Number of Output channels: | | 6 (3 channels at most work<br>together on single patient) | 6 (3 channels at most work<br>together on single patient) | Same | | | | - Synchronous or Alternating? | | Synchronous | Synchronous | Same | | | | - Method of Channel Isolation | | Transformer | Transformer | Same | | 12 | Regulated Current or Regulated Voltage? | | Voltage Control | Voltage Control | Same | | | | | Software/Firmware/Microprocessor Control? | | Yes | Yes | Same | | 13 | | | | | | | | 14 | Automatic Overload Trip? | | No | No | Same | | | 15 | Automatic No-Load Trip? | | No | No | Same | | | 16 | Automatic Shut Off? | | Yes | Yes | Same | | | 17 | User Override Control? | | Yes | Yes | Same | | | 18 | Indicator Display: | | | | | | | | On/Off Status? | | Yes | Yes | Same | | | | Low Battery? | | Yes | Yes | Same | | | | Voltage/Current Level? | | Yes | Yes | Same | | | | 19 | Timer (minutes) Range | | 1-99min | 0-60min | Same<br>Note 2 | | | 20 | Compliance with Voluntary Standards? | | IEC 60601-1, IEC 60601-2-10,<br>IEC 60601-1-2 | IEC 60601-1, IEC 60601-2-10,<br>IEC 60601-1-2 | Same | | 21 | Compliance with 21 CFR 898? | | Yes | Yes | Same | | | 22 | Weight (grams) | | approx. 657g | 740g | Same | | | 23 | Dimensions [W x H x D] | | 238*184*75mm | 230*155*55mm | Same<br>Note 2 | | | 24 | Housing Materials & Construction | | ABS; Injection molded | ABS; Injection molded | Same | | | 25 | Waveform | | Biphasic | Biphasic | Same | | | 26 | Shape | | Asymmetric biphasic square | Asymmetric biphasic square | Same | | | | | | | | | | | | | | wave | wave | | | | 27 | Maximum<br>Output<br>Voltage (volts) | | 27V±10% @500Ω | 27V±10% @500Ω | Same | | | | 60.4V±10% @2kΩ | 60.4V±10% @2kΩ | Same | | | | | | 75V±10% @10kΩ | 75V±10% @10kΩ | Same | | | | | 28 | Maximum | Output | 54mA±10% @500Ω | 54mA±10% @500Ω | Same | | | | Current | (specify | 30.2mA±10% @2kΩ | 30.2mA±10% @2kΩ | Same | | | | units) | | 7.5mA±10% @10kΩ | 7.5mA±10% @10kΩ | Same | | | 29 | Pulse | Positive | 175µs±10% | 200µs±10% | Same | | | | width<br>(µsec) | Negative | 1051µs (6 x (+Phase) ) | 1030µs (5.15 x (+Phase) ) | Note 3 | | | 30 | Pulse<br>Period (msec) | | 10~1000ms | 10~1000ms | Same | | | 31 | Max. pulse frequency<br>(Hz) [or Rate (pps)] | | 1~100Hz±10% | 1~100Hz±10% | Same | | | 32 | Net Charge (µC per<br>pulse) | | 0µC@500Ω, + and - pulses<br>cancel | 0µC@500Ω, + and - pulses<br>cancel | Same | | | 33 | Maximum<br>Phase<br>Charge, (µC) | | 8.225μC @500Ω | 9.4μC @500Ω | Same<br>Note 4 | | | 34 | Maximum<br>Average<br>Current, (mA) | | 0.945mA @500Ω | 1.08mA @500Ω | | | | 35 | Maximum<br>Density,<br>r.m.s.) | Current<br>(mA/cm², | 8.225mA/cm² @500Ω | 13.3mA/cm² @500Ω | | | | 36 | Maximum<br>Power<br>(W/cm²) | Average<br>Density, | 0.1234W/cm²@500Ω | 0.2W/cm²@500Ω | | | | 37 | Biocompatibility | | ISO10993-5, ISO 10993-10 | ISO10993-5, ISO 10993-10 | Same | | | 38 | Accessories | | Lead<br>wires, Alligator<br>type<br>connectors | Lead<br>wires, Alligator<br>type<br>connectors | Same | | {7}------------------------------------------------ {8}------------------------------------------------ # Note 1: Compared with the predicate device, the proposed device can be connected to AC power. Both of the AC power and DC power of the proposed device have passed IEC 60601-1 and EMC tests, and the description of AC power supply has been added to user manual. Therefore, this difference doesn't raise any safety and effectiveness issues. # Note 2: The weight, dimensions and appearance of the proposed device are a little different from the predicate device K130768, but these differences are insignificant in the terms of safety or effectiveness. Besides, there is a difference in treatment time range b…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...