K202791 · Andon Health Co, Ltd. · FLL · Jan 5, 2021 · General Hospital
Device Facts
Record ID
K202791
Device Name
iHealth Clinical Digital Thermometer
Applicant
Andon Health Co, Ltd.
Product Code
FLL · General Hospital
Decision Date
Jan 5, 2021
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Clinical Digital Thermometer is intended to measure the body temperature oral or axillaries (under the arm) and to be used by consumers in household environments. It is intended for use on adults and children ages 4 and up.
Device Story
Hand-held, predictive, thermistor-based digital stick thermometer; measures body temperature via oral or axillary sites. Device uses negative temperature coefficient thermistor in measurement tip; resistance changes correlate to temperature. Predictive algorithm processes raw thermistor data to calculate final temperature; output displayed on integrated LCD. Operated by consumers in home environments. Healthcare providers use displayed temperature to inform clinical decisions regarding fever management. Benefits include rapid, non-invasive temperature monitoring for home users.
Clinical Evidence
Clinical accuracy validation performed per ISO 80601-2-56 on 110 subjects (40 children ages 4-5, 70 adults >5 years). Results showed clinical bias ≤ 0.03 and repeatability ≤ 0.07, demonstrating equivalence to the predicate device.
Technological Characteristics
Hand-held stick thermometer; negative temperature coefficient thermistor sensor. Materials: ABS (shell, battery door, button), PET (LCD lens, film), PP (probe cover), SUS (sensor tip). Powered by 3V CR1632 battery. Connectivity: None. Standards: ASTM E1112-00:2011, ISO 80601-2-56:2017, IEC 60601-1, IEC 60601-1-2, ISO 10993.
Indications for Use
Indicated for oral or axillary body temperature measurement in adults and children ages 4 and up in household environments.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
K180131 — Vicks RapidRead Digital Thermometer · Kaz USA, Inc., A Helen of Troy Company · Aug 14, 2018
K092557 — DIGITAL THERMOMETER,HMCDT-002/KINETIK BRANDED DT1 SERIES · Harvard Medical Devices , Ltd. · Dec 17, 2009
K990168 — MICROLIFE IT2CA1 · Microlife Systems AG · Feb 25, 1999
K173730 — Kinsa QuickCare Thermometer · Kinsa, Inc. · Jan 17, 2018
K163603 — Digital Thermometer · Guangzhou Jinxinbao Electronic Co., Ltd. · Jun 6, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 5, 2021
Andon Health Co., Ltd. Liu Yi President No. 3 Jin Ping Street, Ya An Road, Nankai District Tianiin. Tianiin 300190 China
Re: K202791
Trade/Device Name: iHealth Clinical Digital Thermometer (PT1) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: December 4, 2020 Received: December 8, 2020
Dear Liu Yi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Alan M. Stevens Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202791
#### Device Name
iHealth Clinical Digital Thermometer (PT1)
#### Indications for Use (Describe)
The Clinical Digital Thermometer is intended to measure the body temperature oral or axillaries (under the arm) and to be used by consumers in household environments. It is intended for use on adults and children ages 4 and up.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 20px;">☑</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
# 1.0 Submitter's information
| Name: | Andon Health Co., Ltd. |
|----------------------|--------------------------------------------------------------------------|
| Address: | No 3, Jinping Street Ya An Road, Nankai District, Tianjin,<br>P.R. China |
| Phone number: | 86-22-87611660 |
| Fax number: | 86-22-87612379 |
| Contact: | Liu Yi |
| Date of Preparation: | 9/18/2020 |
#### 2.0 Device information
iHealth Clinical Digital Thermometer (PT1) Trade name: Common name: Clinical Digital Thermometer Classification name: Clinical Electronic Thermometer
## 3.0 Classification
Production code: FLL Regulation number: 880.2910 Classification: II Panel: General Hospital
## 4.0 Predicate device information
Kaz USA, Inc., A Helen of Troy Company Manufacturer: Vicks RapidRead Digital Thermometer Device: 510(k) number: K180131
#### 5.0 Intended use
The Clinical Digital Thermometer is intended to measure the body temperature oral or axillaries (under the arm) and to be used by consumers in household environments. It is intended for use on adults and children ages 4 and up.
{4}------------------------------------------------
## 6.0 Device description
The iHealth PT1 Clinical Digital Thermometer is hand-held, predictive, thermistor-based, stick thermometer capable of measuring temperature in about 30 seconds. The thermometer uses a negative temperature coefficient thermistor embedded in a measurement tip that is in contact with the measurement site. As the thermistor changes temperature, the resistance of the thermistor also changes, which is measured by the thermometer and converted to a measurement of the temperature of the tip of the thermometer. This temperature, following the use of the predictive algorithm, is then displayed to the end user. Because the thermometer displays the measurement for the physiological site at which it is used, it does not need to convert this temperature via clinical offset.
| Item | Subject Device | Predicate Device (K180131) | Comparison |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| Device name | iHealth PT1 Clinical Digital<br>Thermometer | Vicks® VDT972<br>RapidRead™ Digital<br>Thermometer | N/A |
| Manufacturer<br>(legal) | Andon Health Co., Ltd | Kaz USA, Inc., a Helen of<br>Troy Company | N/A |
| Contract<br>manufacturer | Andon Medical Co., Ltd | Microlife Corporation | N/A |
| Thermometer<br>Type | Axillary / Oral, Predictive<br>Digital Thermometer | Axillary / Oral / Rectal,<br>Predictive Digital<br>Thermometer | Different <span style="font-size: 8px;">note 1</span> |
| Models | PT1 | Vicks® VDT972<br>RapidRead™<br>Digital Thermometer | N/A |
| 510K number | K202791 | K180131 | N/A |
| Indications for<br>Use | The Clinical Digital<br>Thermometer is intended to<br>measure the body<br>temperature oral or axillaries<br>(under the arm) and to be<br>used by consumers in<br>household environments. It<br>is intended for use on adults<br>and children ages 4 and up. | The Vicks® VDT972<br>RapidRead™ Digital<br>Thermometer is a handheld,<br>battery-powered, predictive,<br>digital stick thermometer<br>intended for the intermittent<br>determination of human body<br>temperature orally, rectally, or<br>under the arm, in a home-use<br>environment for people of all<br>ages (infants, children, and<br>adults). It is intended to be | Similar <span style="font-size: 8px;">note 2</span> |
| | | | |
| | | used with a single use,<br>disposable probe cover for all<br>measurements. | |
| User population | Children (ages 4 and up) and<br>adults | Infants, children, and adults | Different note 3 |
| Labeling | Instructions for use, package<br>/ box, and rating label | Instructions for use, package<br>/ box, and rating label | Similar note 4 |
| Components | On / Off Button, sensor head,<br>microcontroller, & LCD | On / Off Button, Site<br>Selection Button, sensor<br>head, protective cover,<br>microcontroller, & LCD | Different note 5 |
| Sensor | Thermistor | Thermistor | Same |
| Principles of<br>operation | The thermometer uses a<br>negative temperature<br>coefficient thermistor<br>embedded in a<br>measurement tip that is in<br>contact with the<br>measurement site. As the<br>thermistor changes<br>temperature, the resistance<br>of the thermistor also<br>changes, which is measured<br>by the thermometer and<br>converted to a measurement<br>of the temperature of the tip<br>of the thermometer. This<br>temperature, following the<br>use of the predictive<br>algorithm, is then displayed<br>to the end user. Because the<br>thermometer displays the<br>measurement for the<br>physiological site at which it<br>is used, it does not need to<br>convert this temperature via<br>clinical offset. | The thermometer uses a<br>negative temperature<br>coefficient thermistor<br>embedded in a measurement<br>tip that is in contact with the<br>measurement site. As the<br>thermistor changes<br>temperature, the resistance<br>of the thermistor also<br>changes, which is measured<br>by the thermometer and<br>converted to a measurement<br>of the temperature of the tip<br>of the thermometer. This<br>temperature, following the<br>use of the predictive<br>algorithm, is then displayed to<br>the end user. Because the<br>thermometer displays the<br>measurement for the<br>physiological site at which it<br>is used, it does not need to<br>convert this temperature via<br>clinical offset. | Same |
| Measurement<br>range | $89.6°F-109.4°F (32°C-43°C)$ | $34- 43° C (93.2- 109.4° F)$ | Different note 6 |
| Accuracy | $89.6°F~102.2°F (32°C-39°C) ±0.2°F102.3°F~109.4°F (39.1°C-43°C) ±0.3°F$ | $± 0.1°C / 0.2°F$ within<br>measurement range ( $34.0°C$<br>to $43.0°C /93.2°F$ to $109.4°F$ )<br>at room temperature of $71°F$ | Different note 7 |
| Display resolution | $0.1°F (0.1°C)$ | $0.1°F (0.1°C)$ | Same |
| Response time | Predictive mode: 30 seconds<br>Monitor mode: 3 minutes | 2 to 8 seconds | Different note 8 |
| Signal output and<br>display | LCD, Buzzer | LCD, Buzzer | Same |
| Operating<br>Environment | temperature: +41°F~+104°F<br>(5°C-40°C )<br>humidity: ≤85% RH | 15.0°C to 40.0°C /<br>59.0°F to 104.0°F;<br>≤ 95% Relative Humidity | Different note 9 |
| Storage<br>Environment | temperature: -4°F~+131°F<br>(-20°C-55°C )<br>Humidity: ≤95% RH | -25.0°C to 60.0°C /<br>-13.0°F to 140.0°F;<br>≤ 95% Relative Humidity | Different note 10 |
| Power supply | One 3V CR1632 button | One 3V CR2032 button | Different note 11 |
| Battery life | Approx 1000 measurement<br>for predictive mode<br>About 200 times for other<br>mode | More than 400<br>measurements or<br>approximately 2 years if used<br>every other day. | Different note 12 |
| Materials | User contacting materials<br>include ABS (shell, battery<br>door and On / Off Button),<br>PET(LCD lens and protective<br>film), PP(probe protective<br>cover and SUS(sensor tip) | User contacting materials<br>include ABS (sensor head,<br>battery door and On / Off<br>Button), TPR (probe and Site<br>Selection Button), PMMA<br>(LCD lens and protective<br>cover), and stainless steel<br>(sensor tip) | Different note 13 |
| Performance | Meets ASTM E1112-00:2011<br>and ISO 80601-2-56:2017 | Meets ASTM E1112-00:2011<br>and ISO 80601-2-56:2017 | Same |
| Biocompatibility | Meets ISO 10993 and FDA<br>bluebook memo G95-1 | Meets ISO 10993 and FDA<br>bluebook memo G95-1 | Same |
| Electrical Safety | Meets IEC 60601-1 | Meets IEC 60601-1 | Same |
| EMC | Meets IEC 60601-1-2 | Meets IEC 60601-1-2 | Same |
#### 7.0 Summary comparing technological characteristics with predicate device
{5}------------------------------------------------
{6}------------------------------------------------
Note 1: The subject device is not labeled for rectal use. This technological difference does not raise new questions of safety or efficacy.
Note 2: There is a difference in the indication for use because the subject device only takes intermittent measurement of body temperature on children and adults, and no disposable probe cover is provided for all measurements. Testing was not conducted with a probe cover. The software validation and clinical accuracy test demonstrated the difference does not raise any new performance questions.
Note 3: While the user population is different, the clinical accuracy test demonstrated the difference does not raise any new performance questions because it performs effectively in the labeled population of use.
{7}------------------------------------------------
Note 4: The labeling is similar and both of them comply with consensus standards.
Note 5: There is no separate Site Selection Button on the subject device and the on/off button is used as site and mode selection button, which is different from the predicate device. The performance test and software validation demonstrated the difference does not raise any new performance questions.
Note 6: The Measuring range of the subject device is different from the predicate device. The performance test and the clinical accuracy test demonstrated the difference does not raise any new performance questions.
Note 7: The accuracy of the subject device is different from the predicate device. The performance test and the clinical accuracy test demonstrated the difference does not raise any new performance questions, and it is consistent with ISO 80601-2-56.
Note 8: Response time of the subject is different from the predicate device, the clinical accuracy test demonstrated the difference does not raise any new performance questions.
Note 9: The operating environment of the new device PT1 is changed, the performance test demonstrated the difference does not raise any new performance questions.
Note 10: The storage environment of the new device PT1 is changed, the performance test demonstrated the difference does not raise any new performance questions.
Note 11: Power supply and lens filter function of the subject device is different from the predicate device. The electrical safety and EMC test demonstrated the difference does not raise any new performance questions.
Note 12: Battery life of the new device is different from the predicate device, the electrical safety test, EMC test and performance test demonstrated the difference does not raise any new performance questions.
Note 13: The material contacted is different. The biocompatibility test demonstrated the difference does not raise any new performance questions.
{8}------------------------------------------------
# 8.0 Discussion of non-clinical and clinical test performed
Non-clinical tests were conducted to verify that the proposed device meets all design specifications in order to demonstrate that it is Substantially Equivalent to the predicate device.
#### Non-clinical Tests
Electrical Safety testing was performed according to IEC 60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
EMC testing was performed according to 60601-1-2:2014, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility requirements and tests
Performance testing was performed according to ISO 80601-2-56, Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement. [Including: Amendment 1 (2018)]. This is the applicable standard for this device.
The software verification and validation were conducted according to FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", dated May 11, 2005.
Biocompatibility test (including cytotoxicity, sensitization, irritation) was performed according to (1) ISO 10993-1 Fifth edition 2018-08 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, (2)ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity and (3) ISO 10993-10 Third Edition 2010-08-01 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
## Clinical Accuracy Validation Test
Clinical investigation and data analysis have performed according to ISO 80601-2-56. The test report shows result of 40 subjects in children group (4-5 year) and 70 subjects in adults group (>5 years) (Total 110 subjects) are within acceptable range. Based on the test results, the clinical bias is ≤ 0.03 and the repeatability is ≤ 0.07, which is equivalent to the predicate device.
{9}------------------------------------------------
## 9.0 Conclusion
Basing on the performance testing, comparison and analysis above, the subject device iHealth PT1 Clinical Digital Thermometer is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.