← Product Code [DZE](/productcode/DZE) · K202773

# s-Clean SQ-SL Implant System Mini (K202773)

_Dentis Co., Ltd. · DZE · Mar 18, 2021 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K202773

## Device Facts

- **Applicant:** Dentis Co., Ltd.
- **Product Code:** [DZE](/productcode/DZE.md)
- **Decision Date:** Mar 18, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

s-Clean SQ-SL Implant System Mini is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading.

## Device Story

The s-Clean SQ-SL Implant System Mini is an endosseous dental implant system used to replace missing teeth. The system consists of a thread-type fixture (implanted into alveolar bone) and various abutments. The fixture features a dual screw-joint design and an internal hex anti-rotation connection. The fixture surface is treated with SLA (Sandblasted with Large-grit and Acid-etching) to facilitate osseointegration. The device is used by dental clinicians in surgical procedures. The clinician places the fixture into the mandible or maxilla; after healing, the abutment is connected to the fixture to support a dental prosthesis. The system restores tooth function and aesthetics for the patient. The dual screw-joint design allows for diverse screw connections (1.6M or 2.0M sizes).

## Clinical Evidence

Bench testing only. Evidence includes fatigue testing per ISO 14801:2016 comparing single vs. dual screw-joint designs, demonstrating identical performance limits. Biocompatibility testing (ISO 10993 series), sterilization validation (ISO 11137, ISO 17665), shelf-life testing (ASTM F1980), and bacterial endotoxin testing (USP <161>, <85>) were leveraged from predicate submissions (K153639, K171027, K171694, K192688) based on identical materials, manufacturing processes, and packaging.

## Technological Characteristics

Fixtures and cover screws: Pure titanium (ASTM F67). Abutments: Ti-6Al-4V ELI (ASTM F136). Surface: SLA (Sandblasted with Large-grit and Acid-etching) on fixtures; TiN coating on specific abutments. Connection: Internal hex. Sterilization: Gamma (fixtures) or steam (abutments). Dimensions: Fixtures Ø 3.7-4.1mm, lengths 7.5-13.5mm. Mechanical design: Dual screw-joint.

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Predicate Devices

- s-Clean SQ-SL Implant System Mini ([K200099](/device/K200099.md))

## Reference Devices

- TS Fixture System ([K121995](/device/K121995.md))
- [K153639](/device/K153639.md)
- [K171027](/device/K171027.md)
- [K171694](/device/K171694.md)
- [K192688](/device/K192688.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

K202773

Dentis Co., Ltd. % April Lee Consultant Withus Group Inc. 106 Superior Irvine, California 92620

Re: K202773

Trade/Device Name: s-Clean SQ-SL Implant System Mini Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: February 10, 2021 Received: February 17, 2021

Dear April Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Andrew Steen Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K202773

Device Name s-Clean SQ-SL Implant System Mini

#### Indications for Use (Describe)

s-Clean SQ-SL Implant System Mini is indicated for use in partially or fully edentulous mandibles, in support of single or multiple unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(K) Summary

#### Submitter

Dentis Co., Ltd. Gyu Ri Kim 99, Seongseoseo-ro, Dalseo-gu Daegu, 42718 Korea Email: kgr1026@dentis.co.kr Tel. +82-53-589-3541 Fax. +82-53-289-7922

## Official Correspondent

Withus Group Inc. April Lee 106 Superior, Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122

## Device Information

- Trade Name: s-Clean SQ-SL Implant System Mini ●
- . Common Name: Dental Implant System
- Classification Name: implant, endosseous, root-form
- Product Code: DZE
- Secondary Product Code: NHA
- Panel: Dental
- Regulation Number: 872.3640
- Device Class: Class II
- Date Prepared: 03/18/2021 ●

#### Predicate Devices:

The subject device is substantially equivalent to the following predicate devices:

#### Primary Predicate

- K200099, s-Clean SQ-SL Implant System Mini by Dentis Co., Ltd.
#### Reference Device

- . K121995, TS Fixture System by OSSTEM Implant Co., Ltd
#### Indication for Use:

s-Clean SO-SL Implant System Mini is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one and two stage surgical procedures. This system is intended for delayed loading.

{4}------------------------------------------------

## Device Description:

s-Clean SQ-SL Implant System Mini is composed of Fixture and Abutments. s-Clean SQ-SL Mini Fixture is a thread type implant made of Pure titanium according to ASTM F67 which will be placed in the alveolar bone to replace the function of the missing tooth. This device has connection between the upper prosthesis and the internal Hex.

The s-Clean SO-SL Implant System Mini Abutments are composed of s-Clean Sole Abutment S-Line, s-Clean TiN Half Coating Sole Abutment S-Line, and s-Clean Mini Fixture Cover Screw.

The subject implant body has a two screw-joint (structured with Upper and Lower screw joints) and a hex anti-rotation design connection. The two-screw-ioint feature is for diverse screw connection for 1.6M or 2.0M sizes. The implant bodies are only compatible with subject abutments of S-Clean Sole Abutment s-Line and s-Clean TiN Half Coating Sole Abutment s-Line with 2.0M screw. These abutments are connected with only upper screw of the implant body. The 1.6M size allows the implant to be compatible with potential future cleared abutments.

Image /page/4/Figure/6 description: The image shows three different dental implant configurations. The first configuration shows the "s-Clean SQ-SL Fixture" alone. The second configuration shows the "s-Clean SQ-SL Fixture" with the "s-Clean Cover Screw". The third configuration shows the "s-Clean SQ-SL Fixture" with the "s-Clean TiN Half Coating Sole Abutment".

The surface of fixture is treated with SLA (Sandblasted with Large-grit and Acid-etching). It is only part to be implanted into bone, and to provide connection of prosthetic devices or other components of a dental implant set with human body (mandibular or maxillary bone).

#### The dimensions of fixtures are as following:

| No. | Device Name                   | Dimension Ranges                                    |
|-----|-------------------------------|-----------------------------------------------------|
| 1   | s-Clean SQ-SL Mini<br>Fixture | Ø 3.7 and 4.1mm (D) X 7.5, 9.5, 11.5 and 13.5mm (L) |

Tolerance of dimension shall be within ± 1% range.

#### The dimensions of abutments are as following:

| No. | Device Name                                      | Dimension Ranges                                | Angulation |
|-----|--------------------------------------------------|-------------------------------------------------|------------|
| 1   | s-Clean Mini Fixture Cover Screw                 | Ø 3.2mm (D) X 5.0mm (L)                         | 0°         |
| 2   | s-Clean Sole Abutment S-Line                     | Ø 4.0 and 4.5mm (D) X 11.01, 12.01, 13.01mm (L) | 0°         |
| 3   | s-Clean TiN Half Coating Sole<br>Abutment S-Line | Ø 4.0 and 4.5mm (D) X 11.01, 12.01, 13.01mm (L) | 0°         |

{5}------------------------------------------------

| Name                                             | Uses                                                                                                                                                                    | Surface     | Connection        |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|-------------------|
| s-Clean Mini Fixture Cover<br>Screw              | It is used for protecting inner hole and connecting<br>part with exposed upper part of structure during<br>the healing period after inserting dental implant<br>fixture | N/A         | Screw<br>Retained |
| s-Clean Sole Abutment S-Line                     | The Abutment is connected with fixture and it<br>supports prosthesis which restores tooth function.                                                                     | N/A         | Internal Hex      |
| s-Clean TiN Half Coating Sole<br>Abutment S-Line | The Abutment is connected with fixture and it<br>supports prosthesis which restores tooth function.                                                                     | TiN-Coating | Internal Hex      |

The Abutments have below featured:

Tolerance of dimension for Abutments shall be within ± 1% range.

s-Clean SQ-SL Mini Fixture and s-Clean Mini Fixture Cover Screw are provided sterilized. And the other Abutments are provided non-sterilized.

s-Clean SQ-SL Mini Fixture is enclosed with s-Clean Mini Fixture Cover Screw in a set packing. s-Clean Mini Fixture Cover Screw is also provided separately.

## Materials:

- The fixtures and s-Clean Mini Fixture Cover Screw are fabricated from Pure titanium of ASTM F67 ●
- . The s-Clean Sole Abutment S-Line and s-Clean TiN Half Coating Sole Abutment S-Line are fabricated from Ti-6A1-4V of ASTM F136

{6}------------------------------------------------

# Summaries of Technological Characteristics & Substantial Equivalence Discussion:

| s-Clean SQ-SL Mini Fixture         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Division                           | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Primary Predicate                    | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| 510(k)<br>Number                   | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | K200099                              | K121995                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Trade Name                         | s-Clean SQ-SL Implant<br>System Mini                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | s-Clean SQ-SL Implant<br>System Mini | TS Fixture System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Manufacturer                       | Dentis Co., Ltd                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Dentis Co., Ltd                      | OSSTEM Implant Co., Ltd                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Product Code                       | DZE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | DZE                                  | DZE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Diameter(Ø)                        | 3.7, 4.1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 3.7, 4.1                             | 3.5, 3.75, 3.77, 4.2, 4.25, 4.4, 4.6,<br>4.63, 4.65, 4.9, 5.05, 5.08, 5.1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Length(mm)                         | 7.5, 9.5, 11.5, 13.5                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 7, 7.5, 9.5, 11.5, 13.5              | 7.0~15.0                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Design                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Screw-joint                        | Two screw-joint                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | One screw-joint                      | Two screw-joint                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Surface<br>Treatment               | SLA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | SLA                                  | SLA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Material                           | CP Titanium Grade4<br>(ASTM F67)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | CP Titanium Grade4<br>(ASTM F67)     | CP Titanium Grade4 (ASTM F67)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Abutment<br>Connection<br>Platform | Hex                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Hex                                  | Hex                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Sterilization                      | Gamma Sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Gamma Sterilization                  | Gamma Sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Shelf-Life                         | 8years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 8years                               | 8years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Indication for<br>Use              | s-Clean SQ-SL Implant System Mini is indicated for use<br>in partially or fully edentulous mandibles and maxillae,<br>in support of single or multiple unit restorations<br>including; cemented retained, screw retained, or<br>overdenture restorations, and terminal or intermediate<br>abutment support for fixed bridgework. This system is<br>dedicated for one and two stage surgical procedures.<br>This system is intended for delayed loading.                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                      | The TS Fixture System is designed<br>for dental implant surgery; it is<br>placed on the maxillary or<br>mandibular alveolar bone through a<br>surgical procedure, and after<br>osseointegration with the alveolar<br>bone, it can replace a lost tooth by<br>connecting the abutment post. The<br>TS Fixture System is indicated for<br>use in partially or fully dentulous<br>mandibles and maxillae, in support<br>or single or multiple-unit<br>restorations including: cemented<br>retained screw retained or<br>overdenture restorations and final or<br>temporary abutment support for<br>fixed bridgework. It is intended for<br>delayed loading. |
| SE<br>Comparison                   | The subject device and primary predicate has same characteristics such as indications for Use,<br>outer design, dimensions, surface treatment, material, and sterilization.<br>The difference between two devices is only design of screw joint between fixture and abutment.<br>Subject device has two screw-joint of fixture and abutment but predicate device has only one<br>screw-joint of fixture and abutment. This difference is mitigated through reference device,<br>K121995. Reference device, K121995, has same feature as dual screw joint and screw<br>connection location. Subject device and reference device has dual screw joint and are connected<br>with upper screw joint among upper and low screw joint. Results of dynamic fatigue testing<br>demonstrated that the device is substantially equivalent as the predicates despite the<br>technological differences. |                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |

# s-Clean SO-SL Mini Fixture

{7}------------------------------------------------

| Division      | Subject Device                                                                                                                                                                                                                                                     | Primary Predicate Device          |  |
|---------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|--|
| 510(k) Number | N/A                                                                                                                                                                                                                                                                | K200099                           |  |
| Trade Name    | s-Clean SQ-SL Implant System Mini                                                                                                                                                                                                                                  | s-Clean SQ-SL Implant System Mini |  |
| Manufacturer  | Dentis Co., Ltd                                                                                                                                                                                                                                                    | Dentis Co., Ltd                   |  |
| Product Code  | NHA                                                                                                                                                                                                                                                                | NHA                               |  |
| Diameter(Ø)   | 3.2                                                                                                                                                                                                                                                                | 3.1                               |  |
| Length(mm)    | 5.0                                                                                                                                                                                                                                                                | 5.4                               |  |
| Material      | CP Titanium Grade4 (ASTM F67)                                                                                                                                                                                                                                      | CP Titanium Grade4 (ASTM F67)     |  |
| Sterilization | Gamma Sterilization                                                                                                                                                                                                                                                | Gamma Sterilization               |  |
| Comparison    | The subject device and predicate device have same characteristic such as design, material,<br>sterilization, and indication for Use.<br>The diameter and length are slight difference but this difference is not an important factor to<br>the device performance. |                                   |  |

## s-Clean Mini Fixture Cover Screw

# s-Clean Sole Abutment S-Line

| Division      | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                    | Primary Predicate Device          |  |
|---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|--|
| 510(k) Number | N/A                                                                                                                                                                                                                                                                                                                                                                                                                               | K200099                           |  |
| Trade Name    | s-Clean SQ-SL Implant System Mini                                                                                                                                                                                                                                                                                                                                                                                                 | s-Clean SQ-SL Implant System Mini |  |
| Manufacturer  | Dentis Co., Ltd                                                                                                                                                                                                                                                                                                                                                                                                                   | Dentis Co., Ltd                   |  |
| Product Code  | NHA                                                                                                                                                                                                                                                                                                                                                                                                                               | NHA                               |  |
| Diameter(Ø)   | 4.0, 4.5                                                                                                                                                                                                                                                                                                                                                                                                                          | 4.0, 5.0                          |  |
| Length(mm)    | 11.01, 12.01, 13.01                                                                                                                                                                                                                                                                                                                                                                                                               | 11.2, 12.2, 13.2                  |  |
| Material      | Ti-6Al-4V ELI (ASTM F136)                                                                                                                                                                                                                                                                                                                                                                                                         | Ti-6Al-4V ELI (ASTM F136)         |  |
| Angulation    | N/A                                                                                                                                                                                                                                                                                                                                                                                                                               | N/A                               |  |
| Sterilization | Steam sterilization by End user                                                                                                                                                                                                                                                                                                                                                                                                   | Steam sterilization by End user   |  |
| Comparison    | The subject device and predicate device have same characteristic such as design, material,<br>sterilization, angulation and indication for Use.<br>The subject device has shorter lengths of the abutments such as 11.01mm, however, this<br>difference is minimal and does not impact device performance as demonstrated by fatigue<br>testing. In conclusion, subject device and predicate device are substantially equivalent. |                                   |  |

# s-C lean TiN Half Coating Sole Abutment S-Line

| Division             | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                        | Primary Predicate Device          |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| 510(k) Number        | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K200099                           |
| Trade Name           | s-Clean SQ-SL Implant System Mini                                                                                                                                                                                                                                                                                                                                                                                                                     | s-Clean SQ-SL Implant System Mini |
| Manufacturer         | Dentis Co., Ltd                                                                                                                                                                                                                                                                                                                                                                                                                                       | Dentis Co., Ltd                   |
| Product Code         | NHA                                                                                                                                                                                                                                                                                                                                                                                                                                                   | NHA                               |
| Diameter(ø)          | 4.0, 4.5                                                                                                                                                                                                                                                                                                                                                                                                                                              | 4.0, 5.0                          |
| Length(mm)           | 11.01, 12.01, 13.01                                                                                                                                                                                                                                                                                                                                                                                                                                   | 11.2, 12.2, 13.2                  |
| Material             | Ti-6Al-4V ELI (ASTM F136)                                                                                                                                                                                                                                                                                                                                                                                                                             | Ti-6Al-4V ELI (ASTM F136)         |
| Angulation           | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                   | N/A                               |
| Sterilization        | Steam sterilization by End user                                                                                                                                                                                                                                                                                                                                                                                                                       | Steam sterilization by End user   |
| Surface<br>Treatment | TiN Coating                                                                                                                                                                                                                                                                                                                                                                                                                                           | TiN Coating                       |
| Comparison           | The subject device and predicate device have same characteristic such as design, material,<br>sterilization, angulation, surface treatment, and indication for Use.<br>The subject device has shorter lengths of the abutments such as 11.01mm, however, this<br>difference is minimal and does not affect device performance as demonstrated by fatigue<br>testing. In conclusion, subject device and predicate device are substantially equivalent. |                                   |

{8}------------------------------------------------

## Non-Clinical Test Data:

Below tests were performed for predicate devices and leveraged for the subject device:

- Sterilization Validation Test on Fixtures according to ISO 11137-1.2,3 referenced in K153639
- . End User Sterilization Validation Test Report on Abutments according to ANSI/AAMI ST79, ISO 17665-1. ISO 17665-2. ISO 11737-1. ISO 11737-2. and ISO 11138-1 referenced in K171027
- . Shelf Life Test on Fixtures according to ASTM F1980 referenced in K153639
- Biocompatibility testing on fixtures according to ISO 10993-1:2009, ISO 10993-3:2014.ISO ● 10993-5:2009, ISO 10993-6:2007, ISO 10993-10:2010 and ISO 10993-11:2006 referenced in K153639
- Biocompatibility testing on Abutments according to ISO 10993-1:2009, ISO 10993-3:2014, ISO 10993-5:2009, ISO 10993-6:2007, ISO 10993-10:2010 and ISO 10993-11:2006 referenced in K171027
- . Biocompatibility testing on TiN Coating Abutments according to ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-10:2010 and ISO 10993-11:2006 referenced in K171694
- . Bacterial Endotoxin Test Report on Fixtures according to ANSVAAMI ST72:2011, USP <161>, and USP <85> referenced in K192688

The results of the above tests have met the criteria of the standards, and demonstrated the substantial equivalence with the predicate device.

The surface modification information with SLA (Sandblasted with Large-grit and Acid-etching) for fixtures was provided. To compare surface modification between the subject and predicate devices, K153639, surface roughness, surface composition analysis, and SEM imaging were provided and it demonstrate the substantial equivalence.

The Sterilization validation test and shelf-life test for fixtures were performed for predicate device, K153639 and leveraged for the subject device because the material, sterilization method, packaging methods, and manufacturing process of the both products are exactly same.

The end user sterilization test was performed for predicate device, K171027 and leveraged for the subject device because the product category, material, manufacturing process, facility, and packaging of the both products are exactly same.

The Biocompatibility Test was conducted on the predicate device and leveraged for the subject device because both products are manufactured with same materials and manufacturing process. It demonstrates that the subject device is biocompatible and substantial equivalence with the predicate.

The fatigue testing per ISO 14801 was conducted under the worst-case scenario. To demonstrate the substantial equivalence of the device performance between single screw joint implant and dual screw joint implant, two fatigue tests were performed comparatively, and result say that device performance was substantial equivalent.

| Fatigue Test Report                   | Predicate Device | Subject Device-1 | Subject Device-2 |
|---------------------------------------|------------------|------------------|------------------|
| Fixture Size                          | Ø 3.7 x 13.5mm   | Ø 3.7 x 7.5mm    | Ø 3.7 x 13.5mm   |
| Abutment Size                         | Ø 4.0 x 12.2mm   | Ø 4.5 x 13.01mm  | Ø 4.5 x 13.01mm  |
| Screw Joint Design                    | Single           | Dual             | Dual             |
| Compressive Load Result               | Similar          | Similar          | Similar          |
| Fatigue Test Method<br>(ISO Standard) | ISO14801:2016    | ISO14801:2016    | ISO14801:2016    |
| Fatigue Test Result<br>(Limit)        | Identical        | Identical        | Identical        |
| Gap of connection part                | 0 µm             | 0 µm             | 0 µm             |

{9}------------------------------------------------

The non-clinical testing results demonstrate that the subject device is substantially equivalent to the predicate device.

## Conclusion:

s-Clean SQ-SL Implant System Mini constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate devices. Therefore, s-Clean SQ-SL Implant System Mini and its predicates are substantially equivalent.

---

**Source:** [https://fda.innolitics.com/device/K202773](https://fda.innolitics.com/device/K202773)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
