K202766 · Guangzhou Ajax Medical Equipment Co., Ltd. · FRA · Jun 11, 2021 · General Hospital
Device Facts
Record ID
K202766
Device Name
EOS Air Cleaner
Applicant
Guangzhou Ajax Medical Equipment Co., Ltd.
Product Code
FRA · General Hospital
Decision Date
Jun 11, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6500
Device Class
Class 2
Indications for Use
EOS Air Cleaner is a device intended for medical purposes that is used to destroy bacteria and viruses in the air by exposure to ultraviolet radiation in general medical environment. The core technology components of the EOS Air Cleaner have been demonstrated to destroy MS2 bacteriophage bioaerosol entrained on the filter to achieve an average 4 log reduction over 4 hours in a 30 cubic meter chamber.
Device Story
Free-standing medical air purifier; uses motor to suction ambient air into sealed metal cabinet; air passes through fine filter and HEPA filter; UV-C radiation (255 nm) from two internal lamps irradiates filter media; photocatalytic coated filter media and UV-C light decompose organic structures of bacteria and viruses; purified air expelled from cabinet; operated by buttons for power and air velocity; intended for use in hospitals, nursing homes, and medical facilities; provides reduction of airborne pathogens; benefits patient/staff by reducing bioaerosol load in medical environments.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via MS2 bacteriophage bioaerosol reduction testing in a 30 cubic meter chamber, achieving >4 log reduction over 4 hours. Aging tests confirmed continued performance over service life.
Technological Characteristics
Free-standing unit; sealed metal enclosure; two UV-C Xenon flash lamps (255 nm); fine filter and HEPA filter; photocatalytic coated filter media; centrifugal fan; AC110V/60Hz power; dimensions 12.25" x 10.75" x 40"; manual button controls; electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for destroying bacteria and viruses in the air within general medical environments (e.g., hospitals, nursing homes, medical facilities), excluding specialized settings like operating rooms.
Regulatory Classification
Identification
A medical ultraviolet air purifier is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 11, 2021
Guangzhou Ajax Medical Equipment Co., Ltd. % Iris Fung Specialist SGS-CSTC Standards Technical Services Co., Ltd. 198 KEZHU Road, SCIENTECH Park Guangzhou Economic & Technology Deve Guangzhou, Guangdong China
Re: K202766
Trade/Device Name: EOS Air Cleaner Regulation Number: 21 CFR 880.6500 Regulation Name: Medical Ultraviolet Air Purifier Regulatory Class: Class II Product Code: FRA Dated: May 10, 2021 Received: May 14, 2021
Dear Iris Fung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Clarence W. Murray, III, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202766
Device Name EOS Air Cleaner
Indications for Use (Describe)
EOS Air Cleaner is a device intended for medical purposes that is used to destroy bacteria and viruses in the air by exposure to ultraviolet radiation in general medical environment.
The core technology components of the EOS Air Cleaner have been demonstrated to destroy MS2 bacteriophage bioaerosol entrained on the filter to achieve an average 4 log reduction over 4 hours in a 30 cubic meter chamber.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;"> </span> |
|----------------------------------------------|---------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="text-decoration: overline;">X</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
### 510(k) Number: 202766
This 510(K) Summary is being submitted in accordance with the requirement of 21 CFR 807.92. This is a Traditional 510(K) submission, and there were no prior submissions for the subject device.
#### ー Submitter's Information Sponsor
- � Company Name: Guangzhou Ajax Medical Equipment Co., Ltd.
- � Address: Building No.2 Dagang industrial zone, Shilou Town, Panyu District, Guangzhou City, Guangdong Province, 511447, P.R. China
- � Phone: +86 20 84847938
- � Email: ella@ajaxdent.com
- � Contact Person (including title): Atwood Lee (General Manager)
### Application Correspondent:
- ♦ SGS-CSTC Standards Technical Services Co., Ltd.
- ♦ Address: 198 KEZHU Road, SCIENTECH Park Guangzhou Economic & Technology Development
District, Guangzhou, Guangdong, CHINA
- ♦ Contact Person: Ms. Iris Fung
- ♦ Tel: +86-20-32136908
- ♦ Email: Iris.Fung@sgs.com
#### ﻨ Subject Device Information
- � Type of 510(k) submission: Traditional
- Regulation Name: Medical Ultraviolet Air Purifier �
- � Trade Name: EOS Air Cleaner
- � Review Panel: General Hospital
- � Product Code: FRA
- � Regulation Number: 21 CFR 880.6500
- � Regulation Class: II
#### 3. Date Prepared:
June 9, 2021
#### 4. Predicate Device
- � 510(k) number: K200500
- � Sponsor: Molekule, Inc.
- � Regulation Name: Medical Ultraviolet Air Purifier
- � Trade Name: Molekule Air ProRX
- � Review Panel: General Hospital
- � Product Code: FRA
- � Regulation Number: 21 CFR 880.6500
{4}------------------------------------------------
- � Regulation Class: II
### 5. Device Description
EOS Air Cleaner employs an ultraviolet air purification technology that destroys bacteria and viruses in air in hospital, nursing homes, medical facilities, but not in specialized medical place such as operating rooms.
EOS Air Cleaner is a free-standing device, which has a well-sealed metal box including a fine filter, motor, two UV-C lamps, and a HEPA filter. By sucking surrounding air into the device, EOS Air cleaner can suction the air into the device and filtering the air by fine filter with UV-C light radiation, then the purified air is expelled from the cabinet.
The bacteriophage deposited onto the surface of EOA Air Cleaner's proprietary photocatalytic coated filter media will be killed after exposure to EOS Air Cleaner's UV-C light
UV-C light in 255 nm which is commonly called germicidal kills any remaining bacteria and viruses, as the radiation penetrates the cell walls of bacteria and is absorbed by the organic structures within the bacterial and virus, causing them to decompose and the cell to die.
### 6. Indications for Use
EOS Air Cleaner is a device intended for medical purposes that is used to destroy bacteria and viruses in the air by exposure to ultraviolet radiation in general medical environment.
The core technology components of the EOS Air Cleaner have been demonstrated to destroy MS2 bacteriophage bioaerosol entrained on the filter to achieve an average 4 log reduction over 4 hours in a
| | Subject Device | Primary Predicate Device | Comparison |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 510(k)<br>Number | K202766 | K200500 | |
| Manufacturer | GUANGZHOU AJAX MEDICAL<br>EQUIPMENT CO., LTD. | Molekule, Inc. | -- |
| Product Name | EOS Air Cleaner | Molekule Air Pro RX | -- |
| Classification | Class II, FRA (21 CFR880.6500) | Class II, FRA (21 CFR880.6500) | Same |
| Indications for Use | EOS Air Cleaner is a device intended for medical purposes that<br>is used to destroy bacteria and viruses in the air by exposure to<br>ultraviolet radiation in general medical environment.<br>The core technology components of the EOS Air Cleaner have been<br>demonstrated to destroy MS2 bacteriophage bioaerosol entrained<br>on the filter to achieve an average 4 log reduction over 4 hours in a 30<br>cubic meter chamber. | The Molekule Air Pro RX air purifier is a device intended for medical purposes<br>that is used to destroy bacteria and viruses in the air by exposure to<br>ultraviolet radiation.<br>The core technology components of the Molekule Air Pro RX air purifier have<br>been demonstrated to destroy the following MS2 bacteriophage bioaerosol<br>entrained on the filter of the subject device under the following exposure<br>conditions: | Similar |
| Light Source | 2 UV-C Xenon Flash lamps<br>Wavelength 255 nm | UV Light Source, LED Wavelength, 320-<br>400nm<br>Power per Lamp/ String, 11.4W Number<br>of Lamps/ String, 6 Total UV Power,<br>68.4W<br>Filter Irradiance (Minimum), 30W/m² | Different |
| MS2<br>Bacteriophage<br>inactivation | Average Log reduction/exposure<br>time (hours)<br>Speed 1: 4.05 / 4 hours<br>Speed 5: 4.22 / 4 hours<br>Speed 10: 4.45 / 4 hours | Average Maximum log reduction<br>/exposure time (hours) Room temperature<br>test 5.21 / 24 hours | Similar |
| Environment<br>of Use | Hospital, nursing homes, medical<br>facilities (not surgery setting) | Hospital and general surgery setting | Similar |
| User Control | 4 Buttons to control the device:<br>Power button to turn on/off.<br>Start/Stop button to start/stop<br>device. Higher and Lower button to<br>adjust the<br>air suction velocity. | One knob controls the four speed fan<br>setting One button turns the unit on and<br>off. | Similar |
| Safety<br>Features | UV lamps and centrifugal fan are<br>contained inside sealed metal box. | Safety switches exist in the following<br>locations:<br>PECO filter door, pre filter door, PECO<br>filter compartment, and pre filter<br>compartment. If any door is open or if a<br>filter is missing, the unit will not operate.<br>The purpose of these switches is to<br>protect the user from any possibility of<br>exposure to direct contact with UV light.<br>Grill at outlet and inlet of fan with small<br>enough grating to block user from<br>accessing spinning fan without tools. | Similar |
| Power Supply | AC110V, 60Hz 12-20A | 120 Volt (plugs into standard single<br>phase 120 Volt outlet)<br>Up to 3.72 amps | Similar |
| Power<br>Consumption | 1300 Watts | Up to 450 Watts | Different |
| Dimensions | Main unit: 12.25" × 10.75"× 40" | 22" × 22"× 52" | Similar |
| Safety and<br>EMC | IEC 60601-1: Pass<br>IEC 60601-1-2 CISPR Group 1A: Pass | UL 507<br>IEC 60601-1-2 | Similar |
### 7. Comparison of Technological Characteristics
30 cubic meter chamber.
{5}------------------------------------------------
### 8. Summary for non-clinical test
### The non-clinical testing was provided to demonstrate the subject device met the acceptance criteria of the test methodology or standards listed below.
EOS Air Cleaner complies with voluntary standards for electrical safety and electromagnetic
compatibility.
| Name of Test<br>Methodology | Purpose | Acceptance Criteria | Results |
|-----------------------------|--------------------------------------|--------------------------------------|---------|
| IEC 60601-1 | Demonstration of<br>basic safety and | Meets criteria for<br>CISPR Group 1A | Pass |
{6}------------------------------------------------
| IEC 60601-1-2 | essential performance | | |
|---------------------|-------------------------------------------------------------------|-----------------------------------------------------|------|
| Bacteriophage Test | Demonstration of log<br>reduction of viral load | At least 4 log<br>reduction of MS2<br>bacteriophage | Pass |
| Aging Test | Demonstration of<br>continued function<br>throughout service life | At least 4 log<br>reduction of MS2<br>bacteriophage | Pass |
| Software Validation | Demonstration of<br>software security and<br>good design | Meets design criteria | Pass |
### 9. Clinical Testing
Clinical testing was not performed.
### 10. Conclusion
The conclusion drawn from the nonclinical testes demonstrates that the subject device in 510(K) submission K202766, EOS Air Cleaner, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K202766.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.