← Product Code [FLL](/productcode/FLL) · K202753

# iHealth Wireless No-Touch Forehead Thermometer (K202753)

_Andon Health Co, Ltd. · FLL · Mar 8, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K202753

## Device Facts

- **Applicant:** Andon Health Co, Ltd.
- **Product Code:** [FLL](/productcode/FLL.md)
- **Decision Date:** Mar 8, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2910
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Pediatric

## Indications for Use

The Wireless No-Touch Forehead Thermometer is intended for the intermittent measurement of body temperature from the forehead on people of all ages. It can transmit the temperature to a smart device with Bluetooth and can be used by consumers in the household environment and by healthcare providers.

## Device Story

Hand-held, battery-operated, non-contact infrared thermometer; measures natural thermal radiation from central forehead. User positions probe ≤ 3cm from forehead; thermopile sensor detects infrared radiation; internal circuitry amplifies and calculates temperature; converts to oral equivalent. Results displayed on device LED; transmits data via Bluetooth to smart device running iHealth My Vitals Pro App. Used by consumers or healthcare providers in home or clinical settings. Provides rapid temperature readings to assist in clinical decision-making regarding patient health status.

## Clinical Evidence

Clinical investigation conducted per ASTM E1965-98:2016 with 130 subjects: 40 infants (0-1 yr), 40 children (1-5 yrs), 50 adults (>5 yrs). Primary endpoints focused on clinical bias and uncertainties. Results met acceptance criteria, demonstrating compliance with ASTM E1965-98 (2016) requirements.

## Technological Characteristics

Non-contact infrared thermopile sensor; ABS and PMMA materials (ISO 10993-1 compliant); LED display; Bluetooth Low Energy connectivity; 2x1.5V AAA battery power. Standards: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 80601-2-56 (clinical thermometers).

## Regulatory Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

## Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; 
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and 
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

## Predicate Devices

- iHealth Infrared No-Touch Forehead Thermometer (Model: PT3) ([K200531](/device/K200531.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

March 8, 2021

Andon Health Co., Ltd. Liu Yi President No. 3 Jin Ping Street, Ya An Road, Nankai District Tianjin, Tianjin 300190 China

Re: K202753

Trade/Device Name: iHealth Wireless No-Touch Forehead Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical electronic thermometer Regulatory Class: Class II Product Code: FLL Dated: February 1, 2021 Received: February 5, 2021

Dear Liu Yi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202753

Device Name

iHealth Wireless No-Touch Forehead Thermometer (PT3SBT)

#### Indications for Use (Describe)

The Wireless No-Touch Forehead Thermometer is intenittent measurement of body temperature from the forehead on people of all ages. It can transmit the temperature to a smart device with Bluetooth and can be used by consumers in the household environment and by healthcare providers.

| Type of Use (Select one or both, as applicable)                                 |
|---------------------------------------------------------------------------------|
| <input type="checkbox"/> Registration Use (Part 81 CFR 901 Subpart D)           |
| <input checked="" type="checkbox"/> Over-The-Counter Use (81 CFR 901 Subpart C) |

__ Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K202753

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

## 1.0 submitter's information

| Name:                | Andon Health Co., Ltd.                                                   |
|----------------------|--------------------------------------------------------------------------|
| Address:             | No 3, Jinping Street Ya An Road, Nankai District, Tianjin,<br>P.R. China |
| Phone number:        | 86-22-87611660                                                           |
| Fax number:          | 86-22-6052 6162                                                          |
| Contact:             | Liu, Yi                                                                  |
| Date of Preparation: | 2/11/2021                                                                |

## 2.0 Device information

| Trade name:          | iHealth Wireless No-Touch Forehead Thermometer (model PT3SBT) |
|----------------------|---------------------------------------------------------------|
| Common name:         | Infrared forehead thermometer                                 |
| Classification name: | Clinical Electronic Thermometer                               |

## 3.0 Classification

Production code: FLL Regulation number: 880.2910 Classification: II Panel: General Hospital

## 4.0 Predicate device information

Manufacturer: Andon Health Co., Ltd iHealth Infrared No-Touch Forehead Thermometer (Model: PT3) Device: 510(k) number: K200531

## 5.0 Indications for Use

The Wireless No-Touch Forehead Thermometer is intended for the intermittent measurement of body temperature from the forehead on people of all ages. It can transmit the temperature to a smart device with Bluetooth and can be used by consumers in the household environment and by healthcare providers.

{4}------------------------------------------------

# 6.0 Device description

The iHealth PT3SBT Wireless No-Touch Forehead Thermometer is hand-held, reusable, battery operated device, which can measure human body forehead skin temperature without touching patient's skin. Its operation is based on measuring the natural thermal radiation from the central forehead. Put the probe of the thermometer at a measurement distance less than or equal to 1.18 inch (3cm) without touching the skin at the center of the patient's forehead. It uses a thermopile sensor to measure the patient's forehead temperature and convert it to the oral equivalent. Pressing the activation button to start the measurement of proposed infrared radiation. The electrical signal read out from the detector is fed to the circuit for amplification and calculation. The final measured temperature will be appeared on a LED display of the device. The thermometer can also be connected to a smart device through Bluetooth and display the measured temperature result on the smart device.

| Item                                          | Subject Device<br>(K202753)                                                                                                                                                                                                                                                                                                                   | Predicate Device<br>(K200531)                                                                                                                                                                                                                                         | Comparison  |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Name and Model                                | iHealth Wireless No-Touch<br>Forehead Thermometer<br>Model: PT3SBT                                                                                                                                                                                                                                                                            | iHealth Infrared No-Touch<br>Forehead Thermometer<br>Model: PT3                                                                                                                                                                                                       | ---         |
| Product code                                  | FLL                                                                                                                                                                                                                                                                                                                                           | FLL                                                                                                                                                                                                                                                                   | Same        |
| Regulation<br>number                          | 21 CFR 880.2910                                                                                                                                                                                                                                                                                                                               | 21 CFR 880.2910                                                                                                                                                                                                                                                       | Same        |
| Manufacturer                                  | Andon Health Co., Ltd                                                                                                                                                                                                                                                                                                                         | Andon Health Co., Ltd                                                                                                                                                                                                                                                 | Same        |
| Indications for<br>Use                        | The Wireless No-Touch<br>Forehead Thermometer is<br>intended for the intermittent<br>measurement of body<br>temperature from the<br>forehead on people of all<br>ages. It can transmit the<br>temperature to a smart device<br>with Bluetooth and can be<br>used by consumers in the<br>household environment and<br>by healthcare providers. | The Infrared No-Touch<br>Forehead Thermometer is<br>intended for the intermittent<br>measurement of body<br>temperature from the<br>forehead on people of all<br>ages. It can be used by<br>consumers in the<br>household environment and<br>by healthcare providers. | Different 1 |
| Probe contact<br>type                         | Non-contact;<br>measurement distance: ≤<br>1.18 in(3 cm)                                                                                                                                                                                                                                                                                      | Non-contact;<br>measurement distance: ≤<br>1.18 in(3 cm)                                                                                                                                                                                                              | Same        |
| Sensor                                        | Thermopile                                                                                                                                                                                                                                                                                                                                    | Thermopile                                                                                                                                                                                                                                                            | Same        |
| Measurement<br>method                         | Infrared radiation detection                                                                                                                                                                                                                                                                                                                  | Infrared radiation detection                                                                                                                                                                                                                                          | Same        |
| Display type                                  | LED                                                                                                                                                                                                                                                                                                                                           | LED                                                                                                                                                                                                                                                                   | Same        |
| Other feature                                 | 1. Vibration when get<br>measurement<br>2. Can connect to a smart<br>device and display measured<br>temperature result on a smart<br>device.                                                                                                                                                                                                  | 1. Vibration when get<br>measurement                                                                                                                                                                                                                                  | Different 2 |
| APP name and<br>feature                       | iHealth My Vitals Pro App<br>Store and show data received<br>from the thermometer through<br>Bluetooth                                                                                                                                                                                                                                        | No App related                                                                                                                                                                                                                                                        | Different 3 |
| Measuring range                               | 32- 42.9° C (89.6- 109.2° F)                                                                                                                                                                                                                                                                                                                  | 32- 42.9° C (89.6- 109.2° F)                                                                                                                                                                                                                                          | Same        |
| Accuracy                                      | $\pm$ 0.4°F( $\pm$ 0.2°C) within<br>93.2°F-107.6°F(34°C-42°C),<br>and $\pm$ 0.5°F( $\pm$ 0.3°C) for other<br>temperature ranges.                                                                                                                                                                                                              | $\pm$ 0.4°F (0.2°C) within<br>95~107.6°F (35~42°C),<br>$\pm$ 0.5°F (0.3°C) for other<br>range.                                                                                                                                                                        | Different 4 |
| Measuring<br>response time                    | 1 second                                                                                                                                                                                                                                                                                                                                      | 1 second                                                                                                                                                                                                                                                              | Same        |
| Low battery<br>indicator                      | Indicated on LED                                                                                                                                                                                                                                                                                                                              | Indicated on LED                                                                                                                                                                                                                                                      | Same        |
| Temperature unit                              | °C or °F                                                                                                                                                                                                                                                                                                                                      | °C or °F                                                                                                                                                                                                                                                              | Same        |
| Display resolution                            | 0.1 °F (0.1 °C)                                                                                                                                                                                                                                                                                                                               | 0.1 °F (0.1 °C)                                                                                                                                                                                                                                                       | Same        |
| Power<br>requirements                         | 2X1.5V AAA battery                                                                                                                                                                                                                                                                                                                            | 2X1.5V AAA battery                                                                                                                                                                                                                                                    | Same        |
| Operation<br>environment                      | 0-40° C, 15%-95% RH,<br>70-106kPa                                                                                                                                                                                                                                                                                                             | 15-40° C, 15%-95% RH,<br>70-106kPa                                                                                                                                                                                                                                    | Different 5 |
| Storage<br>environment                        | -20-55° C, 15%-95% RH,<br>70-106kPa                                                                                                                                                                                                                                                                                                           | -20-55° C, 10%-95% RH,<br>70-106kPa                                                                                                                                                                                                                                   | Different 6 |
| Materials of<br>skin-contacting<br>components | ABS and PMMA                                                                                                                                                                                                                                                                                                                                  | ABS and PMMA                                                                                                                                                                                                                                                          | Same        |
| biocompatibility                              | Comply with ISO 10993-1                                                                                                                                                                                                                                                                                                                       | Comply with ISO 10993-1                                                                                                                                                                                                                                               | Same        |
| electrical safety                             | Comply with IEC 60601-1                                                                                                                                                                                                                                                                                                                       | Comply with IEC 60601-1                                                                                                                                                                                                                                               | Same        |
| EMC and<br>performance<br>testing             | Comply with IEC 60601-1-2                                                                                                                                                                                                                                                                                                                     | Comply with IEC 60601-1-2                                                                                                                                                                                                                                             | Same        |
| Wireless<br>communication                     | Bluetooth low energy                                                                                                                                                                                                                                                                                                                          | NA                                                                                                                                                                                                                                                                    | Different 7 |

## 7.0 Summary comparing technological characteristics with predicate device

{5}------------------------------------------------

{6}------------------------------------------------

Different 1: The indication for use of the proposed device is slightly different, because it can transmit the temperature to a smart device with Bluetooth. The performance test and the clinical accuracy test complied with standards. The difference does not raise any new performance questions.

Different 2-3: There is App on smart device for the subject device, test result displayed on the thermometer can also be transmited to App on smart device through bluetooth. The software validation demonstrated the difference does not raise any new performance questions.

Different 4: The accuracy of the proposed device is different from the predicate device, the performance test and the clinical accuracy test complied with standards. The difference does not raise any new performance questions.

Different 5: The operation temperature range of the proposed device PT3SBT is different from the predicate device, but the performance test complied with standards. The difference does not raise any new performance questions.

Different 6: The storage humidity range of the proposed device PT3SBT is different from the predicate device, but the performance test complied with standards. The difference does not raise any new performance questions.

Different 7: There is wireless communication with bluetooth low energy on the subject device, the software validation and performance test demonstrated the difference does not raise any new performance questions.

## 8.0 Discussion of non-clinical and clinical test performed

Non-clinical tests were conducted to verify that the proposed device met all design specifications in order to demonstrate that it is Substantially Equivalent to the predicate device.

#### Non-clinical Tests

Electrical Safety test was performed according to IEC 60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.

EMC test was performed according to 60601-1-2:2014, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility requirements and tests.

{7}------------------------------------------------

Performance test was performed according to ISO 80601-2-56:2017 , Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement. [Including: Amendment 1 (2018)]. is the applicable standards for this device.

Biocompatibility test (including cytotoxicity, sensitization, irritation) was performed according to (1) ISO 10993-1 Fifth edition 2018-08 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, (2)ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity and (3) ISO 10993-10 Third Edition 2010-08-01 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.

Software verification and validation were conducted according to FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", dated May 11, 2005. Cybersecurity testing has been evaluated according to guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".

# Clinical Tests

Clinical investigation and data analysis were performed according to ASTM E1965-98:2016. The test was conducted for 40 subjects in infants group (0-1 year), 40 subjects in children group (1-5 years) and 50 subjects in adults group (>5 years) (Total 130 subjects). The clinical test report demonstrated that the clinical data, represented by clinical bias and uncertainties met the acceptance criteria of the clinical study protocol. The iHealth PT3 Infrared No-Touch Forehead Thermometer complies with the requirement of ASTM E1965-98 (2016).

# 9.0 Conclusion

Basing on the performance testing, comparison and analysis, the subject device iHealth PT3SBT Wireless No-Touch Forehead Thermometer is substantially equivalent to the predicate device .

---

**Source:** [https://fda.innolitics.com/device/K202753](https://fda.innolitics.com/device/K202753)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
