Dentilube dry mouth spray is indicated for the symptomatic relief from the effects of chronic or temporary xerostomia (dry mouth), mouth discomfort, mouth odors and other oral symptoms associated with dry mouth.
Device Story
Dentilube dry mouth spray is a viscous, electrolyte-containing solution designed to mimic natural saliva. It functions as an artificial saliva substitute to moisturize and lubricate the oral cavity. The device is supplied in 50mL or 100mL PET bottles with a spray head. It is intended for use in home and clinical settings by patients suffering from xerostomia. The user applies the spray to the oral cavity as needed. By providing a lubricating and moisturizing coating, the spray relieves symptoms of dry mouth, discomfort, and odors, thereby improving patient comfort.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), and irritation (ISO 10993-10). Shelf-life/stability testing supports a 24-month expiration. No clinical trials were conducted; a 'Use Study' was mentioned but no clinical data or results were provided.
Technological Characteristics
Viscous electrolyte-containing solution; ingredients include purified water, sodium carboxymethyl-cellulose (thickener), dipotassium hydrogen phosphate/potassium dihydrogen phosphate (buffers), sorbitol (sweetener), and methylparaben (preservative). Packaged in 50mL/100mL PET bottles. Non-sterile. Biocompatibility conforms to ISO 10993 standards. pH 6.35; viscosity 3500 cps.
Indications for Use
Indicated for symptomatic relief of chronic or temporary xerostomia (dry mouth), mouth discomfort, mouth odors, and associated oral symptoms in patients experiencing dry mouth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
September 3, 2021
Oxy2plus, LLC Chun Lin Co-Founder 4120 Rosemead Blvd Rosemead, California 91770
Re: K202689
Trade/Device Name: Dentilube Spray Regulatory Class: Unclassified Product Code: LFD Dated: June 28, 2021 Received: July 7, 2021
Dear Chun Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adiodha. M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202689
#### Device Name Dentilube dry mouth sprav
Indications for Use (Describe)
Dentilube dry mouth spray is indicated for the symptomatic relief from the effects of chronic or temporary xerostomia (dry mouth), mouth discomfort, mouth odors and other oral symptoms associated with dry mouth.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(K) SUMMARY (K202689)
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
#### SUBMITTER INFORMATION 1.
| Name: | Oxy2plus, LLC |
|------------------------|-----------------------|
| Address: | 4120 Rosemead |
| | Blvd Rosemead, |
| | CA 91770 |
| Contact person: | Chun Nan Lin, DDS, MS |
| Telephone: | 626-286-7800 |
| FAX: | 626-286-7600 |
| Email: | oxy2plus@outlook.com |
| Date Summary Prepared: | June 26, 2021 |
### 2. DEVICE NAME
| Device Name: | Dentilube dry mouth spray |
|----------------------------|-----------------------------------|
| Trade or Proprietary Name: | Dentilube dry mouth spray |
| Common or Usual Name: | Saliva, Artificial Classification |
| Name: | Saliva, Artificial |
| Product Code: | LFD |
| Classification: | Unclassified |
| Panel: | Dental |
## 3. IDENTIFICATION OF EQUIVALENCE (Predicate Device):
BIO-X HEALTHCARE S.A. BioXtra® Moisturizing Gel (K072306)
## 4. DEVICE DESCRIPTION
Dentilube dry mouth spray is a viscous pink electrolyte-containing solution that is designed to mimic the compositions and actions of saliva to moisturize, and refresh the mouth. It provides a lubricating and moisturizing coating inside the mouth, thereby relieving symptoms of dry mouth.
The product is supplied in 50mL or 100 mL PET bottles.
## 5. STATEMENT OF INTENDED USE
{4}------------------------------------------------
Dentilube dry mouth spray is indicated for the symptomatic relief from the effects of chronic or temporary xerostomia (dry mouth), mouth discomfort, mouth odors and other oral symptoms associated with dry mouth.
# 6. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS
| Attributes | Subject Device | Predicate Device | Comparison |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Product | Dentilube dry mouth<br>spray | BioXtra® Moisturizing<br>Gel | - |
| Manufacturer | Oxy2plus, LLC | Bio-X Healthcare S.A. | - |
| 510(K) # | K202689 | K072306 | - |
| Product Code | LFD | LFD | Same |
| Regulation | Pre-Amendment | Pre-Amendment | Same |
| Class | Unclassified | Unclassified | Same |
| Review Panel | Dental | Dental | Same |
| Indications for<br>Use | Dentilube dry mouth spray<br>is indicated for the<br>symptomatic relief from<br>the effects of chronic or<br>temporary xerostomia (dry<br>mouth), mouth discomfort,<br>mouth odors and other oral<br>symptoms associated with<br>dry mouth. | BioXtra® is indicated<br>for the symptomatic<br>relief from the effects of<br>chronic or temporary<br>xerostomia (dry mouth),<br>mouth discomfort,<br>mouth odors and other<br>oral symptoms<br>associated with dry<br>mouth. | Same; both are<br>indicated for<br>relief of<br>symptoms of<br>xerostomia (dry<br>mouth). |
Characteristics of the device compared to the predicate device Substantial Equivalence Comparison Chart
| Conditions of Use | | | | Mode of<br>Action | Moisturizing and<br>lubricating oral dryness | Moisturizing and<br>lubricating oral dryness | Same |
|------------------------|------------------------------------------------------|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dosage Form | Oral spray | Oral spray | Same | Type of<br>Product | Liquid solution | Liquid solution | Same |
| Area of Use | Oral cavity | Oral cavity | Same | Presentation<br>Rx/OTC | Non-sterile<br>OTC | Non-sterile<br>Rx/OTC | same<br>The subject<br>device is OTC<br>only |
| Disease State | Xerostomia | Xerostomia | Same | Composition | | | |
| Method of Use | Ready to use gel spray | Ready to use gel spray | Same | Solvent | Purified water | Purified water | same |
| Application<br>per Day | As needed | As needed | Same | Buffers | Dipotassium hydrogen<br>phosphate<br>Potassium dihydrogen<br>phosphate | Potassium phosphate,<br>dibasic<br>Sodium Chloride | Similar. These<br>different<br>buffering agents<br>are generally<br>recognized as<br>safe (GRAS) and<br>provide similar<br>pH levels for the<br>both devices. |
| Packaging | 50 mL and 100 mL in<br>PET bottle with spray<br>head | 50 mL in PET bottle<br>with spray head | Same | Humectants/<br>Thickeners | Sodium carboxymethyl-<br>cellulose | Hydroxyethyl cellulose<br>Polyacrylic acid<br>Sodium polyacrylate<br>Hydrogenated starch<br>hydrolysates (HSH) | Similar. These<br>different thickener<br>agents are generally<br>recognized as safe<br>(GRAS) and provide<br>similar viscosity and<br>moisture-retaining<br>quality for both<br>devices. |
| Area of Use | Oral cavity | Oral cavity | Same | Preservatives | Methyl P-<br>hydroxybenzoate<br>(Methylparaben) | Sodium methylparaben<br>Sodium propylparaben<br>Sodium benzoate<br>Potassium sorbate<br>Citric acid monohydrate<br>Lactoperoxidase<br>(Hydrogen-peroxide<br>oxidoreductase) | Similar. These<br>preservative agents<br>are generally<br>recognized as safe<br>(GRAS) by the<br>USFDA and are used<br>for maintaining and<br>extending shelf life. |
| Environment<br>of Use | Home and Clinic | Home and Clinic | Same | Colorant | FD&C Red 40 Dye<br>(2%) | None | Colorant is not<br>used in the<br>predicate device |
| Sweeteners | Sorbitol | Sorbitol<br>Xylitol<br>Sodium Saccharin | Similar. The<br>predicate has more<br>sweeteners. Both<br>devices use sorbitol<br>to balance the<br>solution and to<br>provide a denser<br>solution. | | | | |
{5}------------------------------------------------
{6}------------------------------------------------
| Electrolytes | Calcium chloride,<br>anhydrous<br>Magnesium Chloride,<br>anhydrous<br>Potassium Chloride<br>Sodium fluoride | Calcium chloride,<br>dihydrate<br>Magnesium chloride,<br>hexahydrate<br>Potassium Chloride<br>Sodium monofluoro-<br>phosphate | Similar. Both devices<br>contain electrolytes<br>similar to those<br>found in natural<br>saliva. |
|--------------|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
|--------------|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
The subject device, Dentilube dry mouth spray, and the predicate device have the same indications and method of use. They also share many conditions of use. Further, both employ the same fundamental scientific technology (a formulation of water, humectants or moisturizers, thickening/binding agents, buffering agents, sweeteners, flavor, surfactants and preservatives).
| Physical Properties (Non-Clinical Performance) | | | |
|-------------------------------------------------|----------------------|----------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Flavors | Lemon / Mint | Mint | |
| Appearance | Clear Viscous Liquid | Clear Viscous Liquid | Same |
| Color | Reddish Pink | Light Straw | Different color<br>does not affect<br>use |
| pH | 6.35 | 6.82 | Similar. Both<br>subject and<br>predicate<br>devices have pH<br>values near that<br>of natural saliva<br>(5.3 - 7.8) |
| Solubility | Water Soluble | Water Soluble | Same |
| Viscosity<br>@25°C [RVT<br>Spindle<br>#5@10rpm] | 3500 cps (3.5 Pa-s) | 2900 cps (2.9 Pa-s) | Similar. The<br>values are<br>greater viscosity<br>of natural saliva<br>(0.0078 Pa-s). |
| Specific gravity<br>(SPG) | 1.02 | 1.05 | Similar. The<br>values are not<br>significantly<br>different when<br>compared to the<br>SPG of the<br>saliva (1.000-<br>1.010). |
#### SUMMARY OF NON-CLINICAL PERFORMANCE TESTING 7.
{7}------------------------------------------------
| Sterility | Non-sterile | Non-sterile | Same |
|------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|------|
| Shelf Life | 2 years | 2 years | Same |
| Biocompatibility | Conforms with ISO<br>10993-1<br>• cytotoxicity<br>• sensitization<br>• irritation | Conforms with ISO<br>10993-1<br>• cytotoxicity<br>• sensitization<br>• irritation | Same |
Dentilube dry mouth spray has been tested for shelf-life/stability, which provides for a 24-month shelf life, same as the primary predicate device. Biocompatibility assessments has been done in accordance with ISO 10993 for cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), and irritation (ISO 10993-10) and has been shown to be safe for the intended use. No other clinical tests were performed other than a Use Study for this submission.
#### DISCUSSION AND CONCLUSION 8.
Based on the comparison of intended use and technical characteristics, as well as non-clinical performance testing, we conclude that Dentilube dry mouth spray is substantially equivalent to the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.