K202664 · Sota Precision Optics, Inc. Dba Sota Imaging · MUH · Nov 6, 2020 · Dental
Device Facts
Record ID
K202664
Device Name
Clio Prime and Clio Pedo
Applicant
Sota Precision Optics, Inc. Dba Sota Imaging
Product Code
MUH · Dental
Decision Date
Nov 6, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Clio Prime and Clio Pedo are intended for any dental practice that uses x-ray equipment for intraoral diagnostic purposes. It can be used by trained dental professionals for patients receiving intraoral x-ray examinations and produces digital images for patients receiving intraoral x-ray examinations for diagnostic purposes. An image analysis software is not part of this submission. When Clio Prime or Clio Pedo is to be used in a dental practice, an optional software will be necessary.
Device Story
Clio Prime and Clio Pedo are USB-driven digital intraoral x-ray sensors; CMOS-based imager with Cesium Iodide scintillator; captures x-ray doses to generate digital images. Used in dental clinics by trained professionals; positioned in patient mouth similar to film. Device connects to PC via USB; requires separate, optional image analysis software for image optimization, archiving, and printing. Provides immediate digital visualization; aids diagnosis of anatomical structures and pathologies. Benefits include reduced radiation exposure compared to analog film and enhanced diagnostic sensitivity via software tools. Not for extraoral use.
Clinical Evidence
Bench testing performed by Intertek. Clinical images provided to demonstrate system functionality; no formal clinical study with statistical endpoints reported. Substantial equivalence supported by comparative MTF analysis against predicate sensors.
Technological Characteristics
CMOS sensor with Cesium Iodide scintillator; USB 2.0 interface; 19.5 µm pixel size; IPx8 equivalent housing; biocompatible per ISO 10993-1; power 0.8W max. Standards: ISO 14971, ANSI AAMI ES 60601-1, IEC 62304, IEC 60529, IEC 60601-1-2, EN 60601-1-2.
Indications for Use
Indicated for dental intraoral radiography image acquisition. Intended for patients of any age requiring intraoral x-ray examinations. Operated by trained dental professionals. May be used with positioning devices or hand-positioned with patient assistance.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
November 6, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym and name on the right. The Department of Health & Human Services seal is a stylized depiction of an eagle. The FDA acronym is in a blue square, and the words "U.S. FOOD & DRUG ADMINISTRATION" are in blue text.
Sota Precision Optics, Inc. dba SOTA Imaging % Mr. Ed Johansen Official Correspondent W. Edward Johansen 1239 Stanford Street. #205 SANTA MONICA CA 90404
#### Re: K202664
Trade/Device Name: Clio Prime and Clio Pedo Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: MUH Dated: September 12, 2020 Received: September 14, 2020
#### Dear Mr. Johansen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K202664
Device Name
Clio Prime and Clio Pedo
#### Indications for Use (Describe)
Clio Prime and Clio Pedo are USB-driven, digital intraoral x-ray sensors which are intended to acquire dental intraoral radiography images. Clio Prime or Clio Pedo shall be operated by healthcare professionals who are educated and competent to perform the acquisition of dental intraoral radiographs. Clio Prime or Clio Pedo can be used either in combination with special positioning devices to facilitate positioning and alignment with an x-ray beam, or Clio Prime or Clio Pedo can also be positioned by hand with the assistance of the patient.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
510(k) Premarket Notification No. K202664
October 29, 2020
#### I. Corporation Information
Name: Sota Precision Optics, Inc. dba SOTA Imaging Address: 1073 North Batavia Street, Suite B Orange, California 92867 Telephone: (714) 532-6100 Official Correspondent: W. Edward Johansen Telephone: (310) 795-7425
E-mail address: wedjohansen@msn.com
- Main address: <info@namecheap.com>
Address: 1600 Stewart Street, #905, Santa Monica, California
Address: 1239 Stanford Street, #205, Santa Monica, California 90404
#### II. Identification of New Device
Owner/Operator: Sota Precision Optics, Inc. dba SOTA Imaging Establishment Registration Number: 3000190675 Trade Name: Clio Prime and Clio Pedo 510(k) Premarket Notification Number: K202664 Common Name: Dental digital x-ray sensor Classification Name: Extra-oral source x-ray system Product Code: MUH Class: II Panel: Radiology Regulation Number: 21 C.F.R. §872.1800 Manufacturer: SOTA Imaging Contract Manufacturer: BAE Systems Imaging Solutions, a division of BAE Systems Inc. Sterilization Facility: Not Applicable
{4}------------------------------------------------
### III. Identification of Predicate Device
Owner/Operator/Manufacturer: Dental Imaging Technologies Corporation
Establishment Registration Number: 2530069
510(k) Premarket Notification Number: K090458
Contract Manufacturer: BAE Systems Imaging Solutions, a division of BAE Systems Inc.
Trade Name: DEXIS Platinum sensor and GENDEX GXS-700 sensors
Common Name: Dental intraoral x-rav sensor
Classification Name: Extra-oral source x-ray system
Product Code: MUH
Classification: Class II
Panel: Radiology
Requlation Number: 21 C.F.R. §872.1800
### IV. Device Description
Clio Prime and Clio Pedo are USB-driven digital sensors designed for health care professionals already acquainted with the standard procedures for acquiring dental intra-oral radiographs. Digital x-ray imaging is an aide for diagnosis and should always be confirmed by the doctor using additional Clio Prime or Clio Pedo procedures and other diagnostic aides for confirmation. Clio Pedo is a smaller version of Clio Prime and is for pediatric use. The Clio Prime and Clio Pedo design uses advanced ergonomic principles with four beveled corners, a moderate profile, and a rounded casing providing enhanced comfort for patients. Clio Prime or Clio Pedo is positioned in the patient's mouth in the same manner as intra-oral film is positioned. Each of Clio Prime and Clio Pedo has a CMOS x-ray imager that creates a digital imaqe from x-ray doses perceptible by the sensor. The digital image created is immediately visible on the screen of a personal computer connected to Clio Prime or Clio Pedo through the standard USB port. Image analysis software is not part of the submission. For Clio Prime or Clio Pedo to be used in a dental practice, an optional image analysis software will be necessary. Only with image analysis software can acquired images be optimized for specific diagnostic tasks, archived as image files, and printed out on a suitable printer. Software provides drivers and utilities for x-ray dose optimization, sensor activation and settings.
{5}------------------------------------------------
Clio Prime and Clio Pedo capture x-ray imaqes suitable for recognition of normal anatomical structures, dental pathologies, and abnormal conditions. Inadequate images may result in misdiagnosis thereby subjecting the patient to incorrect or unnecessary dental procedures that would present an unacceptable risk to the patient.
# V. Indications for Use
Clio Prime and Clio Pedo are USB-driven, digital intraoral x-ray sensors which is intended to acquire dental intraoral radiography imaqes. Clio Prime or Clio Pedo shall be operated by healthcare professionals, who are educated and competent to perform the acquisition of dental intraoral radiographs. Clio Prime or Clio Pedo can be used either in combination with special positioning devices to facilitate positioning and alignment with an x-ray beam, or Clio Prime or Clio Pedo can also be positioned by hand with the assistance of the patient.
### VI. Intended Use
Clio Prime and Clio Pedo are intended for any dental practice that uses x-ray equipment for intraoral diagnostic purposes. It can be used by trained dental professionals for patients receiving intraoral x-ray examinations and produces digital images for patients receiving intraoral x-ray examinations for diagnostic purposes. An image analysis software is not part of this submission. When Clio Prime or Clio Pedo is to be used in a dental practice, an optional software will be necessary.
# VII. Performance Testing-Animal Study
SOTA Imaging did not perform an animal study performance testing.
# VIII. Performance Testing-Clinical Study
SOTA Imaging a clinical study performance testing to determine that Clio Prime, and Clio Pedo performed as well or better than did the DEXIS Platinum sensor and GENDEX GXS-700 sensors. SOTA Imaging has provided clinical images. These clinical images provide further evidence in addition to the laboratory performance data that shows that the complete system works as intended.
# IX. Performance Testing-Bench Study
On behalf of SOTA Imaging Intertek performed an extensive bench testing on Clio Prime and Clio Pedo.
{6}------------------------------------------------
## X. Comparison Table
The following comparison table compares Clio Prime and Clio Pedo to the predicates, the DEXIS Platinum sensor and the GENDEX GXS-700 sensors with respect to intended use, indications of use, environment of use, limitations of use, technical performance and technological characteristics, and provides more detailed information regarding the basis for the determination of substantial equivalence.
{7}------------------------------------------------
# Comparison Table
| Descriptive<br>Information | DEXIS Platinum<br>sensor and GENDEX<br>GXS-700 sensors | Clio Prime and Clio<br>Pedo |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K090458 | K202664 |
| Manufacturer | Dental Imaging<br>Technologies<br>Corporation | Sota Precision Optics,<br>Inc. doing business as<br>SOTA Imaging |
| Trade Name | DEXIS Platinum<br>sensor and GENDEX<br>GXS-700 sensors | Clio Prime and Clio<br>Pedo |
| Classification Name | System, x-ray,<br>extraoral source | System, x-ray,<br>extraoral source |
| Classification Panel | Radiology | Radiology |
| Classification | Class II | Class II |
| Regulation Number | 21 C.F.R. § 872.1800 | 21 C.F.R. § 872.1800 |
| Indication for Use | The DEXIS Platinum<br>sensor or the GENDEX<br>GXS-700 sensors are<br>USB-driven, digital<br>intraoral x-ray sensors<br>which are intended to<br>acquire dental intraoral<br>radiography images.<br>The DEXIS Platinum<br>sensor or the GENDEX<br>GXS-700 sensors shall<br>be operated by<br>healthcare<br>professionals who are<br>educated and<br>competent to perform<br>the acquisition of dental<br>intraoral radiographs.<br>The DEXIS Platinum<br>sensor or the GENDEX<br>GXS-700 sensors can<br>be used either in<br>combination with | Clio Prime and Clio<br>Pedo are USB-driven,<br>digital intraoral x-ray<br>sensors which are<br>intended to acquire<br>dental intraoral<br>radiography images.<br>Clio Prime or Clio<br>Pedo shall be operated<br>by healthcare<br>professionals who are<br>educated and<br>competent to perform<br>the acquisition of<br>dental intraoral<br>radiographs. Clio<br>Prime or Clio Pedo can<br>be used either in<br>combination with<br>special positioning |
| | GXS-700 sensors can<br>be used either in<br>combination with | devices to facilitate<br>positioning and<br>alignment with an x-ray<br>beam, or Clio Prime or |
| | | |
| | special positioning<br>devices to facilitate<br>positioning and<br>alignment with an x-ray<br>beam or the intraoral<br>radiography images.<br>The DEXIS Platinum<br>sensor or the GENDEX<br>GXS-700 sensors shall<br>be operated by<br>healthcare<br>professionals who are<br>educated and<br>competent to perform<br>the acquisition of<br>dental intraoral<br>radiographs. The<br>DEXIS Platinum<br>sensor or the GENDEX<br>GXS-700 sensors can<br>be used either in<br>combination with<br>special positioning<br>devices to facilitate<br>positioning and<br>alignment with an x-ray<br>beam, or the DEXIS<br>Platinum sensor or the<br>GENDEX GXS-700<br>sensors can also be<br>positioned by hand<br>with the assistance of<br>the patient. | Clio Pedo can also be<br>positioned by hand<br>with the assistance of<br>the patient. |
| Intended Use | The DEXIS Platinum<br>sensor and the<br>GENDEX GXS-700<br>sensors are indirect<br>converting x-ray<br>detectors, e.g.<br>incident x-rays are | Clio Prime and Clio<br>Pedo are USB-driven<br>digital sensors which<br>are intended to<br>acquire dental<br>intraoral radiography |
| converted by a | images. Clio Prime or | |
| scintillating material | Clio Pedo shall be | |
| into (visible) light, this | operated by healthcare | |
| light is coupled | professionals who are | |
| optically to a light | educated and | |
| detection imager | competent to perform | |
| based on CMOS | the acquisition of | |
| technology. The | dental intraoral | |
| design of the sensor | radiographs. Clio | |
| assembly supports the | Prime or Clio Pedo can | |
| automatic detection of | be used either in | |
| the incident x-rays to | combination with | |
| generate digital images | positioners | |
| for dental intra-oral | manufactured to | |
| applications. The | facilitate the | |
| DEXIS Platinum | positioning and | |
| sensor or the | alignment with an x- | |
| GENDEX GXS-700 | ray beam, or it may | |
| sensors supports USB | also be positioned by | |
| 2.0 and USB 1.1 | hand with the | |
| connectivity to | assistance of the | |
| personal computers | patient. Clio Prime or | |
| using a dedicated | Clio Pedo can be used | |
| electronic assembly | with patients of any | |
| and a sensor software | age, providing the | |
| driver. The DEXIS | correct positioning of | |
| Platinum sensor and | the sensor in the | |
| the GENDEX GXS-700 | patient mouth can be | |
| sensors are USB- | realized. Using Clio | |
| driven digital sensors | Prime or Clio Pedo is | |
| which are intended to | a suitable diagnostic | |
| acquire dental intra- | method and may offer | |
| oral radiography | reduced radiation | |
| images. The DEXIS | exposure compared to | |
| Platinum sensor and | analog procedures. | |
| GENDEX GXS-700 | Furthermore, available |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.