The Novapak Nasal Sinus Packing and Stent is intended for use in patients undergoing nasal/sinus surgery as a space occupying packing to: - Separate tissue or structures compromised by surgical trauma. - Separate and prevent adhesions between mucosal surfaces in the nasal cavity. - Control minimal bleeding following surgery or trauma by tamponade effect, blood absorption, and platelet aggregation. - Act as an adjunct to aid in the natural healing process. Novapak is indicated for use as a nasal packing to treat epistaxis.
Device Story
Novapak is a single-use, absorbable nasal packing and stent; composed of chitosan and cellulose sponge. Used post-sinus surgery or for epistaxis; inserted into nasal cavity to provide tamponade, absorb blood, and prevent adhesions. Hydrates with sterile saline to form a gel; dissolves over 7-14 days with irrigation or removed via aspiration. Physician-operated in clinical settings. Benefits include reduced adhesion formation, bleeding control, and antibacterial properties. Acts as a physical barrier to bacteria.
Clinical Evidence
Bench testing only. Evaluated absorption/hydration (saline/blood), degradation, compression, hemolysis, platelet activation, bacterial log reduction, and bacterial barrier performance. All samples met acceptance criteria. Biocompatibility testing performed per ISO 10993. In vitro antibacterial efficacy demonstrated against various strains; bacterial barrier function confirmed for 72 hours.
Technological Characteristics
Materials: Chitosan and cellulose sponge. Form factor: Compressible sponge, cut-to-size. Energy: None. Sterilization: E-beam radiation (20-30kGy, SAL 10^-6). Connectivity: None. Software: None.
Indications for Use
Indicated for patients undergoing nasal/sinus surgery or requiring treatment for epistaxis to separate tissues, prevent adhesions, control minimal bleeding, and aid healing.
Regulatory Classification
Identification
An intranasal splint is intended to minimize bleeding and edema and to prevent adhesions between the septum and the nasal cavity. It is placed in the nasal cavity after surgery or trauma. The intranasal splint is constructed from plastic, silicone, or absorbent material.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 8, 2020
Medtronic Xomed Matthew Harmon Principal Regulatory Specialist 6743 Southpoint Drive North Jacksonville, Florida 32256
Re: K202623
Trade/Device Name: Novapak Nasal Sinus Packing and Stent Regulation Number: 21 CFR 874.4780 Regulation Name: Intranasal splint Regulatory Class: Class I Product Code: LYA Dated: September 9, 2020 Received: September 10, 2020
Dear Matthew Harmon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Malvina B. Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K202623
Device Name
Novapak™ Nasal Sinus Packing and Stent
Indications for Use (Describe)
The Novapak Nasal Sinus Packing and Stent is intended for use in patients undergoing nasal/sinus surgery as a space occupying packing to:
- · Separate tissue or structures compromised by surgical trauma.
- · Separate and prevent adhesions between mucosal surfaces in the nasal cavity.
- · Control minimal bleeding following surgery or trauma by tamponade effect, blood absorption, and platelet aggregation.
- · Act as an adjunct to aid in the natural healing process.
Novapak is indicated for use as a nasal packing to treat epistaxis.
Type of Use (Select one or both, as applicable)
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------|
|------------------------------------------------------------|
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------|
|-----------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary – K202623
- Company: Medtronic Xomed, Inc. 6743 Southpoint Drive North Jacksonville, Florida 32216 USA Telephone Number: 904-332-8186 Fax Number: 904-296-2386
- Contact: Matthew Harmon Principal Regulatory Affairs Specialist Phone: (904) 332-6704 E-mail: matthew.harmon@medtronic.com
Device Trade name: Novapak™ Nasal Sinus Packing and Stent
Common Name: Intranasal Packing and Stent, Intranasal Splint
Classification Name: Intranasal Splint (21 CFR 874.4780)
Classification: Class I Non-exempt (21 CFR 874.4780)
Product Code: LYA
| Predicate Device: | NasoPore® | K052099 |
|--------------------|--------------------------------------|-------------------------------|
| Reference Devices: | Posisep®<br>Novashield™<br>Chitogel™ | K122494<br>K141704<br>K172179 |
#### Device Description:
Novapak™ is a single use, nasal packing and stent for use following sinus surgery to prevent adhesions, control mild bleeding and provide a level of antibacterial effectiveness. Novapak™ is composed of formulated chitosan and cellulose ingredients in a sponge can be compressed for insertion into anatomy and can be cut to size.
Novapak™ hydrates with sterile saline and forms a gel. The sponge dissolves within the nasal cavity with daily irrigation and natural mucus flow over several days. Residence time is typically 7-14 days with adequate irrigation. Alternately, the dressing may be removed through gentle aspiration at the discretion of the physician.
{5}------------------------------------------------
## Indications for Use:
Novapak™ Nasal Sinus Packing and Stent is indicated for use in patients undergoing nasal/sinus surgery as a space occupying packing to:
- Separate tissue or structures compromised by surgical trauma.
- . Separate and prevent adhesions between mucosal surfaces in the nasal cavity.
- . Control minimal bleeding following surgery or trauma by tamponade effect, blood absorption, and platelet aggregation.
- . Act as an adjunct to aid in the natural healing process.
Novapak™ Nasal Sinus Packing and Stent is indicated for use as a nasal packing to treat epistaxis.
## Substantial Equivalence:
Novapak™ Nasal Sinus Packing and Stent is substantially equivalent in intended use and performance characteristics to NasoPore®, 510 (k) Number K052099, clearance date 11/21/2005.
The differences between the indications for the subject device Novapak™ and Nasopore® are not critical to the intended use of the device, and do not affect the safety and effectiveness of the subject device when used as labeled. Both the subject and predicate devices are indicated to separate and prevent adhesions between mucosal surfaces in the nasal cavity, and to control minimal bleeding following surgery or nasal trauma by the tamponade effect and blood absorption.
The subject device Novapak™ has some indications in addition to those for Nasopore®. These indications include separating structures compromised by surgical trauma, controlling minimal bleeding by platelet aggregation, and use as a nasal packing to treat epistaxis. These indications are matched by those in two reference devices Posisep™, Novashield™, and Chitogel.
Novapak™ and Nasopore® have different technological characteristics. They are made from different materials. The subject device Novapak™ is made of a combination of cellulose and chitosan, while the predicate Nasopore® is made of fragmentable poly (DL-Lactide-co-e-caprolactone) urethane. Performance testing (described below) shows that the subject device Novapak™ performs equivalently to the predicate device Nasopore®.
## Biocompatibility:
Biocompatibility testing was performed using ISO 10993 Biological Evaluation of Medical Devices and FDA guidance document "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process,"" issued June 16, 2016. The Novapak™ Nasal Sinus Packing and Stent complies with the biocompatibility requirements for its intended use.
## Sterilization:
{6}------------------------------------------------
The Novapak™ Nasal Sinus Packing and Stent is sterilized using an e-beam radiation dose of 20-30kGy validated to deliver a minimum sterility assurance level (SAL) of 10°.
## Performance Testing:
Performance testing was conducted on the sterile final product. Product testing followed the appropriate standards and guidance documents, with appropriate modifications following risk assessment. Testing plans were based upon FDA guidance documents and international standards. Testing was performed on baseline (non-aged) and aged products. Testing included:
- Absorption and hydration testing (both saline and blood)
- Degradation and compression testing.
- Hemolysis and platelet activation testing. ●
- Bacterial Log Reduction Testing
- . Bacterial barrier testing
All samples passed testing and met the acceptance criteria of the specification.
## Shelf Life and Packaging Integrity Testing:
Shelf life testing was performed on the sterile final product and its packaging according to the applicable standards and guidance documents. The packaging was subjected to both accelerated and real-time aging. The sterile final product was subjected to real-time aging. The packaging was also subjected to packaging integrity testing, including testing after transportation simulation, high altitude testing, and environmental conditioning.
All samples passed testing and met the acceptance criteria of the specification.
## Antibacterial Information:
The antibacterial effectiveness of Novapak™ was tested against the following bacterial strains. The table below outlines the timeframe in which Novapak™ demonstrated antibacterial activity to the tested bacterial strain. Note: in vitro efficacy is not correlated to clinical effectiveness.
| Bacterial Strain | ATCC | 24 hours | 48 hours | 3 days | 7 days |
|----------------------------|-------|----------|----------|--------|--------|
| Pseudomonas aeruginosa | 9027 | | | | |
| Staphylococcus aureus | 25923 | | | | |
| Staphylococcus epidermidis | 12228 | | | | |
| Escherichia coli | 8739 | | | | |
| Citrobacter freundii | 8090 | | | | |
| Enterobacter aerogenes | 13048 | | | | |
{7}------------------------------------------------
| Klebsiella pneumonia | 4352 | | |
|------------------------------------|-------|--|--|
| Proteus mirabilis | 4630 | | |
| Serratia marcescens | 13880 | | |
| Haemophilus influenzae | 53782 | | |
| Moraxella catarrhalis | 8193 | | |
| Staphylococcus aureus (MRSA) | 33591 | | |
| Staphylococcus saprophyticus | 15305 | | |
| Micrococcus luteus | 49732 | | |
| Streptococcus mutans | 25175 | | |
| Streptococcus pneumoniae | 10015 | | |
| Corynebacterium diphtheriae | 296 | | |
| Corynebacterium tuberculostearicum | 35693 | | |
The Novapak Nasal Sinus and Stent was also tested as a bacterial barrier. The testing shows that the Novapak Nasal Sinus and Stent acts as a protective barrier by preventing bacterial growth through the hydrated Novapak sponge for a 72 hour period.
## Conclusion:
Utilizing FDA's Guidance for Industry and FDA Staff "Format for Traditional and Abbreviated 510(k)s" issued on September 13, 2019, a comparison of key characteristics demonstrates that the proposed device Novapak™ Nasal Sinus Packing and Stent is substantially equivalent to the predicate device in terms of performance characteristics. The Novapak™ Nasal Sinus Packing and Stent is as safe, as effective, and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.