K202564 · Philips Medical Systems Dmc GmbH · MQB · Sep 30, 2020 · Radiology
Device Facts
Record ID
K202564
Device Name
DigitalDiagnost C90
Applicant
Philips Medical Systems Dmc GmbH
Product Code
MQB · Radiology
Decision Date
Sep 30, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The DigitalDiagnost C90 is intended to acquire, process, store, display and export digital radiographic images. The DigitalDiagnost C90 is suitable for all routine radiography examinations, including specialist areas like intensive care, trauma or pediatric work, excluding fluoroscopy, angiography and mammography.
Device Story
DigitalDiagnost C90 is a stationary digital radiography system used in clinical settings by healthcare professionals. It consists of a height-adjustable patient table, ceiling-suspended X-ray tube, and vertical stand. The system acquires X-ray images using a fixed flat-panel detector (Pixium 4343RCE) or wireless portable detectors (SkyPlate family, including the new SkyPlate E). The X-ray generator and tube form the image chain. The Eleva Workstation processes, stores, displays, and exports images. The device supports automatic image stitching and detector sharing. Clinicians use the output images for diagnostic assessment. The system benefits patients by providing high-quality digital radiographs for routine and specialized examinations, including trauma and pediatric care.
Clinical Evidence
No clinical data. Substantial equivalence was demonstrated through non-clinical performance testing, including system verification, image quality testing, and human factors/usability engineering, alongside compliance with consensus standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-54, IEC 62220-1, IEC 62304, IEC 60601-1-6, ISO 14971).
Technological Characteristics
Stationary X-ray system; CsI scintillator; TFT readout; 65kW/80kW generator; motorized ceiling suspension and table. Connectivity via Ethernet/WLAN. Software lifecycle per IEC 62304. SkyPlate E detector: 384.5mm x 460.5mm x 16.0mm, 160 μm pixel size, 3.125 lp/mm Nyquist frequency.
Indications for Use
Indicated for routine radiography examinations, including intensive care, trauma, and pediatric patients. Excludes fluoroscopy, angiography, and mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
September 30, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Philips Medical Systems DMC GmbH % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive. Suite 510k SAINT PAUL MN 55114
## Re: K202564
Trade/Device Name: DigitalDiagnost C90 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB, KPR, LLZ Dated: September 2, 2020 Received: September 4, 2020
# Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "PHILIPS" in a sans-serif font. The letters are a light blue color. The word is written in all capital letters and is centered in the image. The background is white.
### 5. Statement of Indication for Use/Intended Use
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
**Indications for Use**
| 510(k) Number (if known) | K202564 |
|--------------------------------|---------------------|
| Device Name | DigitalDiagnost C90 |
| Indications for Use (Describe) | |
The DigitalDiagnost C90 is intended to acquire, process, store, display and export digital radiographic images. The
DigitalDiagnost C90 is suitable for all routine radiography examinations, including specialist areas like intensive care,
trauma or pediatric work, excluding fluoroscopy, angiography and mammography.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are sans-serif and evenly spaced. The background is plain white, which makes the blue letters stand out.
### K202564
#### 6. 510(k) Summary
### SPECIAL 510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Date Prepared: | September 23, 2020 |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Philips Medical Systems DMC GmbH<br>Röntgenstraße 24<br>22335 Hamburg, GERMANY<br>Establishment registration number: 3003768251 |
| Contact Person: | Ming Xiao<br>Regulatory Affairs Manager<br>Phone: +49 40 34971-2306<br>Fax: +49 40 5078-2147<br>E-mail: ming.xiao@philips.com |
| Device Name: | DigitalDiagnost C90 |
| Classification: | |
| | Trade Name: DigitalDiagnost C90 |
| | Classification Name: Solid State X-Ray Imager (Flat Panel/Digital Imager) |
| | Classification Regulation: 21CFR 892.1680 |
| | Regulation Description: Stationary x-ray system |
| | Classification Panel: 90 -- Radiology |
| | Device Class: Class II |
| | Classification Product Code: MQB, KPR, LLZ |
| Predicate Device: | |
| | Trade Name: DigitalDiagnost C90 |
| | Manufacturer: Philips Medical Systems DMC GmbH |
| | 510(k) Clearance: K182973- January 11, 2019 |
| | Classification Regulation: 21CFR 892.1680 |
| | Classification Name: Solid State X-Ray Imager (Flat Panel/Digital Imager) |
| | Classification Panel: 90 - Radiology |
| | Device Class: Class II |
| | Product Code: MQB, KPR, LLZ |
| Reference Device for SkyPlate E Detector: | |
| | Trade Name : MobileDiagnost wDR 2.2 |
| | Manufacturer: Sedecal S.A |
| | 510(K) Clearance: K191813- August 2, 2019 |
| Classification Regulation: | 21CFR 892.1720 |
| Classification Name: | Mobile X-ray Systems |
| Classification Panel: | 90- Radiology |
| Device Class: | Class II |
| Product Code: | IZL, MQB |
DigitalDiagnost C90 Premarket Notification – Special 510(k)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and appear to be a sans-serif font. The background is plain white, which makes the blue letters stand out.
| Device<br>Description | The proposed DigitalDiagnost C90 is same as the 510(K) approved predicate<br>(K182973) as mentioned below | Predicate Device: | Proposed Device: | Discussion & Conclusion | |
|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | It is a high-end digital radiography system consisting of a height adjustable patient<br>support table and a ceiling suspension consisting of a tube including a control handle<br>used to acquire images with a flat panel fixed RAD detector. Additionally, different<br>vertical stands for the radiography examinations are available. The ceiling suspension<br>can be moved in longitudinal and lateral directions and additionally the tube can be<br>tilted and rotated as well. The proposed DigitalDiagnost C90 is configured with a<br>Philips x-ray generator and a flat panel fixed RAD detector, Pixium 4343RCE,<br>together with the tube these components form the radiography Image Chain.<br>The proposed DigitalDiagnost C90 introduces a new wireless portable detector,<br>pixium 3543DR i.e. SkyPlate E (cleared in reference device 510(K) K191813) and<br>their relevant grids, in addition to the family of SkyPlate detectors (pixium 3543EZ<br>i.e. Large and pixium 2430EZ i.e. Small) that are already integrated in the currently<br>marketed (predicate) DigitalDiagnost C90 (K182973). | DigitalDiagnost<br>C90 (K182973-<br>January 11,<br>2019) | DigitalDiagnost<br>C90 (K202564) | | |
| Indications for<br>Use/Intended<br>Use: | The Indication for Use of the proposed DigitalDiagnost C90 is identical to that of the<br>currently marketed and predicate device, DigtalDiagnost C90 (K182973- January 11,<br>2019) and is as follows: | Table Features | | | |
| | The DigitalDiagnost C90 is intended to acquire, process, store, display and export<br>digital radiographic images. The DigitalDiagnost C90 is suitable for all routine<br>radiography examinations, including specialist areas like intensive care, trauma or<br>pediatric work, excluding fluoroscopy, angiography and mammography. | Height adjustable table (TH-2) | | | |
| Fundamental<br>Scientific<br>Technology: | The proposed DigitalDiagnost C90 employs the same basic construction and<br>fundamental scientific technology as the currently marketed predicate DigtalDiagnost<br>C90 (K182973, January 11, 2019), with regards to the functionality of all its<br>components | | | | |
| | The proposed DigitalDiagnost C90 employs the same cleared features as predicate<br>DigitalDiagnost C90 as follows | | | | |
| | fixed RAD detector (Pixium 4343RCE, K170113, February 9, 2017), wireless detectors (SkyPlate Detector, cleared by Eleva Workspot for<br>DigitalDiagnost, K141736, July 25, 2014), image chain acquisition-station and workflow (Eleva Workspot with<br>SkyFlow, K153318, December 22, 2015). | | | | |
| Height<br>adjustment | 51.5 cm to 91.5<br>cm above floor,<br>motorized<br>adjustment | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Table<br>weight | 335 kg | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Max. patient<br>weight | • Static load center:<br>375 kg<br>• Dynamic load<br>center: 318 kg<br>• Dynamic load off<br>center: 210 kg | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Table top<br>Type | Floating table top<br>of sandwich<br>design with<br>Getalit overlay | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Table top<br>Dimension | 240 cm x 75 cm | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Table top<br>travel | • longitudinal: ±56<br>cm<br>• transverse: ±12.8<br>cm,<br>electromagnetic<br>brakes | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Single side suspended table (TH-S) | | | | | |
| Height<br>adjustment | 51 cm to 91 cm<br>motorized<br>adjustment | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Table<br>weight | 214 kg | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Max. patient<br>weight | • table top center:<br>225kg<br>• end of table top:<br>135kg | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Table top<br>Type | Floating table top<br>of sandwich<br>design with<br>Kevlar overlay,<br>flat top | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Table top<br>Dimension | 260 cm x 75 cm | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Thickness of<br>table top | 4.7 cm | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Table top<br>travel | • longitudinal: ±20<br>cm, hydraulic<br>brakes<br>• transverse: ±20<br>cm, hydraulic<br>brakes | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Patient<br>coverage<br>with fixed<br>RAD<br>detector | • longitudinal: 208<br>cm<br>• Transversal: 83<br>cm | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Celling Suspension CSM | | | | | |
| Type | Four-part<br>aluminum<br>telescopic column<br>with spring<br>counter balanced<br>holder for X-ray<br>tube assembly;<br>adaptable to<br>individual room<br>heights | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Ceiling<br>height at<br>source<br>image<br>distance 110<br>cm | 2.83 to 3.21 m | Same | Equivalent; No impact to the<br>safety and effectiveness of the<br>device. | | |
| Movement | 3.26 m to 3.41m | Same | Equivalent; No impact to the<br>safety and effectiveness of the<br>device. | | |
| Transverse<br>travel | 1.49 m to 3.21 m | Same | Equivalent; No impact to the<br>safety and effectiveness of the<br>device. | | |
| Vertical<br>travel | 1.65 m | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| X-ray tube<br>assembly<br>rotation | • around vertical<br>axis: 360° (±180°)<br>with lock position<br>every 45°<br>• around horizontal<br>axis: ±115°, lock | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| | positions 0° and | | | | |
| | ±90° | | | | |
| Length of<br>rails | 4.3 m | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Collimator | • Ralco P 225 ACS<br>DHHS<br>• Motorized<br>automatic<br>collimation<br>• Manual overrule<br>possible<br>• With light field<br>indicator<br>• With 2 Lasers and<br>Camera | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Vertical Stand | | | | | |
| | Vertical moveable stand (VM) | | | | |
| Hardware | Counterbalanced<br>rugged column<br>for motorized<br>vertical<br>movement of the<br>detector unit | Same | Equivalent; No impact to<br>safety and effectiveness of the<br>device. | | |
| Vertical<br>travel | 35 cm to 185 cm…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.