EliA Rib-P

K202540 · Phadia AB · MQA · Sep 13, 2021 · Immunology

Device Facts

Record IDK202540
Device NameEliA Rib-P
ApplicantPhadia AB
Product CodeMQA · Immunology
Decision DateSep 13, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5100
Device ClassClass 2

Intended Use

EliA Rib-P is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Rib-P in human serum as an aid in the diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA Rib-P uses the EliA IgG method.

Device Story

EliA Rib-P is an automated, solid-phase fluoroenzymeimmunoassay for detecting IgG autoantibodies against ribosomal P-proteins (P0, P1, P2). The device uses human recombinant antigens coated on polystyrene wells. Patient serum is incubated in the wells; if present, anti-Rib-P antibodies bind to the antigen. After washing, enzyme-labeled anti-human IgG (mouse monoclonal) is added, followed by a development solution (4-Methylumbelliferyl-β-D-galactoside). The instrument measures the resulting fluorescence, which is proportional to the antibody concentration. The device is used in clinical laboratories on Phadia 250, 2500, and 5000 series instruments. Results are compared to a 6-point calibration curve to provide semi-quantitative values in EliA U/mL. Clinicians use these results alongside other diagnostic findings to support an SLE diagnosis. The automated process reduces manual handling, improves throughput, and provides standardized, reproducible results compared to traditional manual ELISA methods.

Clinical Evidence

Clinical validation involved 560 serum samples, including 146 SLE patients and 414 disease controls (e.g., RA, Sjögren's, infections). Using a cut-off of >10 EliA U/mL, the device demonstrated 28.1% sensitivity (95% CI: 21.0%–36.1%) and 99.8% specificity (95% CI: 98.7%–100%). Method comparison with the predicate (n=323) showed 98.1% total agreement. Analytical performance included precision (total CV 4.0%–13.3% depending on concentration), linearity, and interference testing.

Technological Characteristics

Solid-phase fluoroimmunoassay; uses human recombinant ribosomal P-proteins (P0, P1, P2) immobilized on molded wells. Employs β-Galactosidase labeled anti-IgG conjugate and 4-Methylumbelliferyl-β-D-galactoside development solution. Automated system for Phadia 250/2500/5000 instruments. Traceable to WHO IRP 67/86. Shelf-life 36 months at 2–8°C.

Indications for Use

Indicated for the semi-quantitative measurement of IgG antibodies to Rib-P in human serum to aid in the diagnosis of systemic lupus erythematosus (SLE).

Regulatory Classification

Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 13, 2021 Phadia AB % Sheryl Skinner Associate Director RA/QA Phadia US Inc. 4169 Commercial Avenue Portage, Michigan 49002 Re: K202540 Trade/Device Name: EliA Rib-P Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear Antibody Immunological Test System Regulatory Class: Class II Product Code: MOA Dated: September 1, 2020 Received: September 2, 2020 Dear Sheryl Skinner: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Ying Mao, Ph.D. Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K202540 Device Name EliA Rib-P Indications for Use (Describe) EliA Rib-P is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to Rib-P in human serum as an aid in the diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA Rib-P uses the EliA IgG method. X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo of Thermo Fisher Scientific. The logo is composed of the words "Thermo Fisher" in a bold, red font on the top line, and the word "SCIENTIFIC" in a smaller, black font on the second line. The logo is simple and modern, and it is likely used to represent the company's brand. #### 510(k) Summary This 510(k) Summary is prepared in accordance with the requirements of 21 CFR Part 807.92. #### Premarket Notification 510(k) No: K202540 Date of Summary Preparation: September 3, 2021 | Manufacturer: | Phadia AB<br>Rapsgatan 7P<br>P.O. Box 6460<br>751 37 Uppsala, Sweden | |-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------| | Distributor: | Phadia US Inc.<br>4169 Commercial Avenue<br>Portage, MI 49002 | | Company Contact Person: | Sheryl Skinner<br>Associate Director, Regulatory and Quality<br>Phadia US Inc.<br>4169 Commercial Avenue<br>(269) 568-3603<br>sheryl.skinner@thermofisher.com | #### Proprietary and Established Device Name: EliA Rib-P ## Regulatory Information: | Product Code: | MQA | |-----------------|---------------------------------------------------------------------| | Classification: | Class II | | Regulation: | 21 CFR 866.5100 – Antinuclear Antibody Immunological<br>Test System | | Panel: | Immunology | # Purpose of Submission: New Device #### Measurand: IgG autoantibodies specific to Rib-P proteins #### Type of Test: Automated semi-quantitative solid phase fluoroenzymeimmunoassay {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for Thermo Fisher Scientific. The words "Thermo Fisher" are in red, with "SCIENTIFIC" in black below it. The logo is simple and modern. #### Intended Use: EliA Rib-P is intended for the in vitro semi-quantitative measurement of lgG antibodies directed to Rib-P in human serum as an aid in the diagnosis of systemic lupus erythematosus (SLE) in conjunction with other laboratory and clinical findings. EliA Rib-P uses the EliA IgG method. #### Indication(s) for Use: Same as intended use #### Special Conditions for Use: Rx - For Prescription Use Only #### Special Instrument Requirements: For use on the Phadia 250 instrument and the Phadia 2500 and Phadia 5000 instrument series (E-modules). #### Device Description: EliA Rib-P is a semi-quantitative solid-phase fluoroenzymeimmunoassay, for the determination of autoantibodies against Rib-P. The EliA Rib-P test System is fully integrated and automated system which comprises of assay-specific reagents, EliA method-specific reagents, and general reagents. Assav-Specific Reagents include: - EliA Rib-P Wells: coated with human recombinant ribosomal P-proteins P0, P1 . and P2 - 2 carriers (12 wells each), ready to use; - . EliA ANA 3 Positive Control 250 or 2500/5000: Human monoclonal antibodies in Tris buffer containing IgG antibodies to Ro52, Rib-P and RNA Pol III – 6 single use vials, 0.3 mL each, ready to use; - . EliA IgG/IgM/IgA Negative Control 250 or 2500/5000: Human blood preparation from healthy donors in PBS containing BSA, detergent and 0.095% sodium azide - 6 single-use vials, 0.3 mL each, ready to use; EliA Method-Specific Reagents include: - EliA Sample Diluent: PBS containing BSA, detergent and 0.095% sodium azide . - 6 bottles, 48 mL each, ready to use; or 6 bottles, 400 mL each, ready to use; - I EliA IqG Conjuqate 50 or 200: ß-Galactosidase labeled anti-lgG (mouse monoclonal antibodies) in PBS containing BSA and 0.06% sodium azide - 6 wedge shaped bottles, 5 mL each, ready to use; or 6 wedge shaped bottles, 19 mL each, ready to use: - . EliA IgG Calibrator Strips: Human IqG (0, 4, 10, 20, 100, 600 uq/L) in PBS containing BSA, detergent and 0.095% sodium azide - 5 strips, 6 single-use vials per strip, 0.3 mL each, ready to use; - EliA IgG Curve Control Strips: Human IgG (20 ug/L) in PBS containing BSA, . detergent and 0.095% sodium azide - 5 strips, 6 single-use vials per strip, 0.3 mL each, ready to use: {5}------------------------------------------------ - EliA IgG Calibrator Well: coated with mouse monoclonal antibodies 4 carriers . (12 wells each), ready to use; General Reagents include: - Development Solution: 0.01% 4-Methylumbelliferyl-β-D-galactoside, <0.0010% ■ preservative - 6 bottles (11 mL, 17 mL, or 112 mL each), sufficient for 6x >110, 6x >170, or 6x >1165 determinations; - I Stop Solution: 4% Sodium Carbonate - 6 bottles (65 mL, 119 mL, or 2800 mL each), sufficient for 6x >292, 6x >560, or 6x >13100 determinations; - 트 Washing Solution Additive: detergent, preservative <0.13% - 6x 17.2 mL, 2x 86mL, or 4x 850 mL; - I Washing Solution Concentrate: phosphate buffer - 6x 80 mL, 2x 400 mL, or 1x 2800 mL. #### Instrument System EliA Rib-P is run on the Phadia 250 instrument and the Phadia 2500 and 5000 instrument series. The instruments are automated platforms for EliA test procedures from sample and reagent handling to the processing of results. #### Substantial Equivalence Quanta Lite Ribosome P ELISA, INOVA Diagnostics, Inc., (K981237) | | Similarities | | |-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Feature | Proposed Device<br>EliA Rib-P | Predicate Device<br>Quanta Lite Ribosome P ELISA | | Intended Use | EliA Rib-P is intended for the in<br>vitro semi-quantitative<br>measurement of IgG antibodies<br>directed to Rib-P in human<br>serum as an aid in the diagnosis<br>of systemic lupus<br>erythematosus (SLE) in<br>conjunction with other laboratory<br>and clinical findings. EliA Rib-P<br>uses the EliA IgG method. | QUANTA Lite Ribosome P is an<br>enzyme-linked immunosorbent<br>assay (ELISA) for the semi-<br>quantitative detection of<br>Ribosome P antibodies in<br>human serum. The presence of<br>Ribosome P antibodies can be<br>used in conjunction with clinical<br>findings and other laboratory<br>tests to aid in the diagnosis of<br>Systemic Lupus Erythematosus<br>(SLE) and other related<br>connective tissue diseases. | | Internal Controls | Positive and negative Control<br>provided with EliA ANA 3<br>Positive Control 250 and<br>2500/5000 and EliA IgG/IgM/IgA<br>Negative Control 250 and<br>2500/5000, resp. | Low Positive, High Positive and<br>Negative Control<br>included in the kit | Comparison with Predicate Device: {6}------------------------------------------------ #### Thermoris SCIENTI Traditional 510(k) for EliA Rib-P A.6 510(k) Summary_V6 | Assay technique | ELISA | Same | |--------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|-------------------| | Type of test | Semi-quantitative | Same | | Sample Dilution<br>(taking a 1%<br>pipetting<br>imprecision into<br>consideration,<br>this sample<br>dilution is<br>regarded as a<br>similarity) | 1:100 | 1:101 | | Reported Unit | EliA U/mL (arbitrary) | Units (arbitrary) | | | Differences | | |--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------| | Feature | Proposed Device | Predicate Device | | | EliA Rib-P | Quanta Lite Ribosome P ELISA | | Antigen | Human recombinant P-proteins<br>P0, P1 and P2 | Synthetic Ribosome P antigen | | Instrumentation | EliA Rib-P uses the EliA IgG<br>method on the instruments<br>Phadia 250 and the E-modules<br>of the Phadia 2500 and Phadia<br>5000 series. | ELISA-Reader needed | | Reaction<br>temperature | 37°C controlled | Room temperature, 20-26°C | | Detection<br>antibody<br>(conjugate) | IgG conjugate: anti-human IgG<br>β-Galactosidase (mouse<br>monoclonal antibodies) | IgG conjugate: HRP IgG<br>Conjugate, (goat), anti-human<br>IgG | | Signal | Fluorescence | Optical density | | Calibration | 6-point total IgG Calibration<br>6 vials of human IgG at<br>concentrations of<br>0 - 4 - 10 - 20 - 100 - 600<br>µg/L | One-point calibration | | Calibration curve | Option to store curve for up to<br>28 days and run curve controls<br>in each assay for calibration | N/A | | Interpretation of<br>results | Negative < 7 EliA U/mL<br>Equivocal 7-10 EliA U/mL<br>Positive > 10 EliA U/mL | Negative < 20 Units<br>Weak positive 20-39 Units<br>Moderate positive 40-80 Units<br>Strong positive > 80 Units | | Substrate | Development Solution 0.01 %<br>4-Methylumbelliferyl-β-D-<br>galactoside. | TMB Chromogen | {7}------------------------------------------------ | <0.0010% preservative* | |----------------------------------| | *Preservative: mixture of 5- | | chloro-2-methyl-2H-isothiazol-3- | | one [EC no. 247-500-7] and 2- | | methyl-2H-isothiazol-3-one [EC | | no. 220-239-6] (3:1) | ## Standard/Guidance Document Referenced - CLSI EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents, ■ September 2009 - I CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures, September 2014 - 트 CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach. April 2003 - . CLSI EP07-ED3, Interference Testing in Clinical Chemistry, September 2018 - . CLSI EP37-ED1, Supplemental Tables for Interference Testing, September 2018 - . CLSI EP09c-ED3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples, June 2018 - . CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures, June 2012 - . CLSI EP28-A3c, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory, October 2010 # Test Principle The EliA tests are fluorescence immunoassays for the detection and measurement of human antibodies based on EliA solid-phase components, which contain specific antigens for the antibodies to be measured. The specific antigen for the antibodies to be detected is bound to the EliA solid phase component (EliA Well). The EliA wells are molded cups comparable to excised wells from a microtiter plate. They are made of polystyrene and are coated with the respective antigen. The wells are at the same time a holder of the coupled antigen for convenient automation and a reaction chamber with reaction/washing solution handling based on pipetting to add and aspiration to remove liquids. If present in the patient's specimen, antibodies to these proteins bind to their specific antigen. After washing away non-bound antibodies, enzyme-labeled antibodies against human IgG antibodies (EliA IgG Conjugate) are added to form an antibody-conjugate complex. After incubation, non-bound conjuqate is washed away, and the bound complex is incubated with a Development Solution. After stopping the reaction, the fluorescence in the reaction mixture is measured. The assay directly measures the amount of antibody of interest bound to the antigen coating the EliA well, therefore the higher the value of fluorescent signal detected by the instrument, the higher the amount of antibody bound and detected in the sample tested. To evaluate test results, the response for patient samples is compared directly to the response for calibrators. Phadia AB, Rapsgatan 7P, P.O. Box 6460, 751 37 Uppsala, Sweden Confidential Page 5 (14) {8}------------------------------------------------ #### Performance Characteristics - 1. Analytical performance: - a) Precision/Reproducibility: To determine the precision of the assay on the Phadia 250 instrument and the Phadia 2500 and Phadia 5000 instrument series, the variability was assessed on 5 samples. Three lots were used to determine the precision of the assay on Phadia 250 (totaling 252 replicate determinations per sample). One lot was used to determine the precision of the assay on Phadia 2500E, which is a representative of the Phadia 2500 and Phadia 5000 instrument series. #### EliA Rib-P on Phadia 250: To determine the precision of the assay on the Phadia 250 instrument, the variability was assessed in a study with 21 runs by examining the samples in 252 replicates across 3 lots and 3 Phadia 250 instruments over 7 days. The data was calculated against the calibration curve from Day 1. The statistical evaluation was performed by Analysis of Variance. The results are given in the table below. | Mean<br>EliA<br>U/mL | Within-<br>Run<br>SD | Within-<br>Run<br>%CV | Between-<br>Run<br>SD | Between-<br>Run<br>%CV | Between-<br>Instrument<br>SD | Between-<br>Instrument<br>%CV | Lot-to-Lot<br>SD | Lot-to-Lot<br>%CV | Total<br>Imprecision<br>SD | Total<br>Imprecision<br>%CV | |----------------------|----------------------|-----------------------|-----------------------|------------------------|------------------------------|-------------------------------|------------------|-------------------|----------------------------|-----------------------------| | 2.8 | 0.2 | 5.9 | 0.2 | 6.6 | 0.2 | 8.5 | 0.1 | 5.0 | 0.4 | 13.3 | | 7.8 | 0.3 | 3.2 | 0.3 | 4.1 | 0.3 | 4.2 | 0.0 | 0.0 | 0.5 | 6.7 | | 9.5 | 0.4 | 3.8 | 0.3 | 2.9 | 0.5 | 5.3 | 0.4 | 3.9 | 0.8 | 8.1 | | 56.8 | 1.8 | 3.2 | 2.9 | 5.1 | 1.0 | 1.8 | 0.6 | 1.0 | 3.6 | 6.4 | | 316.4 | 9.3 | 2.9 | 7.5 | 2.4 | 2.6 | 0.8 | 3.0 | 1.0 | 12.5 | 4.0 | #### Within-lab Imprecision To determine the within-lab precision of the assay, the variability was assessed in a study with 40 runs by examining the samples in 80 replicates on 1 instrument over 20 davs. The data was calculated against the calibration curve from Day 1. The statistical evaluation was performed by Analysis of Variance. The results are given in the table below. | Mean<br>(EliA U/mL) | Within-Run | | Between-<br>Run | | Between-Day | | Within-Lab<br>Imprecision | | |---------------------|------------|-----|-----------------|------|-------------|-----|---------------------------|------| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | 3.0 | 0.2 | 7.2 | 0.4 | 11.8 | 0.1 | 3.6 | 0.4 | 13.8 | | 7.6 | 0.3 | 3.9 | 0.4 | 5.9 | 0.0 | 0.0 | 0.5 | 7.0 | | 11.0 | 0.2 | 2.2 | 0.5 | 4.2 | 0.3 | 2.3 | 0.5 | 4.7 | | 87.6 | 2.0 | 2.2 | 2.3 | 2.7 | 1.6 | 1.8 | 3.0 | 3.5 | Confidential Page 6 (14) {9}------------------------------------------------ EliA Rib-P on the Phadia 2500 and Phadia 5000 instrument series: To determine the precision of the assay on the of the Phadia 2500 and Phadia 5000 instrument series (E-module), the variability was assessed in a study with 21 runs by examining the samples in 84 replicates on 3 Phadia 2500E instruments over 7 days. The data was calculated against the calibration curve from Day 1. The statistical evaluation was performed by Analysis of Variance. The results are given in the table below. | Mean | | Within-Run | | Between-Run | | Between-Instrument | | Total Imprecision | | |--------------|-----|------------|-----|-------------|------|--------------------|-----|-------------------|------| | EliA<br>U/mL | n | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | 2.1 | 83* | 0.2 | 8.1 | 0.4 | 16.5 | 0.0 | 1.0 | 0.4 | 18.4 | | 7.1 | 84 | 0.4 | 5.4 | 0.5 | 6.6 | 0.1 | 1.9 | 0.6 | 8.7 | | 8.8 | 84 | 0.5 | 5.5 | 0.8 | 8.7 | 0.4 | 4.3 | 1.0 | 11.2 | | 48.7 | 84 | 1.5 | 3.1 | 1.7 | 3.4 | 1.7 | 3.4 | 2.8 | 5.7 | | 362.1 | 84 | 31.6 | 8.7 | 20.9 | 5.8 | 0.0 | 0.0 | 37.9 | 10.5 | * One sample result is missing due to an instrument error. ## b) Linearity/Assay Reportable Range: 3 serum samples were diluted in EliA Sample Diluent and tested on Phadia 250 and Phadia 2500E. The ratios of observed/expected values were calculated. The results are summarized below. #### Phadia 250 | Dilution Range<br>(EliA U/mL) | Slope | Intercept | R² | |-------------------------------|-------|-----------|------| | 45.1 - 452.6 | 0.95 | 9.01 | 1.00 | | 2.7 - 122.9 | 1.00 | 1.28 | 1.00 | | 0.5 - 32.2 | 1.00 | 0.24 | 1.00 | ## Phadia 2500E | Dilution Range<br>(EliA U/mL) | Slope | Intercept | R2 | |-------------------------------|-------|-----------|------| | 8.5 - 427.6 | 1.03 | 2.96 | 0.99 | | 0.5 - 34.3 | 1.00 | 0.44 | 1.00 | | 0.7 - 22.6 | 1.02 | 0.30 | 0.99 | Linearity was shown for the entire measuring range. Hook Effect/Over the Range Results: Not applicable. Results above the upper limit of the measuring range are reported as ">403". No recommendations are made for dilution of samples outside measuring range in the Directions For Use. {10}------------------------------------------------ ## c) Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): Traceability: The IgG calibrators are traceable (via unbroken chain of calibrations) to the International Reference Preparation (IRP) 67/86 of Human Serum Immunoglobulins A, G and M from WHO. New batches of IgG calibrators are compared to a secondary standard (standardized with the IRP) or the IRP directly and adjusted accordingly to meet the correct concentration. The instrument measures specific IgG concentrations in ug/L. By using a conversion factor given by the lot-specific code of the EliA test well, the results are automatically converted to EliA U/mL. #### Stability: Data for open and closed real-time stability and on-board stability of EliA IgG reagents and general EliA reagents on Phadia 250 as well as on the Emodule of the Phadia 2500 and Phadia 5000 series were already cleared with several other EliA tests, e.q. under K141375 (EliA M2 on Phadia 250). For the Phadia 2500 and Phadia 5000 instrument series, they were already cleared under K061165/A003 (EliA CCP). #### Shelf-life: The stability of EliA Rib-P Wells was evaluated with a real-time study. The results support stability of the test under the recommended storage of 2 – 8°C for up to 36 months. ## On-board stability: The on-board stability EliA Rib-P carriers (containing the antigen coated wells) was tested over 8 weeks using 3 positive and 2 negative samples only on the Phadia 250 instrument. As the storage conditions in the Emodule of the Phadia 2500 and Phadia 5000 series are similar to the Phadia 250, the results can also be used for stability claims for these instruments. The on-board stability for the Phadia 250 was determined to be 28 days at 2-8°C. ## Open Stability: Stability after first opening of the foil bag containing the EliA Rib-P wells was tested with a real-time study. According to this study, a shelf-life of 9 months at 2-8°C after first opening can be assigned to EliA Rib-P wells. ## d) Detection Limit: Four blank and four low level samples were measured with two different reagent sets (two lots of antigen wells). The four blank samples were created from depleted IqG sera, each diluted with EliA Sample Diluent. The blank samples and the low-level samples were assayed in three runs using two different sets of EliA Rib-P Well lots over three different days on a Phadia 250 and Phadia 2500E each in 5-fold determination. Confidential Page 8 (14) {11}------------------------------------------------ For each instrument type, the total number of combined observations for blank and low-level samples is 120 (60 per reagent set, 15 per sample and reagent set). The results are summarized in the table below: | Instrument | LoB | LoD | LoQ | |-------------------------------------------------------|-----------|-----------|-----------| | | EliA U/mL | EliA U/mL | EliA U/mL | | Phadia 250 | 0.0 | 0.5 | 1.9 | | E-module of the Phadia 2500<br>and Phadia 5000 series | 0.1 | 0.4 | 1.1 | A harmonized LoB of 0.1 EliA U/mL, LoD of 0.5 EliA U/mL, and LoQ of 1.9 EliA U/mL for the immunoassay was used. The LoD for EliA Rib-P is 0.5 EliA U/mL, determined consistent with the quidelines in CLSI document EP17-A2 and with proportions of false positives (α) less than 5% and false negatives (β) less than 5%: based on 240 determinations with 120 blank and 120 low-level replicates per instrument type; and LoB of 0.1 EliA U/mL. ## e) Analytical specificity: Endogenous and Exogenous Interference: A study was run to investigate whether high concentrations of potentially interfering substances in serum, like bilirubin, hemoglobin, lipemic factor, rheumatoid factor, Ibuprofen, Losartan, Hydroxychloroquine, Azathioprine, Prednisone, Rituximab and Infliximab adversely affect the results of the new device. Three serum samples (one negative sample, one sample with a concentration within the equivocal range, and one high positive sample) were prediluted in EliA Sample Diluent and spiked with the different interfering substances or blank solution. The samples were tested in triplicates. A calibration curve was run in duplicate. The runs were repeated twice. One batch of EliA antigen wells and one batch of system reagents were used throughout the studies. The ratio of blank/spiked sample ranged from 0.90 – 1.10 for EliA Rib-P. No interference was observed up to the concentrations listed in the table below: | Potential Interfering<br>Compound | Concentration in<br>undiluted sample | |-----------------------------------|--------------------------------------| | Bilirubin F | 40 mg/dL | | Bilirubin C | 40 mg/dL | | Hemoglobin | 1000 mg/dL | | Lipemic factor | 2000 mg/dL | | Rheumatoid factor | 550 IU/mL | Phadia AB, Rapsgatan 7P, P.O. Box 6460, 751 37 Uppsala, Sweden {12}------------------------------------------------ Image /page/12/Picture/0 description: The image contains the logo of Thermo Fisher Scientific. The words "Thermo Fisher" are in red, and the word "SCIENTIFIC" is in a smaller font size and in black. The logo is simple and clean, and it is likely used on the company's website, products, and marketing materials. | Ibuprofen | 21.9 mg/dL | |--------------------|------------| | Losartan | 1.14 mg/dL | | Hydroxychloroquine | 0.23 mg/dL | | Azathioprine | 0.26 mg/dL | | Prednisone | 0.01 mg/dL | | Rituximab | 109 mg/dL | | Infliximab | 26.4 mg/dL | #### Reference Sera: Externally defined sera should be measured according to their target values as mentioned by the institution CDC. Using EliA Rib-P, all 12 CDC samples were found according to their target. #### f) Assay Cut-Off: To define the cut-off, a study was performed using a cohort consisting of 70 apparently healthy blood donors and 30 samples from SLE patients. The samples were measured on a Phadia 250 instrument. The cut-off was set as follows for EliA Rib-P: | <7 EliA U/mL | Negative | |----------------|-----------| | 7-10 EliA U/mL | Equivocal | | >10 EliA U/mL | Positive | In case of equivocal results, it is recommended to retest the patient after 8-12 weeks. - 2. Comparison Studies: - a) Method Comparison with Predicate Device: A total of 323 patient samples with concentrations covering the measuring range were tested. The samples were analyzed with the EliA Rib-P and Quanta Lite Ribosome P ELISA assay. The test was run in single determination and evaluated according to the Directions for Use. The results are summarized in the tables below: {13}------------------------------------------------ | N = 323 | Quanta Lite<br>Ribosome P<br>positive:<br>> 20 Units | Quanta Lite<br>Ribosome P<br>negative:<br>< 20 Units | Total | |----------------------------------------|------------------------------------------------------|------------------------------------------------------|-------------| | EliA Rib-P positive:<br>> 10 EliA U/mL | 36 | 4 | 40 | | EliA Rib-P negative:<br>< 10 EliA U/mL | 2 | 281 | 283 | | Total | 38 | 285 | 323 | | | Calculation | Agreement (%) | 95% CI | | Positive Percent<br>Agreement | $100% x 36 / 38$ | 94.7 | 82.3 - 99.4 | | Negative Percent<br>Agreement | $100% x 281 / 285$ | 98.6 | 96.4 - 99.6 | | Total Agreement | $100% x (36 + 281) / 323$ | 98.1 | 96.0 - 99.3 | EliA Rib-P: equivocal results considered negative EliA Rib-P: equivocal results considered positive | N = 323 | Quanta Lite<br>Ribosome P<br>positive:<br>> 20 Units | Quanta Lite<br>Ribosome P<br>negative:<br>< 20 Units | Total | |---------------------------------------|------------------------------------------------------|------------------------------------------------------|-------------| | EliA Rib-P positive:<br>>7 EliA U/mL | 38 | 34 | 72 | | EliA Rib-P negative:<br>< 7 EliA U/mL | 0 | 251 | 251 | | Total | 38 | 285 | 323 | | | Calculation | Agreement (%) | 95% CI | | Positive Percent<br>Agreement | $100% x 38 / 38$ | 100 | 90.7 - 100 | | Negative Percent<br>Agreement | $100% x 251 / 285$ | 88.1 | 83.7 - 91.6 | | Total Agreement | $100% x (38 + 251) / 323$ | 89.5 | 85.6 - 92.6 | - b) Instrument Comparison: Performance of EliA Rib-P was evaluated on the Phadia 250 and Phadia 2500E instrument using 47 positive, 10 equivocal and 28 negative samples. The samples were analyzed in single determination on one Phadia 250 and one Phadia 2500E instrument each. The regression analysis results are summarized as follows: Confidential Page 11 (14) {14}------------------------------------------------ | | Intercept | Slope | |----------|---------------|-------------| | Estimate | -0.76 | 0.94 | | 95% Cl | -1.20 - -0.48 | 0.93 - 0.98 | - 3. Clinical Studies: - a) Clinical Sensitivity and Specificity: 560 clinically defined serum samples with a diagnosis from patients with systemic lupus erythematosus (n = 146), Celiac disease (n = 13), Crohn's disease (n = 12), CTD overlap Non-MCTD (n = 10), Dermatomyositis (n = 4), Polymyositis (n = 6), Graves' disease (n = 12), Primary antiphospholipid syndrome (n = 12), Primary biliary cirrhosis (n = 21), Sjögren's syndrome (n = 23), Type 1 Diabetes (n = 12), Ulcerative colitis (n = 11) cancer (n = 10), Mixed connective tissue disease (n = 10), Rheumatoid arthritis (n = 30), bacterial infections (n = 36), viral infections (n = 56), Hashimoto's disease (n = 10), Granulomatosis with Polyanqiitis (n = 4), Autoimmune Hepatitis (n = 16), Polymyalgia Rheumatica (n = 25), Systemic sclerosis, diffuse (n = 48) and Systemic sclerosis, limited (n = 33) were used to determine sensitivity and specificity of the assay. The results are summarized in the tables below. | n=560 | Diagnostic<br>Group | Disease<br>Controls | Total | |--------------------------------|---------------------|---------------------|-------| | Positive test<br>≥ 7 EliA U/mL | 51 | 3 | 54 | | Negative test<br>< 7 EliA U/mL | 95 | 411 | 506 | | Total | 146 | 414 | 560 | EliA Rib-P - equivocal results evaluated as positive: Sensitivity (95% CI): 34.9% (27.2% - 43.3%) Specificity (95% CI): 99.3% (97.9% - 99.9%) EliA Rib-P - equivocal results evaluated as neqative: | n=560 | Diagnostic<br>Group | Disease<br>Controls | Total | |---------------------------------|---------------------|---------------------|-------| | Positive test<br>> 10 EliA U/mL | 41 | 1 | 42 | | Negative test<br>≤ 10 EliA U/mL | 105 | 413 | 518 | | Total | 146 | 414 | 560 | Sensitivity (95% CI): 28.1% (21.0 % - 36.1%) Phadia AB, Rapsgatan 7P, P.O. Box 6460, 751 37 Uppsala, Sweden {15}------------------------------------------------ Image /page/15/Picture/0 description: The image contains the logo for Thermo Fisher Scientific. The words "Thermo Fisher" are in red, stacked on top of the word "SCIENTIFIC" which is in black. The logo is simple and clean, with a focus on the company name. Specificity (95% CI): 99.8% (98.7% - 100%) The table below shows the results for each clinical subgroup: | | total | positive | positive | |--------------------------------------------------------------------------|-------|----------|----------| | Diagnostic Groups | n | n | % | | Systemic lupus erythematosus | 136 | 38 | 28% | | Systemic lupus erythematosus with<br>secondary antiphospholipid syndrome | 10 | 3 | 30% | | Target disease (Total) | 146 | 41 | | | Celiac disease | 13 | 0 | 0% | | Crohn's disease | 12 | 0 | 0% | | CTD overlap Non-MCTD | 10 | 0 | 0% | | Dermatomyositis | 4 | 0 | 0% | | Polymyositis | 6 | 0 | 0% | | Graves' disease | 12 | 0 | 0% | | Primary antiphospholipid syndrome | 12 | 0 | 0% | | Primary Biliary Cholangitis | 21 | 0 | 0% | | Sjögren's syndrome | 23 | 0 | 0% | | Type 1 Diabetes | 12 | 0 | 0% | | Ulcerative colitis | 11 | 0 | 0% | | Varied Cancer | 10 | 0 | 0% | | Mixed connective tissue disease | 10 | 0 | 0% | | Rheumatoid arthritis | 30 | 0 | 0% | | Bacterial infections | 36 | 0 | 0% | | Viral infections | 56 | 0 | 0% | | Hashimoto's disease | 10 | 0 | 0% | | Granulomatosis with Polyangiitis | 4 | 0 | 0% | | Systemic sclerosis, diffuse | 48 | 1 | 2% | | Autoimmune Hepatitis | 16 | 0 | 0% | | Polymyalgia Rheumatica | 25 | 0 | 0% | | Systemic sclerosis, limited | 33 | 0 | 0% | | Disease Controls (Total) | 414 | 1 | | | Total | 560 | | | - b) Other Clinical Supportive Data: Not applicable. {16}------------------------------------------------ - 4. Clinical Cut-Off: Same as assay cut-off. - 5. Expected Values/Reference Range: Antibody prevalence in autoimmune patients varies widely depending on disease area. The proportion of sera from a normal population found positive for the antinuclear antibodies covered by the EliA Rib-P test is below 1%. Expected values may vary depending on the population tested. The frequency distribution for antinuclear antibodies was investigated in a group of apparently healthy subjects equally distributed by age and gender, using sera from Caucasian, African American, Hispanic and Asian population obtained from a blood bank. The results are given in the table below: | Test | n | Median<br>EliA U/mL | 95th percentile<br>EliA U/mL | 99th percentile<br>EliA U/mL | |------------|-----|---------------------|------------------------------|------------------------------| | EliA Rib-P | 638 | 1.6 | 3.4 | 5.0 | #### Proposed Labeling The labeling is drafted in accordance with the requirements of 21 CFR Part 809.10. ## Conclusion All available data support that both immunoassays, the new device EliA Rib-P and its predicate device Quanta Lite Ribosome P ELISA perform substantially equivalent.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...