← Product Code [DSY](/productcode/DSY) · K202471

# exGraft ePTFE Vascular Graft, exGraft Carbon ePTFE Vascular Graft (K202471)

_Peca Labs · DSY · Dec 11, 2020 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K202471

## Device Facts

- **Applicant:** Peca Labs
- **Product Code:** [DSY](/productcode/DSY.md)
- **Decision Date:** Dec 11, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3450
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The exGraft and exGraft Carbon ePTFE Vascular Grafts are indicated for use as vascular prostheses. exGraft and exGraft Carbon ePTFE Vascular Grafts are intended for use as vascular prostheses for replacement or bypass of diseased vessels in patients suffering occlusive or aneurysmal diseases, in trauma patients requiring vascular replacement, for dialysis access, or for other vascular procedures.

## Device Story

Single-use, sterile vascular grafts constructed of expanded polytetrafluoroethylene (ePTFE). Used as vascular prostheses for vessel replacement or bypass. Features include radiopaque ink on the outer surface for visualization and, in the Carbon variant, a carbon coating on the inner luminal surface. Devices are implanted by surgeons during vascular procedures. They function as conduits to restore blood flow in patients with occlusive or aneurysmal disease, trauma, or those requiring dialysis access. Benefit is the provision of a biocompatible conduit for vascular reconstruction.

## Clinical Evidence

Bench testing only. Performance data includes probe burst strength, longitudinal and circumferential tensile strength, kink radius, microscopic porosity, suture retention strength, strength after repeated puncture, water entry pressure, and conduit dimensioning. Historical data from reference device (K183613) also utilized.

## Technological Characteristics

Expanded polytetrafluoroethylene (ePTFE) vascular graft. Features radiopaque ink on abluminal surface. Carbon-impregnated luminal surface in Carbon variant. Single-use, sterile. No outer ePTFE wrap.

## Regulatory Identification

A vascular graft prosthesis is an implanted device intended to repair, replace, or bypass sections of native or artificial vessels, excluding coronary or cerebral vasculature, and to provide vascular access. It is commonly constructed of materials such as polyethylene terephthalate and polytetrafluoroethylene, and it may be coated with a biological coating, such as albumin or collagen, or a synthetic coating, such as silicone. The graft structure itself is not made of materials of animal origin, including human umbilical cords.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance Document for Vascular Prostheses 510(k) Submissions.”

## Predicate Devices

- Gore-Tex Vascular Grafts and Gore-Tex Stretch Vascular Grafts ([K013250](/device/K013250.md))

## Reference Devices

- exGraft and exGraft Carbon ePTFE Vascular Grafts ([K183613](/device/K183613.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 11, 2020

PECA Labs Doug Bernstein Chief Executive Officer 4424 Penn Ave, Suite 201 Pittsburgh, Pennsylvania 15224

Re: K202471

Trade/Device Name: exGraft ePTFE Vascular Graft, exGraft Carbon ePTFE Vascular Graft Regulation Number: 21 CFR 870.3450 Regulation Name: Vascular Graft Prosthesis Regulatory Class: Class II Product Code: DS Y Dated: August 26, 2020 Received: August 28, 2020

Dear Doug Bernstein:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Carmen Gacchina Johnson, PhD Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202471

Device Name

exGraft ePTFE Vascular Graft, exGraft Carbon ePTFE Vascular Graft

### Indications for Use (Describe)

The exGraft and exGraft Carbon ePTFE Vascular Grafts are indicated for use as vascular prostheses.

exGraft and exGraft Carbon ePTFE Vascular Grafts are intended for use as vascular prostheses for replacement or bypass of diseased vessels in patients suffering occlusive or aneurysmal diseases, in trauma patients requiring vascular replacement, for dialysis access, or for other vascular procedures.

| Type of Use (Select one or both, as applicable)                                                 |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label>             |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for PECA Labs. The text is in a dark purple color. The font is sans-serif and the letters are spaced closely together.

# K202471

# 510(k) Summary

(as required by 21 CFR 807.92)

### l. SUBMITTER

PECA Labs, Inc. 4424 Penn Avenue Suite 201 Pittsburgh, PA 15224 Phone: (412) 482-3755 Establishment Registration Number: 3013718163

Contact Person: Doug Bernstein, Chief Executive Officer Date Prepared: December 11, 2020

### II. DEVICE

| Name of Device:       | exGraft ePTFE Vascular Graft, exGraft Carbon ePTFE Vascular Graft |
|-----------------------|-------------------------------------------------------------------|
| Common or Usual Name: | Vascular Graft                                                    |
| Classification Name:  | 21 CFR 870.3450 Prosthesis, Vascular Graft                        |
| Regulatory Class:     | Class II                                                          |
| Product Code:         | DSY                                                               |

#### III. PREDICATE DEVICE

Gore-Tex Vascular Grafts and Gore-Tex Stretch Vascular Grafts [K013250].

exGraft and exGraft Carbon ePTFE Vascular Grafts [K183613] are used as reference device.

The predicate and reference devices have not been subject to a design-related recall.

- IV. DEVICE DESCRIPTION
The exGraft and exGraft Carbon ePTFE Vascular Grafts are single use sterile vascular grafts constructed of expanded polytetrafluoroethylene (ePTFE) with a radiopaque ink applied to the surface. The exGraft Carbon ePTFE vascular grafts also contain a carbon coating impregnated into the inner surface of the graft wall.

- INDICATIONS FOR USE V.
The exGraft and exGraft Carbon ePTFE Vascular Grafts are indicated for use as vascular prostheses.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for PECA labs. The logo is in a dark purple color. The text is simple and modern.

exGraft and exGraft Carbon ePTFE Vascular Grafts are intended for use as vascular prostheses for replacement or bypass of diseased vessels in patients suffering occlusive or aneurysmal diseases, in trauma patients requiring vascular replacement, for dialysis access, or for other vascular procedures.

### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE VI.

# Compared to the predicate device Gore-Tex Vascular Grafts and Gore-Tex Stretch Vascular Grafts [K013250]:

The proposed exGraft and exGraft Carbon ePTFE Vascular Grafts and the predicate device have the following similarities:

- Same intended use
- . Same basic design and material
- Same basic configurations
- Same basic packaging configuration

The proposed exGraft and exGraft Carbon ePTFE Vascular Grafts and the predicate device have the following differences:

- Addition of radiopaque ink pad printed on the outer (abluminal) surface of the exGraft and exGraft Carbon ePTFE Vascular Grafts
- . No additional outer ePTFE wrap/coating on the abluminal surface of the exGraft and exGraft Carbon ePTFE Vascular Grafts
- Addition of carbon impregnation on the inner (luminal) surface of the exGraft Carbon ePTFE Vascular Grafts.

The differences stated above are considered technological differences. However, there are no new questions of safety and effectiveness raised by these differences.

#### VII. PERFORMANCE DATA

The proposed exGraft and exGraft Carbon ePTFE Vascular Grafts leverage historical data from the reference device, the exGraft and exGraft Carbon ePTFE Vascular Grafts, cleared under K183613.

In addition, the following testing was conducted to support substantial equivalence:

- . Probe Burst Strength
- Longitudinal Tensile Strength
- Circumferential Tensile Strength
- Kink Radius
- Microscopic Porosity ●
- Suture Retention Strength ●
- . Strength After Repeated Puncture

{5}------------------------------------------------

- Water Entry Pressure ●
- Conduit Dimensioning (including Relaxed Internal Diameter, Wall Thickness, and Length)

# CONCLUSION

Compared to the predicate device, the exGraft Carbon ePTFE Vascular Grafts has the same intended use and the technological differences do not raise different questions of safety and effectiveness. Based on historical data and performance testing, the exGraft and exGraft Carbon ePTFE Vascular Grafts are substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K202471](https://fda.innolitics.com/device/K202471)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).