Li-Battery powered OTC TENS/EMS, model EM-6000, EM-6100, EM-6200 & EM-6300
K202470 · Everyway Medical Instruments Co.,Ltd · NUH · Nov 25, 2020 · Neurology
Device Facts
Record ID
K202470
Device Name
Li-Battery powered OTC TENS/EMS, model EM-6000, EM-6100, EM-6200 & EM-6300
Applicant
Everyway Medical Instruments Co.,Ltd
Product Code
NUH · Neurology
Decision Date
Nov 25, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Intended Use
TENS (For model EM-6000, EM-6100 & EM-6300) : intended for temporary relief of pain associated with sore and aching muscles in the low back as well as upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities. (Choose TENS B/N/M/S1/S2 adjustable mode & TENS Modes P to P6 for EM-6300). EMS (For model EM-6000, EM-6200 & EM-6300): intended to stimulate healthy muscles in order to improve or facilitate muscle performance. (Choose EMS C/S/A adjustable mode & EMS Modes P7 to P12 for EM-6300).
Device Story
Li-Battery powered OTC TENS/EMS devices (models EM-6000, EM-6100, EM-6200, EM-6300) provide transcutaneous electrical nerve stimulation for pain relief and electrical muscle stimulation for muscle performance. Devices operate via user-selected modes (adjustable B/N/M/S1/S2 for TENS; C/S/A for EMS; preset P-modes for EM-6300). Intended for OTC use by patients for home-based muscle stimulation and pain management. Output is delivered via electrodes to the skin; intensity and mode are controlled by the user to manage muscle soreness or facilitate performance. Benefits include non-pharmacological pain relief and muscle conditioning.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Li-battery powered; transcutaneous electrical nerve stimulation (TENS) and electrical muscle stimulation (EMS) circuitry; multiple adjustable and preset stimulation modes; handheld form factor; output via surface electrodes.
Indications for Use
Indicated for OTC use for temporary relief of pain associated with sore/aching muscles in low back and extremities due to strain from exercise or normal activities (TENS); and for stimulation of healthy muscles to improve or facilitate muscle performance (EMS).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Related Devices
K202317 — Li-Battery powered OTC TENS/EMS, model EV-804, model EV-805, & model EV-806 · Everyway Medical Instruments Co.,Ltd · Nov 13, 2020
K212184 — TENS and EMS Stimulation (OTC) · Changsha Yuwen Medical Equipment Co., Ltd. · Apr 1, 2022
K243856 — TENS AND EMS (TENS and Muscle Stimulator) ( Model AS8012C,AS8016,AS8019,AS9020E,AS9026) · Shenzhen Astec Technology Co., Ltd. · May 29, 2025
K201335 — Li-Battery powered OTC TENS/EMS Combination Stimulator · Everyway Medical Instruments Co.,Ltd · Oct 24, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 25, 2020
Everyway Medical Instruments Co., Ltd. Paul Hung Official Correspondent 3FL., No. 5, Lane 155, Section 3, Peishen Rd. Shenkeng District, New Taipei City, Taiwan 22203 China
Re: K202470
Trade/Device Name: Li-Battery powered OTC TENS/EMS, model EM-6000, EM-6200 & EM-6300 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NUH, NGX Dated: August 27, 2020 Received: August 28, 2020
Dear Paul Hung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Heather L. Dean -S
Heather Dean, Ph.D. Assistant Director. Acute Injurv Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202470
#### Device Name
Li-Battery powered OTC TENS/EMS, model EM-6000, EM-6100, EM-6200 & EM-6300.
## Indications for Use (Describe)
TENS (For model EM-6000, EM-6100 & EM-6300) : intended for temporary relief of pain associated with sore and aching muscles in the low back as well as upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities. (Choose TENS B/N/M/S1/S2 adjustable mode & TENS Modes P to P6 for EM-6300).
EMS (For model EM-6000, EM-6200 & EM-6300): intended to stimulate healthy muscles in order to improve or facilitate muscle performance. (Choose EMS C/S/A adjustable mode & EMS Modes P7 to P12 for EM-6300).
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|----------------------------------------------|----------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span style="text-decoration: overline;">X</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.