DAILIES AquaComfort Plus and DAILIES AquaComfort Plus FreshTech (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in not-aphakic persons with non-diseased eyes with up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity. DAILIES AquaComfort Plus Toric (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in not-aphakic persons with non-diseased eyes with 6.00 diopters (D) or less of astigmatism. DAILIES AquaComfort Plus Multifocal (nelfilcon A) One-Day Contact Lenses are indicated for the optical correction of refractive ametropia) and/or presbyopia in not-aphakic persons with non-diseased eves who may require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity. All DAILIES AquaComfort Plus (nelfilcon A) One-Day Contact Lenses are to be prescribed for single use, daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
Device Story
Soft, daily disposable contact lenses (nelfilcon A) for refractive error correction. Lenses feature spherical, toric, or multifocal designs; tinted with phthalocyanine blue for handling. Supplied sterile in foil-blister packs with phosphate-acetate buffered saline containing HPMC, PEG 400, and Poloxamer 108. Prescribed by eye care professionals; patient self-applied for single-day use. Modification involves extended shelf-life (84 months). Benefits include vision correction without cleaning/disinfection requirements.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical stability and biocompatibility testing (cytotoxicity via CGI and Modified ISO/USP Elution methods).
Indicated for daily wear optical correction of refractive ametropia (myopia/hyperopia) and/or presbyopia in non-aphakic, non-diseased eyes. Includes patients with up to 6.00D astigmatism (Toric) or up to 2.00D astigmatism (Multifocal) and reading additions up to +3.00D.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
September 22, 2020
Alcon Laboratories, Inc. Dr. Andreas Friese Regulatory Project Director Industriering 1 Grosswallstadt, Bavaria 63868 Germany
Re: K202448
Trade/Device Name: DAILIES AquaComfort Plus / DAILIES AquaComfort Plus FreshTech / DAILIES AquaComfort Plus Toric / DAILIES AquaComfort Plus Multifocal Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: August 24, 2020 Received: August 26, 2020
Dear Dr. Andreas Friese:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K202448
#### Device Name
DAILIES AquaComfort Plus, DAILIES AquaComfort Plus FreshTech, DAILIES AquaComfort Plus Toric, DAILIES AquaComfort Plus Multifocal
#### Indications for Use (Describe)
DAILIES AquaComfort Plus and DAILIES AquaComfort Plus FreshTech (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in not-aphakic persons with non-diseased eyes with up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity.
DAILIES AquaComfort Plus Toric (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in not-aphakic persons with non-diseased eyes with 6.00 diopters (D) or less of astigmatism.
DAILIES AquaComfort Plus Multifocal (nelfilcon A) One-Day Contact Lenses are indicated for the optical correction of refractive ametropia) and/or presbyopia in not-aphakic persons with non-diseased eves who may require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity.
All DAILIES AquaComfort Plus (nelfilcon A) One-Day Contact Lenses are to be prescribed for single use, daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
Type of Use (Select one or both, as applicable)
| <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------|
| <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K202448
#### 510(k) Summary 8
This 510(k) summary document has been prepared in accordance with section 21 CFR 807.92.
#### I. Submitter Information
| Company: | Alcon Laboratories, Inc.<br>6201 South Freeway<br>Fort Worth, TX 76134-2099, USA |
|-----------------|----------------------------------------------------------------------------------|
| Contact Person: | Dr. Andreas Friese, Regulatory Project Director |
| Phone: | +49 6022-240-514 |
| Fax: | +49 6022-240-512 |
| Email: | andreas.friese@alcon.com |
| Date Prepared: | August 20, 2020 |
#### Devices Subject to this 510(k) II.
| Trade Names: | DAILIES AquaComfort Plus,<br>DAILIES AquaComfort Plus FreshTech,<br>DAILIES AquaComfort Plus Toric,<br>DAILIES AquaComfort Plus Multifocal |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Soft Contact Lenses |
| Classification Name: | Soft (Hydrophilic) Contact Lenses |
| Device Classification: | Class II [21 CFR 886.5925] |
| Product Code: | LPL, MVN |
## III. Predicate Device
The 510(k) devices are a modification of the same predicate devices, i.e. DAILIES AquaComfort Plus family (nelfilcon A) soft contact lenses, which are legally commercialized devices in the US per the following most recent US FDA 510(k) clearance: K181454.
{4}------------------------------------------------
# IV. Device Description
DAILIES AquaComfort Plus are soft contact lenses, intended for the optical correction of refractive error. The lenses are available in spherical, toric and multifocal designs.
The DAILIES AquaComfort Plus lens material is nelfilcon A, a high water, non-ionic hydrophilic lens material consisting of approximately 31% PVA (polyviny) alcohol partially acetalized with N-formylmethyl acrylamide) and 69% water.
The lens material is considered a Group II high water, non-ionic contact lens material per the 1994 FDA Premarket Notification (510(k)) guidance document for daily wear contact lenses. The lens material further contains non-functionalized high and ultra-high molecular weight PVA and the color additive phthalocyanine blue to create a light blue edge to edge tint (Visitint™) to make the lenses easier to see when handling.
DAILIES AquaComfort Plus soft contact lenses are supplied sterile. The lenses immersed in buffered saline solution are packaged in individual foil-blister-packs primary packaging system and are terminally sterilized in a validated autoclave (moist heat, steam under pressure).
The foil-blister pack primary packaging system consists of an injection molded polypropylene blister shell sealed with a polyester coated aluminum foil lidding material top.
The lenses are supplied in strips of five foil sealed blister packs each containing approximately 0.65 ml phosphate-acetate buffered saline solution. The packaging saline may contain up to 0.05% Poloxamer 108. The package saline additionally contains the comfort additives hydroxypropylmethyl cellulose (HPMC) and polyethylene glycol 400 (PEG 400). Sealed blister strips are provided in secondary packaging carton boxes containing 5, 30 or 90 lenses each.
#### Indications for Use V.
The 510(k) devices are a modification of the same predicate devices. The Indications for Use remain the same:
DAILIES AquaComfort Plus and DAILIES AquaComfort Plus FreshTech (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in not-aphakic persons with non-diseased eyes with up to approximately 1.50 diopters (D) of astigmatism that does not interfere with visual acuity.
DAILIES AquaComfort Plus Toric (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia and hyperopia) in not-aphakic persons with non-diseased eyes with 6.00 diopters (D) or less of astigmatism.
{5}------------------------------------------------
DAILIES AquaComfort Plus Multifocal (nelfilcon A) One-Day Contact Lenses are indicated for daily wear for the optical correction of refractive ametropia (myopia or hyperopia) and/or presbyopia in not-aphakic persons with non-diseased eyes who may require a reading addition of +3.00 diopters (D) or less and who may have 2.00 diopters (D) or less of astigmatism that does not interfere with visual acuity.
All DAILIES AquaComfort Plus (nelfilcon A) One-Day Contact Lenses are to be prescribed for single use, daily disposable wear. The lenses are not intended to be cleaned or disinfected and should be discarded after a single use.
# VI. Comparison of Technological Characteristics with the Predicate Device
The proposed device modification involves a labeling change in terms of an extension of the expiration dating for DAILIES AquaComfort Plus soft contact lenses.
The following matrix (Table 1) summarizes the characteristics of the modified devices as compared to the predicate devices.
| Element of<br>Comparison | Predicate Device(s) | Modified Device(s) |
|--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Administrative / Regulatory Information | | |
| 510(k) Number | K181454 | To be assigned |
| Product Name(s) | DAILIES AquaComfort Plus,<br>DAILIES AquaComfort Plus FreshTech, DAILIES<br>AquaComfort Plus Toric, DAILIES AquaComfort Plus<br>Multifocal | Same |
| Device<br>Classification<br>Information | Class II, Soft (Hydrophilic)<br>Contact Lenses, 21 CFR<br>886.5925 | Same |
| Indications For Use Information | | |
| Intended Use | One-day contact lenses for the<br>optical correction of refractive<br>error. Single use, daily<br>disposable wear. | Same |
| Technology Information | | |
| Lens Material | nelfilcon A | Same |
| Material<br>Classification | FDA Group II (>50% H2O,<br>nonionic) | Same |
| Element of<br>Comparison | Predicate Device(s) | Modified Device(s) |
| Water Content | 69% | Same |
| Visibility Tint | Light blue | Same |
| Manufacturing<br>Method | Lightstream Technology:<br>Full mold cast | Same |
| Lens Designs | Spherical, aspherical, toric,<br>multifocal | Same |
| Sterilization | Steam sterilization, validated<br>autoclave | Same |
| Primary<br>Packaging System | Foil blister pack:<br>polypropylene blister shell<br>sealed with a polyester coated<br>aluminum foil lidding | Same |
| Primary<br>Packaging Blister<br>Shell | Injection-molded poly-<br>propylene (PP) blister shell | Same |
| Primary<br>Packaging Foil<br>Lidding | Multi-layer laminate structure<br>polyester-coated aluminum<br>foil lidding supplied by:<br>Constantia-Pirk Folien Huhtamaki | Same |
| Package Storage /<br>Saline Solution | Phosphate-acetate buffered<br>saline (with PEG and HPMC<br>additives for DAILIES Aqua-<br>Comfort Plus products) and up<br>to 0.05% Poloxamer 108 | Same |
| Performance Specifications including any Testing | | |
| Refractive Index | 1.38 | Same |
| Light<br>Transmittance | $\ge$ 92%T | Same |
| Oxygen<br>Permeability (Dk) | 26 | Same |
| Elastic Modulus | 0.9 MPa | Same |
| Biocompatibility | Biocompatible as confirmed<br>by biocompatibility testing | Same |
| Shelf-life | 60 months as confirmed by<br>shelf-life stability testing | 84 months as confirmed by<br>shelf-life stability testing |
Table 1: Substantial Equivalence Comparison
{6}------------------------------------------------
# Table 1: Substantial Equivalence Comparison
{7}------------------------------------------------
In accordance with the criteria for claims of substantial equivalence in the FDA guidance Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses, May 1994, the information provided supports the claim of substantial equivalence to a lens with an existing USAN and the same manufacturing process.
## VII. Performance Data
Performance testing was conducted in consideration of the May 1994 FDA guideline titled Premarket Notification 510(k) Guidance Document for Class II Contact Lenses. The following performance data are provided in support of the substantial equivalence determination:
## Non-clinical Testing
Successful stability testing including biocompatibility testing, i.e. cytotoxicity testing by Cell Growth Inhibition (CGI) and Modified ISO/USP Elution (MEL) methods, was completed for the modified device to verify equivalence to the predicate device. This resulted in all acceptance criteria being met.
## Clinical Testing
The scope of the device modification did not require clinical testing to establish safety and effectiveness of the modified device.
## VIII. Conclusions
The cumulative results of performance testing demonstrate the safety, efficacy and performance of the modified device(s) and, thus, substantial equivalence to the predicate device(s).
DAILIES AquaComfort Plus (nelfilcon A) soft contact lenses with extended shelf-life are substantially equivalent to the predicate lenses of material properties, biocompatibility, clinical performance, and indications for use.
Any differences which may exist between the modified and the predicate device do not adversely affect the established performance characteristics and safety and effectiveness profile.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.