Infrared Forehead Thermometer

K202420 · Conmo Electronic Company Limited · FLL · Feb 25, 2021 · General Hospital

Device Facts

Record IDK202420
Device NameInfrared Forehead Thermometer
ApplicantConmo Electronic Company Limited
Product CodeFLL · General Hospital
Decision DateFeb 25, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2910
Device ClassClass 2
AttributesPediatric

Intended Use

The infrared forehead Thermometer DPT-IFT100 is a non-contact thermometer intended for the intermittent measurement of human body temperature from forehead for people of one month old and above. The device is reusable for home use and clinical use.

Device Story

The DPT-IFT100 is a non-contact infrared forehead thermometer. It uses a thermopile sensor to collect infrared thermal radiation from the forehead at a distance of 3-5 cm. The sensor converts radiant power into an electrical signal, which is processed by an application-specific integrated circuit (ASIC) featuring a 24-bit analog-to-digital converter. An internal algorithm (CM3.1) calculates the temperature, compensating for ambient conditions, and displays the result on an LCD. The device is powered by two AAA batteries and operated via a single button for power and measurement. It is intended for use by clinicians or patients in home and clinical settings. The output provides a digital temperature reading, assisting healthcare providers in assessing patient body temperature. The device is reusable and includes a backlight for visibility.

Clinical Evidence

Clinical study conducted with 150 subjects (50 infants, 50 children, 50 adults) per ASTM E1965-98. Results demonstrated that clinical bias and repeatability met protocol acceptance criteria. Bench testing confirmed compliance with ISO 80601-2-56, ASTM E1965-98, and electrical safety/EMC standards (IEC 60601-1, 60601-1-2, 60601-1-11). Biocompatibility testing (cytotoxicity, skin irritation, sensitization) per ISO 10993-5 and 10993-10 confirmed compliance.

Technological Characteristics

Non-contact infrared thermometer using a thermopile sensor (MRT511). Materials: ABS housing and battery cover. Energy source: 3.0V DC (2x AAA batteries). Connectivity: Standalone. Dimensions: 145x38x35mm. Standards: ASTM E1965-98, ISO 80601-2-56, AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-1. Signal processing: 24-bit ADC. Algorithm: CM3.1.

Indications for Use

Indicated for intermittent measurement of human body temperature via forehead in patients one month of age and older. Suitable for home and clinical use.

Regulatory Classification

Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; (2) Device is not a clinical thermometer with continuous temperature measurement functions; and (3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size below. Conmo Electronic Company Limited % Charles Mack Principal Engineer Irc 2950 E Lindrick Drive Chandler, Arizona 85249 Re: K202420 Trade/Device Name: Infrared Forehead Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: January 18, 2021 Received: January 26, 2021 Dear Charles Mack: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) Device Name Infrared Forehead Thermometer, DPT-IFT100 Indications for Use (Describe) The infrared forehead Thermometer DPT-IFT100 is a non-contact thermometer intended for the intermittent measurement of human body temperature from forehead for people of one month old and above. The device is reusable for home use and clinical use. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for CONMO. The logo is in blue and consists of the word "CONMO" in a sans-serif font. The word is enclosed in an oval shape, with the top and bottom of the oval curving around the word. The logo is simple and clean, and the blue color gives it a professional look. ### K202420 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92. ### Date: February 25, 2021 - 1. Company and Correspondent submitting: Name - Conmo Electronic Company Limited Address - Suiyi Town, Xincheng County, Laibin City Guangxi Province, China 546100 Tel: 86-755-28397469 Fax: 86-755-82922769 Contact - Ms. Ying Qi General Manager Email: aliceqy@vip.163.com, aliceqy@163.com FDA Establishment Registration No.: 3010968402 Designated Submission Correspondent and US Agent: IRC USA 2950 E Lindrick Dr., Chandler, Arizona 85249, USA Mr. Charles Mack Principal Engineer Tel: 931-6254938 Email: charliemack@irc-us.com ### 2. Device : Trade/proprietary name Infrared Forehead Thermometer, Model DPT-IFT100 Common Name Clinical Electronic Thermometer Regulation Name Clinical Electronic Thermometer Classification Number 21 CFR 880.2910 Product Code FLL Regulatory Class Class II Device Panel General Hospital {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "CONMO" in large, bold, blue letters. The word is framed by two curved lines, one above and one below, that are also blue. The overall design is simple and clean, with a focus on the text. - 3. Predicate Device: | Manufacturer | Predicate Device | 510(k) Number | |--------------------------------------------|---------------------------------------------------------------------------------|---------------| | Microlife<br>Intellectual<br>Property GmbH | Microlife Non-Contact Infrared<br>Forehead Thermometer,<br>Model FR1DG1 (NC200) | K191829 | 4. Device Description: The infrared forehead thermometer, Model DPT-IFT100, is an electronic thermometer using an infrared temperature sensor to measure infrared energy radiated from the forehead. This energy is collected through the infrared temperature sensor convert to a voltage signal. The signal is measured by the main microcontroller, calculated by the internal algorithm, finally converted into a digital temperature value to display on the LCD. The design consists of a lens to focus the infrared thermal radiation on to a detector, which converts the radiant power to an electrical signal displayed in units of temperature after being compensated for ambient temperature. This permits temperature measurement from a 3~5 cm distance without contact with the object to be measured. The Infrared forehead thermometer, Model DPT-IFT100, consists of the following parts: - a) Thermopile Sensor - b) Application-Specific Integrated Circuit - c) Housing - d) LCD and Backlight - e) 1 button (Power on/Power off/Measurement) - f) Alkaline batteries; size AAA, 2 x 1.5 V - g) LCD cover 5. Indications for use: The infrared forehead Thermometer DPT-IFT100 is a non-contact thermometer intended for the intermittent measurement of human body temperature from forehead for people of one month old and above. The device is reusable for home use and clinical use. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the word "CONMO" in blue, with a curved line above and below the word. The word is in all capital letters and appears to be a logo. The background is white. The font is a serif font. - 6. Comparison with the predicate device: Conmo Electronic Company Limited believes that the Infrared Forehead Thermometer, DPT-IFT100 is substantially equivalent to the (K191829) Microlife Non-Contact Infrared Forehead Thermometer (Microlife Intellectual Property GmbH). | | | Predicate Device<br>Microlife Non-Contact Infrared<br>Forehead Thermometer, Model<br>FR1DG1(NC200) (K191829) | | |-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------| | Characteristics | Subject Device<br>Infrared Forehead Thermometer, DPT-<br>IFT100 (K202420) | | Remark | | Manufacturer | CONMO ELECTRONIC COMPANY<br>LIMITED | Microlife Intellectual Property GmbH,<br>Switzerland | - | | Indication for Use | The infrared forehead Thermometer<br>DPT-IFT100 is a non-contact<br>thermometer intended for the intermittent<br>measurement of human body<br>temperature from forehead for people of<br>one month old and above. The device is<br>reusable for home use and clinical use. | The Microlife Non-Contact Infrared<br>Thermometer, Model FR1DG1(NC200) is<br>intended for the intermittent<br>measurement and monitoring of human<br>body temperature. The device is<br>indicated for use by people of all ages in<br>the home. | Note 1 | | Measurement<br>Method | Infrared radiation detection | Infrared radiation detection | Identical | | Measurement<br>range | 34.0°C to 43.0°C (93.2°F to 109.4°F) | 34.0°C to 43.0°C (93.2°F to 109.4°F) | Identical | | Accuracy<br>(Laboratory) | ±0.2°C : 35.0°C ~ 42.0°C<br>±0.3°C : 34.0°C ~ 34.9°C , 42.1°C ~ 43°C<br><br>±0.4°F: 95°F ~ 107.6°F<br>±0.5°F: 93.2°F ~ 94.8°F, 107.8°F ~<br>109.4°F | ±0.2°C : 35.0°C ~ 42.0°C<br>±0.3°C : 34.0°C ~ 34.9°C , 42.1°C ~ 43°C<br><br>±0.4°F: 95°F ~ 107.6°F<br>±0.5°F: 93.2°F ~ 94.8°F, 107.8°F ~<br>109.4°F | Identical | | Accuracy(clinic<br>repeatability) | 0.1°C | Unknown | Note 2 | | Temperature<br>Measurement<br>distance | 3~5cm | Appropriate within 5cm | Note 3 | | Measurement site | Forehead | Forehead | Identical | | Sensor type | Thermopile | Thermopile | Identical | | Temperature unit | °C or °F | °C or °F | Identical | | Operating<br>Environment | 10°C to 40°C (50°F to 104°F)<br>15%~85%RH | 15°C to 40°C (59°F to 104°F)<br>15%~95% RH | Note 4 | | Storage<br>Environment | -25°C to 55°C (-13°F to 131°F)<br>90% RH or less | -25°C to 55°C (-13°F to 131°F)<br>15-95% RH | | | Auto power-off<br>while no<br>operation | Yes | Yes | Identical | | Memory capacity | 1 measurement | 30 measurements | Note 5 | | Battery type | 3.0V DC with 2 AAA batteries | 3.0V DC with 2 AAA batteries | Identical | # Comparison to Predicate Devices {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the word "CONMO" in large, bold, blue letters. The word is centered and appears to be a logo or brand name. There are curved lines above and below the word, creating a visual frame around it. | Characteristics | Subject Device<br>Infrared Forehead Thermometer, DPT-<br>IFT100 (K202420) | Predicate Device<br>Microlife Non-Contact Infrared<br>Forehead Thermometer, Model<br>FR1DG1(NC200) (K191829) | Remark | |--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------| | LCD Backlight<br>feature | Available | Available | Identical | | Display type &<br>components | LCD | LCD | Identical | | Display resolution | 0.1°C/0.1 °F | 0.1°C/0.1 °F | Identical | | Reference site | Oral | Oral | Identical | | Physical<br>dimensions | 145*38*35mm | 156.7 x 43 x 47 mm | Note 6 | | Weight | 78.2g (with batteries),56.6g (w/o batteries) | Unknown | | | Skin contacting<br>materials: | Housing and battery cover: ABS<br>Key: ABS | Housing and battery cover: ABS<br>Key: PMMA | Note 7 | | Response time | 1s | Unknown | - | | Measurement<br>mode<br>(adjusted/direct<br>mode) | Adjusted | Unknown | Note 8 | | Algorithm | CM3.1 Algorithm | PH15.0 Algorithm | Note 9 | | Signal processing | 24-bit analog-to-digital<br>converter | 24-bit analog-to-digital<br>converter | Identical | | Sensor model | MRT511 | TPS336 | Note 10 | | IC model | SN8P2977 | HY11P14 | Note 11 | | Touch IC | No | Yes | Note 12 | | Conformance<br>standard | AAMI/ANSI ES60601-1<br>IEC60601-1-2<br>IEC 60601-1-11<br>ISO10993-1<br>ASTM E1965-98 (Performance)<br>ISO 80601-2-56(Performance) | AAMI/ANSI ES60601-1<br>IEC60601-1-2<br>IEC 60601-1-11<br>ISO10993-1<br>ASTM E1965-98 (Performance)<br>ISO 80601-2-56(Performance) | Identical | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the word "CONMO" in blue font. The word is surrounded by two curved lines, one above and one below, also in blue. The font appears to be a bold sans-serif typeface. The overall design is simple and clean. #### Note 1: The wording for the subject device is slightly different from the predicate device but they are both non-contact thermometers for the intermittent measurement of human body temperature via forehead. The subject device addresses the target population based on the clinical study conducted, and they both conform to the same performance standard - ISO 80601-2-56, Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement. The subject device added the device's description as reusable, the predicate device is also reusable based on the device manual. The subject device added the "clinical use" claim, they conform to the same safety and performance standards. The above wording differences between the subject device and predicate device will not affect the subject device's safety and effectiveness. #### Note 2: The result of clinical accuracy characteristics – clinic repeatability of subject device is 0.1℃. Although the predicate results are unknown, they conform to the same performance standard requirements of ASTM E1965-98. The differences between the predicate device and the subject device will not affect the subject device's performance. #### Note 3: The subject device's measurement distance is included in the predicate device range. The subject device's clinical study report demonstrates the clinical accuracy requirements of the standards ISO 80601-2-56 and ASTM E1965-98 within the distance range. The differences between the subject device and predicate will not affect the subject device's safety and effectiveness. #### Note 4: Although the subject device's operating and storage conditions are slightly different from predicate device, they conform to the same standard requirements of ISO 80601-2-56 and IEC 60601-1-11. Therefore, operating and storage conditions do not influence product safety and effectiveness. #### Note 5: Memory function is only used to store data and has no effect on product safety and effectiveness. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the word "CONMO" in blue, with a curved line above and below the word. The word is in a bold, sans-serif font. The curved lines are also blue and are symmetrical, creating an oval shape around the word. The background is white. #### Note 6: The physical dimension & weight difference does not raise new safety and effectiveness issues. The subject device conforms to the applicable safety, EMC, and performance standards, including ANSI AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-56, ASTM E1965-98, etc. same as predicate device and reference device. #### Note 7: Although the skin-contacting materials are different, both conform to the same biocompatibility standard requirements of ISO 10993-1. #### Note 8: The subject device's measurement mode is adjusted mode only, and the measurement reference site is oral. The predicate device's information is unknown, but they both conform to the same performance standards requirements of ASTM E1965 and ISO 80601-2-56. #### Note 9: The subject device's algorithm is CM3.1, while the predicate device's is PH15.0. The difference is the algorithm version. It does not affect performance and accuracy, which was evaluated in the performance testing. #### Note 10: The subject device sensor model is MRT511, whereas the predicate device is TPS336. Both sensor models have the same operating principles. The difference does not affect the safety, EMC, and performance standards, including ANSI AAMI ES 60601-1-2, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-56, ASTM E1965-98, etc. same as the predicate device. #### Note 11: The main IC of the subject device is SN8P2977, whereas the predicate device is HY11P14. Both main ICs have the same 24-bit analog to digital converter, just the model is different. The difference does not affect performance and accuracy, which was evaluated in the performance testing, {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the word "CONMO" in blue font. The word is surrounded by two curved lines, one above and one below, also in blue. The curved lines create an oval shape around the word. #### Note 12: The subject device has no touch IC, whereas the predicate device has a touch IC. The function of the touch IC is to power the device ON. The difference does not affect the safety, EMC, and performance standards, including ANSI AAMI ES 60601-1-2, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-56, ASTM E1965-98, etc. same as the predicate device. ### 7. Testing Summary: The following performance data is provided in support of the substantial equivalence determination. # Comparative Clinical Study: Clinical tests were conducted to verify that the proposed device meets the same design specifications as the Microlife Intellectual Property GmbH predicate. The testing was conducted according to the standard: - -ASTM E1965-98 (Reapproved 2016) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature This clinical study consists of 150 subjects, of which 1/3 (50 subjects) are infants, 1/3 (50 subjects) are children and the rest 1/3 (50 subjects) are adults (NOTE: Infants---newborn to one year; Children--- greater than one to five years; Adults---greater than five years old.). The test report demonstrated that the clinical data, represented by clinical bias and clinical repeatability met the acceptance criteria of the clinical study protocol. # Safety and EMC - Testing was performed to verify the basic safety and essential performance of the Infrared Forehead Thermometer, DPT-IFT100. The following tests were performed: - AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and । A2:2010/(R)2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005. MOD) - IEC 60601-1-2:2014 Medical devices part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances -Requirements and tests {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the word "CONMO" in blue font, with a curved line above and below the word. The curved lines create an oval shape around the word. The font is a bold sans-serif typeface. - IEC 60601-1-11 Edition 2.0 2015-01, Medical electrical equipment part 1--11: General requirements for basic safety and essential performance -Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. # Performance Data: The subject Infrared Forehead Thermometer, DPT-IFT100, was subjected to the following tests and passed all test criteria: - ISO 80601-2-56 Second edition 2017-03 Medical electrical equipment --Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement. [Including Amendment 1 (2018)] - ASTM E1965-98 (Reapproved 2016) Standard Specification for Infrared -Thermometers for Intermittent Determination of Patient Temperature - Premarket Notification [510(K)] Submissions For Clinical Electronic -Thermometers # Software Verification and Validation Software documentation, including verification & validation, was provided following FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for software with a moderate level of concern. # Cleaning Validation The device was cleaned following the procedure defined in the User's Manual. The device was checked for performance following the cleaning and met all performance requirements following the cleaning. # Biocompatibility The subject Infrared Forehead Thermometer, DPT-IFT100, uses similar material to the predicate Microlife Intellectual Property GmbH device. Biocompatibility testing was performed to demonstrate compliance with the same biocompatibility standards as performed by the predicate device. {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the word "CONMO" in blue font. The word is surrounded by two curved lines, one above and one below, that form an oval shape. The font is a simple, sans-serif style. The Infrared Forehead Thermometer, DPT-IFT100 device is classified per ISO 10993-1: 2009 Annex 1 Biological evaluation tests as follows: Surface Device – Intact skin- Contact Duration A – Iimited (≤24 h) Based on this classification, the following cytotoxicity, skin irritation, and sensitization tests were conducted per the following standards: - -ISO 10993-5: 2009 Biological evaluation of medical devices - Part 5 Tests for In Vitro Cytotoxicity - ISO 10993-10: 2010 Biological evaluation of Medical Devices Part 10: -Tests for Irritation- Skin Irritation - ISO 10993-10: 2010 Biological evaluation of Medical Devices Part 10: -Tests for Irritation - Skin sensitization The test results confirm compliance with the requirements of the standards. All the labeling and characteristics of the submitted Infrared Forehead Thermometer, DPT-IFT100, is the same as the predicate device and most infrared forehead thermometers currently on the market. The proposed device and predicate use similar measuring methodologies and components to achieve the measurements. The design of the submitted device specifications are similar to the predicate, with minor differences in the storage humidity range. - 8. Conclusions: Based on the performance testing, comparison and analysis provided it was concluded that the subject device, Infrared Forehead Thermometer, DPT-IFT100 is substantially equivalent to predicate device, Microlife Non-Contact Infrared Forehead Thermometer, Model FR1DG1(NC200) cleared under K191829.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...