K202391 · Aesculap, Inc. · IZI · Jan 21, 2021 · Radiology
Device Facts
Record ID
K202391
Device Name
DIR 800
Applicant
Aesculap, Inc.
Product Code
IZI · Radiology
Decision Date
Jan 21, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1600
Device Class
Class 2
Indications for Use
The DIR 800 is an accessory for the Aesculap Aeos and is used in viewing intra-operative blood flow in the cerebral vascular area including, but not limited to, assessing cerebral aneurysm and vessel branch occlusion, as well as patency in neurosurgery. It also aids in the visual assessment of intraoperative blood flow and vessel patency in bypass surgical procedures in neurosurgery.
Device Story
DIR 800 is an accessory for Aesculap Aeos surgical microscope; enables intra-operative fluorescence imaging using Indocyanine Green (ICG) dye. Device illuminates surgical site with near-infrared excitation light (740-800 nm); CMOS sensor captures resulting fluorescence signal (820-900 nm) from blood vessels. System provides real-time visualization of dye distribution on a monitor; includes picture-in-picture capability. Used by neurosurgeons in OR to assess vessel patency and aneurysm occlusion. Output assists surgeons in immediate intra-operative decision-making regarding bypass graft success or vessel flow status. Benefits include enhanced visualization of cerebrovascular anatomy compared to standard white-light microscopy.
Clinical Evidence
No human clinical trials. Bench testing verified excitation bandwidth, fluorescence pixel intensity, and system safety (IEC 60601-1/2). Animal studies (porcine and rabbit elastase models) compared DIR 800 imaging to Digital Subtraction Angiography (DSA). Neurosurgeons performed independent scoring of comparative image sets (non-occluded vs. occluded vessels). Results confirmed DIR 800 provides equivalent visualization of blood flow and aneurysms to DSA, meeting all pre-established acceptance criteria for image quality, brightness, and alignment.
Technological Characteristics
Accessory to Aeos surgical microscope; CMOS infrared-sensitive camera; 13x IR LED excitation source (740-800 nm); 820-900 nm bandpass detection filter. Motorized 8.9:1 zoom. Connectivity: PC-based control system with HDD storage. Sterilization: Not specified (accessory to microscope). Software: Major level of concern. Complies with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for neurosurgical patients requiring intra-operative visualization of blood flow in the cerebral vascular area, including assessment of cerebral aneurysms, vessel branch occlusion, and vessel patency during bypass procedures.
Regulatory Classification
Identification
An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Carl Zeiss Surgical GmbH, INFRARED 800 with FLOW Option (K100468)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 21, 2021
Aesculap, Inc. Ms. Kathy A. Racosky Project Manager I 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K202391
Trade/Device Name: DIR 800 Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic X-ray system Regulatory Class: Class II Product Code: IZI Dated: August 21, 2020 Received: August 21, 2020
Dear Ms. Racosky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202391
Device Name DIR 800
Indications for Use (Describe)
The DIR 800 is an accessory for the Aesculap Aeos and is used in viewing intra-operative blood flow in the cerebral vascular area including, but not limited to, assessing cerebral aneurysm and vessel branch occlusion, as well as patency in neurosurgery. It also aids in the visual assessment of intraoperative blood flow and vessel patency in bypass surgical procedures in neurosurgery.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 201 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 201 Subpart C) |
|------------------------------------------------------------|-----------------------------------------------------------|
|------------------------------------------------------------|-----------------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY (as required by 21 CFR 807.92)
DIR 800 January 21, 2021
| COMPANY: | Aesculap ®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 291671 |
|----------------------|---------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Kathy A. Racosky<br>610-984-9291 (phone)<br>610-791-6882 (fax)<br>kathy.racosky@aesculap.com |
| TRADE NAME: | DIR 800 |
| COMMON NAME: | Angiographic x-ray system |
| CLASSIFICATION: | Class II |
| CLASSIFICATION NAME: | System, X-Ray, Angiographic |
| REGULATION NUMBER: | 892.1600 |
| PRODUCT CODE: | IZI |
### SUBSTANTIAL EQUIVALENCE
DIR 800 is substantially equivalent to Carl Zeiss Surgical GmbH, INFRARED 800 with FLOW Option (K100468). DIR 800 has similar indications for use and is similar in design and operating principle to the predicate device.
### INDICATIONS FOR USE
The DIR 800 is an accessory for the Aesculap Aeos and is used in viewing intra-operative blood flow in the cerebral vascular area including, but not limited to, assessing cerebral aneurysm and vessel branch occlusion, as well as patency in neurosurgery. It also aids in the visual assessment of intraoperative blood flow and vessel patency in bypass surgical procedures in neurosurgery.
### DEVICE DESCRIPTION
The DIR 800 is an accessory to the Aeos Digital Surgical Microscope Class I 510(k) exempt surgical operating microscope. The DIR 800 allows the Aeos to produce excitation light to illuminate the fluorescence properties of the Indocyanine Green (ICG). The generated fluorescence signal depicts the distribution of the infrared dye in the patient's blood vessels during surgery.
{4}------------------------------------------------
## TECHNOLOGICAL CHARACTERISTICS (compared to Predicate(s))
As established in this submission, the DIR 800 is substantially equivalent to other predicate devices cleared by FDA. The subject device is shown to be substantially equivalent and has the same technological characteristics to its predicate device through comparison in function, design and intended use. The device characteristics comparing the DIR 800 to the predicate device are summarized below.
The subject and predicate device each utilizes a surgical stereo microscope imaging system and has the same basic functions for viewing, recording, and replaying fluorescent images. Similar cameras are utilized as sensors and the infrared options are equivalent. Each device utilizes the same type of fluorescent agent, Indocyanine Green (ICG), to visualize blood flow. Using near infared excitation, real-time visualization of images takes place in equivalent transmission ranges.
The indications for use for DIR 800 are similar to the indications for use for the predicate device cited in this file. A technological comparison and testing demonstrate that the DIR 800 is functionally equivalent to the predicate device. An evaluation performed on the DIR 800 supports the indications for use statement and demonstrates that the device is substantially equivalent to the predicate device and does not raise new questions regarding safety and effectiveness. Review of the comparison table below shows that the DIR 800 has similar characteristics compared to the predicate and supports the substantial equivalence determination.
| | Aesculap, Inc.<br>DIR 800 | Predicate<br>Carl Zeiss Meditec AG<br>INFRARED 800 with FLOW 800<br>option |
|-------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K# | Pending | K100468 |
| Indications | The DIR 800 is an accessory for the<br>Aesculap Aeos and is used in viewing<br>intra-operative blood flow in the<br>cerebral vascular area including, but<br>not limited to, assessing cerebral<br>aneurysm and vessel branch<br>occlusion, as well as patency in<br>neurosurgery. It also aids in the<br>visual assessment of intraoperative<br>blood flow and vessel patency in<br>bypass surgical procedures in<br>neurosurgery. | The Carl Zeiss Surgical INFRARED<br>800 with the FLOW 800 option is a<br>surgical microscope accessory used in<br>viewing and visual assessment of<br>intraoperative blood flow in the<br>cerebral vascular area including, but<br>not limited to, assessing cerebral<br>aneurysm and vessel branch<br>occlusion, as well as patency of very<br>small perforating vessels. It also aids<br>in the realtime visualization of blood<br>flow and visual assessment of vessel<br>types before and after Arteriovenous<br>Malformation (AVM) surgery.<br>Likewise, INFRARED 800 with the<br>FLOW 800 option used during<br>fluorescence guided surgery aids in<br>the visual assessment of<br>intraoperative blood flow as well as<br>vessel patency in bypass surgical<br>procedures in neurosurgery, plastic<br>and reconstructive procedures and<br>coronary artery bypass graft surgery. |
{5}------------------------------------------------
# K202391, Page 3 of 5
| Regulation number | 892.1600 | 892.1600 |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product code | IZI | IZI |
| Regulation Medical Specialty | Radiology | Radiology |
| Device Description | Accessory to surgical stereo<br>microscope for intraoperative<br>imaging. | Accessory to surgical stereo<br>microscope for intraoperative<br>imaging. |
| Basic System Function | View, record, and replay fluorescent<br>images of blood vessels in the<br>cerebral area, and view blood flow<br>and vessel patency in bypass surgical<br>procedures during neurosurgery. | View, record, and replay fluorescent<br>images of blood vessels in the<br>cerebral area, view bypass grafts<br>during coronary artery bypass<br>(CABG) surgery, and view blood<br>flow during plastic and reconstructive<br>surgery. |
| Accessory to | The DIR 800 is an accessory to the<br>Aesculap Aeos [Class I Exempt] | INFRARED 800 is an accessory to<br>the OPMI Pentero surgical<br>microscope. [Class I Exempt]<br>FLOW 800 is an accessory to<br>INFRARED 800. |
| Sensor | CMOS video camera, infrared-<br>sensitive | CCD video camera<br>Infrared-sensitive video camera<br>Black & white camera |
| Fluorescent Agent | Indocyanine Green (ICG) | Indocyanine Green (ICG) |
| Activation of the ICG option | The press of a single button on the<br>surgical microscope activates the ICG<br>function. | The press of a single button on the<br>surgical microscope activates the ICG<br>function. |
| Result | Fluorescent image of the distribution<br>of the infrared dye in the patient's<br>blood vessels during the operation. | Fluorescent image of the distribution<br>of the infrared dye in the patient's<br>blood vessels during the operation.<br>FLOW 800 provides overview and<br>comparison functions of display and a<br>summary of the dynamic of the<br>fluorescent dye. |
| Visualization of Real-time<br>Images | Yes | Yes |
| Display | Video and images are presented on<br>monitor. Picture-in-picture available. | Video and images are presented on<br>monitor. Image may also be displayed<br>in the eyepiece. Picture-in-picture<br>available. |
| Termination of the ICG<br>option | Determined by surgeon or will turn<br>off automatically after 100 seconds. | Determined by surgeon or will turn<br>off automatically after 5 minutes. |
| Light Source | 2x white light LEDs and 13x IR<br>LEDs | 2x 300W Xenon lamps |
| White Light Application | 400 nm - 700 nm | 400 nm - 700 nm |
| ICG Excitation | 740 nm - 800 nm bandpass filter | 700 nm - 780 nm bandpass filter |
| ICG Detection | Filter with high transmission from<br>820 nm - 900 nm | Filter with high transmission from<br>820 nm - 900 nm |
| Test Card | Pre-operative check test card | Pre-operative check test card |
| Distance of Imaging Head to<br>Patient | 200 mm - 300 mm | 200 mm - 500 mm |
| Camera adaptation | Integrated into the microscope head | Integrated into the microscope head |
| Zoom | Motorized 8.9:1 | Motorized 6:1 |
| Auto detection of fluorescence<br>influx | Yes | Yes |
| Autofocus | Yes | Yes |
| Control System | Personal Computer | Personal Computer |
| Storage | HDD | HDD, DVD |
{6}------------------------------------------------
## PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Bench testing
The following bench testing was performed to demonstrate that the DIR 800 performs as intended and is substantially equivalent to the predicate device.
| Test | Test Summary | Results |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|---------|
| Verification that the reference<br>target fluoresces | No fluorescence is detected by the<br>Aeos when DIR 800 mode is<br>disabled | Pass |
| Verification of the software /<br>hardware accessory visualization<br>feature | The excitation light bandwidth and<br>peak fall within 700nm-802nm<br>range at all working distances | Pass |
| | Total excitation absorption power<br>over the entire ICG excitation<br>spectrum | Pass |
| | Visualization of the Aeos DIR 800<br>mode | Pass |
| | Fluorescence pixel intensity shows<br>the optics and detectors adequately<br>visualize a fluorescent agent | Pass |
### Biocompatibility
None of the components come into contact with a patient; therefore, no biocompatibility data is required.
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Aeos System. The DIR 800 module was tested as part of the Aeos System. The system complies with the IEC 60601-1. and IEC 60601-2 standards for safety and the IEC 60601-1-2 standard for EMC.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
### Animal Study
Animal studies using a porcine model and a rabbit model were conducted to confirm the DIR 800 enabled viewing of intra-operative blood flow in the cerebrovascular area. Testing was completed by Neurosurgeons using test cases and independent scoring assessments which were predefined in the protocol. Comparative image sets of the same cerebrovascular anatomy in either a non-occluded (before clipping), occluded (clipped) and non-occluded (after removal of clip) were visualized by the DIR 800 and digital subtraction angiography (DSA).
{7}------------------------------------------------
Data was collected using Indocyanine Green (ICG) introduced into the vascular system of a porcine model and a rabbit elastase model. Evidence regarding the capability to visualize blood flow, vascular structures and aneurysms was collected and compared directly to data collected with a DSA on the same porcine and rabbit elastase model.
The device performance was evaluated for overall image quality, brightness and illumination, visual acuity, visual artifacts, depth perception, contrast and alignment of visible light images versus fluorescent images.
The testing confirmed that the DIR 800 reflects the intended use and provides intraoperative visualization and visual assessment of blood flow when compared to the DSA. The individual assessments of comparative images verify that the DIR 800 allowed visualization of intraoperative blood flow in the cerebrovascular area and imaging of aneurysms in small sized vessels in an equivalent manner to the DSA.
The individual assessments of comparative images to verify the device performance met their pre-established acceptance criteria.
### CONCLUSION
The subject device and predicate device have the same intended use, for viewing internal surgical sites during general surgical procedures. The specifications and functionality of the subject device is similar to the predicate device. Any differences between the devices do not raise new questions of safety and effectiveness. Performance data demonstrated that the device meets its specifications and intended uses and support the device. The hardware and software verification and validation demonstrate that the DIR 800 should perform as intended in the specified use conditions. Therefore, the subject device can be found substantially equivalent to the Carl Zeiss Surgical INFRARED 800 with the FLOW 800 option as cleared in K100468.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.