K202376 · Gi Supply, Inc. · PLL · Sep 30, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K202376
Device Name
EverLift Submucosal Lifting Agent
Applicant
Gi Supply, Inc.
Product Code
PLL · Gastroenterology, Urology
Decision Date
Sep 30, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
EverLift™ Submucosal Lifting Agent is indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early-stage cancers, or other gastrointestinal lesions prior to excision with a snare or other appropriate endoscopic device.
Device Story
EverLift is a sterile, injectable liquid composition supplied in a prefilled 10mL plastic syringe with a luer lock connection. It is used by endoscopists in hospital or clinic settings during endoscopic mucosal resection (EMR), endoscopic mucosal dissection (ESD), and polypectomy procedures. The agent is injected into the submucosal layer via a standard endoscopic injection needle inserted through the endoscope's working channel. The injection creates an in situ cushion, lifting the gastrointestinal mucosa from the submucosal layer to facilitate safe and easy resection of lesions in the esophagus, stomach, small intestine, sigmoid colon, and rectum. The device is for single-use only.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included syringe tip cap removal force, product color, injection flow rate, graduation marking tolerance (ISO 7886-1:2017), container closure integrity, tray lid peel force, sterilization validation (ISO 17665-1), LAL testing (USP <85>, USP <161>), distribution testing (ASTM D7386-16), and accelerated aging (ASTM F1980-16).
Technological Characteristics
Prefilled 10mL COC syringe; contains water, hydroxyethyl cellulose, glycerin, methylene blue, benzyl alcohol, sodium phosphate, and potassium phosphate. Sterilized by moist-heat (steam). Shelf life: 2 years. Single-use. Luer lock interface for standard endoscopic needles. Packaging: thermoformed tray with lid stock.
Indications for Use
Indicated for patients undergoing gastrointestinal endoscopic procedures requiring submucosal lift of polyps, adenomas, early-stage cancers, or other lesions prior to excision.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 30, 2020
Gi Supply, Inc. Erika Parry Manager, Quality and Regulatory 5069 Ritter Road Suite 104 Mechanicsburg, Pennsylvania 17055
Re: K202376
Trade/Device Name: EverLift Submucosal Lifting Agent Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: PLL Dated: September 1, 2020 Received: September 2, 2020
Dear Erika Parry:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202376
Device Name EverLift Submucosal Lifting Agent
Indications for Use (Describe)
EverLift Submucosal Lifting Agent is indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early-stage cancers, or other gastrointestinal lesions with a snare or other appropriate endoscopic device.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <path d="M2 2 L14 14 M2 14 L14 2" stroke="black" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over The Counter Use (21 CFR 801 Subpart C) </div> |
||X| Prescription Use (Part 21 CFR 801 Subpart D)
_ | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### I. SUBMITTER [Per 807.92(a)(1)]
Sponsor/Manufacturer GI Supply 5069 Ritter Road Suite 104 Mechanicsburg, PA 17055 USA Phone: (800)-451-5797
Contact Person Erika Parry Manager, Quality and Regulatory Phone: (717)-562-7580 Email: e.parry@gi-supply.com
### Date Prepared
August 19, 2020 Amended September 29, 2020
#### II. DEVICE [Per 807.92(a)(2)]
| Device Trade/Proprietary Name | EverLift™ Submucosal Lifting Agent |
|-------------------------------|------------------------------------|
| Device Common or Usual Name | Submucosal Injection Agent |
| Regulation Number | 21 CFR 876.1500 |
| Regulation Name | Endoscope and Accessories |
| Regulatory Class | Class II |
| Submission Type | Special 510(k) |
| Product Code | PLL |
| Classification Panel | Gastroenterology/Urology |
### III. PREDICATE DEVICE [Per 807.92(a)(3)]
The EverLift™ Submucosal Lifting Agent (10mL) [subject device] is substantially equivalent in terms of its intended use to the claimed predicate device, the 5mL EverLift™ Submucosal Lifting Agent (5mL) (K191923) with respect to device design, fundamental technology, physical characteristics, performance and intended use.
The device design and fundamental technology of the subject device are nearly identical to that of the predicate device. The formulation and composition of the primary packaging materials are identical. There are slight differences in the dimensions of the subject device and predicate device which do not impact safety or effectiveness. Both the subject and predicate devices are delivered sterile and are indicated for single-use only. The Sterilization Method and Shelf Life for the subject device are identical as that for the predicate device.
| Predicate<br>Device | The EverLift™ Submucosal Lifting Agent (10mL) [subject device] is substantially equivalent to<br>the following predicate device manufactured by GI Supply:<br>• EverLift™ Submucosal Lifting Agent (5mL) (K191923) |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{4}------------------------------------------------
# IV. DEVICE DESCRIPTION [Per 807.92(a)(4)]
The GI Supply EverLift™ Submucosal Lifting Agent (10mL) is a prefilled plastic syringe with attached plunger rod containing 10mL of lifting agent. The syringe has a luer lock connection capable of interfacing with a standard, commercially available endoscopic injection needle.
EverLift™ Submucosal Lifting Agent is an injectable liguid composition for use as a submucosal injection agent during endoscopic mucosal resection (EMR), endoscopic mucosal dissection (ESD), and polypectomy procedures in the gastrointestinal tract. The device is intended for use in endoscopic resection procedures in the upper and the lower gastrointestinal tract, including the esophagus, the stomach, the small intestine, the sigmoid colon, and the rectum, as a submucosal injectable agent during the removal of polyps, adenomas, early-stage cancers, and other pathological lesions by EMR, ESD, or polypectomy.
EverLift™ Submucosal Lifting Agent is injected into the submucosal layer by means of a standard, commercially available endoscopic injection needle, which is inserted into the working channel of the endoscope. The composition, when injected, creates a cushion in situ by lifting the gastrointestinal mucosa from the submucosal layer, allowing the endoscopist to perform an easy and safe resection procedure (EMR, ESD, or polypectomy).
# V. INTENDED USE / INDICATIONS FOR USE [Per 807.92(a)(5)]
### Intended Use / Indications for Use
EverLift™ Submucosal Lifting Agent is indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early-stage cancers, or other gastrointestinal lesions prior to excision with a snare or other appropriate endoscopic device.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE [Per 807.92(a)(6)]
The EverLift™ Submucosal Lifting Agent (10mL) [subject device] is substantially equivalent to the Gl Supply EverLift™ Submucosal Lifting Agent (5mL) (K191923) [predicate device] based on the same indication for use as the predicate device and the similar or identical functional and performance characteristics of the subject device when compared to the predicate device. The differences between the subject device and predicate device do not raise different issues of safety and effectiveness.
The detailed substantial equivalence comparison of the similarities and differences between the Gl Supply EverLift™ Submucosal Lifting Agent (10ml) [subject device] and the GI Supply EverLift™ Submucosal Lifting Agent (5mL) (K191923) [predicate device] is provided in the table below.
| Regulatory<br>Information | [Subject Device] | [Predicate Device] | Similarities /<br>Differences |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | GI Supply | GI Supply | Same |
| Device Trade<br>or<br>Proprietary<br>Name | EverLift™ Submucosal Lifting Agent<br>(10mL) | EverLift™ Submucosal Lifting Agent<br>(5mL) | --- |
| Regulatory<br>Information | [Subject Device] | [Predicate Device] | Similarities /<br>Differences |
| 510(k)<br>Number | K202376 | K191923 | --- |
| Device Class | Class II | Class II | Same |
| Device<br>Classification<br>Name | Endoscope and Accessories | Endoscope and Accessories | Same |
| Device<br>Common<br>Name | Submucosal Injection Agent | Submucosal Injection Agent | Same |
| Product Code | PLL | PLL | Same |
| Regulation<br>Number | 21 CFR 876.1500 | 21 CFR 876.1500 | Same |
| Design Features and Capabilities of the Device | | | |
| Indications<br>for Use | EverLift™ Submucosal Lifting Agent<br>is indicated for use in<br>gastrointestinal endoscopic<br>procedures for submucosal lift of<br>polyps, adenomas, early-stage<br>cancers, or other gastrointestinal<br>lesions prior to excision with a<br>snare or other appropriate<br>endoscopic device. | EverLift™ Submucosal Lifting Agent<br>is indicated for use in<br>gastrointestinal endoscopic<br>procedures for submucosal lift of<br>polyps, adenomas, early-stage<br>cancers, or other gastrointestinal<br>lesions prior to excision with a<br>snare or other appropriate<br>endoscopic device. | Same |
| Intended Use | Same as Indications for Use | Same as Indications for Use | Same |
| Prescription<br>or Over-the-<br>Counter<br>(OTC) Use | Prescription Use | Prescription Use | Same |
| Use<br>Environment | Hospital / Clinic | Hospital / Clinic | Same |
| Sterile | Sterile | Sterile | Same |
| Single-Use | Single-Use Only | Single-Use Only | Same |
| Design Features | | | |
| Packaging | Each individual 10mL COC syringe is<br>packed inside a thermoformed tray.<br>The tray is sealed with a labeled lid<br>stock with three (3) patient labels<br>affixed to the lid stock. Ten (10) of<br>the sealed trays are packed into a<br>shelf box with a copy of the IFU<br>placed between the trays and the<br>shelf box wall. The shelf box is<br>shipped in a regular die cut and<br>scored 200-pound B-flute kraft<br>shipper box sealed with a shipper<br>box label. | Each individual 5mL COC syringe is<br>packed along with three (3) patient<br>labels inside a propionate tube with<br>a friction fit press on closure. Ten<br>(10) tubes are packed into a shelf<br>box on top of a copy of the IFU. The<br>shelf box is shipped in a regular die<br>cut and scored 200-pound B-flute<br>kraft shipper box sealed with a<br>shipper box label. | Different. The<br>differences in<br>packaging do<br>not raise<br>different<br>questions of<br>safety or<br>effectiveness.<br>A Distribution<br>Study was<br>performed on<br>the new<br>packaging<br>configuration. |
| Regulatory<br>Information | [Subject Device] | [Predicate Device] | Similarities /<br>Differences |
| Images<br>of Device | Image: Subject Device | Image: Predicate Device | --- |
| Sterilization<br>Method | Sterilized by moist-heat (steam)<br>sterilization | Sterilized by moist-heat (steam)<br>sterilization | Same |
| Shelf Life | 2-Year Shelf Life | 2-Year Shelf Life | Same |
| Composition | Each steam sterilized syringe<br>contains 10mL of lifting agent with<br>the following ingredients:<br>• Water<br>• Hydroxyethyl Cellulose<br>• Glycerin<br>• Methylene Blue<br>• Benzyl Alcohol<br>• Sodium Phosphate<br>• Potassium phosphate | Each steam sterilized syringe<br>contains 5mL of lifting agent with<br>the following ingredients:<br>• Water<br>• Hydroxyethyl Cellulose<br>• Glycerin<br>• Methylene Blue<br>• Benzyl Alcohol<br>• Sodium Phosphate<br>• Potassium phosphate | Same<br>concentration;<br>however,<br>volume of<br>lifting agent<br>provided<br>differs. |
{5}------------------------------------------------
{6}------------------------------------------------
# VII. SUMMARY OF PERFORMANCE DATA AND PERFORMANCE TEST CONCLUSIONS [Per 807.92(b)(1)(2)(3)]
The determination of substantial equivalence is based on an assessment of non-clinical performance data. To verify that the device design meets its functional and performance requirements, the EverLift™ Submucosal Lifting Agent (10mL) underwent performance testing. Biocompatibility testing was performed only on the 5mL variant as it represented worst-case surface area to volume ratio.
The EverLift™ Submucosal Lifting Agent was developed under GI Supply's risk management process in accordance with ISO 14971:2012 - Medical Devices-Application of Risk Management to Medical Devices. The identified risks were adequately mitigated and verified by means of non-clinical performance testing.
### Summary of Performance Testing
A series of non-clinical performance testing was conducted on the subject device. Please refer to the table below for a summary of all non-clinical performance testing conducted on the subject device in support of substantial equivalence.
| Test Description | Conforming Standard(s) | Conclusion in Support of Substantial Equivalence |
|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Syringe Tip Cap<br>Removal Force<br>Testing | None | Syringe tip cap removal force testing was performed on<br>the subject device using the same method and<br>acceptance criteria as the predicate device. Results of<br>the testing confirmed that the syringe tip cap removal<br>force is equivalent between the subject and predicate<br>devices, thereby demonstrating substantial equivalence. |
| Test Description | Conforming Standard(s) | Conclusion in Support of Substantial Equivalence |
| Product Color<br>Testing | None | Product color testing was performed on the subject device using the same method and acceptance criteria as the predicate device. Results of the testing confirmed that product color is identical between the two devices, thereby demonstrating substantial equivalence. |
| Injection Flow<br>Rate Testing | None | Injection flow rate testing was performed on the subject device using the same method and acceptance criteria as the predicate device. Results of the testing confirmed that the injection flow rate is equivalent between the subject and predicate devices, thereby demonstrating substantial equivalence. |
| Graduation<br>Marking<br>Tolerance | ISO 7886-1:2017, Sterile<br>hypodermic needles for single<br>use – Part 1: Syringes for manual<br>use | Graduation mark tolerance testing was performed on the subject device using the same method as the predicate device, and acceptance criteria as defined in ISO 7886-1. Results of the testing confirmed that the marking tolerance meets the acceptance criteria defined in ISO 7886-1, thereby demonstrating substantial equivalence. |
| Container<br>Closure<br>Integrity Testing | None | Container closure integrity testing was performed on the subject device using the same method and acceptance criteria as the predicate device. Results of the testing confirmed that the container closure integrity is identical between the subject and predicate devices, thereby demonstrating substantial equivalence. |
| Tray Lid Peel<br>Force Testing | None | The secondary packaging of the subject device required a minor design change to accommodate the larger 10mL syringe. Tray lid peel force test results post-distribution simulation met acceptance criteria, confirming that the differences in secondary packaging do not raise different questions of safety or effectiveness. |
| Sterility<br>Assurance Level | ISO 17665-1, Sterilization of<br>healthcare products – Moist<br>heat – Part 1: Requirements for<br>the development, validation and<br>routine control of a sterilization<br>process for medical devices | A sterilization validation was performed on the subject device using the same sterilization process and sterilization parameters as the predicate device. Validation results confirmed that the Sterility Assurance Level (SAL) is identical for both devices (10-6), thereby demonstrating substantial equivalence. |
| LAL Testing | USP <85>, Bacterial Endotoxins<br><br>USP <161>, Transfusion and<br>Infusion Assemblies and Similar<br>Medical Devices | LAL testing was performed on the subject device using the same method and acceptance criteria as the predicate device. Results of the testing confirmed that the LAL level is identical between the subject and predicate devices, thereby demonstrating substantial equivalence. Additionally, LAL testing will be performed on each lot prior to final product release. |
| Distribution and<br>Post-<br>Distribution<br>Testing | D7386-16, ASTM Standard<br>Practice for Performance Testing<br>of Packages for Single Parcel<br>Delivery Systems | Post-distribution testing was performed on the subject device using the same method and acceptance criteria as the predicate device. Results of the testing confirmed that package integrity is identical between the subject and predicate devices, thereby demonstrating substantial equivalence. |
| Test Description | Conforming Standard(s) | Conclusion in Support of Substantial Equivalence |
| Accelerated and<br>Post-<br>Accelerated<br>Aging Studies | F1980-16, ASTM Standard Guide<br>for Accelerated Aging of Sterile<br>Barrier Systems for Medical<br>Devices | Post-accelerated aging testing was performed on the<br>subject device using the same method and acceptance<br>criteria as the predicate device. Results of the testing<br>was identical between the subject and predicate devices,<br>thereby demonstrating substantial equivalence. |
{7}------------------------------------------------
{8}------------------------------------------------
# VIII. CONCLUSIONS
Based on the results of the non-clinical performance testing conducted on the subject device, it has been concluded that the proposed EverLift™ Submucosal Lifting Agent is as safe and effective and performs as well as the legally marketed predicate device (K191923). The similar indications for use, intended use, technological characteristics, and performance characteristics for the proposed EverLift™ Submucosal Lifting Agent have been assessed to be substantially equivalent to the predicate device, and any differences do not raise different issues of safety and effectiveness when compared to the predicate device. Therefore, the EverLift™ Submucosal Lifting Agent is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.