K202375 · ZOLL Medical Corporation · MWI · Mar 9, 2021 · Cardiovascular
Device Facts
Record ID
K202375
Device Name
ZOLL Propaq M
Applicant
ZOLL Medical Corporation
Product Code
MWI · Cardiovascular
Decision Date
Mar 9, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Electrocardiogram (ECG) Monitoring The Propaq M system is indicated to monitor and/or record 3-, 5-, or 12-lead electrocardiogram (ECG) waveform and heart rate, and to alarm when heart rate is above or below limits set by the operator. ECG monitoring is indicated for patients from newborn (neonate) to adult, with and without heart dysfunction. Non-Invasive Blood Pressure Monitoring The Propaq M system is indicated for use to make non-invasive measurements of arterial pressure and to alarm if either parameter is outside of the limits set by the user. The non-invasive blood pressure is indicated for patients from newborn (neonate) to adult. Temperature Monitoring The Propaq M system is indicated for use to make continuous temperature measurements of rectal, or surface temperatures, and to alarm if the temperature is outside of the user. The temperature monitoring feature is indicated for use in patients from newborn (neonate) to adult. SpO2 Monitoring The Propaq M system is indicated for use for continuous nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and/or carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin (SpHb), oxygen content (SpOC), pleth variability index (PVI), and perfusion index (PI) via the pulse Co-oximeter and accessories. The pulse Co-oximeter and accessories are indicated for use on adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused, in hospitals, hospital-type facilities, or in mobile environments. Respiration Monitoring The Propaq M system is indicated for use to continuously monitor respiration rate and to alarm if the rate falls outside of the range set by the operator. Because the measurement method actually measures respiratory effort, apnea episodes with continued respiratory effort (such as obstructive apnea) may not be detected. It is not intended to be used as an apnea monitor. The respiration monitoring feature is indicated for use on patients from newborn (neonate) to adult. CO2 Monitoring The Propaq M system is indicated for use in continuous noninvasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and breath rate. The CO2 monitoring feature is indicated for use on patients from newborn (neonate) to adult. Invasive Pressure Monitoring The Propaq M system is indicated for use to display and make continuous invasive pressure measurements via a compatible pressure transducer. The invasive pressure is indicated for use on patients from newborn (neonate) to adult. TBI Dashboard The Propaq M system is indicated to provide graphical trend data for SpO2, Systolic BP (SBP) and EtCO2 as well as ventilation assistance relevant to the management of a TBI patient. 12-Lead Analysis The Propaq M system is indicated for use in acquiring, and reporting physiological data via 12-lead ECG Analysis, and to provide interpretation of the data for consideration by caregivers. The 12-lead ECG Analysis feature is indicated for use on adults (> 18 years of age). Web Console The Propaq M system is indicated for the remote display of physiological data displayed on connected Propaq M systems via the Web Console feature, including electrocardiogram (ECG), non-invasive blood pressure (NIBP), temperature, and heart rate.
Device Story
Portable patient monitor; provides continuous monitoring of ECG, SpO2, NIBP, IBP, CO2, temperature, respiration. Inputs: physiological sensors/transducers. Transforms inputs into real-time waveforms, numerics, and graphical trends. Output: LCD display of vital signs; TBI Dashboard for graphical trends of SBP, SpO2, EtCO2; ventilation assistance panel. Used in hospital, EMS, and military environments by trained medical personnel. Data transfer via USB or wireless to ZOLL server/RescueNet/ePCR. TBI Dashboard assists clinicians in monitoring TBI patients by grouping parameters and ventilation prompts. Device is a depopulated version of ZOLL X Series (defibrillator/pacer modules removed).
Clinical Evidence
No clinical data. Substantial equivalence supported by software verification, system-level validation, and regression testing on the X Series platform. Summative usability evaluations performed for the new user interface.
Technological Characteristics
Portable monitor; LCD display; rechargeable lithium-ion battery (SurePower II). Connectivity: USB, wireless. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62304, ISO 10993-1, ISO 14971. Software: v02.34. Depopulated X Series hardware (no defibrillator/pacer).
Indications for Use
Indicated for monitoring physiological parameters (ECG, NIBP, IBP, SpO2, CO2, temperature, respiration) in patients from neonate to adult. TBI Dashboard indicated for TBI patient management. 12-lead ECG analysis indicated for adults >18 years. Used in hospitals, EMS, and military environments.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 9, 2021
ZOLL Medical Corporation Elizabeth McMeniman Director, Regulatory Affairs 269 Mill Road Chelmsford, Massachusetts 01824
Re: K202375/S002
Trade/Device Name: ZOLL Propaq M Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm). Regulatory Class: Class II Product Code: MWI, DRT, DXN, DSK, CCK, DQA, DPS, FLL Dated: August 19, 2020 Received: August 20, 2020
Dear Elizabeth McMeniman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K202375
Device Name ZOLL® Propaq® M
### Indications for Use (Describe)
Electrocardiogram (ECG) Monitoring
The Propaq M system is indicated to monitor and/or record 3-, 5-, or 12-lead electrocardiogram (ECG) waveform and heart rate, and to alarm when heart rate is above or below limits set by the operator. ECG monitoring is indicated for patients from newborn (neonate) to adult, with and without heart dysfunction.
### Non-Invasive Blood Pressure Monitoring
The Propaq M system is indicated for use to make non-invasive measurements of arterial pressure and to alarm if either parameter is outside of the limits set by the user. The non-invasive blood pressure is indicated for patients from newborn (neonate) to adult.
### Temperature Monitoring
The Propaq M system is indicated for use to make continuous temperature measurements of rectal, or surface temperatures, and to alarm if the temperature is outside of the user. The temperature monitoring feature is indicated for use in patients from newborn (neonate) to adult.
### SpO2 Monitoring
The Propaq M system is indicated for use for continuous nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and/or carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin (SpHb), oxygen content (SpOC), pleth variability index (PVI), and perfusion index (PI) via the pulse Co-oximeter and accessories. The pulse Co-oximeter and accessories are indicated for use on adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused, in hospitals, hospital-type facilities, or in mobile environments.
Type of Use (Select one or both, as applicable)
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------|
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K202375
Device Name ZOLL® Propaq® M
### Indications for Use (Describe)
### Respiration Monitoring
The Propaq M system is indicated for use to continuously monitor respiration rate and to alarm if the rate falls outside of the range set by the operator. Because the measurement method actually measures respiratory effort, apnea episodes with continued respiratory effort (such as obstructive apnea) may not be detected. It is not intended to be used as an apnea monitor. The respiration monitoring feature is indicated for use on patients from newborn (neonate) to adult.
### CO2 Monitoring
The Propaq M system is indicated for use in continuous noninvasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and breath rate. The CO2 monitoring feature is indicated for use on patients from newborn (neonate) to adult.
### Invasive Pressure Monitoring
The Propaq M system is indicated for use to display and make continuous invasive pressure measurements via a compatible pressure transducer. The invasive pressure is indicated for use on patients from newborn (neonate) to adult.
### TBI Dashboard
The Propaq M system is indicated to provide graphical trend data for SpO2, Systolic BP (SBP) and EtCO2 as well as ventilation assistance relevant to the management of a TBI patient.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <svg height="15px" width="15px"> <rect fill="none" height="15px" stroke="black" width="15px" x="0" y="0"></rect> <line stroke="black" x1="2" x2="13" y1="2" y2="13"></line> <line stroke="black" x1="13" x2="2" y1="2" y2="13"></line> </svg> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <svg height="15px" width="15px"> <rect fill="none" height="15px" stroke="black" width="15px" x="0" y="0"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202375
Device Name ZOLL® Propaq® M
### Indications for Use (Describe)
#### 12-Lead Analysis
The Propaq M system is indicated for use in acquiring, and reporting physiological data via 12-lead ECG Analysis, and to provide interpretation of the data for consideration by caregivers. The 12-lead ECG Analysis feature is indicated for use on adults (> 18 years of age).
#### Web Console
The Propaq M system is indicated for the remote display of physiological data displayed on connected Propaq M systems via the Web Console feature, including electrocardiogram (ECG), non-invasive blood pressure (NIBP), temperature, and heart rate.
| Type of Use ( <i>Select one or both, as applicable</i> ) |
|----------------------------------------------------------|
|----------------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
ZOLL®
269 Mill Road Chelmsford, Massachusetts 01824-4105
978.421.9655 (main) 978.421.0025 (fax) www.zoll.com
An Asahi Kasei Group Company
# 510(k) Summary
# Sponsor Information:
ZOLL Medical Corporation 269 Mill Road Chelmsford, MA 01824
Contact Person: Elizabeth McMeniman Director, Regulatory Affairs Phone: (978) 569-6862 Fax: (978) 421-0010 Email: emcmeniman@zoll.com
### Date of Summary: September 1, 2020
# Device Name and Classification:
Device Name: ZOLL® Propaq® M Common Name: Portable Patient Monitor Regulatory Class: Class II Classification Name: Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) (21 CFR 870.2300)
Product Code: Monitor, Physiological, Patient (Without Arrhythmia Detection or Alarms) (MWI) Cardiac Monitors - including Cardiotachometer and Rate Alarms (DRT) Noninvasive Blood Pressure Measurement System (DXN) Blood Pressure Computer (DSK) Carbon Dioxide Gas Analyzer (CCK) Oximeter (DQA) Electrocardiograph (DPS) Clinical electronic thermometer (FLL)
{6}------------------------------------------------
### Predicate Devices:
Trade Name: ZOLL® Propaq® M Regulation Number: 21 CFR 870.2300 Regulatory Class: Class II Product Code: MWI, DRT, DXN, DSK, CCK, DQA, DPS, FLL Identification Predicate Device: ZOLL® Propaq® M, K180482
Trade Name: ZOLL® X Series® Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III Product Code: MKJ, LIX, LDD, DRT, DRO, DXN, DSK, CCK, DQA Identification Predicate Device: ZOLL® X Series®, K141774 (K141774 introduced the TBI Dashboard)
# Device Description
The proposed ZOLL® Propag® M is a portable monitor designed for use by trained medical personnel who are familiar with basic monitoring and vital signs assessment. In the currently cleared configuration, the ZOLL® Propaq® M provides monitoring capabilities for ECG, Pulse CO-Oximetry (SpO2), Non-Invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), End Tidal CO2, Temperature, and Respiration. The device is powered by an auxiliary power supply or an easily replaced battery pack that is quickly recharged in the device when it is connected to the auxiliary power supply.
The product is designed for use in hospital. EMS, and rugged military environments. The unit has a large colorful LCD display of numerics and waveform data that provides easy visibility from across the room and at any angle. ECG, plethysmograph, and respiration waveform traces can be displayed simultaneously, providing easy access to all patient monitoring data at once. The display screen is configurable to allow the best visual layout according to monitoring needs.
The Propaq® M has a patient data review and collection system that allows the user to view, store, and transfer patient data. The Propaq® M unit contains a USB port, which the user can use to transfer data to a PC and, optionally, a printer, which can be used to print patient data.
The Propaq® M unit can send data through a wireless connections. The unit can send 12lead report snapshots (including trend data) or disclosure logs to a recipient via a ZOLL® server. In addition, full disclosure cases, which also contain trend data, can be automatically retrieved from the Propaq® M unit using ZOLL® RescueNet® or ePCR software.
Propaq® M models use an easily replaced rechargeable lithium-ion battery pack (the SurePower II Battery Pack). A new, fully charged battery pack typically delivers more than 8 hours of ECG monitoring. Use of other functions (such as higher screen brightness or shorter NIBP intervals) reduces this time
A clinical dashboard for monitoring patients with suspected traumatic brain injury (TBI) has been added as part of Software Version 02.34 of the proposed Propaq® M device. The TBI Dashboard was initially cleared for ZOLL® X Series® (K141774) to be displayed on a separate mobile device that connects to the main device (ZOLL® X Series®) via Bluetooth. The proposed feature is to display the TBI Dashboard directly on the Propag® M device via Software 02.34. The TBI Dashboard presents the user with graphical trend plots of 3 different parameters - (1) 15 minutes of systolic blood pressure (SBP), (2) 3
{7}------------------------------------------------
minutes of oxygen saturation of arterial hemoglobin (SpO2), and (3) 3 minutes of carbon dioxide concentration of the expired breath (EtCO2). The proposed ZOLL® Propaq® M device already provides this information in tabular format and in the live parameter boxes. The TBI Dashboard groups the information on one screen in a graphical format. The trend graphs display supervisor-configurable limits to allow the user to monitor patient parameters based on their local protocols. These protocol limits only appear in Adult mode. There is also a ventilation panel available on the TBI Dashboard to assist the user with manual ventilations. The TBI Dashboard allows the user to modify the target ventilation rate, the scale of each graphical trend, and to turn the ventilation timer and prompt on or off.
The Indications for use section in the Propaq® M Operator's Guide was updated to include the following statement for introducing of the TBI Dashboard:
### TBI Dashboard
The Propaq® M system is indicated to provide graphical trend data for SpO2, Systolic BP (SBP) and EtCO2 as well as ventilation assistance relevant to the management of a TBI patient.
The lower half of Figure 1, identified by the red box, depicts the TBI Dashboard on the host device's screen and includes the SBP (teal), SpO2 (yellow), and EtCO2 (purple) trend graphs.
Image /page/7/Figure/6 description: The image shows a medical monitor displaying various vital signs and readings. The heart rate is 80 bpm, NIBP is 117/80 mmHg, EtCO2 is 30 mmHg, SpO2 is 97%, and the temperature is 72.3 *F. There is also a TBI section showing trends for SBP, EtCO2, and SpO2 over time.
Figure 1: TBI Dashboard displayed on front panel user interface
Software Version 02.34 introduces automatic retries of failed 12-Lead report transmissions. While legacy Software of the Propaq® M attempts transmission of the 12-Lead report only once, the new proposed Software Version 02.34 attempts re-transmission up to five times if the initial attempt fails. This feature is a non-clinical enhancement since it affects only an external communication feature of the device with no impact to the acquiring, analyzing, or reporting process of ECG data.
The Propaq® M device is an alternate configuration of the ZOLL® X Series® device that depopulates the defibrillator/pacer module circuitry. Removal of this circuitry also removes CPR Monitoring capability from the device as the sensor that enables the CPR feature resides in the defibrillation electrodes. Physical
{8}------------------------------------------------
changes to the proposed Propaq® M device are limited to removing the defibrillator and pacer modules, removing the defibrillator and pacer controls from the front keypad, and changing the color scheme of the device. The remaining features and functions of the proposed Propaq® M device are identical to the existing ZOLL® X Series® model. In order to simulate the functioning of the Propag® M device, a regression testing was conducted on the X Series device by disabling the defibrillator/pacer module, which ensures the effective functioning of the proposed Propag® M device. The TBI Dashboard was introduced and cleared for the ZOLL® X Series® (K141774) to be displayed on a separate mobile device that connects to the main device (ZOLL® X Series®) via Bluetooth. The new Software Version 02.34 (TBI Dashboard installed on device) was introduced to ZOLL® X Series® and submitted to the agency under P160022/S013 on August 28, 2019.
The ZOLL® Propag® M is a depopulated configuration of the ZOLL® X Series® device. The defibrillator/pacer module (circuit boards) in the Propaq® M software configuration is removed as Propaq® M is a monitoring device. The TBI Dashboard was introduced and cleared for the ZOLL® X Series® (K141774) to be displayed on a separate mobile device that connects to the main device via Bluetooth. The new Software Version 02.34 (TBI Dashboard installed on device) was introduced to ZOLL® X Series® and submitted to the agency under P160022/S013 on August 28, 2019.
# Indications for Use
### Electrocardiogram (ECG) Monitoring
The Propaq® M system is indicated to monitor and/or record 3-, 5-, or 12-lead electrocardiogram (ECG) waveform and heart rate, and to alarm when heart rate is above or below limits set by the operator. ECG monitoring is indicated for patients from newborn (neonate) to adult, with and without heart dysfunction.
### Non-Invasive Blood Pressure Monitoring
The Propaq® M system is indicated for use to make non-invasive measurements of arterial pressure and heart rate, and to alarm if either parameter is outside of the limits set by the user. The non-invasive blood pressure monitoring feature is indicated for patients from newborn (neonate) to adult.
### Temperature Monitoring
The Propaq® M system is indicated for use to make continuous temperature measurements of rectal, esophageal, or surface temperatures, and to alarm if the temperature is outside of the limits set by the user. The temperature monitoring feature is indicated for use in patients from newborn (neonate) to adult.
### SpO2 Monitoring
The Propaq® M system is indicated for use for continuous nonitoring of functional oxygen saturation of arterial hemoglobin (SpOz), pulse rate, and/or carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin (SpHb), oxygen content (SpOC), pleth variability index (PVI), and perfusion index (PI) via the pulse Co-oximeter and accessories. The pulse Co-oximeter and accessories are indicated for use on adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused, in hospital-type facilities, or in mobile environments.
### Respiration Monitoring
The Propaq® M system is indicated for use to continuously monitor respiration rate and to alarm if the rate falls outside of the range set by the operator. Because the measurement method actually measures respiratory effort, apnea episodes with continued respiratory effort (such as obstructive apnea) may not be
{9}------------------------------------------------
detected. It is not intended to be used as an apnea monitor. The respiration monitoring feature is indicated for use on patients from newborn (neonate) to adult.
### CO2 Monitoring
The Propaq® M system is indicated for use in continuous noninvasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath rate. The CO2 monitoring feature is indicated for use on patients from newborn (neonate) to adult.
### TBI Dashboard
The Propaq® M system is indicated to provide graphical trend data for SpO2, Systolic BP (SBP) and EtCO2 as well as ventilation assistance relevant to the management of a TBI patient.
### Invasive Pressure Monitoring
The Propaq® M system is indicated for use to display and make continuous invasive pressure measurements via a compatible pressure transducer. The invasive pressure is indicated for use on patients from newborn (neonate) to adult.
### 12-Lead Analysis
The Propaq® M system is indicated for use in acquiring and reporting physiological data via 12lead ECG Analysis, and to provide interpretation of the data for consideration by caregivers. The 12-lead ECG Analysis feature is indicated for use on adults (> 18 years of age).
### Web Console
The Propaq® M system is indicated for the remote display of physiological data displayed on connected Propaq® M systems via the Web Console feature, including electrocardiogram (ECG), non-invasive blood pressure (NIBP), temperature, and heart rate.
# Comparison of Technological Characteristics:
The principles of operation are identical between the predicate ZOLL® Propaq® M (K180482) and subject device ZOLL® Propaq® M. The subject device is an easy-to-use portable monitor that has the following monitoring capabilities: ECG, CO-Oximeter, Non-invasive Blood Pressure, IBP, CO2 SpO2 Temperature, and Respiration. The technological difference between the ZOLL® Propaq® M subject device and the predicate device ZOLL® Propag® M (K180482) is the introduction of Software 02.34/ TBI Dashboard. Therefore, the ZOLL® X Series® (K141774) will be referenced as a predicate for the TBI Dashboard.
# Substantial Equivalence - Non-Clinical Evidence:
The following performance data were provided in support of substantial equivalence determination:
### Software Verification and Validation
The Propaq® M device is an alternate configuration of the ZOLL® X Series® device that depopulates the defibrillator/pacer module circuitry. Removal of this circuitry also removes CPR Monitoring capability from the device as the sensor that enables the CPR feature resides in the defibrillation electrodes. Physical changes to the proposed Propaq® M device are limited to removing the defibrillator and pacer modules, removing the defibrillator and pacer controls from the front keypad, and changing the color scheme of the device. The remaining features and functions of the proposed Propag® M device are identical to the existing ZOLL® X Series® model. In order to simulate the functioning of the Propag® M device, a regression testing
{10}------------------------------------------------
was conducted on the X Series device by disabling the defibrillator/pacer module, which ensures the effective functioning of the proposed Propaq® M device.
As X Series® encompasses a larger set of functionalities than the Propaq® M device, Software verification and validation testing were conducted for the ZOLL® X Series® (P160022) and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submission for Software Contained Medical Devices". Thus, the verification and validation relevant to X Series® device is also applicable to proposed Propag® M device.
Performance testing in the form of the software verification and system level validation for the ZOLL® X Series® (P160022) ensured that the Propaq® M performs as well as the indicated predicate devices and meets all of its functional requirements and performance specifications.
### Safety testing per the International Recognized Standards
The device was evaluated and found to be in compliance with the standards identified in Table 1.
| Table 1: Propaq® M Standards | |
|------------------------------|--------------------------------------|
| FDA Consensus # | Standard Designation |
| 19-4 | IEC 60601-1: 2005/A1:2012/COR 1 2014 |
| 19-8 | IEC 60601-1-2: 2014 |
| 5-89 | IEC 60601-1-6: 2013 |
| 5-76 | IEC 60601-1-8:2006/A1:2012 |
| 3-105 | IEC 60601-2-25: 2011 |
| 3-126 | IEC 60601-2-27: 2011 |
| 3-115 | IEC 60601-2-34: 2011 |
| N/A | IEC 60601-2-49: 2011 |
| 13-79 | IEC 62304: 2006 AMD 1: 2015 |
| 5-114 | IEC 62366-1:2015 COR 1 2016 |
| 3-152 | ISO 80601-2-30: 2009 |
| N/A | ISO 80601-2-55: 2011/AC: 2012 |
| N/A | ISO 80601-2-56: 2009 |
| N/A | ISO 80601-2-61: 2011 |
| 2-220 | EN ISO 10993-1: 2009 AC: 2010 |
| 5-117 | EN ISO 15223-1: 2016 COR 2017 |
| N/A | BS EN 1041: 2008+A1:2013 |
| 5-40 | EN ISO 14971: 2012 |
# Table 1: Propaq® M Standards
Safety testing ensures that the device complies with applicable sections of recognized industry and safety standards. No new safety testing was conducted with the introduction of the changes proposed in this application.
### Electrical Safety and electromagnetic compatibility (EMC)
At this time ZOLL® proposes to revise the Propaq® M device software to introduce TBI Dashboard and 12-Lead re-transmission. As the proposed modification involves a software-only change, the Electromagnetic Compatibility and Electrical Safety aspect of the Propag® M device has not been impacted by the proposed change.
{11}------------------------------------------------
### Usability Testing
Summative usability evaluations were performed for the ZOLL® X Series® to validate the usability of the new Software revision 02.34 user interface. All identified use-related hazards have been found to be mitigated to acceptable levels.
The 12-Lead transmission retry enhancement does not change the user interface or the user interaction with the device. The process of capturing a 12-Lead, selecting the 12-Lead to send, selecting the transmission destination and sending the file will remain the same. If the transmission fails due to a connection issue, the device will attempt to automatically re-try the transmission up to 5 times. After 5 retries the user would revert to the previous method of manually transmitting the user would need to manually transmit the case file upon each transmission failure. Therefore, usability testing is not required for the 12-Lead transmission feature.
# Substantial Equivalence - Clinical Evidence:
Clinical evidence was not necessary to show substantial equivalence. However, relevant clinical literature is provided in Section 20. Performance Testing - Clinical.
# Conclusion:
The information provided in this 510(k) demonstrates that the proposed Propaq® M is substantially equivalent to its predicate ZOLL® Propaq® (K180482) and ZOLL® X Series® (K141774) for TBI Dashboard for the new introduced Software Release 02.34.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.