← Product Code [PGY](/productcode/PGY) · K202374

# 4MP/8MP Color LCD Monitors C44W+/C82W+, C83W+, C85W+ (K202374)

_Shenzhen Beacon Display Technology Co., Ltd. · PGY · Oct 19, 2020 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K202374

## Device Facts

- **Applicant:** Shenzhen Beacon Display Technology Co., Ltd.
- **Product Code:** [PGY](/productcode/PGY.md)
- **Decision Date:** Oct 19, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

These products are intended to be used in displaying and viewing digital images for review and analysis by trained medical practitioners. They do not support the display of mammography images for diagnosis.

## Device Story

TFT color LCD monitors (30", 31.5", 27") for displaying high-definition medical images (Ultrasound, General Radiography). Inputs via DVI-D, DisplayPort, VGA. Features 14-bit image processing, brightness stability control, and multi-display modes. Used in clinical settings by physicians/radiologists for image review. Output is visual display of medical images; supports DICOM 3.14 calibration. Benefits include consistent, high-precision image viewing for diagnostic and interventional radiology applications.

## Clinical Evidence

Bench testing only. No clinical or animal data provided. Performance verified via MTF, DICOM GSDF conformance, luminance/chromaticity non-uniformity, temporal response, and pixel defect analysis per AAPM TG18 guidelines.

## Technological Characteristics

TFT color LCD panels (IPS) with LED backlighting. Resolutions: 2560x1600 (30") or 3840x2160 (27", 31.5"). Connectivity: DVI-D, DisplayPort, VGA. Features: 14-bit image processing, brightness stabilization sensors, DICOM 3.14 calibration support. Power: DC 24V. Operating temp: 0-40°C. Software: Beacon Monitor Manage.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- RadiForce RX440 ([K130070](/device/K130070.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Shenzhen Beacon Display Technology Co., Ltd. % Fu Ailing Document Engineer 12F, Block B1, Nanshan Zhiyuan, No.1001 Xueyuan Road Shenzhen, Guangdong 518055 CHINA

Re: K202374

Trade/Device Name: 4MP/8MP Color LCD Monitors C44W+/C82W+, C83W+, C85W+ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: PGY Dated: August 15, 2020 Received: August 19, 2020

October 19, 2020

Dear Fu Ailing:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

#### K202374

Device Name

4MP/8MP Color LCD Monitors C44W+/C82W+, C83W+, C85W+

#### Indications for Use (Describe)

These products are intended to be used in displaying digital images for review and analysis by trained medical practitioners. They do not support the display of mammography images for diagnosis.

| Type of Use (Select one or both, as applicable)                                                                                           |                                                                                                                                          |
|-------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="margin-right:5px;">☑</span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="margin-right:5px;">☐</span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K202374

# 510(k) Summary

[As required by 21 CFR 807.92]

## 1. Date Prepared [21 CFR807.92 (a) (1)]

July 1, 2020

## 2. Submitter's Information [21 CFR807.92 (a) (1)]

| Name of Sponsor: | Shenzhen Beacon Display Technology Co., Ltd.                                                      |
|------------------|---------------------------------------------------------------------------------------------------|
| Address:         | 12F, Block B1, Nanshan Zhiyuan, No.1001 Xueyuan Road, Nanshan<br>District, Shenzhen, 518055 China |
| Contact Name:    | Fu Ailing                                                                                         |
| Telephone No.:   | +86-0755-36868363-5150                                                                            |
| Fax No.:         | +86-0755-26995755                                                                                 |
| Email Address:   | fuailing@beacon-display.cn                                                                        |

## 3. Trade Name, Common Name, Classification [21 CFR807.92 (a) (2)]

Trade Name/Model: 4MP/8MP Color LCD Monitors C44W+/C82W+, C83W+, C85W+ Common Name: LCD Monitors C44W+, C82W+, C83W+, C85W+ Classification Name: Picture archiving and communications system Regulation Number: 21 CFR 892.2050 Product code: PGY Classification Panel: Radiology Device Class: II

## 4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]

The identified predicate within this submission is as follows: Eizo Nanao Corporation, RadiForce RX440 has been cleared by FDA through 510(k) No. 

{4}------------------------------------------------

K130070 (Decision Date - February 08, 2013).

### 5. Description of the Device [21 CFR 807.92(a) (4)]

C44W+ is a 30-inch TFT color LCD monitor. It is specifically designed to display high quality, high definition images required for Ultrasound and General Radiography. With built-in multi-display modes, it facilitates convenient viewing by doctors of various images, data, and layout of information. The product is designed to be able to meet DICOM 3.14 calibration standards. With features such as brightness stability control. 14 bit image processing technology, and interfaces with external calibration software, it ensures optimal viewing conditions over the life of the display. The specially designed base supports height and tilt adjustments for erqonomic comfort.

C82W+ is a 31.5-inch TFT color LCD monitor. It is specifically designed to display high quality, high definition images required for diagnostic, interventional radiology and other medical applications. With built-in multi-display modes, it facilitates convenient viewing by doctors of various images, data, and layout of information. The product is designed to be able to meet DICOM 3.14 calibration standards. With features such as brightness stability control, 14 bit image processing technology, and interfaces with external calibration software, it ensures optimal viewing conditions over the life of the display. The specially designed base supports height and tilt adjustments for ergonomic comfort.

C83W+ and C85W+ are 27-inch TFT color LCD monitors. They are specifically designed to provide the high definition images output for general Radiography. The products have been strictly calibrated so they meet DICOM Part 3.14 and other standards. They use the latest generation of LED backlight panel, supporting resolution 3840 x 2160. With built-in brightness stabilization control circuit, make sure the brightness of this monitors are stable in their life, and the products meet the demand of high precision medical imaging.

For C44W+, C82W+, C83W+, C85W+, the only difference is the screen. C83W+ and C85W+ have the same screen, but their customers are different, which results in the two models.

{5}------------------------------------------------

## 6. Intended Use [21 CFR 807.92(a)(5)]

These products are intended to be used in displaying and viewing digital images for review and analysis by trained medical practitioners. They do not support the display of mammography images for diagnosis.

## 7. Technological Characteristics [21 CFR 807.92(a)(6)]

| Panel                              | 30", TFT, color, LCD screen, AG type, 3H hard coating                             |
|------------------------------------|-----------------------------------------------------------------------------------|
| Brightness (Typ.)                  | 700 cd/m²                                                                         |
| CR (Typ.)                          | 1000:1                                                                            |
| Viewing angle                      | R/L 178°, U/D 178° Typ. (CR > 10)                                                 |
| Pixel Pitch                        | (H) 0. 2505 × (V) 0.2505 mm                                                       |
| Native resolution                  | 2,560 x 1,600                                                                     |
| Display area                       | 641.28 mm (H) x 400.8 mm (V)                                                      |
| Aspect ratio                       | 16:10                                                                             |
| Screen size                        | 30" real diagonal                                                                 |
| Power                              | DC 24 V/9 A                                                                       |
| Power consumption                  | 80 W (max.)<br>< 5 W (standby)                                                    |
| Refresh rate                       | 60 Hz                                                                             |
| Backlighting                       | LED                                                                               |
| Response time (Gray to gray)       | 6 ms (typ.)                                                                       |
| Color support                      | 1.073G                                                                            |
| Input signals                      | DVI-D, Display Port, VGA                                                          |
| Dimension (W x H x D)              | 699.3 x 513~633 x 251.4 mm (without packing)<br>864 x 600 x 334 mm (with packing) |
| Weight                             | 13 ± 0.5 kg (wet)<br>15 ± 0.5 kg (gross)                                          |
| Operating temperature and humidity | Temperature: 0°C ~ 40°C<br>Humidity: 20% ~ 80%                                    |
| Storage temperature and humidity   | Temperature: -20°C ~ 60°C<br>Humidity:10% ~ 90%                                   |

### C44W+ Color LCD monitor

{6}------------------------------------------------

| Panel                                    | 31.5", TFT, color, LCD screen, AG type                                        |
|------------------------------------------|-------------------------------------------------------------------------------|
| Brightness (Typ.)                        | 1000 cd/m²                                                                    |
| CR (Typ.)                                | 1300:1                                                                        |
| Viewing angle                            | R/L 178°, U/D 178° Typ. (CR > 10)                                             |
| Pixel Pitch                              | (H) 0. 182 × (V) 0.182 mm                                                     |
| Native resolution                        | 3,840 x 2,160                                                                 |
| Display area                             | 697.31 mm (H) x 392.23 mm (V)                                                 |
| Aspect ratio                             | 16:9                                                                          |
| Screen size                              | 31.5" real diagonal                                                           |
| Power                                    | DC 24 V/9.0 A                                                                 |
| Power consumption                        | 160 W (max.)<br>< 5 W (standby)                                               |
| Refresh rate                             | 60 Hz                                                                         |
| Backlighting                             | LED                                                                           |
| Response time (Gray<br>to Gray)          | 14 ms (typ.)                                                                  |
| Color support                            | 1.07B                                                                         |
| Input signals                            | DVI-D, Display Port, VGA                                                      |
| Dimension (W x H x D)                    | 753 x 503~583 x 251 mm (without packing)<br>850 x 580 x 320 mm (with packing) |
| Weight                                   | 18.5 kg ± 0.5 kg (wet)<br>23.5 ± 0.5 kg (gross)                               |
| Operating<br>temperature and<br>humidity | Temperature: 0°C ~ 40°C<br>Humidity: 20% ~ 80%                                |

#### C82W+ Color LCD monitor

### C83W+/C85W+ Color LCD monitor

| Panel             | 27", TFT, color, LCD screen, AG type, 3H hard coating |
|-------------------|-------------------------------------------------------|
| Brightness (Typ.) | 800 cd/m²                                             |
| CR (Typ.)         | 1000:1                                                |
| Viewing angle     | R/L 178°, U/D 178° Typ. (CR > 10)                     |
| Pixel Pitch       | (H) 0. 1553 × (V) 0.1553 mm                           |

{7}------------------------------------------------

| Native resolution                        | 3,840 x 2,160                                                                   |
|------------------------------------------|---------------------------------------------------------------------------------|
| Display area                             | 596.16 mm (H) x 335.34 mm (V)                                                   |
| Aspect ratio                             | 16:9                                                                            |
| Screen size                              | 27" real diagonal                                                               |
| Power                                    | DC 24 V/5.0 A                                                                   |
| Power consumption                        | 100 W (max.)<br>< 5 W (standby)                                                 |
| Refresh rate                             | 60 Hz                                                                           |
| Backlighting                             | LED                                                                             |
| Response time ( TR + TF )                | 16 ms (typ.)                                                                    |
| Color support                            | 1.07B                                                                           |
| Input signals                            | DVI-D, Display Port, VGA                                                        |
| Dimension (W x H x D)                    | 651 x 463~583 x 251.4 mm (without packing)<br>764 x 556 x 344 mm (with packing) |
| Weight                                   | 10 ± 0.5 kg (wet)<br>13 ± 0.5 kg (gross)                                        |
| Operating<br>temperature and<br>humidity | Temperature: 0°C ~ 40°C<br>Humidity: 20% ~ 80%                                  |
| Storage temperature<br>and humidity      | Temperature: -20°C ~ 60°C<br>Humidity:10% ~ 90%                                 |

{8}------------------------------------------------

## 8. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92]

#### 8.1 Intended uses:

| ID | Comparison<br>Item | Proposed Devices<br>C44W+, C82W+, C83W+,<br>C85W+                                                                                                                                                                                   | Predicate Device<br>RadiForce RX440                                                                                                                                                                                             |
|----|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1  | Intended<br>Use    | These products are intended to<br>be used in displaying and<br>viewing digital images for review<br>and analysis by trained medical<br>practitioners. They do not<br>support the display of<br>mammography images for<br>diagnosis. | The product is intended to be<br>used in displaying and viewing<br>digital images for review and<br>analysis by trained medical<br>practitioners. They do not<br>support the display of<br>mammography images for<br>diagnosis. |

Table 1 Intended Use Comparison of C44W+, C82W+, C83W+, C85W+

### 8.2 Comparison table

{9}------------------------------------------------

D Monitors C44W+, C82W+, C83W+, C83W+, C85W

510(k) Submission

| Table 2 | General Comparison of |
|---------|-----------------------|
| C44W+,  | C82W+, C83W+, C85W+   |

| ID   | Comparison<br>Item                          | Proposed Devices                                        |                                                         |                                                         |                                                         | Predicate<br>Device                                                         | Explanation of<br>Differences                                                                                                                                                                                                           |                            |
|------|---------------------------------------------|---------------------------------------------------------|---------------------------------------------------------|---------------------------------------------------------|---------------------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
|      |                                             | C44W+                                                   | C82W+                                                   | C83W+                                                   | C85W+                                                   | RadiForce RX440                                                             |                                                                                                                                                                                                                                         |                            |
| 2    |                                             | Display Performance/Specifications                      |                                                         |                                                         |                                                         |                                                                             |                                                                                                                                                                                                                                         |                            |
| 2.1  | Screen<br>technology                        | TFT Color LCD<br>Panel (IPS)                            | TFT Color LCD<br>Panel (IPS)                            | TFT Color LCD<br>Panel (IPS)                            | TFT Color LCD<br>Panel (IPS)                            | TFT Color LCD<br>Panel (IPS)                                                |                                                                                                                                                                                                                                         |                            |
| 2.2  | Viewing angle<br>(H, V)                     | H: 178°, V:178°<br>@CR>10                               | H: 178°, V:178°<br>@CR>10                               | H: 178°, V:178°<br>@CR>10                               | H: 178°, V:178°<br>@CR>10                               | H: 176°, V:176°<br>@CR=10                                                   | Different screen<br>provided by the<br>different<br>manufacturer                                                                                                                                                                        |                            |
| 2.3  | Active screen<br>size                       | 641.28 mm x<br>400.8 mm                                 | 697.31 mm x<br>392.23 mm                                | 596.16 mm x<br>335.34 mm                                | 596.16 mm x<br>335.34 mm                                | 641.28 mm x<br>400.8 mm                                                     | Different screen<br>provided by the<br>different<br>manufacturer                                                                                                                                                                        |                            |
| 2.4  | Resolution                                  | 4MP (2,560 x<br>1,600)                                  | 8MP (3, 840 x<br>2,160)                                 | 8MP (3, 840 x<br>2,160)                                 | 8MP (3, 840 x<br>2,160)                                 | 4MP (2,560 x<br>1,600)                                                      | Different screen<br>provided by the<br>different<br>manufacturer                                                                                                                                                                        |                            |
|      |                                             |                                                         |                                                         |                                                         |                                                         |                                                                             |                                                                                                                                                                                                                                         |                            |
| 2.5  | Aspect ratio                                | 16:10                                                   | 16:9                                                    | 16:9                                                    | 16:9                                                    | 16:10                                                                       | Different screen provided by the different manufacturer                                                                                                                                                                                 |                            |
| 2.6  | Pixel pitch                                 | 0.2505 mm x<br>0.2505 mm                                | 0.182 mm x<br>0.182 mm                                  | 0.1553 mm x<br>0.1553 mm                                | 0.1553 mm x<br>0.1553 mm                                | 0.2505 mm x<br>0.2505 mm                                                    | Different screen provided by the different manufacturer                                                                                                                                                                                 |                            |
| 2.7  | Maximum luminance                           | 700 cd/m²                                               | 1000 cd/m²                                              | 800 cd/m²                                               | 800 cd/m²                                               | 750 cd/m²                                                                   | Different screen provided by the different manufacturer                                                                                                                                                                                 |                            |
| 2.8  | DICOM calibrated luminance                  | 500 cd/m²                                               | 500 cd/m²                                               | 500 cd/m²                                               | 500 cd/m²                                               | 400 cd/m²                                                                   | Different screen provided by the different manufacturer                                                                                                                                                                                 |                            |
| 2.9  | Contrast ratio                              | 1000:1                                                  | 1300:1                                                  | 1000:1                                                  | 1000:1                                                  | 1100:1                                                                      | Different screen provided by the different manufacturer                                                                                                                                                                                 |                            |
| 2.10 | Backlighting                                | LED                                                     | LED                                                     | LED                                                     | LED                                                     | LED                                                                         | -                                                                                                                                                                                                                                       |                            |
| 2.11 | Display Colors                              | 10-bit,<br>1,073,741,824<br>color                       | 1.07 Billion<br>colors                                  | 1.07 Billion<br>colors, 10Bit<br>(8Bit + A-FRC)         | 1.07 Billion<br>colors, 10Bit<br>(8Bit + A-FRC)         | 8-bit colors:<br>16.77 million<br>from a palette of<br>68 billion colors    | Tone between<br>the predicate<br>device and our<br>proposed devices<br>are different. But<br>they pass the<br>exams in<br>AAPM-TG18 4.3<br>"Luminance<br>Response".<br>Therefore, they<br>are equivalent to<br>the predicate<br>device. |                            |
| 2.12 | Luminance<br>non-uniformity<br>compensation | -                                                       | -                                                       | -                                                       | -                                                       | -                                                                           | Digital<br>Uniformity<br>Equalizer                                                                                                                                                                                                      | Different design<br>scheme |
| 3    | Video Signal Input                          |                                                         |                                                         |                                                         |                                                         |                                                                             |                                                                                                                                                                                                                                         |                            |
| 3.1  | Input video<br>signals                      | DVI-D (dual link)<br>x 2,<br>DisplayPort x 2<br>VGA x 1 | DVI-D (dual link)<br>x 2,<br>DisplayPort x 2<br>VGA x 2 | DVI-D (dual link)<br>x 2,<br>DisplayPort x 2<br>VGA x 1 | DVI-D (dual link)<br>x 2,<br>DisplayPort x 2<br>VGA x 1 | DVI-D (dual link)<br>x 1,<br>DVI-D (single<br>link) x 1,<br>DisplayPort x 1 |                                                                                                                                                                                                                                         | Different design<br>scheme |

510(k) Summary

{10}------------------------------------------------

D Monitors C44W+, C82W+, C83W+, C85W

510(k) Submission

510(k) Summary

{11}------------------------------------------------

D Monitors C44W+, C82W+, C83W+, C85V

510(k) Submission

510(k) Summary

{12}------------------------------------------------

| 1<br>AAE!<br>ﮐﺎ ﮐﺎ ﮐﮧ                                                                                                                                                                         |  |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| -<br>0000<br>้                                                                                                                                                                                |  |
| 1<br>AAALA                                                                                                                                                                                    |  |
| 3<br>-<br>Company Company                                                                                                                                                                     |  |
| )<br>ANITARO<br>ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘ |  |
| ﺭ<br>œ<br>(<br>I                                                                                                                                                                              |  |

510(k) Submission

|     |                              |                                                                                                          | Power Related Specifications |           |                                                                                      | Power<br>Management | QC software        | Sensors                                 | USB<br>Ports/Standard | Dimensions<br>w/o stand (W x<br>H x D | Miscellaneous Features/Specifications                                                                                                                                                                 |                               |                          |                                                                                                       |                                           |                             |                            |
|-----|------------------------------|----------------------------------------------------------------------------------------------------------|------------------------------|-----------|--------------------------------------------------------------------------------------|---------------------|--------------------|-----------------------------------------|-----------------------|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|--------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------|-----------------------------|----------------------------|
| 3.2 | Scanning<br>Frequency (H, V) | 31 - 140 kHz, 29<br>- 61 Hz<br>Frame synchronous mode: 59 - 61 Hz, 29.5 - 30.5 Hz                        | 3.3                          | Dot Clock | 268 MHz                                                                              | 4.1                 | Power Requirements | DC 24 V / 9 A                           | 4.2                   | Power Consumption/<br>Save Mode       | 80 W / Less than 5 W                                                                                                                                                                                  | DVI DMPM,<br>DisplayPort 1.2  | Beacon Monitor<br>Manage | Backlight<br>Sensor,<br>Integrated Front<br>Sensor,<br>Ambient Light<br>Sensor                        | 1 upstream,<br>2 downstream /<br>Rev. 2.0 | 699.3 x 459.8 x<br>67.7 mm  |                            |
|     |                              | 31 - 140 kHz, 29<br>- 61 Hz<br>Frame synchronous mode: 59 - 61 Hz, 29.5 - 30.5 Hz                        |                              |           | 533 MHz                                                                              |                     |                    | DC 24 V / 9 A                           |                       |                                       | 160 W / Less than 5 W                                                                                                                                                                                 | DVI DMPM,<br>DisplayPort 1.2  | Beacon Monitor<br>Manage | Backlight<br>Sensor                                                                                   | 1 upstream,<br>2 downstream /<br>Rev. 2.0 | 752.7 x 449.2 x<br>73.05 mm |                            |
|     |                              | 31 - 140 kHz, 29<br>- 61 Hz<br>Frame synchronous mode: 59 - 61 Hz, 29.5 - 30.5 Hz                        |                              |           | 533 MHz                                                                              |                     |                    | DC 24 V / 5 A                           |                       |                                       | 100 W / Less than 5 W                                                                                                                                                                                 | DVI DMPM,<br>DisplayPort 1.2  | Beacon Monitor<br>Manage | Backlight<br>Sensor                                                                                   | -                                         | 650.7 x 389.5 x<br>67.8 mm  |                            |
|     |                              | 31 - 140 kHz, 29<br>- 61 Hz<br>Frame synchronous mode: 59 - 61 Hz, 29.5 - 30.5 Hz                        |                              |           | 533 MHz                                                                              |                     |                    | DC 24 V / 5 A                           |                       |                                       | 100 W / Less than 5 W                                                                                                                                                                                 | DVI DMPM,<br>DisplayPort 1.2  | Beacon Monitor<br>Manage | Backlight<br>Sensor                                                                                   | -                                         | 650.7 x 389.5 x<br>67.8 mm  |                            |
|     |                              | 31 - 159 kHz, 29<br>- 61 Hz (VGA Text: 69 - 71 Hz)<br>Frame synchronous mode: 59 - 61 Hz, 29.5 - 30.5 Hz |                              |           | 280 MHz                                                                              |                     |                    | AC 100 - 120 V, 200 - 240 V; 50 / 60 Hz |                       |                                       | 167 W / Less than 0.7 W                                                                                                                                                                               | DVI DMPM,<br>DisplayPort 1.1a | RadiCS                   | Backlight<br>Sensor,<br>Integrated Front<br>Sensor,<br>Presence<br>Sensor,<br>Ambient Light<br>Sensor | 1 upstream,<br>2 downstream /<br>Rev. 2.0 | 720 x 498 x 119<br>mm       | Different design<br>scheme |
|     |                              | Different design scheme                                                                                  |                              |           |                                                                                      |                     |                    |                                         |                       |                                       | Difference between built-in power supply and built-out power supply<br>Compared with the predicate device, the proposed devices consume less power normally, and more power in the power saving mode. |                               |                          |                                                                                                       |                                           |                             | Different design<br>scheme |
|     |                              |                                                                                                          |                              |           | Different design<br>scheme                                                           |                     |                    |                                         |                       |                                       |                                                                                                                                                                                                       |                               |                          |                                                                                                       |                                           |                             |                            |
|     |                              |                                                                                                          |                              |           | Same for C44W+<br>and C82W+;<br>different design<br>scheme for<br>C83W+ and<br>C85W+ |                     |                    |                                         |                       |                                       |                                                                                                                                                                                                       |                               |                          |                                                                                                       |                                           |                             |                            |
|     |                              |                                                                                                          |                              |           | Different housing<br>design due to the<br>different panel<br>size                    |                     |                    |                                         |                       |                                       |                                                                                                                                                                                                       |                               |                          |                                                                                                       |                                           |                             |                            |

510(k) Summary

{13}------------------------------------------------

510(k) Submission

D Monitors C44W+, C82W+, C83W+, C85V

510(k) Summary

{14}------------------------------------------------

It is clear that the technological characteristic differences discussed above do not affect the safety and the effectiveness of the C44W+, C82W+, C83W+, C85W+.

#### 8.3 Performance Testing

The bench tests were performed on C44W+, C82W+, C83W+, C85W+ as below.

- Measure the spatial resolution expressed as modulation transfer function (MTF).
- Verify the conformance to DICOM GSDF in accordance with Assessment of Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline).
- Measure the luminance non-uniformity characteristics of the display screen in accordance with TG18 guideline.
- Measure the chromaticity non-uniformity characteristics of the display screen in ● accordance with TG18 quideline.
- Visually check the presence or absence of miscellaneous artifacts on the display screen in accordance with TG18 quideline.
- O Meausure the maximum, minimum, achievable, and recommended luminance.
- Measure the temporal response using the typical data provided by the panel manufacturer.
- 0 Maximum number allowed for each type of pixel defects/faults
- Measure the chromaticity at the center of the display screen at 5%, 50% and 95% of the maximum luminance as specified in Guidance for Industry and FDA Staff: Display Accessories for Full-Field Digital Mammography Systems-Premarket Notification (510(k)) Submissions.
- Measure the color tracking (primary colors and color gamut).

The test results showed that C44W+, C82W+, C83W+ are with display characteristics equivalent to those of the predicate device, RadiForce RX440 except some items, each of which was determined that it would not affect observer's performance.

No animal or clinical testing is needed for C44W+, C82W+, C83W+, C85W+.

{15}------------------------------------------------

## 9. Conclusion [21 CFR 807.92(b) (3)]

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Shenzhen Beacon Display Technology Co., Ltd. concludes that:

- The intended use of C44W+, C82W+, C83W+, C85W+ is totally same as that of the predicate device.
- The technological characteristic differences between C44W+, C82W+, C83W+, C85W+ and the predicate device do not affect the safety and effectiveness, so no new risk is raised.
- Demonstrated by the bench tests, the display characteristics of C44W+, C82W+, C83W+, C85W+ are equivalent to those of the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K202374](https://fda.innolitics.com/device/K202374)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
