Protective Face Mask for Medical Use

K202358 · Hunan Heng Chang Pharmaceutical Co., Ltd. · FXX · Jan 29, 2021 · General, Plastic Surgery

Device Facts

Record IDK202358
Device NameProtective Face Mask for Medical Use
ApplicantHunan Heng Chang Pharmaceutical Co., Ltd.
Product CodeFXX · General, Plastic Surgery
Decision DateJan 29, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Intended Use

The Protective Face Mask for Medical Use is intended to worn by the healthcare personnel during procedure to protect both the patient and the healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile/sterile.

Device Story

Protective Face Mask for Medical Use; flat-folded, three-layer disposable mask. Input: physical barrier against airborne/fluid-borne contaminants. Operation: mechanical filtration via polypropylene layers (spun-bond outer/inner, melt-blown middle). Output: physical protection for wearer and patient. Used in clinical/surgical settings by healthcare personnel. Benefits: reduces potential exposure to blood, body fluids, and microorganisms during procedures. Features conformable nose piece and elastic ear loops; non-sterile or sterile options.

Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM F1862M-17 (fluid resistance: 32/32 pass at 120mmHg), ASTM F2299 (particulate filtration: 98.9-99.9%), ASTM F2101-19 (bacterial filtration: 98.9-99.9%), EN 14683:2019 (differential pressure: 4.0-5.6 mmH2O/cm²), and 16 CFR 1610 (flammability: non-flammable). Biocompatibility confirmed per ISO 10993 series.

Technological Characteristics

Three-layer construction: polypropylene spun-bond (outer/inner), polypropylene melt-blown (middle). Includes conformable nose piece and ultrasonically welded elastic ear loops (latex-free). Flat-fold form factor. Single-use, disposable. Non-sterile or sterile. Complies with ASTM F2100-19 Level 2, ISO 10993, and 16 CFR 1610.

Indications for Use

Indicated for healthcare personnel use during procedures to protect patients and staff from transfer of microorganisms, body fluids, and particulate material; intended for infection control to reduce exposure to blood and body fluids.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 29, 2021 Hunan Heng Chang Pharmaceutical Co., Ltd. % Shelley Li Director Shanghai Landlink Medical Information Technology Co., Ltd. Room 703, 705, Baohua International Plaza, West Guangzhong Road 555, Jingan Shanghai, 200071 China Re: K202358 Trade/Device Name: Protective Face Mask for Medical Use Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: December 23, 2020 Received: December 28, 2020 Dear Shelley Li: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K202358 Device Name Protective Face Mask for Medical Use #### Indications for Use (Describe) The Protective Face Mask for Medical Use is intended to worn by the healthcare personnel during procedure to protect both the patient and the healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the posure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile/sterile. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--------------------------------------| | Prescription Use (Part 21 CFR 801 Subpart D) | | | Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span> ☒ </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) summary: K202358 ## I. Submitter Hunan Heng Chang Pharmaceutical Co., Ltd. Floor 2,3,102 of Building 16, Standard Plant of Shanhe Pharmaceutical and Health Industrial Park, No. 1048, Zhongqing Road, Kaifu District, Changsha 410000, China Contact person: Tao jingfeng Position: QC Manager Tel.: +86-18774961605 E-mail: taojingfeng@hcyy.com Preparation date: Dec.23, 2020 # II. Proposed Device | Trade Name of Device: | Protective Face Mask for Medical Use | |-----------------------|--------------------------------------| | Common name: | Surgical Mask | | Regulation Number: | 21 CFR 878.4040 | | Regulatory Class: | Class II | | Product code: | FXX | | Review Panel: | General Hospital | | 510(k) Number: | K202358 | # III. Predicate Devices and Reference Device ### Predicate device | 510(k) Number: | K143287 | |-----------------|-------------------------------------------------| | Trade name: | FLUIDSHIELD Surgical Mask with Expanded Chamber | | Common name: | Surgical Mask | | Classification: | Class II | | Product Code: | FXX | | Manufacturer: | Halyard Health, Inc. | # Reference Device | 510(k) Number: | K160271 | |----------------|------------------------------------------------------------------| | Trade name: | N95 Particulate and Surgical Mask, Models TN01-11 and<br>TN01-12 | | Common name: | Surgical Respirator | {4}------------------------------------------------ | Classification: | Class II | |-----------------|-------------------------| | Product Code: | MSH | | Manufacturer | San-M Package Co., Ltd. | ## IV. Device description The Protective Face Masks for Medical Use are flat-folded masks are three layers of materials consisting of polypropylene spun-bond (outer layer and inner layer), polypropylene melt-blown (middle layers). The masks contain a conformable nose piece enclosed in a binding tape welding the top edge to conform to the contours of the face. In addition, the masks contain an ultrasonically welded, elastic ear loops not made with natural rubber latex, to secure the masks in place on the wearer. The masks are offered in white. The Protective Face Masks for Medical Use are single use, disposable device, provided non-sterile/sterile. ## V. Indication for use The Protective Face Mask for Medical Use is intended to worn by the healthcare personnel during procedure to protect both the patient and the healthcare personnel from the transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile/sterile. ### VI. Comparison of technological characteristics with the predicate devices | Item | Proposed device<br>(K202358) | Predicate device<br>(K143287) | Reference device<br>(K160271) | Discussion | |----------------|-----------------------------------------|-------------------------------------------------------|------------------------------------------------------------------------|------------| | Product name | Protective Face Mask<br>for Medical Use | FLUIDSHIELD<br>Surgical Mask with<br>Expanded Chamber | N95 Particulate and<br>Surgical Mask,<br>Models TN01-11 and<br>TN01-12 | - | | Product Code | FXX | FXX | MSH | Similar | | Regulation No. | 21 CFR 878.4040 | 21 CFR 878.4040 | 21 CFR 878.4040 | Same | | Class | Class II | Class II | Class II | Same | Table 1-Comparison of proposed device and predicate devices {5}------------------------------------------------ | Mask style | | Expanded Chamber<br>(Duckbill) Flat-fold,<br>headband ties , 4<br>Layers | Flat-fold, Headband, 4<br>Layers | Similar1 | | |--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------| | Indication<br>for use | The Protective Face<br>Mask for Medical Use<br>is intended to worn by<br>the healthcare<br>personnel during<br>procedure to protect<br>both the patient and<br>the healthcare<br>personnel from the<br>transfer of<br>microorganisms, body<br>fluids, and particulate<br>material. These face<br>masks are intended<br>for use in infection<br>control practices to<br>reduce the potential<br>exposure to blood and<br>body fluids. This is a<br>single use, disposable<br>device(s), provided<br>non-sterile/sterile. | The Expanded<br>Chamber Surgical<br>Face Mask is<br>intended to be worn to<br>protect both the<br>patient and healthcare<br>personnel from<br>transfer of<br>microorganisms, body<br>fluids and particulate<br>material. These face<br>masks are intended<br>for use in infection<br>control practices to<br>reduce the potential<br>exposure to blood and<br>body fluids. This is a<br>single use, disposable<br>device(s), provided<br>non-sterile. | The N95 Particulate<br>Respirator and<br>Surgical Mask,<br>Models TN01-11 and<br>TN01-12 are single<br>use , disposable<br>devices, provided<br>non-sterile and are<br>intended to be worn<br>by operating room<br>personnel or other<br>healthcare workers to<br>protect both the<br>patient and operating<br>room personnel from<br>of microorganisms,<br>body fluids and<br>particulate material. | Similar to<br>predicate<br>device<br>Different<br>from<br>reference<br>device | | | Material | Inner<br>layer | Spun-bond<br>polypropylene | Polyethylene/Polyester | Polypropylene and<br>polyethylene | Different2 | | | Middle<br>layer | Melt blown<br>polypropylene filter | Middle layer #1:<br>Polypropylene<br>Spunbond<br>Middle layer #2:<br>Polypropylene<br>Meltblown | Two filter layers<br>Polypropylene<br>meltblown | Similar | | | Outer<br>layer | Spun-bond<br>polypropylene | Top Half: Blue<br>Polypropylene<br>Spunbond (w Print)<br>Bottom Half: White<br>Polypropylene | Polypropylene<br>spunbond | Similar | | | Ear loop/head band | Spandex + Polyester | Polyester Spunlace or<br>Polypropylene<br>Spunbond | Polyurethane, not<br>made with natural<br>rubber latex | Different2 | | | Nose piece | Outer plastic, inner<br>aluminum wire,<br>Sponge strips | Unknown | Polyethylene coated<br>steel wire | Different2 | | Offered with Visor | | No | Yes | No | Different | | Color | | White | Top Half: Blue<br>Bottom Half: White | Orange / White | Different2 | | Dimensions | | Length: 16cm<br>Width: 20cm | Length: 19cm<br>Width: 21cm | TN01-12 (Small):<br>Length: 205 ± 5 mm<br>Width: 75 ± 5 mm<br>Band length: 205 ± 5 mm<br>TN01-11 (Medium):<br>Length: 240 ± 5 mm<br>Width: 75 ± 5 mm<br>Band length: 240 ± 5 mm | Different3 | | OTC use | | Yes | Yes | Yes | Same | | Sterility | | Non-sterile/sterile | Non-sterile | Non-sterile | Different4 | | For single use | | Yes | Yes | Yes | Same | | ASTM<br>F2100 Level | | ASTM F 2100-19,<br>Level 2 | ASTM F2100-11,<br>Level 2 | ASTM F2100-11,<br>Level 3 | Similar4 | | Fluid Resistance<br>Performance | | pass at 120mmHg | pass at 120mmHg | pass at 160mmHg | Same as the<br>predicate device | | Particulate Filtration<br>Efficiency | | ≥98% | ≥98% | NIOSH Certification<br>#TC 84A-3348<br>(includes TN01-11 &<br>TN01-12) | Same as the<br>predicate device | | Bacterial | | ≥98% | ≥98% | NIOSH Certification | Same as the | | Filtration<br>Efficiency | | | #TC 84A-3348<br>(includes TN01-11 &<br>TN01-12) | predicate<br>device | | | Differential<br>Pressure | <6.0mmH2O/cm² | <5.0mmH2O/cm² | NIOSH Certification<br>#TC 84A-3348<br>(includes TN01-11 &<br>TN01-12) | Similar4 | | | Flammability | Class I<br>flammable | non | Class I non flammable | Same | | | Biocompatibility | Confirm to the<br>requirements of ISO<br>10993 series<br>standards | Confirm to the<br>requirements of ISO<br>10993 series<br>standards | Confirm to the<br>requirements of ISO<br>10993 series<br>standards | Same | | {6}------------------------------------------------ {7}------------------------------------------------ 1 The physical feature of the proposed device has minor difference from the predicate device. But the proposed device has similar mask shape with the reference device, which it is a surgical respirator. Both of them have the similar indications. The conclusion can be drawn out that the different mask style does not affect the indications of the device. 2 The difference in the materials does not raise additional questions for safety and effectiveness of the device. The biocompatibility evaluation test of the subject devices have been performed on the final finished device. The test results shows pass the requirements. 3 Compare with the predicate and reference device, the different of the physical feature or size does not affect the intended use of the subject device. 4 The subject device (sterile or non-sterile) has performed performance test according to the ASTM F 2100-19. The test results demonstrate that meet the acceptance criteria. # VII. Non-Clinical Testing Non clinical performance tests were conducted to verify that the proposed device met all design specifications. The below table shows the test results of tests article, which demonstrated that the proposed device complies with the following standards: {8}------------------------------------------------ | Methodology | Purpose | Acceptance Criteria | Results | |--------------------------|--------------------------------------|---------------------------------|---------------------------------| | ASTM<br>F1862M-17 | Fluid Resistance<br>Performance | 29 out of 32 pass<br>at 120mmHg | 32 out of 32 pass<br>at 120mmHg | | ASTM F2299 | Particulate<br>Filtration Efficiency | $\geq$ 98% | 98.9~99.9% | | ASTM F2101-19 | Bacterial Filtration<br>Efficiency | $\geq$ 98% | 98.9~99.9% | | EN 14683:2019<br>Annex C | Differential<br>Pressure | <6.0mmH2O/cm² | 4.0~5.6<br>mmH2O/cm² | | 16 CFR 1610 | Flammability | Class I non<br>flammable | non flammable | - ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For ● In Vitro Cytotoxicity - ISO 10993-10: 2010 Biological Evaluation Of Medical Devices Part 10: Tests ● For Irritation And Skin Sensitization. - ASTM F2100-19, Standard Specification for Performance of Materials Used In ● Medical Face Masks. - ASTM F1862M-17, Standard Test Method For Resistance Of Medical Face . Masks To Penetration By Synthetic Blood (Horizontal Projection Of Fixed Volume At A Known Velocity) - EN 14683:2019, Medical Face Mask-Test-Requirements and Test Methods ● - ASTM F2101-19, Standard Test Method For Evaluating The Bacterial Filtration . Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol Of Staphylococcus Aureus; - ASTM F2299-03. Stand test method for determining the initial efficiency of . materials used in medical face masks to penetration by particulates using latex spheres: - 16 CFR 1610, Standard for the Flammability of clothing textiles: . # VIII. Clinical Testing No clinical study is included in this submission. # IX. Conclusion The conclusion drawn from the non-clinical performance testing data demonstrates that the subject device is as safe, as effective, and performs as well as the predicate {9}------------------------------------------------ device, FLUIDSHIELD Surgical Mask with Expanded Chamber (K143287).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...