K202340 · Htl-Strefa S.A · FMI · Oct 15, 2020 · General Hospital
Device Facts
Record ID
K202340
Device Name
Droplet Pen Needle 30G & 33G
Applicant
Htl-Strefa S.A
Product Code
FMI · General Hospital
Decision Date
Oct 15, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The DROPLET PEN NEEDLES 30G & 33G are intended for use with pen injector devices for the subcutaneous injection of drugs.
Device Story
Droplet Pen Needles 30G & 33G are sterile, single-use, prescription/OTC devices designed for subcutaneous drug delivery via pen injectors. The assembly features a double-ended stainless steel cannula secured in an injection-molded plastic hub with adhesive. The hub includes internal threads for attachment to pen injectors, allowing the cartridge-end cannula to pierce the cartridge septum. The patient-end cannula is silicone-lubricated to facilitate skin penetration. An inner shield protects the needle point, and an outer cover maintains sterility until use. The device is operated by patients or caregivers in home or clinical settings. By providing a reliable interface between the drug cartridge and the patient, the needle ensures accurate subcutaneous delivery, supporting patient adherence to prescribed injection regimens.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including compliance with ISO 11608-2 (flow rate, bond strength, needle point sharpness, functional compatibility) and ISO 10993-1 biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility).
Technological Characteristics
Materials: Stainless steel AISI 304L cannula, plastic resin hub, medical-grade silicone lubricant. Dimensions: 30G (8mm) and 33G (4mm). Connectivity: None (mechanical). Sterilization: Ethylene Oxide (EtO) to SAL 10^-6. Standards: ISO 11608-2, ISO 10993-1, ISO 9626, ISO 7864.
Indications for Use
Indicated for use with pen injector devices for the subcutaneous injection of drugs in patients requiring such therapy.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
K252908 — Easydrip Classic Pen Needle; Easydrip Plus Pen Needle; Easydrip Classic Pro Pen Needle; Easydrip Plus Pro Pen Needle · Sandstone Medical (Suzhou), Inc. · Mar 25, 2026
K213478 — BD Pen Needle · Becton, Dickinson and Company · Jun 28, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 15, 2020
HTL-Strefa S.A. Aleksandra Prazmowska-Wilanowska OA/RA Director Adamowek 7 Ozorkow. 95-035 Poland
Re: K202340
Trade/Device Name: Droplet Pen Needle 30G & 33G Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: August 4, 2020 Received: August 17, 2020
Dear Aleksandra Prazmowska-Wilanowska:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For CPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202340
Device Name Droplet Pen needle 30G &33G
Indications for Use (Describe)
The DROPLET PEN NEEDLES 30G & 33G are intended for use with pen injector devices for the subcutaneous injection of drugs.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K202340
| Date Prepared: | October, 7th, 2020 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Submitted By: | HTL-STREFA S.A.<br>ul. Adamówek 7<br>95-035 Ozorków<br>POLAND |
| Primary Contact: | Aleksandra Prażmowska-Wilanowska<br>RA Director<br>Phone: +48 42 270 00 16<br>ola.prazmowska-wilanowska@htl-strefa.pl |
| Secondary Contact: | Izabela Banaś<br>Senior RA Specialist<br>Phone: +48 42 270 02 93<br>izabela.banas@htl-strefa.pl |
| Trade Name: | DROPLET® PEN NEEDLE 30G & 33G |
| Common Name: | Pen Needle |
| Regulation Number: | 21 CFR § 880.5570 |
| Product Code: | FMI |
| Device Classification: | II |
| Review Panel: | 80 General Hospital |
| Predicate Device | |
| [807.92(a)(3)] | |
# The legally marketed device to which substantial equivalence is claimed is:
Predicate device:
| Manufacturer Name | Trade Name | 510(k) Number |
|-------------------|------------------------|---------------|
| HTL-Strefa S.A. | Droplet pen needle 34G | K192082 |
Description of Device:
{4}------------------------------------------------
## [807.92(a)(4)]
DROPLET® PEN NEEDLES 30G & 33G 4G are sterile, single use needles intended for use with pen injector devices for the subcutaneous injection of drugs. The pen needles are OTC devices.
The pen needle assembly consists of a double-ended cannula that is assembled into an injection molded hub using adhesive. The hub has internal threads, which allow it to be screwed onto the pen injector device. This allows the cartridge end of the cannula to penetrate through the rubber septum of the cartridge. The patient end and the cartridge end of the cannula are lubricated using a silicone -based lubricant for ease of injection and rubber septum penetration.
There is an inner needle shield assembled over the patient end of the cannula to protect the needle point from damage and accidental needle sticks. There is also an outer cover. Each pen needle assembly is protected with a peel away seal to provide a sterility barrier.
### Indications for Use:
### [807.92(a)(5)]
The DROPLET® PEN NEEDLES 30G & 33G are intended for use with pen injector devices for the subcutaneous injection of drugs.
### Technological Characteristics:
### [807.92(a)(6)]
A comparison of characteristics of DROPLET® PEN NEEDLE 30G, 33G and the predicate device is shown in the table below:
## Device Comparison between Subject Device and Predicate Device Including Indications for Use
| Feature | Subject Device<br>DROPLET® PEN<br>NEEDLE 30G | Subject Device<br>DROPLET® PEN<br>NEEDLE 33G | Predicate Device<br>DROPLET® PEN<br>NEEDLE 34G | Conclusion |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| 510(k)<br>Number | K202340 | K202340 | K192082 | |
| Product Code | FMI | FMI | FMI | Same |
| Feature | Subject Device<br>DROPLET® PEN<br>NEEDLE 30G | Subject Device<br>DROPLET® PEN<br>NEEDLE 33G | Predicate Device<br>DROPLET® PEN<br>NEEDLE 34G | Conclusion |
| Primary<br>Container | Image: Primary Container for DROPLET® PEN NEEDLE 30G | Image: Primary Container for DROPLET® PEN NEEDLE 33G | Image: Primary Container for DROPLET® PEN NEEDLE 34G | Same |
| Design<br>Needle Shield | Image: Needle Shield for DROPLET® PEN NEEDLE 30G | Image: Needle Shield for DROPLET® PEN NEEDLE 33G | Image: Needle Shield for DROPLET® PEN NEEDLE 34G | Same |
| Needle Tube and<br>Hub | Image: Needle Tube and Hub for DROPLET® PEN NEEDLE 30G | Image: Needle Tube and Hub for DROPLET® PEN NEEDLE 33G | Image: Needle Tube and Hub for DROPLET® PEN NEEDLE 34G | Different<br>No impact<br>on safety<br>or<br>effective |
| Indications<br>for use | The DROPLET®<br>PEN NEEDLE<br>30G is intended<br>for use with a pen<br>injector device<br>for the<br>subcutaneous<br>injection of | The DROPLET®<br>PEN NEEDLE<br>33G is intended<br>for use with a pen<br>injector device<br>for the<br>subcutaneous<br>injection of | The DROPLET®<br>PEN NEEDLE<br>34G is intended<br>for use with a pen<br>injector device<br>for the<br>subcutaneous<br>injection of | Same |
| Length | 8 mm | 4 mm | 3.5 mm | Different<br>No<br>impact<br>on safety |
| Gage | 30G | 33G | 34G | Different<br>No<br>impact<br>on safety |
| Method of<br>attachment<br>to pen<br>injector | Screw threads | Screw threads | Screw threads | Same |
| Feature | Subject Device<br>DROPLET® PEN<br>NEEDLE 30G | Subject Device<br>DROPLET® PEN<br>NEEDLE 33G | Predicate Device<br>DROPLET® PEN<br>NEEDLE 34G | Conclusion |
| Biocompatibi<br>lity | Conforms to ISO<br>10993-1 | Conforms to ISO<br>10993-1 | Conforms to<br>ISO 10993-1 | Same |
| Sterility | SAL = 10-6 | SAL = 10-6 | SAL = 10-6 | Same |
| Sterilization<br>method | EtO | EtO | EtO | Same |
| Unit<br>Packaging | Polypropylene<br>container with seal<br>made of medical<br>grade paper | Polypropylene<br>container with seal<br>made of medical<br>grade paper | Polypropylene<br>container with seal<br>made of medical<br>grade paper | Same |
| User<br>Packaging | Cardboard sales box | Cardboard sales box | Cardboard sales box | Substantially<br>equivalent |
| | Materials | | | |
| Needle Tube | Stainless steel<br>AISI 304L | Stainless steel<br>AISI 304L | Stainless steel<br>AISI 304L | Same |
| Hub<br>Primary<br>Container<br>Needle<br>Shield | Plastic resins | Plastic resins | Plastic resins | Same |
| Lubricant | Medical grade<br>silicone | Medical grade<br>silicone | Medical grade<br>silicone | Same |
{5}------------------------------------------------
{6}------------------------------------------------
HTL-STREFA S.A. has determined that the Droplet® Pen Needle 30G & 33G is substantially equivalent to a predicate device currently cleared for marketing in the United States.
The subject device and the predicate device are classified under 21 CFR 880.5570, which states: "A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin."
The Droplet® Pen Needle 30G&33G is substantially equivalent to the Droplet® Pen Needle 34G cleared under K192082 in terms of indications for use, compositions, material, design and performance. Specifically, the following performance comparisons were made to determine equivalence to the predicate device pen needles referenced: length, gauge, biocompatibility,
{7}------------------------------------------------
materials, shelf life and sterility. Based on the comparisons above of the predicate devices the Droplet® Pen Needle 30G & 34G has raised no different questions of safety and effectiveness.
The difference in needle hub color between the proposed device and the predicate device - 30G x 8 mm and 33G x 4 mm vs. 34G x 3.5 mm has been analyzed. The different colored hubs were introduced for easy needle size identification (color coding for the user). The hubs, irrespective of their color, are made of the same raw materials and were assessed with regard to biocompatibility. No negative biological reaction is expected to occur as a result of the different color. Consequently, the hub color is considered not to have any impact on safety or performance of the proposed device.
The difference in needle gage and length between the proposed device and the predicate device - 30G x 8 mm and 33G x 4 mm vs. 34G x 3.5 mm has also been analyzed. Based on all available design, technical and technological information as well as the published scientific and clinical data and based on information on pen needles by other manufacturers available in the USA, we have concluded that the two proposed needle gages and lengths ensure the same level of confidence for device safety and performance with regard to ensuring subcutaneous injection. They are being introduced to meet individual needs of patients based on their skin type. BMI, and drug dosing and treatments.
Based on the above analysis of the needle length, it has been concluded that difference in needle gage or length between the Droplet 30G & 33G and the Droplet 34G device does not impact safety or performance of the former device and raises no different questions of safety and performance.
The proposed device and the predicate device are classified under 21 CFR 880.5570, which states: "A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin."
The subject device and the predicate device use statements similar to the underlined portion of the excerpt from the regulation.
HTL-STREFA S.A. has determined that the Droplet® Pen Needle 30G and 33G is substantially equivalent to a predicate device currently cleared for marketing in the United States.
### Non-Clinical Performance Data:
## [(807.92(b)(1)]
DROPLET® PEN NEEDLE 30G & 33G successfully passed all the required non-clinical testing which included the following:
- Testing for compliance with the requirements of 11608-2:2012 Needle-based injection . systems for medical use -- Requirements and test methods -- Part 2: Needles
{8}------------------------------------------------
- The table below presents the requirements of the 11608-2:2012 Needle-based injection . systems for medical use -- Requirements and test methods -- Part 2: Needles standard and the result of the testing conducted.
| Test Parameter | Clause no. & requirement of ISO 11608-2:2012 | Result |
|-------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Materials | 4.1 The needle shall be made of tubing materials specified in<br>ISO 9626. | Meets<br>requirements |
| Dimensions | 4.2 The needles shall fit the test apparatus specified in item 7.3<br>of ISO 11608-2. | Meets<br>requirements |
| Determination of<br>flow rate through<br>the needle | 4.3 The needle was tested in accordance with Annex A to ISO<br>11608-2 to determine flow rate through the needle. | Meets<br>requirements |
| Bond between<br>hub and needle<br>tube | 4.4 The union of the hub and needle tube shall not break when<br>tested in accordance with Clause 9 of ISO 11608-2. | Meets<br>requirements |
| Needle points | 4.5 When examined under a magnification of x2,5, needle<br>points shall appear sharp and free from feather edges, burrs and<br>hooks. | Meets<br>requirements |
| Freedom from<br>defects | 4.6 The needle tube shall fulfill the requirements of ISO 7864,<br>11.3. | Meets<br>requirements |
| Lubrication | 4.7 The needle tube should be lubricated at both the patient end<br>and the cartridge end. The lubricant shall not, under normal<br>or corrected-to-normal vision, be visible as droplets of fluid<br>on the outside surface of the needle tube. | Meets<br>requirements |
| Dislocation of<br>measuring point at<br>patient end | 4.8 Dislocation of the cannula point at the patient end shall be in<br>accordance with Table 2 below when tested as per Clause 8<br>(of ISO 11608-2). | Meets<br>requirements |
| Determination of<br>functional<br>compatibility with<br>needle-based<br>injection systems | 4.9 Compatibility with any NIS shall be claimed only after<br>testing in accordance with Clause 11. | Meets<br>requirements |
| Ease of assembly<br>and disassembly | 4.10 Attachment of the needle shall be possible without<br>removing the needle from its opened unit packaging.<br>Compliance is checked according to the requirements of<br>Clause 11. | Meets<br>requirements |
| Sterility | 4.11 The needle in its unit packaging shall has been subjected<br>to a validated sterilization process. | Meets<br>requirements |
| Pre-conditioning<br>of needles | 6 All requirements of the standard related to preconditioning of<br>needles were met. | Meets<br>requirements |
{9}------------------------------------------------
### Biocompatibility testing:
### Selection of Biological Evaluation Tests
The Biological Tests selected to be performed on the sterile final product that has direct contact with the end-user according to the 2016 FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" considering contact type and duration were as follows:
- 1. Cytotoxicity
- 2. Sensitization
- 3. Irritation or Intracutaneous Reactivity
- 4. Acute Systemic Toxicity
- 5. ISO Two Week Systemic Toxicity Study in the Rat, Repeated Parenteral Administration of Two Extracts
- 6. Material-Mediated Pyrogenicity
- 7. Hemocompatibility
Biocompatibility tests selected as per the requirements of 2016 FDA guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" considering contact type and duration for the DROPLET® PEN NEEDLE 30G & 33G did not show any adverse biological / biocompatibility reactions.
### Clinical Performance Data:
### [(807.92(b)(2)]
Clinical data is not required.
### Conclusion:
### [(807.92(b)(3)]
DROPLET® PEN NEEDLE 30G & 33G concluded to be substantially equivalent in the intended use, technology/principle of operation, materials and performance to the legally market predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.