← Product Code [JXI](/productcode/JXI) · K202234

# NervAlign Nerve Cuff (K202234)

_Renerve, Ltd. · JXI · Feb 10, 2022 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K202234

## Device Facts

- **Applicant:** Renerve, Ltd.
- **Product Code:** [JXI](/productcode/JXI.md)
- **Decision Date:** Feb 10, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5275
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The NervAlign® Nerve Cuff is indicated for the repair of peripheral nerve injuries in which there is no gap or where a gap closure is achieved by flexion of the extremity.

## Device Story

NervAlign Nerve Cuff is a sterile, freeze-dried, acellular collagen membrane matrix derived from porcine pericardium. Implanted by surgeons during peripheral nerve repair procedures; device acts as a protective scaffold around damaged nerves. Scaffold becomes infiltrated by patient cells and remodels into native tissue over time. Provides physical protection to nerve site during healing process. Single-use device available in multiple sizes. Benefits include nerve protection and support for natural tissue regeneration. No electronic components or software.

## Clinical Evidence

No human clinical data. Evidence consists of extensive non-clinical bench and animal testing. Bench testing: suture retention, thickness, tensile strength, endotoxin levels, and ISO 10993 biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, genotoxicity). Animal studies: 4-week and 13-week rat subcutaneous implantation (biocompatibility/degradation), 6-month rabbit sciatic nerve wrap study (safety/compatibility), and rat transected sciatic nerve model (performance comparison vs. predicate). Results confirmed biocompatibility, appropriate degradation profile, and nerve healing outcomes similar to predicate.

## Technological Characteristics

Collagen matrix derived from porcine pericardium. Sterile, freeze-dried, pre-cut flat sheet. Dimensions: 10x10mm, 20x30mm, 30x40mm. Sterilization: Gamma irradiation. Mechanical properties: Suture retention and tensile strength sufficient for nerve repair. Bioresorbable scaffold. No energy source, connectivity, or software.

## Regulatory Identification

A nerve cuff is a tubular silicone rubber sheath used to encase a nerve for aid in repairing the nerve (e.g., to prevent ingrowth of scar tissue) and for capping the end of the nerve to prevent the formation of neuroma (tumors).

## Predicate Devices

- AxoGuard Nerve Protector ([K132660](/device/K132660.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 10, 2022

Renerve Ltd % Chris Sloan President Sloan Regulatory Consulting LLC 322 Hart Road Gaithersburg, Maryland 20878

Re: K202234

Trade/Device Name: NervAlign Nerve Cuff Regulation Number: 21 CFR 882.5275 Regulation Name: Nerve Cuff Regulatory Class: Class II Product Code: JXI Dated: January 10, 2022 Received: January 11, 2022

Dear Chris Sloan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202234

Device Name NervAlign® Nerve Cuff

Indications for Use (Describe)

The NervAlign® Nerve Cuff is indicated for the repair of peripheral nerve injuries in which there is no gap closure is achieved by flexion of the extremity.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary NervAlign® Nerve Cuff K202234

#### Submitter

| Name           | ReNerve Ltd                                                            |
|----------------|------------------------------------------------------------------------|
| Address        | Suite 3, 21 Vale Street<br>North Melbourne, Victoria 3051<br>Australia |
| Contact Person | David Rhodes                                                           |
| Phone          | +61 (0)418598835                                                       |
| Email Address  | drhodes@renerve.com.au                                                 |
| Date Prepared  | February 2, 2022                                                       |

### Device

| Trade Name            | NervAlign® Nerve Cuff |
|-----------------------|-----------------------|
| Common Name           | Nerve Cuff            |
| Classification Name   | Nerve Cuff            |
| Classification Number | 21 CFR 882.5275       |
| Product Code          | JXI                   |
| Regulatory Class      | II                    |

### Predicate Device

| Name [510(k) Number] | AxoGuard Nerve Protector [K132660] |
|----------------------|------------------------------------|
|----------------------|------------------------------------|

### Device Description

The NervAlign® Nerve Cuff is a collagen membrane matrix derived from porcine pericardium. It is a sterile, whitish to light beige colored, freeze-dried, pre-cut, flat sheet of acellular collagen. The NervAlign® Nerve Cuff is available in three (3) different sizes: 10x10mm, 20x30mm and 30x40mm.

The collagen material that comprises the Nerve Cuff is derived from the same species as that of the predicate nerve cuff (AxoGuard Nerve Protector; K132660) manufactured by Cook Biotech Incorporated.

Like the predicate, the NervAlign® Nerve Cuff is implanted providing a scaffold which becomes infiltrated by the patient's cells and is remodelled into native tissue. The Nerve Cuff provides protection of the damaged nerve while the nerve heals.

The NervAlign® Nerve Cuff is packaged in a dried state, is for single use and provided sterile.

{4}------------------------------------------------

## Indications for Use

The NervAlign® Nerve Cuff is indicated for the repair of peripheral nerve injuries in which there is no gap or where a gap closure is achieved by flexion of the extremity.

## Comparison of Technological Characteristics with the Predicate Device

The proposed device is substantially equivalent to the design and materials in the predicate device. The table below summarizes the comparison between the predicate device and the NervAlign® Nerve Cuff.

| Feature                           | NervAlign® Nerve Cuff<br>[Proposed Device]                                                                                                                                                 | AxoGuard Nerve Protector<br>[Predicate Device]                                                                                                                                                                                                      |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)                            | K202234                                                                                                                                                                                    | K132660                                                                                                                                                                                                                                             |
| Device Class                      | Class II                                                                                                                                                                                   | Class II                                                                                                                                                                                                                                            |
| Classification Name and<br>Number | Nerve Cuff; 21 CFR 882.5275                                                                                                                                                                | Nerve Cuff; 21 CFR 882.5275                                                                                                                                                                                                                         |
| Product Code                      | JXI                                                                                                                                                                                        | JXI                                                                                                                                                                                                                                                 |
| Indications for Use               | The NervAlign® Nerve Cuff is<br>indicated for the repair of<br>peripheral nerve injuries in which<br>there is no gap or where a gap<br>closure is achieved by flexion of the<br>extremity. | The Nerve Cuff is indicated for<br>the repair of peripheral nerve<br>injuries in which there is no gap<br>or where a gap closure is<br>achieved by flexion of the<br>extremity. The device is provided<br>sterile and intended for one-time<br>use. |
| Principles of Operation           | Collagen matrix intended to wrap<br>and repair damaged peripheral<br>nerves                                                                                                                | Collagen matrix intended to wrap<br>and repair damaged peripheral<br>nerves                                                                                                                                                                         |
| Material                          | Collagen Matrix - Porcine<br>Pericardium                                                                                                                                                   | Collagen Matrix – Porcine Small<br>Intestinal Submucosa (SIS)                                                                                                                                                                                       |
| Sterile/Single Use                | Sterile; single use                                                                                                                                                                        | Sterile; single use                                                                                                                                                                                                                                 |
| Sterilization Method              | Gamma Irradiation                                                                                                                                                                          | Ethylene Oxide (EtO)                                                                                                                                                                                                                                |

# NervAlign® Nerve Cuff Comparison to AxoGuard Nerve Protector

# Performance Data

The following testing was performed on the ReNerve NervAlign® Nerve Cuff.

{5}------------------------------------------------

| Test                                                                   | Test Method Summary                                                                                                                                                                                                                      | Results                                                                                                                                                                                                                                                                                                                                                                                              |
|------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Suture retention<br>strength                                           | A suture was placed through aged and<br>unaged devices and the force required to<br>pull free was measured.                                                                                                                              | Test was completed and met specification.<br>Like predicate the device has sufficient<br>strength for its intended use.                                                                                                                                                                                                                                                                              |
| Thickness                                                              | Direct measurement of the device with<br>micrometer when hydrated according to<br>IFU.                                                                                                                                                   | Thickness measurements of device showed<br>substantially equivalent to predicate.                                                                                                                                                                                                                                                                                                                    |
| Tensile Strength                                                       | Direct measurement of the device.<br>Device placed between two grips and the<br>separation force required to reach device<br>failure was measured.                                                                                       | Test was completed and met specification.<br>Like predicate device is suitable for<br>intended use.                                                                                                                                                                                                                                                                                                  |
| Cytotoxicity                                                           | The device was evaluated for potential<br>cytotoxic effects using a mammalian cell<br>line following ISO 10993-5 guidelines.                                                                                                             | The test article extract showed no cytotoxic<br>potential.                                                                                                                                                                                                                                                                                                                                           |
| Sensitization                                                          | The device was evaluated for the<br>potential to cause delayed dermal<br>contact sensitization in guinea pigs<br>based on ISO 10993-10.                                                                                                  | Extracts of the test article showed no<br>evidence of inducing delayed contact<br>sensitization in the guinea pig.                                                                                                                                                                                                                                                                                   |
| Acute intracutaneous<br>reactivity (Irritation)                        | The device was evaluated for the<br>potential to cause irritation following<br>intracutaneous injection in rabbits based<br>on ISO 10993-10.                                                                                             | Extracts of the test article show no evidence<br>of irritation.                                                                                                                                                                                                                                                                                                                                      |
| Acute systemic<br>toxicity                                             | The device was evaluated for acute<br>systemic toxicity in mice based on ISO<br>10993-11.                                                                                                                                                | No mortality or evidence of systemic<br>toxicity from both extracts, injected into<br>mice.                                                                                                                                                                                                                                                                                                          |
| Pyrogenicity                                                           | The device was evaluated for the<br>potential to induce a pyrogenic response<br>following intravenous injection in<br>rabbits. Study conducted according to<br>the United States Pharmacopeia (USP 41<br>- NF36, General Chapter <151>). | The test article met the requirements and is<br>judged non-pyrogenic.                                                                                                                                                                                                                                                                                                                                |
| Hemolysis                                                              | The device was evaluated for hemolytic<br>potential when in contact with blood<br>based on ASTM F756 and requirements<br>of ISO 10993-4.                                                                                                 | The direct contact of the test article was<br>slightly hemolytic.<br>The extracts at 25, 12.5 and 6.25% were<br>non-hemolytic.<br>Device is suitable for intended indication.                                                                                                                                                                                                                        |
| Genotoxicity<br>(AMES)                                                 | The device was evaluated for the<br>potential to induce reverse mutations in<br><i>Salmonella typhimurium</i> and                                                                                                                        | Extracts of the device were considered to be<br>non-mutagenic to tester strains.                                                                                                                                                                                                                                                                                                                     |
| Test                                                                   | Test Method Summary                                                                                                                                                                                                                      | Results                                                                                                                                                                                                                                                                                                                                                                                              |
| Escherichia coli tester strains per ISO<br>10993-3.                    |                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                      |
| Genotoxicity (mouse<br>lymphoma assay)                                 | The device was evaluated to determine<br>its mutagenic potential using the mouse<br>lymphoma forward gene mutation assay<br>per ISO 10993-3.                                                                                             | Test article considered non-mutagenic.                                                                                                                                                                                                                                                                                                                                                               |
| Endotoxin                                                              | Bacterial endotoxin testing is conducted<br>per USP 85 and European<br>Pharmacopoeia 2.6.14                                                                                                                                              | Device is produced and released with<br>endotoxin level <2.15EU/device.                                                                                                                                                                                                                                                                                                                              |
| Subacute Systemic<br>Toxicity & Local<br>effects of<br>Implantation    | The device was surgically implanted in<br>the subcutaneous tissue of rats for 4<br>weeks to evaluate its potential systemic<br>toxicity and local tissue response.                                                                       | The device showed no evidence of systemic<br>toxicity. Device appeared as a well-<br>integrated scaffold with remodelling<br>responses. Microscopically the test article<br>demonstrated minimal reaction to the tissue.                                                                                                                                                                             |
| Sub-chronic Systemic<br>Toxicity & Local<br>effects of<br>implantation | The device was surgically implanted in<br>the subcutaneous tissue of rats for 8 and<br>13 weeks to evaluate potential systemic<br>toxicity after 13 weeks and local tissue<br>responses after 8 and 13 weeks.                            | On subcutaneous implantation the device<br>showed no evidence of systemic toxicity<br>after 13 weeks.<br>Microscopically the test article<br>demonstrated minimal tissue reaction<br>compared to control. Device was<br>bioresorbing at 13 weeks as expected.                                                                                                                                        |
| 6-Month Nerve Wrap<br>Study in New-<br>Zealand White<br>Rabbits        | The device was surgically implanted<br>around the sciatic nerve of rabbits for 1,<br>2 & 6 months. The effects and<br>compatibility of the wrapping material<br>on the nerve and surrounding tissues<br>were assessed.                   | The device was well tolerated with no<br>device related clinical signs, changes in<br>body weight, food consumption,<br>neurological parameters, NCV, clinical<br>pathology, or organ weights up to 6 months<br>post-implantation.<br>No adverse macroscopic or microscopic<br>changes in the nerve and surrounding<br>tissues were observed. No device related<br>axonal degeneration was observed. |
| Rat transected sciatic<br>nerve model                                  | The device was surgically implanted<br>around the transected sciatic nerve of the<br>rat and compared to nerves wrapped<br>with marketed predicate control<br>material. Data were collected at 1, 4 and<br>13 weeks.                     | At all time points the changes observed in<br>the nerve were similar between device and<br>predicate control material and were typical<br>of changes in a nerve after transection.<br>At all time points the device was considered<br>to elicit no or minimal reaction in<br>comparison to the predicate control                                                                                     |
| Test                                                                   | Test Method Summary                                                                                                                                                                                                                      | Results                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                        | Local tissue responses and device<br>degradation were evaluated, and motor<br>and sensory neurological assessments<br>were conducted.                                                                                                    | The reactivity scores for device and<br>predicate control decreased at later time<br>points indicating healing.<br>The device was markedly or completely<br>degraded at 13 weeks, compared with the<br>control device which was not degraded.<br>Neuromas were not observed in any animal.                                                                                                           |

{6}------------------------------------------------

{7}------------------------------------------------

# Conclusions

The NervAlign® Nerve Cuff has the following similarities to the predicate device:

- . Same intended use,
- Same basic design,
- Similar material, ●
- Similar safety and performance characteristics, and ●
- Same operating principles. ●

Based on the similarities of the intended use/indications for use, device design, principles of operation, technological characteristics and the results of the non-clinical performance testing, the subject device, NervAlign® Nerve Cuff, is substantially equivalent to the legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/device/K202234](https://fda.innolitics.com/device/K202234)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
