← Product Code [FQH](/productcode/FQH) · K202222

# Irrisept Wound Debridement and Cleansing System (K202222)

_Irrimax Corporation · FQH · Dec 14, 2020 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K202222

## Device Facts

- **Applicant:** Irrimax Corporation
- **Product Code:** [FQH](/productcode/FQH.md)
- **Decision Date:** Dec 14, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5475
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Irrisept Wound Debridement and Cleansing System is a wound cleansing delivery system. The mechanical action effectively loosens and removes wound debris.

## Device Story

Irrisept is a single-use, manual, self-contained wound irrigation system. It consists of a 450mL bottle containing 0.05% Chlorhexidine Gluconate (CHG) in 99.95% sterile water and a multi-port applicator (Irriprobe). The device is used by clinicians to cleanse wounds; the operator manually compresses the Blow-Fill-Seal (BFS) bottle to deliver the solution through the applicator tip. The mechanical force of the fluid stream loosens and removes debris from the wound bed. The CHG acts as a preservative. The device is provided sterile and is intended for single use to assist in wound management, potentially reducing microbial load through mechanical irrigation.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including USP <51> antimicrobial effectiveness, ISO 13408 aseptic processing, ISO 11135/10993-7 sterilization validation, USP <85>/<151>/<161> endotoxin/pyrogen testing, ISO 11607 packaging validation, ASTM F1980 accelerated aging, and ISO 10993 biocompatibility testing.

## Technological Characteristics

Manual, single-use irrigation system. Components: Polypropylene bottle, polycarbonate applicator. Solution: 0.05% CHG in 99.95% sterile water. Sterilization: EO (applicator and bottle) and aseptic BFS processing (solution). Biocompatibility: ISO 10993-1 compliant. Packaging: Tyvek header pouch and CSR wraps. Shelf-life: 2 years.

## Regulatory Identification

A jet lavage is a device used to clean a wound by a pulsatile jet of sterile fluid. The device consists of the pulsing head, tubing to connect to a container of sterile fluid, and a means of propelling the fluid through the tubing, such as an electric roller pump.

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.

## Predicate Devices

- Irrisept Wound Debridement and Cleansing System ([K080779](/device/K080779.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

December 14, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

Irrimax Corporation Christy Coleman VP Regulatory 1665 Lakes Parkway - Suite 102 Lawrenceville, Georgia 30043

Re: K202222

Trade/Device Name: Irrisept Wound Debridement and Cleansing System Regulation Number: 21 CFR 880.5475 Regulation Name: Jet Lavage Regulatory Class: Class II, unclassified Product Code: FOH, FRO Dated: November 12, 2020 Received: November 13, 2020

#### Dear Christy Coleman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Kimberly Ferlin, Ph.D. Assistant Director (acting) DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K202222

Device Name

Irrisept Wound Debridement and Cleansing System

#### Indications for Use (Describe)

The Irrisept Wound Debridement and Cleansing System is a wound cleansing delivery system. The mechanical action effectively loosens and removes wound debris.

| Type of Use (Select one or both, as applicable)                                                                                  |                                                                                                                                 |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

| 510(K) NUMBER:         | K202222                                                                                                                                                                                                                                          |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PREPARATION DATE:      | December 11, 2020                                                                                                                                                                                                                                |
| APPLICANT:             | Irrimax® Corporation<br>1665 Lakes Parkway, Suite 102<br>Lawrenceville, GA 30043                                                                                                                                                                 |
| CONTACT PERSON:        | Christy Coleman, OD, MPH<br>Vice President, Regulatory Affairs, Irrimax®<br>Corporation<br>Tel: (770) 807-3355                                                                                                                                   |
| DEVICE TRADE NAME:     | Irrisept® Wound Debridement and<br>Cleansing System                                                                                                                                                                                              |
| CLASSIFICATION NAME:   | Jet Lavage                                                                                                                                                                                                                                       |
| DEVICE CLASSIFICATION: | Class II (21 CFR 880.5475)<br>Unclassified (Pre-amendment)                                                                                                                                                                                       |
| PRODUCT CODE:          | FQH; FRO                                                                                                                                                                                                                                         |
| PREDICATE DEVICE:      | Irrisept® Wound Debridement and Cleansing System<br>(K080779)<br>• Product Code FQH (Jet Lavage);<br>Class II (21 CFR 880.5475)<br>• Product Code FRO (Dressing, Wound,<br>Drug); Unclassified (pre-amendment)<br>Applicant: Irrimax Corporation |

#### DEVICE DESCRIPTION:

Irrisept® Wound Debridement and Cleansing System is a single-use, manual, self-contained irrigation device comprised of a 450mL bottle of 0.05% Chlorhexidine Gluconate (CHG) in 99.95% Sterile Water for Irrigation, United States Pharmacopeia (USP) and an applicator (labeled Irriprobe®). The solution is aseptically-filled in a Blow-Fill-Seal (BFS) bottle. The CHG acts as a preservative to inhibit microbial growth in the solution.

### INTENDED USE/INDICATIONS FOR USE:

The Irrisept Wound Debridement and Cleansing System is a wound cleansing delivery system. The mechanical action effectively loosens and removes wound debris.

{4}------------------------------------------------

### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS:

The Irrisept that is the subject of this 510(k) is substantially equivalent to its predicate, predecessor K080779 Irrisept. This is a catch-up 510(k) submission intended to update the Irrisept record at FDA with the changes that have been implemented in the subject product. As such, the intended use/indications remain identical to that for the cleared product. Further, the implemented changes were primarily to be responsive to user feedback, capture the now available longer-term shelf-life data, or reflect a need to change a supplier or manufacturing contractor. These are captured in the comparison table below.

| COMPARISON OF TECHNOLOGICAL CHARACTERISTICS |                                                                                                                                                                                                                             |                                                                                                                                                                                                        |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison Feature                          | Cleared Irrisept Product<br>(K080779)                                                                                                                                                                                       | Subject Irrisept Product<br>(K202222)                                                                                                                                                                  |
| Intended Use/<br>Indications for Use        | The Irrisept Wound<br>Debridement and Cleansing<br>System is a wound cleansing<br>delivery system. The<br>mechanical action effectively<br>loosens and removes wound<br>debris.                                             | Identical                                                                                                                                                                                              |
| Rx only                                     | Prescription Product                                                                                                                                                                                                        | Identical                                                                                                                                                                                              |
| Single Use                                  | Single Use                                                                                                                                                                                                                  | Identical                                                                                                                                                                                              |
| Mechanism of Action                         | The mechanical action of fluid<br>across the wound removes<br>wound debris.                                                                                                                                                 | Identical.                                                                                                                                                                                             |
| Components and<br>Materials                 | Polypropylene bottleSolutionPolycarbonate applicator                                                                                                                                                                        | New supplier for<br>Chlorhexidine GluconateNew supplier for<br>polycarbonate                                                                                                                           |
| Solution Composition                        | 0.05% Chlorhexidine Gluconate<br>in 99.95% Sterile Water for<br>Irrigation, USP                                                                                                                                             | Identical                                                                                                                                                                                              |
| Container Design                            | 450 mL BFS bottle delivers fluid<br>by manual compression                                                                                                                                                                   | Identical                                                                                                                                                                                              |
| Applicator Design                           | Multi-port applicator that<br>threads onto bottle.                                                                                                                                                                          | Multi-port applicator that<br>threads onto bottle and with<br>a longer tip.Change to a higher capacity<br>manufacturing mold                                                                           |
| Biocompatibility                            | Conforms to ISO 10993-1 for a<br>surface device with breached or<br>compromised surface contact<br>and a limited contact duration<br>(≤ 24 hours)                                                                           | Identical                                                                                                                                                                                              |
| COMPARISON OF TECHNOLOGICAL CHARACTERISTICS |                                                                                                                                                                                                                             |                                                                                                                                                                                                        |
| Comparison Feature                          | Cleared Irrisept Product<br>(K080779)                                                                                                                                                                                       | Subject Irrisept Product<br>(K202222)                                                                                                                                                                  |
| Manufactured using<br>Aseptic Techniques    | Solution is manufactured using Blow-Fill-Seal (BFS) aseptic processing. Solution labeled as Sterile.                                                                                                                        | Identical BFS process  New contract manufacturer Changed to a higher capacity BFS manufacturing mold                                                                                                   |
| Non-Pyrogenic                               | Solution                                                                                                                                                                                                                    | Solution and applicator<br>Endotoxin Limit ≤ 20 EU/device<br>(<0.045 EU/mL for 450 mL)                                                                                                                 |
| Packaging                                   | 1. Packaged applicator in Tyvek® film pouch<br>2. Protective PETG tube added over bottle cap to protect bottle cap during shipment<br>3. Sterilized applicator and bottle-filled solution are then packaged in a shelf box. | Changed manufacturing final step for protective PETG tube Bottle-filled solution and applicator are packaged together in CSR wraps and Tyvek® header pouch for sterilization New contract manufacturer |
| Sterilization                               | Packaged applicator conforms to ISO 11135 and ISO 10993-7 for EO sterilization, whereas the solution is aseptically processed                                                                                               | The bottle and applicator are EO-sterilized and conform to ISO 11135 and ISO 10993-7, whereas the solution is aseptically processed                                                                    |
| Shelf-Life                                  | Six-month shelf-life.                                                                                                                                                                                                       | Two-year shelf-life                                                                                                                                                                                    |
| Labeling                                    | Labeling consists of:  Bottle Label Applicator label Package Label - Shelf Box Label Shipping Box Label                                                                                                                     | Labeling consists of:  Bottle Label Package Label - Front Package Label - Back Shipping Box Label                                                                                                      |

{5}------------------------------------------------

The differences between the two versions of Irrisept do not raise different questions of safety and effectiveness and have been confirmed through testing.

{6}------------------------------------------------

## PERFORMANCE TESTING:

Performance testing demonstrates that the subject Irrisept is as safe and as effective as the predicate, predecessor Irrisept product. These tests include:

- Preservative Antimicrobial Effectiveness ●
	- o USP <51> Antimicrobial Effectiveness Testing
- Aseptic Processing
	- o ISO 13408 Aseptic Processing of Health Care Products
- Sterilization
	- ISO 11135 Sterilization of Health-Care Products Ethylene Oxide O
	- ANSI AAMI ISO 10993-7 Ethylene oxide sterilization residuals O
	- USP <71> Sterility Tests O
	- AAMI TIR 28 Product adoption and process equivalence for ethylene oxide O sterilization
- Endotoxins and Pyrogens ●
	- USP <85> Bacterial Endotoxins Test O
	- USP <151> Pyrogen Test (USP Rabbit Test) O
	- USP <161> Medical Devices Bacterial Endotoxin and Pyrogen Tests O

#### ● Packaging and Shelf-Life

- ISO 11607 Packaging for Terminally Sterilized Medical Devices O
- ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier O Systems for Medical Devices
- ASTM F2096-11 Standard Test Method for Detecting Gross Leaks in Packaging о by Internal Pressurization (Bubble Test)
- ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous o Medical Packaging by Dye Penetration
- ASTM D4169-16 Standard Practice for Performance Testing of Shipping o Containers and Systems
- Biocompatibility ●
	- o ISO 10993- Biological Evaluation of Medical Devices

#### CONCLUSIONS:

In conclusion, Irrimax has demonstrated through performance testing that the subject Irrisept is substantially equivalent to its predicate Irrisept product in intended use and technological characteristics.

---

**Source:** [https://fda.innolitics.com/device/K202222](https://fda.innolitics.com/device/K202222)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
