← Product Code [QEW](/productcode/QEW) · K202218

# ZelanteDVT THrombectomy System, ZelanteDVT ClotHunter Helical Rotation Device (K202218)

_Boston Scientific Corporation · QEW · Oct 2, 2020 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K202218

## Device Facts

- **Applicant:** Boston Scientific Corporation
- **Product Code:** [QEW](/productcode/QEW.md)
- **Decision Date:** Oct 2, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5150
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The ZelanteDVT™ Thrombectomy System is intended for removal of thrombus from and infusion of fluids into the peripheral vasculature

## Device Story

System comprises single-use ZelanteDVT Thrombectomy Set and ClotHunter Helical Rotation Device; used with AngioJet Ultra Console. Thrombectomy set uses high-velocity saline jets to create low-pressure zone at catheter tip; thrombus drawn into catheter, fragmented, and evacuated to waste bag. ClotHunter device features distal helical-shaped wire attached to control handle; deflects catheter tip to improve luminal coverage and clot removal; provides transition between catheter and vessel wall. Used in clinical setting by physicians. Output is physical removal of thrombus or infusion of fluids. Benefits include effective thrombus removal from large peripheral veins.

## Clinical Evidence

Bench, animal, and biocompatibility testing performed. Biocompatibility testing included cytotoxicity, hemolysis, sensitization, complement activation, intracutaneous reactivity, PTT, acute systemic toxicity, in vitro hemocompatibility, materials-mediated pyrogenicity, and platelet/leukocyte counts. In-vitro performance testing covered clot removal, wire torque, catheter withdrawal, radiopacity, distal emboli, and mechanical integrity (tensile, fatigue, column strength). GLP animal study conducted to evaluate safety and performance.

## Technological Characteristics

System consists of catheter and helical rotation device. Materials evaluated for biocompatibility. Operates via high-velocity saline jet (hydrodynamic principle). Connectivity: requires AngioJet Ultra Console. Sterilization: not specified. Form factor: catheter-based system with manual control handle for helical wire rotation.

## Regulatory Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

## Predicate Devices

- ZelanteDVT Thrombectomy Set ([K151313](/device/K151313.md))

## Reference Devices

- Amplatz Super Stiff Guidewire ([K843012](/device/K843012.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 2, 2020

Boston Scientific Corporation Vanessa Fowler Senior Regulatory Affairs Specialist 3 Scimed Place Maple Grove, Minnesota 55311

## Re: K202218

Trade/Device Name: ZelanteDVT™ Thrombectomy System ZelanteDVT™ ClotHunter™ Helical Rotation Device Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW, KRA Dated: August 5, 2020 Received: August 6, 2020

Dear Vanessa Fowler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K202218

#### Device Name

ZelanteDVT™ Thrombectomy System ZelanteDVT™ ClotHunter™ Helical Rotation Device

#### Indications for Use (Describe)

The ZelanteDVT Thrombectomy System, which includes the ZelanteDVT Thrombectomy Set and the ClotHunter Helical Rotation Device, is intended for use with the AngioJet Ultra Console to break apart and remove thrombus, including deep vein thrombus (DVT), from:

· Iliofemoral and lower extremity veins ≥ 6.0 mm in diameter and

· Upper extremity peripheral veins ≥ 6.0 mm in diameter.

The ZelanteDVT Thrombectomy System is also intended for use with the AngioJet Ultra Power Pulse technique for the controlled and selective infusion of physician specified fluids, including thrombolytic agents, into the peripheral vascular system.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| Sponsor                         | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, Massachusetts 01752<br>USA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name and<br>Information | Vanessa Fowler<br>Sr Regulatory Affairs Specialist<br>3 Scimed Place<br>Maple Grove, MN 55311-1566<br>Phone: 763-494-2537<br>Fax: 763-494-2222<br>e-mail: Vanessa.Fowler@bsci.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Date Prepared                   | August 5, 2020                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Proprietary Name                | ZelanteDVT™ Thrombectomy System<br>ZelanteDVT™ ClotHunter™ Helical Rotation Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Common Name                     | Catheter, Embolectomy; Catheter, Continuous Flush                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Product Code                    | QEW, KRA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Classification                  | Class II, 21 CFR Part 870.5150                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Predicate Device                | ZelanteDVT Thrombectomy Set (K151313), cleared 21 September 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Reference Device                | Amplatz Super Stiff Guidewire (K843012), cleared 29 November 1984                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device Description              | The AngioJet™ ZelanteDVT Thrombectomy System is a component of the<br>AngioJet Ultra Thrombectomy System (AngioJet Ultra System or System).<br>The ZelanteDVT Thrombectomy System can only be used in conjunction with<br>the AngioJet Ultra Console. The AngioJet ZelanteDVT Thrombectomy<br>System is comprised of the single-use ZelanteDVT Thrombectomy Set and<br>the single-use ClotHunter Helical Rotation Device. Both components are<br>packaged and sold separately. The ClotHunter Helical Rotation Device can<br>only be used in conjunction with the ZelanteDVT Thrombectomy Set.<br><br>The Thrombectomy Set uses this pressurized, high-velocity saline to create a<br>low-pressure zone at the Catheter tip. Thrombus is drawn into the Catheter<br>where it is fragmented by the jets and evacuated from the body. The waste<br>tubing transports the thrombus debris from the Catheter to the waste bag for<br>ultimate disposal.<br><br>The ClotHunter Helical Rotation Device consists of a distal helical-shaped<br>wire portion attached to a control handle that is used with the ZelanteDVT<br>Thrombectomy Set. The Helical Rotation Device deflects the distal tip of the<br>Catheter and improves clot removal due to increased luminal coverage. The<br>ClotHunter Helical Rotation Device extends beyond the distal tip of the<br>Catheter to provide a transition between the Catheter and the vessel wall. |
| Intended Use of<br>Device       | The ZelanteDVT™ Thrombectomy System is intended for removal of<br>thrombus from and infusion of fluids into the peripheral vasculature                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Indications for Use             | The ZelanteDVT Thrombectomy System, which includes the ZelanteDVT<br>Thrombectomy Set and the ClotHunter Helical Rotation Device, is intended<br>for use with the AngioJet Ultra Console to break apart and remove thrombus,<br>including deep vein thrombus (DVT), from:<br>• Iliofemoral and lower extremity veins ≥ 6.0 mm in diameter and<br>• Upper extremity peripheral veins ≥ 6.0 mm in diameter.<br>The ZelanteDVT Thrombectomy System is also intended for use with the<br>AngioJet Ultra Power Pulse technique for the controlled and selective infusion<br>of physician specified fluids, including thrombolytic agents, into the peripheral<br>vascular system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

### 510(k) Summary for K202218 Per 21 CFR §807.92

{4}------------------------------------------------

| Comparison of<br>Technological<br>Characteristics | The ZelanteDVT Thrombectomy System incorporates substantially equivalent<br>device materials and design, packaging materials and design, fundamental<br>technology, manufacturing processes, sterilization process and intended use<br>as those featured in the ZelanteDVT Thrombectomy Set (K151313).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance Data                                  | Bench, animal, and biocompatibility testing were performed to support a<br>determination of substantial equivalence. The results of these tests provide<br>reasonable assurance that the proposed device has been designed and<br>tested to assure conformance to the requirements for its intended use. No<br>new safety or performance issues were raised during the testing.<br><br>The following biocompatibility tests were completed on the ZelanteDVT<br>ClotHunter Helical Rotation Device or leveraged from the reference device,<br>the Amplatz Super Stiff Guidewire:<br>Cytotoxicity Extract and Direct Contact Hemolysis Sensitization Complement Activation Intracutaneous Reactivity Partial Thromboplastin Time Acute Systemic Toxicity In Vitro Hemocompatibility Materials Mediated Pyrogenicity Platelet and Leukocyte Counts   The following <i>in-vitro</i> performance tests were completed for the ZelanteDVT<br>Thrombectomy System.<br>Clot Removal Wire Whip Force (Wire Torque) Catheter Withdrawal Infused Volume Handle Attach/Detach Radiopacity Distal Emboli Particulate Evaluation Catheter Advance Rotator Disengagement Force Insertion Force Rotational Fatigue Rotation Force Exposed Wire Length Wire Tip Tensile Wire Shape Retention Wire to Handle Torsional Force Trapped Wire Rotation Handle / Wire Tensile Strength Thumbwheel Breakage Gear Slip Force Column Strength Corrosion Handle Function Post Fluid Exposure Coating Integrity  Additionally a GLP animal study was performed to evaluate the safety and |

AddRonally, a Jan ZelanteDVT Thrombectomy System.

{5}------------------------------------------------

| Conclusion | Based on the indications for use, technological characteristics, and safety and<br>performance testing, the ZelanteDVT Thrombectomy System, consisting of<br>the ZelanteDVT Thrombectomy Set and the ZelanteDVT ClotHunter Helical<br>Rotation Device, has been shown to be appropriate for its intended use and is<br>considered to be substantially equivalent to the ZelanteDVT Thrombectomy<br>Set (K151313). |
|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

---

**Source:** [https://fda.innolitics.com/device/K202218](https://fda.innolitics.com/device/K202218)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
