The intended use of Philips EPIQ series diagnostic ultrasound systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intra-luminal, Intraoperative (Vascular), Intraoperative (Cardiac). Musculoskeletal (Conventional). Musculoskeletal (Superficial). Other: Urology. Pediatric. Peripheral Vessel. Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal. The clinical environments where Philips EPIQ diagnostic ultrasound systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients. When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo guidance. The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information, and only for the purposes for which it was designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
Device Story
EPIQ Series Diagnostic Ultrasound Systems are software-controlled capital equipment for high-resolution imaging and fluid flow analysis. System console generates electrical current sent to compatible transducers/intravascular catheters; piezoelectric elements convert current to soundwaves; reflected echoes are received and processed into images for display. Used by trained professionals in clinics, hospitals, and point-of-care settings. When integrated with Philips EchoNavigator, provides live echo guidance alongside X-ray for cardiovascular interventional procedures. Output assists clinicians in diagnosis and image-guided treatment. Benefits include non-invasive visualization of anatomy and physiology, and real-time guidance during interventional procedures.
Clinical Evidence
Bench testing only. No clinical data required. Testing included system-level verification against design specifications and compliance with safety standards: ANSI/AAMI 60601-1, IEC 60601-1-2, and IEC 60601-2-37. All pre-determined acceptance criteria were met.
Technological Characteristics
Capital equipment console with software-controlled imaging. Compatible with various transducers (TEE, pencil, curved, linear, sector arrays) and single-use intravascular catheters. Principle: Piezoelectric stimulation for ultrasound wave generation and echo reception. Connectivity: Standalone/networked. Standards: ANSI/AAMI 60601-1, IEC 60601-1-2, IEC 60601-2-37.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis of the human body, including abdominal, cardiac (adult/pediatric/fetal), vascular (cerebral/peripheral), cephalic, fetal/obstetric, gynecological, intra-luminal, intraoperative, musculoskeletal, urological, and small organ applications. Used in clinics, hospitals, and point-of-care settings.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Philips EPIQ Series Diagnostic Ultrasound Systems (K200304)
{1}------------------------------------------------
September 2, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure, while the FDA acronym is in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Philips Ultrasound, Inc. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k SAINT PAUL MN 55114
Re: K202216
Trade/Device Name: EPIQ Series Diagnostic Ultrasound Systems Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX, OBJ Dated: August 5, 2020 Received: August 6, 2020
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202216
Device Name EPIQ Series Diagnostic Ultrasound Systems
### Indications for Use (Describe)
The intended use of Philips EPIQ series diagnostic ultrasound systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use:
Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intra-luminal, Intraoperative (Vascular), Intraoperative (Cardiac). Musculoskeletal (Conventional). Musculoskeletal (Superficial). Other: Urology. Pediatric. Peripheral Vessel. Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal.
The clinical environments where Philips EPIQ diagnostic ultrasound systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo guidance.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information, and only for the purposes for which it was designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
| CONTINUE ON A SEPARATE PAGE IF NEEDED | |
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
K202216
This summary of safety and effectiveness information is submitted in accordance with 21 CFR § 807.92. Date Prepared: August 31, 2020
#### I. Submitter
| Manufacturer Name<br>and Address | Philips Ultrasound, Inc.<br>22100 Bothell Everett Hwy<br>Bothell, WA 98021-8431 |
|----------------------------------|---------------------------------------------------------------------------------------------------------------|
| Contact<br>Information | Colin S. Jacob<br>Senior Regulatory Affairs Specialist<br>TEL: 425.908.1209<br>EMAIL: colin.jacob@philips.com |
### II. Device
| Trade Name | EPIQ Series Diagnostic Ultrasound Systems |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Diagnostic ultrasound system and transducers |
| Regulation Description | Ultrasonic pulsed doppler imaging system<br>Ultrasonic pulsed echo imaging system<br>Diagnostic ultrasonic transducer<br>Diagnostic intravascular catheter |
| Regulation Number | 892.1550<br>892.1560<br>892.1570<br>870.1200 |
| Primary Product Code | IYN |
| Secondary Product Codes | IYO<br>ITX<br>OBJ |
| Device Class | Class II |
| Classification Panel | Radiology |
{5}------------------------------------------------
#### Predicate Device III.
- K200304 - Philips EPIQ Series Diagnostic Ultrasound Systems
#### IV. Device Description
Philips EPIQ Diagnostic Ultrasound Systems are durable capital equipment / medical devices which are software-controlled and intended for high-resolution general imaging, interventional radiology, cardiology, vascular and OB/GYN applications and fluid flow analysis. They are intended to be used by trained professionals at various settings of patient care such as clinical admission, periodic evaluations, prior to hospitalization discharge, and/or academic research, via maneuverable caster wheels.
Software modes/applications, scanning protocols, and pre-installed settings or functionality to create dedicated settings for imaging of specific anatomy are available with the subject Philips EPIQ Diagnostic Ultrasound Systems and may vary among model configurations.
The systems are manufactured with hardware components which consist of:
- a primary console (e.g., workstation, tablet) with built-in software components, features, and various 1) clinical applications, and
- 2) a range of compatible ultrasound transducers and ultrasound intravascular catheters.
A suite of reusable compatible transducer types are offered such as transesophageal echocardiography (TEE) transducers, non-imaging (pencil) probes, curved array and sector/phased array. Philips ultrasound transducers may be bagged into compatible transducer/probe cover sheaths, designed by other manufacturers, for each procedure to prevent cross-contamination and reduce the risk of healthcareassociated infections. Philips EPIQ Diagnostic Ultrasound Systems are also compatible with single-use diagnostic intravascular ultrasound catheters (i.e., Philips VeriSight Intracardiac Echocardiography (ICE) Catheter, VeriSight Pro Intracardiac Echocardiography (ICE) Catheter) that are intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart for image guidance during cardiac interventional procedures.
Some configurations may have additional previously-cleared accessories, components and software features which are manufactured by Philips Ultrasound, Inc. or other manufacturers.
Image /page/4/Picture/14 description: The image shows the Philips logo. The logo is a blue shield with the word "PHILIPS" at the top. Inside the shield, there are wavy lines and four stars. The logo is simple and recognizable.
{6}------------------------------------------------
#### V. Indications for Use
| Device | Indications for Use |
|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Philips EPIQ<br>Diagnostic<br>Ultrasound<br>System | The intended use of Philips EPIQ series diagnostic ultrasound systems is diagnostic ultrasound imaging and fluid flow<br>analysis of the human body, with the following indications for use:<br><br>Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic<br>(Neonatal), Fetal/Obstetric, Gynecological, Intra-cardiac Echo, Intra-luminal, Intraoperative (Vascular),<br>Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric,<br>Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal.<br><br>The clinical environments where Philips EPIQ diagnostic ultrasound systems can be used include clinics, hospitals,<br>and clinical point-of-care for diagnosis of patients.<br><br>When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image<br>guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo<br>guidance.<br><br>The systems are intended to be installed, used, and operated only in accordance with the safety procedures and<br>operating instructions given in the product user information, and only for the purposes for which it was designed.<br>However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best<br>clinical procedure. |
{7}------------------------------------------------
### Comparison of Technological Characteristics between Proposed Subject Device and Predicate Device VI.
| Standard Feature | Philips EPIQ Series Diagnostic<br>Ultrasound Systems<br>K202216<br>(Subject Device) | Philips EPIQ Series Diagnostic<br>Ultrasound Systems<br>K200304<br>(Predicate Device) | Comparison |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation<br>Number | 892.1550 | 892.1550 | Same |
| Device<br>Classification<br>Name | System, Imaging, Pulsed Doppler,<br>Ultrasonic | System, Imaging, Pulsed Doppler,<br>Ultrasonic | Same |
| Product Code | IYN | IYN | Same |
| Secondary Product<br>Code | IYO, ITX, OBJ | IYO, ITX | Subject of submission |
| Indications for Use | The intended use of Philips EPIQ series<br>diagnostic ultrasound systems is<br>diagnostic ultrasound imaging and fluid<br>flow analysis of the human body, with<br>the following indications for use:<br>Abdominal, Cardiac Adult, Cardiac other<br>(Fetal), Cardiac Pediatric, Cerebral<br>Vascular, Cephalic (Adult), Cephalic<br>(Neonatal), Fetal/Obstetric,<br>Gynecological, Intra-cardiac Echo, Intra-<br>luminal, Intraoperative (Vascular),<br>Intraoperative (Cardiac), Musculoskeletal<br>(Conventional), Musculoskeletal<br>(Superficial), Other: Urology, Pediatric, | The intended use of Philips EPIQ series<br>diagnostic ultrasound systems is<br>diagnostic ultrasound imaging and fluid<br>flow analysis of the human body, with<br>the following indications for use:<br>Abdominal, Cardiac Adult, Cardiac other<br>(Fetal Echo), Cardiac Pediatric, Cerebral<br>Vascular, Cephalic (Adult), Cephalic<br>(Neonatal), Fetal/Obstetric,<br>Gynecological, Intraoperative (Vascular),<br>Intraoperative (Cardiac), Musculoskeletal<br>(Conventional), Musculoskeletal<br>(Superficial), Other: Urology, Pediatric, | Subject of submission<br>Indications for use of<br>EPIQ is being updated to<br>include "Intra-cardiac<br>echo, Intra-luminial" to<br>reflect compatibility with<br>ultrasound intravascular<br>catheters (K200812) |
| Standard Feature | Philips EPIQ Series Diagnostic<br>Ultrasound Systems<br>K202216<br>(Subject Device) | Philips EPIQ Series Diagnostic<br>Ultrasound Systems<br>K200304<br>(Predicate Device) | Comparison |
| | Thyroid, Testicle), Transesophageal | Thyroid, Testicle), Transesophageal | |
| | (Cardiac), Transrectal, Transvaginal. | (Cardiac), Transrectal, Transvaginal. | |
| | The clinical environments where Philips<br>EPIQ diagnostic ultrasound systems can<br>be used include clinics, hospitals, and<br>clinical point-of-care for diagnosis of<br>patients. | The clinical environments where Philips<br>EPIQ diagnostic ultrasound systems can<br>be used include clinics, hospitals, and<br>clinical point-of-care for diagnosis of<br>patients. | |
| | When integrated with Philips<br>EchoNavigator, the systems can assist<br>the interventionalist and surgeon with<br>image guidance during treatment of<br>cardiovascular disease in which the<br>procedure uses both live X-ray and live<br>echo guidance. | When integrated with Philips<br>EchoNavigator, the systems can assist<br>the interventionalist and surgeon with<br>image guidance during treatment of<br>cardiovascular disease in which the<br>procedure uses both live X-ray and live<br>echo guidance. | |
| | The systems are intended to be installed,<br>used, and operated only in accordance<br>with the safety procedures and operating<br>instructions given in the product user<br>information, and only for the purposes | The systems are intended to be installed,<br>used, and operated only in accordance<br>with the safety procedures and operating<br>instructions given in the product user<br>information, and only for the purposes | |
| | for which it was designed. However, | for which it was designed. However, | |
| | nothing stated in the user information | nothing stated in the user information | |
| | reduces your responsibility for sound | reduces your responsibility for sound | |
| Standard Feature | Philips EPIQ Series Diagnostic<br>Ultrasound Systems<br>K202216<br>(Subject Device) | Philips EPIQ Series Diagnostic<br>Ultrasound Systems<br>K200304<br>(Predicate Device) | Comparison |
| | clinical judgement and best clinical<br>procedure. | clinical judgement and best clinical<br>procedure. | |
| Reusable? | Yes | Yes | Same |
| Duration of Use | Limited (≤ 24 hours) | Limited (≤ 24 hours) | Same |
| Scientific<br>Technology | Ultrasound Imaging | Ultrasound Imaging | Same |
| Principles of<br>Operation | System console generates electrical<br>current and sends to a connected,<br>compatible transducer to stimulate its<br>piezoelectric elements at the distal end.<br>Stimulation causes the elements to<br>expand and contract which creates a<br>high-pressured wave (i.e., soundwave). A<br>series of soundwaves then propagate<br>toward tissue medium (e.g., mucus<br>membrane, bone). The transducer then<br>receives echoed soundwaves that are<br>reflected from the tissue medium and<br>transmits it to the system. The system<br>processes the echoed soundwaves into<br>an image which is displayed on the<br>display monitor screen of the system for<br>user interpretation. | System console generates electrical<br>current and sends to a connected,<br>compatible transducer to stimulate its<br>piezoelectric elements at the distal end.<br>Stimulation causes the elements to<br>expand and contract which creates a<br>high-pressured wave (i.e., soundwave). A<br>series of soundwaves then propagate<br>toward tissue medium (e.g., mucus<br>membrane, bone). The transducer then<br>receives echoed soundwaves that are<br>reflected from the tissue medium and<br>transmits it to the system. The system<br>processes the echoed soundwaves into<br>an image which is displayed on the<br>display monitor screen of the system for<br>user interpretation. | Same |
| Type of<br>Compatible<br>Components | TEE Probes<br>Non-imaging (pencil) probes<br>Curved Array Transducers | TEE Probes<br>Non-imaging (pencil) probes<br>Curved Array Transducers | Subject of submission |
| Standard Feature | Philips EPIQ Series Diagnostic<br>Ultrasound Systems<br>K202216<br>(Subject Device) | Philips EPIQ Series Diagnostic<br>Ultrasound Systems<br>K200304<br>(Predicate Device) | Comparison |
| | Linear Array Transducers<br>Sector Array Transducers<br>Ultrasound Intravascular Catheters* | Linear Array Transducers<br>Sector Array Transducers | |
| Acoustic Outputs<br>Within Range? | Yes | Yes | Same |
| Previously cleared<br>Imaging Modes? | Yes | Yes | Same |
{8}------------------------------------------------
## w / VeriSight ICE / Pro ICE Catheters
Page 5 of 8
{9}------------------------------------------------
### w/ VeriSight ICE / Pro ICE Catheters
Page 6 of 8
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the Philips logo on the left side of the image. To the right of the logo is the text "TRADITIONAL 510(k) EPIQ Series Diagnostic Ultrasound Systems w/ VeriSight ICE / Pro ICE Catheters". On the right side of the image is the text "Page 7 of 8".
*Compatible ultrasound intravascular catheter in K200812.
{11}------------------------------------------------
#### VII. Performance Data
Relevant non-clinical verification testing and compatibility information has been performed and included in this submission to address system level requirements according to system and design specifications, and risk control measures. The proposed Philips EPIQ Diagnostic Ultrasound Systems did not require clinical data for determination of substantial equivalence.
The proposed Philips EPIQ Series Diagnostic Ultrasound Systems with the VeriSight ICE / PRO ICE Catheters is an electronic product as defined by 21 CFR Subject Chapter J regulations, and is Track 3 Device and complies with the referenced standard as the FDA ultrasound guidance document, Guidance for Industry and FDA Staff - Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers, issued on June 27, 2019.
Philips Ultrasound, Inc. performed the following testing to ensure the safety and effectiveness of the proposed Philips EPIQ Series Diagnostic Ultrasound Systems with the VeriSight ICE / PRO ICE catheters:
- 트 ANSI/AAMI 60601-1: Medical electrical equipment. General requirements for basic safety and essential performance, 2005, Amendment 1, 2012,
- I IEC 60601-1-2 Medical Electrical Equipment - Part 1-2, General Requirements for Basic Safety and Essential Performance – Collateral Standard Electromagnetic Compatibility, 2014,
- IEC 60601-2-37: Medical electrical equipment. Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment, 2015.
#### VIII. Conclusion
For testing, all pre-determined acceptance criteria were met. Results of these tests show that the proposed subject Philips EPIQ Series Diagnostic Ultrasound Systems with VeriSight ICE / PRO ICE Catheters meet their intended use. The results of the relevant performance data and compatibility support a determination that the proposed subject device does not rasse new questions of safety or effectiveness and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.