← Product Code [LNH](/productcode/LNH) · K202210

# Vantage Orian 1.5T, MRT-1550, V6.0 (K202210)

_Canon Medical Systems Corporation · LNH · Sep 22, 2020 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K202210

## Device Facts

- **Applicant:** Canon Medical Systems Corporation
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Sep 22, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

Vantage Orian 1.5T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are: ·Proton density (PD) (also called hydrogen density) - ·Spin-lattice relaxation time (T1) ·Spin-spin relaxation time (T2) ·Flow dynamics ·Chemical Shift Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

## Device Story

Vantage Orian 1.5T is a magnetic resonance imaging (MRI) system utilizing a 1.5 Tesla magnet. It acquires cross-sectional images (transaxial, coronal, sagittal, oblique) based on proton density, T1/T2 relaxation times, flow dynamics, and chemical shift. Modifications include integration of a TN150 magnet (from Vantage Titan), addition of IR pulse to MultiBand SPEEDER on SEEPI2D sequences, and combination of QuickStar with Real Time Motion Correction (RMC) on SSFP3D sequences. Operated by trained clinicians in clinical settings; output is interpreted by physicians to aid diagnosis. System includes Pianissimo/Pianissimo Zen noise reduction. Benefits include high-quality diagnostic imaging for head/body anatomy. Device uses standard MRI safety parameters (SAR, dB/dt) and emergency ramp-down for safety.

## Clinical Evidence

Bench testing only. No clinical data provided. Verification and validation testing included image quality metrics (SNR, geometric distortion, uniformity, slice thickness) and software validation, confirming the device meets predetermined acceptance criteria.

## Technological Characteristics

1.5T MRI system; TN150 magnet; 55x50 cm FOV. Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-33, IEC 60825-1, IEC 62304, IEC 62366, NEMA MS 1-4. Software: Moderate level of concern. Connectivity: Standard MRI imaging modality.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Vantage Orian 1.5T, MRT-1550, V6.0 ([K193021](/device/K193021.md))

## Reference Devices

- Vantage Titan 1.5T, MRT-1510, M-Power GX ([K170412](/device/K170412.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

September 22, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Canon Medical Systems Corporation % Ms. Janine Reyes Manager, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive TUSTIN CA 92780

Re: K202210

Trade/Device Name: Vantage Orian 1.5T, MRT-1550, V6.0 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: August 20, 2020 Received: August 24, 2020

Dear Ms. Reyes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K202210

Device Name Vantage Orian 1.5T, MRT-1550, V6.0

#### Indications for Use (Describe)

Vantage Orian 1.5T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.

MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:

·Proton density (PD) (also called hydrogen density)

- ·Spin-lattice relaxation time (T1)
·Spin-spin relaxation time (T2)

·Flow dynamics

·Chemical Shift

Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

Type of Use (Select one or both, as applicable)

| <span style="font-size: 10pt;"> <svg height="15" width="15"> <rect height="15" style="fill:lightgrey;stroke-width:1;stroke:black" width="15"></rect> <path d="M2,2 L13,13 M13,2 L2,13" stroke="black" stroke-width="2"></path> </svg>   Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <svg height="15" width="15"> <rect height="15" style="fill:lightgrey;stroke-width:1;stroke:black" width="15"></rect> </svg>   Over-The-Counter Use (21 CFR 801 Subpart C)</span>                                                                            |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92

# 1. CLASSIFICATION and DEVICE NAME

| Classification Name:    | Magnetic Resonance Diagnostic Device |
|-------------------------|--------------------------------------|
| Regulation Number:      | 90-LNH (Per 21 CFR § 892.1000)       |
| Trade Proprietary Name: | Vantage Orian 1.5T, MRT-1550, V6.0   |
| Model Number:           | MRT-1550                             |

# 2. SUBMITTER'S NAME

Canon Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550

# 3. OFFICIAL CORRESPONDENT

Naofumi Watanabe Senior Manager, Regulatory Affairs and Vigilance Canon Medical Systems Corporation

# 4. CONTACT PERSON, U.S. AGENT and ADDRESS

#### Contact Person

Janine Reyes Manager, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 669-7853 Fax: (714) 730-1310 jfreyes@us.medical.canon

# Official Correspondent/U.S. Agent

Paul Biggins Senior Director, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 730-7808 Fax: (714) 730-1310 pbiggins@us.medical.canon

{4}------------------------------------------------

#### 5. MANUFACTURING SITE

Canon Medical Systems Corporation 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan

- 6. ESTABLISHMENT REGISTRATION 9614698
- 7. DATE PREPARED August 5, 2020
- 8. DEVICE NAME Vantage Orian 1.5T, MRT-1550, V6.0

# 9. TRADE NAME

Vantage Orian 1.5T, MRT-1550, V6.0

#### 10. CLASSIFICATION NAME

Magnetic Resonance Diagnostic Device (MRDD)

#### 11. CLASSIFICATION PANEL

Radiology

- 12. DEVICE CLASSIFICATION
Class II (per 21 CFR 892.1000, Magnetic Resonance Diagnostic Device)

# 13. PRODUCT CODE

90-LNH

#### 14. PREDICATE DEVICE

Predicate Device (system): Vantage Orian 1.5T, MRT-1550, V6.0 (K193021) Reference Device (magnet): Vantage Titan 1.5T, MRT-1510, M-Power GX (K170412)

|                | Subject Device                         | Predicate Device                       | Reference Device                             |
|----------------|----------------------------------------|----------------------------------------|----------------------------------------------|
| System         | Vantage Orian 1.5T, MRT-<br>1550, V6.0 | Vantage Orian 1.5T, MRT-<br>1550, V6.0 | Vantage Titan 1.5T, MRT-<br>1510, M-Power GX |
| Marketed By    | Canon Medical Systems USA              | Canon Medical Systems USA              | Canon Medical Systems USA                    |
| 510(k) Number  | This Submission                        | K193021                                | K170412                                      |
| Clearance Date | This Submission                        | June 3, 2020                           | June 7, 2017                                 |

#### 15. REASON FOR SUBMISSION

Modification of a cleared device

{5}------------------------------------------------

#### 16. SUBMISSION TYPE

Special 510(k) Premarket Notification

#### 17. DEVICE DESCRIPTION

The Vantage Orian (Model MRT-1550) is a 1.5 Tesla Magnetic Resonance Imaging (MRI) System. The Vantage Orian uses 1.4 m short and 4.1 tons light weight magnet. It includes the Canon Pianissimo™ and Pianissimo Zen technology (scan noise reduction technology). The design of the gradient coil and the whole body coil of the Vantage Orian provides the maximum field of view of 55 x 50 cm. The Model MRT-1550/A3, A4, A7, A8 includes the standard gradient system.

This system is based upon the technology and materials of previously marketed Canon Medical Systems MRI systems and is intended to acquire and display cross-sectional transaxial, coronal, sagittal, and oblique images of anatomic structures of the head or body. The Vantage Orian MRI System is comparable to the current 1.5T Vantage Orian MRI System (K193021), cleared June 3, 2020 with the following modifications.

# 18. SUMMARY OF CHANGE(S)

- Magnet (for MRT-1550/A3, MRT-1550/A4, MRT-1550/A7, MRT-1550/A8): The Vantage Orian 1.5T, MRT-1550/A3, MRT-1550/A4, MRT-1550/A7, and MRT-1550/A8 utilize a 1.5T magnet (TN150), the same magnet used with the Vantage Titan, MRT-1510, M-Power GX (K170412) cleared June 7, 2017.
- . IR pulse can be used in MultiBand SPEEDER: Added combination of MultiBand SPEEDER and IR pulse on SEEPI2D sequence.
- . RMC (Real Time Motion Correction) can be used in QuickStar: Added combination of QuickStar and RMC on SSFP3D sequence.

| Item                                      | Subject Device:<br>Vantage Orian 1.5T, MRT-1550, V6.0                                                | Predicate Device:<br>Vantage Orian 1.5T, MRT-1550, V6.0<br>K193021                                | Notes |
|-------------------------------------------|------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|-------|
| Static field strength                     | 1.5T                                                                                                 | 1.5T                                                                                              | Same  |
| Operational Modes                         | Normal and 1st Operating Mode                                                                        | Normal and 1st Operating Mode                                                                     | Same  |
| i. Safety parameter<br>display            | SAR, dB/dt                                                                                           | SAR, dB/dt                                                                                        | Same  |
| ii. Operating mode<br>access requirements | Allows screen access to 1st level<br>operating mode                                                  | Allows screen access to 1st level<br>operating mode                                               | Same  |
| Maximum SAR                               | 4W/kg for whole body (1st<br>operating mode specified in IEC<br>60601-2-33:<br>2010+A1:2013+A2:2015) | 4W/kg for whole body (1st operating<br>mode specified in IEC 60601-2-33:<br>2010+A1:2013+A2:2015) | Same  |

#### 19. SAFETY PARAMETERS

{6}------------------------------------------------

| Item                                                                | Subject Device:                                                                           | Predicate Device:                                                                         | Notes |
|---------------------------------------------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|-------|
|                                                                     | Vantage Orian 1.5T, MRT-1550, V6.0                                                        | Vantage Orian 1.5T, MRT-1550, V6.0                                                        |       |
|                                                                     |                                                                                           | K193021                                                                                   |       |
| Maximum dB/dt                                                       | 1st operating mode specified in IEC<br>60601-2-33: 2010+A1:2013+A2:2015                   | 1st operating mode specified in IEC<br>60601-2-33: 2010+A1:2013+A2:2015                   | Same  |
| Potential emergency<br>condition and means<br>provided for shutdown | Shutdown by Emergency Ramp<br>Down Unit for collision hazard for<br>ferromagnetic objects | Shutdown by Emergency Ramp<br>Down Unit for collision hazard for<br>ferromagnetic objects | Same  |

#### 20. IMAGING PERFORMANCE PARAMETERS

No change from the previous predicate submission, K193021.

#### 21. INDICATIONS FOR USE

Vantage Orian 1.5T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.

MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:

- Proton density (PD) (also called hydrogen density)
- Spin-lattice relaxation time (T1)
- Spin-spin relaxation time (T2) ●
- Flow dynamics
- Chemical Shift

Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

#### 22. SUMMARY OF DESIGN CONTROL ACTIVITIES

Hazard analysis has been performed and documentation is included in this submission. The test methods used are the same as those submitted in the previously cleared submissions of the predicate device, Vantage Orian 1.5T, MRT-1550, V6.0 (K193021). A declaration of conformity with design controls is included in this submission.

#### 23. SAFETY

This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards.

This device is based upon the same technologies, materials and software as the predicate device. Risk activities were conducted in concurrence with established medical device development

{7}------------------------------------------------

standards and guidance. Additionally, testing was done in accordance with applicable recognized consensus standards published by the International Electrotechnical Commission (IEC) for medical devices and the National Electrical Manufacturers Association (NEMA):

# LIST OF APPLICABLE STANDARDS

- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012
- . IEC60601-1-2 (2014)
- IIEC60601-1-6 (2010), Amd.1 (2013)
- IEC60601-2-33 (2010), Amd.1 (2013), ● Amd.2 (2015)
- IEC60825-1 (2007)
- . IEC62304 (2006), Amd.1(2015)
- IEC62366 (2007), Amd.1 (2014)
- NEMA MS 1 (2008) ●
- NEMA MS 2 (2008)
- NEMA MS 3 (2008)
- NEMA MS 4 (2010)

# 24. TESTING

Risk analysis and verification/validation testing conducted through bench testing are included in this submission which demonstrate that the system requirements have been met. Additionally, image quality testing was completed which demonstrated that the subject device meets predetermined acceptance criteria. Image quality metrics were performed to assess signal-to-noise ratio, twodimensional geometric distortion, image uniformity, and slice thickness.

Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.

# 25. SUBSTANTIAL EQUIVALENCE

Canon Medical Systems Corporation believes that the Vantage Orian 1.5T, MRT-1550, V6.0 Magnetic Resonance Imaging (MRI) System is substantially equivalent to the previously cleared predicate device referenced in this submission.

Canon Medical Systems Corporation believes that the changes incorporated into the Vantage Orian 1.5T, MRT-1550, V6.0 are substantially equivalent to the previously cleared predicate device.

# 26. CONCLUSION

The modifications incorporated into the Vantage Orian 1.5T, MRT-1550, V6.0 do not change the indications for use or the intended use of the device. Based upon bench testing, successful completion of software validation and application of risk management and design controls, it is concluded that the subject device is safe and effective for its intended use.

---

**Source:** [https://fda.innolitics.com/device/K202210](https://fda.innolitics.com/device/K202210)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
