The TrueRelief Device is intended for use under the supervision of a Healthcare Professional. TrueRelief is intended for temporary symptomatic relief of chronic intractable pain, and as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.
Device Story
TrueRelief is an AC-powered transcutaneous electrical nerve stimulator (TENS) used in clinical settings (physician offices, physical therapy clinics, hospitals, nursing homes, chiropractic clinics) under healthcare professional supervision. The device comprises a pulse generator, a primary probe, and a secondary probe. It delivers electromagnetic energy transcutaneously to painful tissue using a proprietary waveform with small voltage fluctuations. The healthcare provider operates the device to apply electrical pulses to the patient's tissue. By stimulating nerves, the device aims to provide temporary symptomatic relief of chronic intractable pain and manage acute post-surgical or post-traumatic pain, potentially reducing patient discomfort.
Clinical Evidence
Bench testing only. Performance testing included electrical safety (IEC 60601-1, IEC 60601-2-10, IEC 60601-1-6), electromagnetic compatibility (IEC 60601-1-2), and usability/human factors evaluation. No clinical data was provided.
Technological Characteristics
AC-powered (115 VAC, 60Hz) TENS device. Patient contact material: 316L stainless steel. Output: single channel, transformer-isolated, regulated current. Waveform: rectangular. Pulse rate: 1-400 pps (low), 4,000-20,000 pps (high). Pulse duration: 0.24-0.74 ms (low), 15-25 µs (high). Max current: 8.9 mA. Max charge per pulse: 7 µC. Electrical safety: Type BF. Non-sterile.
Indications for Use
Indicated for adult patients requiring temporary symptomatic relief of chronic intractable pain or adjunctive treatment for post-surgical and post-traumatic acute pain. Intended for use under healthcare professional supervision.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
TrueRelief % Kristen Allen Founder and Principal Consultant AllenBridge Consulting 2221 Oleander Drive Wilmington, North Carolina 28403
March 23, 2021
Re: K202186
Trade/Device Name: TrueRelief Device Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZJ Dated: March 16, 2021 Received: March 17, 2021
Dear Kristen Allen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name TrueRelief Device
Indications for Use (Describe)
The TrueRelief Device is intended for use under the supervision of a Healthcare Professional.
TrueRelief is intended for temporary symptomatic relief of chronic intractable pain, and as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for TrueRelief. Above the name is a graphic of a sound wave in shades of orange, purple, and blue. The name TrueRelief is written in a blue font, with the word True in a darker shade of blue than the word Relief.
## 510(k) Summary
## TrueRelief Device
| Submitter: | True Relief<br>100 Wilshire Boulevard, Suite 700<br>Santa Monica, CA 90401<br>310.918.9064 (P) |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Kristen Allen<br>Principal Consultant, Regulatory, Quality and<br>Compliance<br>910-612-4153 (P)<br>kallen@truerelief.com |
| Date Prepared: | March 23, 2021 |
| Trade Name: | TrueRelief Device |
| Common Name: | Transcutaneous Electrical Nerve Stimulator (TENS) for pain<br>relief |
| Device Product Code<br>and Classification: | Regulation Number: 21 CFR 882.5890<br>GZJ, Class II, Transcutaneous Electrical Nerve Stimulator<br>(TENS) for pain relief |
| Primary Predicate: | TrueRelief Device (originally NewLife Sciences TMR device)<br>K070474 |
| Additional Predicates: | DJO Chattanooga Revolution Wireless, K153696 |
#### Device Description:
The TrueRelief device is an AC power operated device that consists of three main components and associated power cables: the pulse generator, a primary probe and a secondary probe. The signal generator applies electromagnetic energy transcutaneously through the primary and secondary probes into painful tissue, similar to Transcutaneous Electrical Nerve Stimulation (TENS) device. The TrueRelief device uses a proprietary waveform with small voltage fluctuations to achieve pain relief.
#### Indications and Intended use:
The TrueRelief Device is intended for use under the supervision of a Healthcare Professional.
TrueRelief is intended for temporary symptomatic relief of chronic intractable pain, and as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for TrueRelief. Above the name is a graphic of four diamonds in a row, each a different color. The name TrueRelief is written in two different colors, with True in blue and Relief in purple.
## Comparison of Technological Characteristics With The Predicate Device:
The subject and predicate devices have similar technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. The following characteristics are similar between the subject device and predicate devices:
- . Indications for use
- . Principle of operation
- Technological characteristics .
- . Performance testing
### Comparison with Predicate Devices
| Item | Primary Predicate Device<br>K070474 TMR | Reference Device<br>K153696 | Subject Device<br>TrueRelief |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for<br>Use | TrueRelief is intended for<br>temporary symptomatic relief<br>of chronic intractable pain, and<br>as an adjunctive treatment in<br>the management of post-<br>surgical and post-traumatic<br>acute pain. | The Chattanooga<br>Revolution Wireless is a<br>clinical electrotherapy<br>device intended for use<br>under the supervision of<br>a Healthcare<br>Professional.<br>As a TENS device,<br>indications are for the<br>following conditions:<br>- Symptomatic relief and<br>management of chronic,<br>intractable pain - Post-<br>surgical and post-trauma<br>acute pain | The TrueRelief Device is<br>intended for use under<br>the supervision of a<br>Healthcare Professional.<br><br>TrueRelief is intended for<br>temporary symptomatic<br>relief of chronic<br>intractable pain, and as<br>an adjunctive treatment in<br>the management of post-<br>surgical and post-<br>traumatic acute pain. |
| Prescription/<br>Professional Use<br>Only or OTC | Prescription | Prescription/<br>Professional Use Only | Prescription/<br>Professional Use Only |
| Where used | Physician office, physical<br>therapy clinic, hospital nursing<br>home, post acute care,<br>chiropractic clinic | Physician office,<br>physical therapy clinic,<br>hospital nursing home,<br>post acute care,<br>chiropractic clinic | Physician office, physical<br>therapy clinic, hospital<br>nursing home, post acute<br>care, chiropractic clinic |
| Target population | Adult population | Adult population | Adult population |
| Battery or mains<br>powered | Mains | Mains and battery<br>operated | Mains |
| Principle of<br>Operation | External nerve stimulator that<br>generates an electrical pulse<br>that is delivered to the tissue<br>through the primary and<br>secondary probes | External nerve stimulator<br>that generates an<br>electrical pulse that is<br>delivered to the tissue<br>through the primary and<br>secondary probes | External nerve stimulator<br>that generates an<br>electrical pulse that is<br>delivered to the tissue<br>through the primary and<br>secondary probes |
| Sterility | Non-Sterile | Non-sterile | Non-sterile |
| Materials | Patient contact material: 316L<br>stainless steel | Patient contact material:<br>stainless steel | Patient contact material:<br>316L stainless steel |
| Connection of<br>device to<br>electrodes | The device uses two<br>electrodes(probes) using a<br>single channel. At one end of<br>each probe is a 316 stainless | Stimulation Module is<br>directly connected to the<br>custom male SNAP<br>assembled in the | The device uses two<br>electrodes(probes) using<br>a single channel. At one<br>end of each probe is a |
| Item | Primary Predicate Device<br>K070474 TMR | Reference Device<br>K153696 | Subject Device<br>TrueRelief |
| | steel tip that touches the<br>patient and the other end of<br>the stainless steel tip screws<br>into the probe. On the other<br>end of the probe is a cable<br>which connects each probe to<br>the device using a connector<br>with pins. | electrode. User Interface<br>(LCD and buttons) is<br>physically separated<br>(Remote Control) and<br>communicates wirelessly<br>with up to four (4)<br>stimulation modules.<br>Stimulation safety<br>remains fully managed<br>by Stimulation Module<br>electronic circuit itself. | 316 stainless steel tip that<br>touches the patient and<br>the other end of the<br>stainless steel tip screws<br>into the probe. On the other<br>end of the probe is<br>a cable which connects<br>each probe to the device<br>using a connector with<br>pins. |
| Power Supply | 115 VAC, 60Hz input<br>12 volt DC output | Rechargeable battery | 115 VAC, 60Hz input<br>12 volt DC output |
| Electrical Type | Type BF | NA<br>Battery operated device | Type BF |
| Patient Leakage<br>Current - Normal<br>Condition (µA) | <100uA patient leakage | NA<br>Battery operateddevice | <100uA patient leakage |
| Patient Leakage<br>Current - Single<br>Fault Condition<br>(μΑ) | <300uA line leakage | NA<br>Battery operateddevice | <300uA line leakage |
| Number of<br>Output Channels | 1 | 0, 2, or 4 | 1 |
| Method of<br>Channel Isolation | Transformer isolated. | Each channel is the<br>middle of a H Bridge.<br>Except when it is<br>activated, each<br>channel is always in high<br>impedance state | Transformer isolated. |
| Regulated<br>Current or<br>Requlated<br>Voltage (output<br>signals only) | Regulated current on only<br>channel | Regulated current on all<br>channels | Regulated current on<br>only channel |
| Maximum<br>Current (RMS)<br>Density<br>(mA/cm2) | 14mA/cm2<br>Average of 2.1mA/cm2 | 1.34mA/cm2 | 14mA/cm2<br>Average of 2.1mA/cm2 |
| Maximum Power<br>Density<br>[mW/cm²] | 79 mW/cm2 @5000<br>Average of 11mW/cm2<br>@5000 | 14.4 [mW/cm²] @500Ω | 79 mW/cm2 @5000<br>Average of 11mW/cm2<br>@5000 |
| Output Voltage | Range of normal use: 50-60 V<br>Peak pulse amplitude: 200 V | 60 V @ 500 Ω<br>180V @ 2 kΩ<br>180 V @ 10 kΩ | Range of normal use: 50-<br>60 V<br>Peak pulse amplitude:<br>200 V |
| Pulse Rate | 1 to 400 (+10%) Pulse /<br>Second (Low)<br>4,000 (-5%) to 20,000 (+10%)<br>Pulse/Second (High) | 35 - 80 Hz | 1 to 400 (+10%) Pulse /<br>Second (Low)<br>4,000 (-5%) to 20,000<br>(+10%) Pulse/Second<br>(High) |
| Item | Primary Predicate Device<br>K070474 TMR | Reference Device<br>K153696 | Subject Device<br>TrueRelief |
| Pulse Duration | 0.24-0.74 millisecond (Low)<br>15-25 microsecond (High) | 300 to 400 [μs]<br>(microseconds) | 0.24-0.74 millisecond<br>(Low)<br>15-25 microsecond (High) |
| Output Current<br>(maximum) | 8.9 milliamps (at body<br>resistance > 10.11K ohms) | 120 mA @ 500 Ω<br>90 mA @ 2 kΩ<br>18 mA @ 10 kΩ | 8.9 milliamps (at body<br>resistance > 10.11K<br>ohms) |
| Maximum Charge<br>Per Pulse | 7 μC | 36 [μC] @ 500Ω | 7 μC |
| Wave Form<br>Shape | Rectangular | Rectangular | Rectangular |
| Maximum<br>Amplitude | No load - 200 V Peak<br>With 50K ohm load - 175 V<br>Peak | 60 V @ 500 Ω<br>180V @ 2 kΩ<br>180 V @ 10 kΩ | No load - 200 V Peak<br>With 50K ohm load - 175<br>V Peak |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for TrueRelief. The logo consists of the word "TrueRelief" in a sans-serif font, with "True" in a dark blue color and "Relief" in a lighter blue color. Above the word is a graphic of four diamond shapes in a row, with the colors of the diamonds being orange, pink, purple, and blue.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for TrueRelief. Above the name is a graphic of a sound wave in different shades of purple and orange. The name TrueRelief is written in a combination of blue and purple.
## Summary of Performance Testing:
Electrical Safety: The TrueRelief Device was tested and certified to comply with recognized standards for electrical safety (IEC 60601-1, IEC 60601-2-10, IEC 60601-1-6).
Electromagnetic Compatibility: The TrueRelief Device was tested and certified to comply with recognized standards for electromagnetic compatibility (IEC 60601-1-2).
Usability/Human Factors: Usability/Human Factors were evaluated, which demonstrated that the established requirements for usability were met, and the device's design is appropriate for the intended users and use environment. The result of this evaluation substantiates the acceptability of the use-related risks identified during the risk assessment activities.
## Conclusion:
Based on the performance testing and the similarities of the indications for use and the technological characteristics, it can be concluded that the TrueRelief Device is as safe and effective as, and substantially equivalent to, the predicate device(s).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.