Python catheters are indicated for removal of thromboemboli from the peripheral arterial system, and for occlusion and infusion of fluids into a vessel. The Over-the-Wire Latis Graft Cleaning Catheter is indicated for the removal of thromboemboli from vascular grafts, and for occlusion and infusion of fluids into a graft.
Device Story
Python and Over-the-Wire Latis catheters are sterile, single-use vascular balloon catheters. Devices feature dual-lumen design: one lumen for balloon inflation, one for guidewire insertion or fluid infusion. Latis Graft Cleaning Catheter includes braided mesh over balloon to increase wall engagement in artificial vascular grafts. Used in operating rooms by clinicians. Physician inserts catheter into arterial vessel or graft; inflates balloon to remove thromboemboli or occlude vessel; performs fluid infusion as needed. Output is mechanical removal of thrombus or vessel occlusion. Benefits include restoration of blood flow and localized fluid delivery.
Clinical Evidence
Bench testing only. Performance evaluated per ISO 10555-1:2017. Testing included balloon pull force, inflation/hub pull force, tensile strength, balloon cycling, burst pressure, diameter inspection, leak testing, simulated use bend test, torsion, and kink testing (EN 13868). Biocompatibility testing included cytotoxicity, irritation, sensitization, acute systemic toxicity, hemolysis, pyrogenicity, platelet/leukocyte counts, PTT, complement activation, and thrombogenicity.
Technological Characteristics
Dual-lumen vascular balloon catheter. Materials: polymers and metal. Braided mesh on Latis model. Sterilization: Ethylene Oxide. ISO 10555-1:2017 and EN 13868 compliance. External communicating device, blood contact < 24 hours.
Indications for Use
Indicated for removal of thromboemboli from peripheral arterial system or vascular grafts, and for occlusion and infusion of fluids into vessels or grafts. For use in patients requiring embolectomy or fluid infusion in arterial/graft settings.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 7, 2021
Applied Medical Resources Corporation Jessica Piovarcsik Director, Regulatory Affairs 22872 Avenida Empresa Rancho Santa Margarita, California 92688
# Re: K202049
Trade/Device Name: Python Catheter/Over-the-Wire Latis Graft Cleaning Catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: DXE, KRA Dated: August 6, 2021 Received: August 9, 2021
# Dear Jessica Piovarcsik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or post marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202049
Device Name Python Catheter
Indications for Use (Describe)
Python catheters are indicated for removal of thromboemboli from the peripheral arterial system, and for occlusion and infusion of fluids into a vessel.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202049
#### Device Name
Over-the-Wire Latis Graft Cleaning Catheter
Indications for Use (Describe)
The Over-the-Wire Latis Graft Cleaning Catheter is indicated for the removal of thromboemboli from vascular grafts, and for occlusion and infusion of fluids into a graft.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary – K202049
| 510(k) Submitter: | Applied Medical Resources Corporation<br>22872 Avenida Empresa<br>Rancho Santa Margarita, CA 92688<br>(949) 713 - 8000 |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jessica Piovarcsik<br>Director, Regulatory Affairs<br>Applied Medical Resources Corporation<br>Tel: (949) 713-7958<br>Fax: (949) 713-8205<br>Email: jessica.piovarcsik@appliedmedical.com |
| Preparation Date: | September 7, 2021 |
| Trade Name: | Python® Embolectomy Catheter & Over-the-Wire Latis® Graft Cleaning<br>Catheter |
| Common Name: | Embolectomy Catheter |
| Device<br>Classification: | Product codes<br>DXE - Catheter, Embolectomy (21 CFR 870.5150)<br>KRA - Catheter, Continuous Flush (21 CFR 870.1210)<br>Device Class: Class II |
| Predicate<br>Devices: | Python Catheter<br>510(k)#: K970762<br>Product Code: DXE |
| | Over-the-Wire Latis Graft Cleaning Catheter<br>510(k)#: K973465<br>Produce Code: DXE |
| Model Numbers: | A4E01 CE0340DR A4GW6<br>A4E02 CE0380DR<br>A4E03 CE0440DR<br>A4E04 CE0480DR<br>A4E05 CE0540DR<br>A4E06 CE0580DR<br>A4E07 CE0680DR<br>A4E08<br>A4E09 |
| Sterilization<br>Method:<br>Device<br>Descriptions: | Ethylene Oxide<br>The Python Embolectomy Catheter & Over-the-Wire Latis Graft Cleaning<br>Catheters are sterile, single use vascular balloon catheters intended for use in the |
| | removal of thromboemboli, occlusion of vessels, and the infusion of fluids. The<br>devices are intended for use in arterial vessels and artificial vascular grafts<br>respectively. |
| | The body of the catheters utilize a dual lumen design with one inflation lumen<br>used to inflate the balloon, and one through lumen allowing for insertion over a<br>guide wire or infusion of fluids. The outer diameter of the body is designed to<br>accommodate use with a standard percutaneous access sheath. The Over-the-wire<br>Latis Graft Cleaning Catheter includes a braided mesh which surrounds the<br>balloon and is designed to increase engagement with wall of artificial vascular<br>graft. |
| General type of<br>material used: | Polymers, metal |
| Indications for<br>Use: | Python catheters are indicated for removal of thromboemboli from the peripheral<br>arterial system and for occlusion and infusion of fluids into a vessel. |
| | The Over-the-Wire Latis Graft Cleaning catheter is indicated for the removal of<br>thromboemboli from vascular grafts and for occlusion and infusion of fluids into<br>a graft. |
| Environment of<br>use: | Operating rooms |
| Duration and type<br>of contact: | External communicating device in contact with circulating blood for less than 24<br>hours. |
{5}------------------------------------------------
## Summary of Technological Characteristics between Subject and Predicate Devices:
The subject and predicated devices are single use, vascular balloon catheters designed for the removal of thromboemboli from the arterial system and artificial vascular grafts. The devices are offered at in range of sizes varying with respect to lengths and inflated balloon diameter sizes to accommodate use in the vessels within their respective indications for use.
Modifications have been made to the design & materials of the subject devices, however, the underlying dual lumen design remains the same as their respective predicate device.
#### Differences in the indications for use statements:
Minor edits were made to the indications for use statements: however, these were simply for clarification and do not alter the meaning of the indications or the intended use of the devices.
The intended use of the subject Over-the-Wire Latis Graft Cleaning catheter has not changed. However, the indications for use statement was clarified to align with the terminology commonly utilized in the clinical environment by updating the term "thrombus" to "thromboemboli". The terms "thrombus" and "thromboembli" are used interchangeably within the clinical setting. Therefore, the change is not considered an expansion of the indications for use.
{6}------------------------------------------------
The intended use of the Python catheter, has not changed. However, the indications for use were modified slightly to:
- . Further clarify the devices should be used to remove thomboemboli from the peripheral arterial system: and
- Remove the term "temporary" from the intended use of occluding the vessel. ●
Peripheral arterial thromboembolisms are encompassed within the arterial thromboemboli stated in the indications for use of the predicate device. Therefore, the modification was only made to provide further clarity and is not considered an expansion of the intended use of the device.
The term "temporary" was the term is subjective and the duration of use of the device in occluding the vessel is intuitive to the user. Applied Medical has not changed the classification of the patient contact duration with respect to biocompatibility (i.e. in contact with blood < 24 hrs)
# Discussion of Performance Testing:
The FDA recognized consensus standard International Standard Organization (ISO) 10555-1:ed2:2017 Intravascular catheters – Sterile and single -use catheters – Part 1: General Requirements was considered when evaluating the performance of the subject devices. The following test endpoints were utilized to demonstrate substantial equivalence to their respective predicate devices.
## Engineering/Bench Testing
- Balloon Pull Force (ISO 10555-1, Annex B) ●
- Inflation & Guidewire Hub Pull Force (ISO 10555-1, Annex B) ●
- Catheter Body Tensile Strength ●
- Catheter Balloon Cycling Test ●
- Balloon Burst Pressure Test
- Inflated Balloon Diameter Inspection
- Guidewire & Inflation Hub Leak Test (ISO 10555-1, Annex C)
- Simulated Use Bend Test
- Torsion Test
- Kink Test (EN 13868, Annex B) ●
# Biocompatibility Testing:
- Cytotoxicity
- Intracutaneous Irritation
- Sensitization
- Acute Systemic Toxicity ●
- Hemolysis
- Materials Mediated Pyrogenicity
- Platelet and Leukocyte Counts
- Partial Thromboplastin Time ●
- Complement Activation ●
- Thrombogenicity .
{7}------------------------------------------------
# Conclusion:
Both the subject Python Catheters and subject Over-the-wire Latis Graft Cleaning Catheters, with modified materials, are substantially equivalent in performance and intended use to the predicate Dual Lumen Embolectomy Catheters and Dual Lumen Graft Cleaning Catheter, respectively.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.