K202044 · Hanuri Distribution, Inc. · IRP · Oct 21, 2020 · Physical Medicine
Device Facts
Record ID
K202044
Device Name
Powerpress Recovery Unit
Applicant
Hanuri Distribution, Inc.
Product Code
IRP · Physical Medicine
Decision Date
Oct 21, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Powerpress Recovery Unit is indicated for the temporary relief of minor muscle aches and pains, and for temporary increase in circulation to the treated areas in people who are in good health. The Powerpress Recovery Unit simulates kneading and stroking of tissues by using an inflatable garment.
Device Story
Sequential pneumatic compression device; applies gradient pressure to limbs via 4-chamber inflatable garment. User-adjusted manual/analog pump controls pressure (30-100 mmHg); sequential inflation from distal to proximal chambers; constant pressure held until all chambers inflated; simultaneous release. Used in home, athlete training, and clinic environments; operated by patient or clinician. Provides kneading/stroking simulation to increase circulation and relieve muscle aches. Benefits include temporary pain relief and circulation improvement. No software or automated logic; relies on motorized valve for fixed cycle timing.
Clinical Evidence
No clinical data. Bench testing only. Device verified against IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-6 (usability), IEC 60601-1-11 (home healthcare), and ISO 10993 (skin sensitization).
Indicated for temporary relief of minor muscle aches and pains and temporary increase in circulation in healthy individuals. Contraindicated for patients with inflammation, infection, unexplained pain, bleeding, wounds, sensitive skin, poor circulation, acute pulmonary edema, acute thrombophlebitis, acute congestive cardiac failure, acute infections, DVT, pulmonary embolism, tumors, or bone fractures at the application site.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 21, 2020
Hanuri Distribution, Inc Jung Moon President 9601 Owensmouth Ave. # 8 Chatsworth, California 91311
Re: K202044
Trade/Device Name: Powerpress Recovery Unit Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: July 13, 2020 Received: July 23, 2020
Dear Jung Moon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name POWERPRESS RECOVERY UNIT
Indications for Use (Describe)
The Powerpress Recovery Unit is indicated for the temporary relief of minor muscle aches and pains, and for temporary increase in circulation to the treated areas in good health. The Powerpress Recovery Unit simulates kneading and stroking of tissues by using an inflatable garment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
(as required by 21 CFR 807.92)
#### A. 510K Number : K202044
- B. Date: October 20, 2020
#### C. Applicant
Company & Manufacturer Name: Hanuri Distribution, Inc Address: 9601 Owensmouth Ave. # 8 Chatsworth, CA 91311 USA Owner & Contact Person: Jung H Moon Phone: 818-998-1023, Fax: 818-998-0277 Email: info@air1000.com
#### D. Regulatory Information
- 1. Proprietary/Trade Name: Powerpress Recovery Unit
- 2. Regulation Name (Common/Usual Name): Powered inflatable tube massager
- 3. Regulation Number: 21 CFR 890.5650
- 4. Product Code: IRP
- 5. Regulatory Class: Class II
- E. Indications for Use
The Powerpress Recovery Unit is indicated for the temporary relief of minor muscle aches and pains, and for temporary increase in circulation to the treated areas in people who are in good health. The Powerpress Recovery Unit simulates kneading and stroking of tissues by using an inflatable garment.
- F. Environment of Use: Home, Athlete Training and Clinics environments
- G. Warning:
Do not use the compressor to direct pressurized air towards your eyes, nose, mouth, or ears. Doing so may lead to a serious injury
- H. Predicate Device
Powerpress Recovery Unit is substantially equivalent to the following
| Predicate Device | Manufacturer | 510(k)# |
|--------------------------------------------------------------|------------------|---------|
| DJS Massager; Recovery<br>Pump 701-E; Recovery Pump<br>701RA | MEGO AFEK AC LTD | K112479 |
#### I. Device Description
Powerpress Recovery Unit is a sequential pneumatic compression device designed to apply compression to a limb. The device is composed of two components.
{5}------------------------------------------------
- Pneumatic Manual / Analog Pump .
- . Limb Sleeve or garment composes of 4 chambers
Powerpress Recovery Unit enables different treatment pressure (30 ~ 100mmHg). When activated, air flows into chamber, the pump provides gradient pressurization to the chambers (sequential inflation of distal to proximal, with distal chambers inflated to a greater pressure than the proximal ones).
After each chamber is inflated, the pressure is held constant until all chambers are inflated, in order to prevent reverse gradient flow. Once all chambers are inflated, they are then all released simultaneously, and the cycle repeats. Pressure within chambers are adjustable pressure to chamber 1 is controlled by user-adjusted regulator on the pump. Pressure in chamber 2, 3 & 4 are individually lowered according to the default factory setting.
Gradient: appx 7%, example: Foot 60mmHg – Ankle 56mmHg – Calf 52mmHg - Thigh 48mmHg Cycle Time: Inflation 36 sec / Deflation 24 sec Inflation time each chamber : 1* chamber(foot) – 36 sec, 2nd chamber(ankle) – 27 sec, 3rd chamber(calf) - 18 sec, 4th chamber(thigh) - 9 sec
A calibrated dial gauge displays pressure in the range of 0 ~ 160 mmHg
- J. Substantial Equivalence of Technological Characteristics
Powerpress Recovery Unit Model PR1000 Massage Systems and the predicate device all provide sequential inflation pressure from distal to proximal segments. Both the Powerpress Recovery Unit and predicate devices offer adjustable pressure ranges.
The Powerpress Recovery Unit have decreasing gradient pressures from distal to proximal segments.while predicate device does not have decreasing gradient pressures. The Powerpress Recovery Unit provide continuously adjustable pressure in the range between 30 to 100 mmHg. while the predicate device provides three pressure adjustments options in increments of low, medium and high pressure ranges within the 20 to 80 mmHg range. The Powerpress Recovery Unit have fixed cycle times that are fixed by a motorized valve, and are independent of garment size and number (one or two garments). The predicate device does not have a motorized valve and uses back pressure to trigger the cycle to the next chamber of the garment.
A difference in the subject device from the predicate is that the Powerpress recovery unit has a pressure range from 30 mmHg to 100 mmHg while predicate device has from 20 mmHg to 80 mmHg. But there are FDA cleared devices for same intended use, which are shown in the table below, prove that our pressure range raises no new questions of safety and effectiveness. Also Electrical Safety Test (IEC 60601-1:2005+AMD1:2012) verifies that the technological differences in the fuse rating, power consumption, electric power plug, size and net weight of device do not raise different questions of safety and effectiveness.
{6}------------------------------------------------
| MANUFACTURER | 510(k)<br>Number | TRADE NAME | PRESSURE RANGE | PRESCRIPTION OR<br>OTC |
|-------------------------|------------------|------------------------------------------|---------------------|------------------------|
| MEGO AFEK AC LTD | K140519 | Recovery Pump 737R | 20 mmHg to 100 mmHg | OTC |
| NormaTec Industries, LP | K160608 | NormaTec Pulse and<br>NormaTec Pulse Pro | 30 mmHg to 110 mmHg | OTC |
The Powerpress Recovery Unit operate within clinically-established parameters. The differences between the predicate and the applicant devices do not impact safety or effectiveness. A table illustrating the substantial equivalence comparison table is provided below.
## K. Substantial Equivalence Comparison Table
| | Subject Device | Predicate Device | Comment |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Manufacturer | Hanuri Distribution, Inc | MEGO AFEK AC LTD | Different |
| 510(k) Number | K202044 | K112479 | N/A |
| Proprietary /<br>Trade Name | Powerpress Recovery Unit | DJS Recovery Pump 701RA | Different |
| Regulation Name | Powered Inflatable Tube Massager | Powered Inflatable Tube Massager | Identical |
| Regulation<br>Number | 21 CFR 890.5650 | 21 CFR 890.5650 | Identical |
| Product Code | IRP | IRP | Identical |
| Regulatory Class | II | II | Identical |
| Indications for<br>Use | The Powerpress Recovery Unit is indicated for<br>the temporary relief of minor muscle aches and<br>pains, and for temporary increase in circulation<br>to the treated areas in people who are in good<br>health. The Powerpress Recovery Unit<br>simulates kneading and stroking of tissues by<br>using inflatable garment. | The DJS Massager is indicated for the<br>temporary relief of minor muscle aches and<br>pains, and for temporary increase in<br>circulation to the treated areas in people who<br>are in good health. The DJS Massager<br>simulates kneading and stroking of tissues by<br>using inflatable garment. | Identical |
| Contraindications | * Do not use this product if you are<br>experiencing inflammation, an infection, pain<br>of unknown origin, or bleeding (internal or<br>external) at or near the site of application or if<br>you have a wound at or near the site of<br>application.<br>* Do not use this product if you are under the<br>care of a physician or have a condition<br>requiring the use of any medical device.<br>* Do not use this product on sensitive skin or<br>on in the presence of poor circulation.<br>* Do not use this product if you have any of the<br>following conditions:<br>o Acute pulmonary edema<br>o Acute thrombophlebitis<br>o Acute congestive cardiac failure<br>o Acute infections | *This device is intended for use by people in<br>good health. This device is not recommended<br>for people who have heart problems, or<br>vascular problems, have a condition<br>requiring the use of any medical device, or<br>have any condition that might affect their<br>normal well being.<br>*If you are, or may be, pregnant, consult<br>with your physician before use.<br>*Do not use this device over insensitive or<br>numb areas, or phlebitis. This device should<br>not be used over swollen or inflamed areas or<br>skin eruptions. Do not use in the presence of<br>unexplained calf pain.<br>*Consult your physician prior to use | Similar |
| | o Deep Vein Thrombosis (DVT)<br>o Episodes of Pulmonary embolism<br>o Wounds lesions or tumors at or in the<br>vicinity of application<br>o Where increased venous and lymphatic<br>return is undesirable<br>o Bone fractures or dislocations at or in the<br>vicinity of application<br>o Please consult a physician before using this<br>device if:<br>a) If you are pregnant for feel weak<br>b) If you have an implant at the site of<br>application<br>c) If you have a cardiac pacemaker | | |
| Prescriptive or<br>OTC | OTC | OTC | Identical |
| Pressure Range | 30 - 100 mmHg | 20 – 80 mmHg | Similar |
| Time Setting | No Timer, Continuous Mode | No Timer, Continuous Mode | Identical |
| Compression<br>Mode | Gradient Sequential | Gradient Sequential | Identical |
| Number of Mode | 1 Mode | 1 Mode | Identical |
| Chamber of<br>Garment | 4 Chamber (Air Compartments Garment)<br>Overlap Each Chamber | 4 Chamber (Air Compartments Garment)<br>Overlap Each Chamber | Identical |
| Treatment Time | 15 min to 45 min | 15 min to 45 min | Identical |
| Pressure Lock<br>Button | YES | YES | Identical |
| Single or Double<br>Garment<br>Available | YES | YES | Identical |
| Overload<br>Protection Fuses | This device is equipped with two overload<br>fuses, on each of the power lines | This device is equipped with two overload<br>fuses, on each of the power lines | Identical |
| Size of Device | H 5.12" x L 11.81" x W 8.27" | H 3.9" x L 10.2" x W 5.1" | Similar |
| Net Weight | 4.5 LBS | 5.1 LBS | Similar |
| Power<br>Consumption | 20 watts | 11 watts | Similar |
| Power Source(s) | 115V / 50 or 60 Hz | 115V / 50 or 60 Hz | Identical |
| Fuse Rating | 0.5 A | 2 A | Similar |
| Electric Plug | 2 Prong Plug | Three-pin Plug | Similar |
| Software | Manual / Analog (not software driven) | Manual / Analog (not software driven) | Identical |
| Material of<br>Garments | 200 denier nylon with a Polyurethane | 200 denier nylon with a Polyurethane | Identical |
{7}------------------------------------------------
## L. Compression Garment Description
- a) List all types of compression garments
Half Leg Garment Large Item No. 1016 Half Leg Garment X Large Item No. 1016-A Full Leg Garment Large Item No. 1014 Full Leg Garment X Large Item No. 1014-A
{8}------------------------------------------------
- b) Description of every type of compression garment
- 1. Half Leg Garment Large Item No. 1016 4 Chamber Overlapping Sequential Compression Garment Length from Floor to Top 19.5" / Length from Ankle to Top 18" Top (Knee) Circumference Measurement: 20" Mid Calf Circumference Measurement: 18" Ankle Circumference Measurement: 15"
- 2. Half Leg Garment X Large Item No. 1016-A 4 Chamber Overlapping Sequential Compression Garment Length from Floor to Top 19.5" / Length from Ankle to Top 18" Top (Knee) Circumference Measurement: 25" Mid Calf Circumference Measurement: 23" Ankle Circumference Measurement: 19"
- 3. Full Leg Garment Large Item No. 1014 4 Chamber Overlapping Sequential Compression Garment Length from Floor to Top 30" / Length from Ankle to Top 28" Top (Mid Thigh) Circumference Measurement: 24" Knee Circumference Measurement: 20" Mid Calf Circumference Measurement: 18" Ankle Circumference Measurement: 15"
- 4. Full Leg Garment Large Item No. 1014-A 4 Chamber Overlapping Sequential Compression Garment Length from Floor to Top 30" / Length from Ankle to Top 28" Top (Mid Thigh) Circumference Measurement: 30" Knee Circumference Measurement: 25" Mid Calf Circumference Measurement: 23" Ankle Circumference Measurement: 19"
- c) Identify whether each type of compression garment you intend to market with your device has received prior 510(k) clearance or has not received prior 510(k) clearance
#### All of above compression garment has received prior 510(k) clearance. 510(k) number K110276
#### M. Technological Characteristics
The manufacturer believes that the technological characteristic of the Powerpress Recovery Unit is substantially similar to those of the predicate devices.
Powerpress Recovery Unit has very similar components to its predicate devices and very similar principles of operation.
- N. Performance Testing Bench:
{9}------------------------------------------------
The device has been tested to the requirements of the following standards:
IEC 60601-1:2005+AMD1:2012 .
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2: 2014
Collateral standard: Electromagnetic Compatibility - Requirements and Tests
- IEC 60601-1-6:2010+AMD1:2013 ●
Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 60601-1-11: 2015
Collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- ISO 10993 Compression Garment Skin Sensitization Test Report
Animal: No animal testing was performed
Clinical: No clinical testing was performed
#### O. Conclusion
Powerpress recovery unit is substantially equivalent to the predicates in: indications for use, contraindications, patient population, environment of use, technology characteristics, materials, specifications / performance and compliance with international standards. Minor differences as detailed in the Substantial Equivalence of Technological Characteristics above do not raise questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.