K202031 · Remed Co., Ltd. · IPF · May 6, 2021 · Physical Medicine
Device Facts
Record ID
K202031
Device Name
Talent-Pro Electromagnetic Stimulator
Applicant
Remed Co., Ltd.
Product Code
IPF · Physical Medicine
Decision Date
May 6, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Talent-Pro™ Electromagnetic Stimulator is intended to be used under medical supervision for adjunctive therapy for the treatment of medical diseases and conditions. The Talent-Pro™ Electromagnetic Stimulator is indicated for use in stimulating neuromuscular tissue for bulk muscle excitation in the legs or arms for rehabilitative purposes.
Device Story
The Talent-Pro Electromagnetic Stimulator is a powered muscle stimulator used under medical supervision in clinical settings. It utilizes magnetic field technology to induce neuromuscular tissue stimulation. The device consists of a main body and two magnetic coil transducers. The operator interacts with the system via a touch screen interface to control stimulation parameters, including pulse frequency (1-150 Hz) and magnetic field intensity (1.00-2.5T). The device delivers biphasic, sine-wave pulses to the patient's limbs to achieve bulk muscle excitation. This process aids in rehabilitative therapy by relaxing spasms, preventing atrophy, increasing circulation, and re-educating muscles. The device is mounted on castors for mobility. Clinical benefits include improved muscle function and prevention of post-surgical complications like venous thrombosis.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993), design verification, and electrical safety/EMC testing (IEC 60601 series).
Technological Characteristics
Powered muscle stimulator using magnetic field technology. Features two magnetic coil applicators, 8-inch color touch screen, and continuous operation mode. Pulse parameters: 1-150 Hz frequency, 280μs pulse duration, 1.00-2.5T intensity. Electrical safety: Class II, BF. Connectivity: Standalone system. Sterilization: Not applicable (non-invasive).
Indications for Use
Indicated for patients requiring neuromuscular stimulation for bulk muscle excitation in legs or arms for rehabilitation. Specific uses: relaxation of muscle spasms, prevention/retardation of disuse atrophy, increasing local blood circulation, muscle re-education, immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, and maintaining/increasing range of motion.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date May 6, 2021. The text is in a simple, sans-serif font. The date is written in the standard month-day-year format. The text is black on a white background.
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Remed Co., Ltd % Kyungyoon Kang CEO K-Biotech 201 South 4th Street. Suite 727 San Jose. California 95112
Re: K202031
Trade/Device Name: Talent-Pro Electromagnetic Stimulator Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: IPF Dated: January 25, 2021 Received: February 8, 2021
Dear Kyungyoon Kang:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations af fecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equi valence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory to pic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Patrick Antkowiak. PhD Assistant Director (Acting) DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
### 510(k) Number (if known) K202031
Device Name
Talent-ProTM Electromagnetic Stimulator
Indications for Use (Describe)
The Talent-Pro™ Electromagnetic Stimulator is intended to be used under medical supervision for adjunctive therapy for the treatment of medical diseases and conditions.
The Talent-Pro™ Electromagnetic Stimulator is indicated for use in stimulating neuromuscular tissue for bulk muscle excitation in the legs or arms for rehabilitative purposes.
Indications for Use for Muscle Stimulators:
- Relaxation of muscle spasms
- Prevention or retardation of disuse atrophy
- Increasing local blood circulation
- Muscle re-education
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- Maintaining or increasing range of motion
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:10pt"> <b> </b> </span> <span style="font-size:10pt">Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:10pt"> <b> </b> </span> <span style="font-size:10pt">Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Pursuant to Section 510(k) of Chapter V of the Federal Food, Drug, and Cosmetic Act and in accordance with subpart E of Part 807, Title 21 of the Code of Federal Regulations, REMED Co., Ltd. submits the following information as premarket notification for the proposed device, Talent-Pro Electromagnetic Stimulator.
- l. SUBMITTER
- Company: REMED Co., Ltd. ●
- Address: #301-#303 Migun Techno World II, 187, Techno 2-Ro, Yuseong-gu, Daejeon, Republic of Korea, Postal code: 34025
- Tel: 82-42-934-5560
- Fax: 82-42-934-5562 ●
- 510(k) Number: TBD .
510(k) Lead: Seong Hyeon Kim, K-Bio Solutions Manager (sarah.kim @kbiotechsolutions.com, Tel: 82-2-597-2700)
Date Prepared: June 22nd, 2020
#### II. PROPOSED DEVICE
- Type of Submission: Traditional Premarket Notification ●
- Trade Name of Device: Talent-Pro Electromagnetic Stimulator ●
- Classification Name of Device: Stimulator, Muscle, Powered
- Review Panel: Physical Medicine ●
- . Regulation Number: 21 CFR 890.5850
- Requlatory Class: Class II
- Product Code: IPF ●
#### = PREDICATE DEVICE
HPM-6000 (K160992, Manufacturer: BTL Industries, Inc.) The predicate device has not been subject to a design-related recall.
#### DEVICE DESCRIPTION IV.
The Talent-Pro Electromagnetic Stimulator is indicated for use in stimulating neuromuscular tissue for bulk muscle excitation in the legs or arms for rehabilitative purposes, prevention or retardation of disuse atrophy. Talent-Pro Electromagnetic Stimulator consists of the Main body, Transducer for magnetic stimulation
The proposed, Talent-Pro Electromagnetic Stimulator is substantially equivalent to the predicate device. HPM-6000 (K160992. Manufacturer: BTL Industries. Inc.),
{4}------------------------------------------------
The intended use and indications for use of the proposed device are identical to the predicate device. The fundamental scientific, and technological characteristics of the proposed Talent-Pro Electromagnetic Stimulator and principles of Talent-Pro Electromagnetic Stimulator operation are similar to the predicate, except differences primarily shown in the product dimensions, which are assessed not to raise different questions in terms of safety and effectiveness compared to the predicate device.
- V. INDICATIONS FOR USE
The Talent-Pro™ Electromagnetic Stimulator is intended to be used under medical supervision for adjunctive therapy for the treatment of medical diseases and conditions.
The Talent-Pro™ Electromagnetic Stimulator is indicated for use in stimulating neuromuscular tissue for bulk muscle excitation in the legs or arms for rehabilitative purposes.
Indications for Use for Muscle Stimulators:
- Relaxation of muscle spasms -
- -Prevention or retardation of disuse atrophy
- Increasing local blood circulation ।
- -Muscle re-education
- -Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
- Maintaining or increasing range of motion -
- VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The proposed Talent-Pro Electromagnetic Stimulator and predicate device, HPM-6000 (K160992) share the same intended use and indications for use.
The fundamental technological characteristics and critical functional features of the Talent-Pro Electromagnetic Stimulator are similar to the predicate device, HPM-6000 (K160992). The minor differences in the external appearance and dimensions of the products do not raise different questions in safety and effectiveness compared to the predicate device. The favorable results of the design verification and electrical safety testing of the Talent-Pro Electromagnetic Stimulator have demonstrated that the proposed device has met the predetermined design requirements and the FDA's recognized standards, including the electrical safety and performance standards and EMC standards as the same as the predicate device.
The design and function of the Talent-Pro Electromagnetic Stimulator do not involve a new or different technological principle that would be considered to raise different questions in safety and effectiveness compared to the predicate device. The principle of operation, critical function and clinical use and application are the same as the predicate device, HPM-6000 (K160992).
{5}------------------------------------------------
Detailed comparisons of intended use/indications for use, technological characteristics and critical functional features between the proposed and predicate devices are provided in Table 1 below.
| Comparison<br>Item | Proposed Device | Predicate Device | Assessment of<br>Substantial Equivalence |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade/Device<br>Name | Talent-Pro<br>Electromagnetic<br>Stimulator | HPM-6000 | N/A |
| Product Code<br>and Regulation | 21 CFR 890.5850<br>IPF-Stimulator,<br>Muscle, Powered | 21 CFR 890.5850<br>IPF-Stimulator,<br>Muscle, Powered | Same |
| 510k Number | N/A | K160992 | N/A |
| Manufacturer | REMED Co., Ltd. | BTL Industries, Inc. | N/A |
| Application | Muscle Stimulation | Muscle Stimulation | Same |
| Intended<br>Use/Indications<br>for Use | The Talent-Pro™<br>Electromagnetic<br>Stimulator is<br>intended to be used<br>under medical<br>supervision for<br>adjunctive therapy<br>for the treatment of<br>medical diseases<br>and conditions.<br><br>The Talent-Pro™<br>Electromagnetic<br>Stimulator is indicated<br>for use in stimulating<br>neuromuscular tissue<br>for bulk muscle<br>excitation in the legs<br>or arms for<br>rehabilitative<br>purposes. | The HPM-6000 device<br>is intended to be used<br>under medical<br>supervision for<br>adjunctive therapy for<br>the treatment of<br>medical diseases and<br>conditions.<br><br>The HPM-6000 device<br>is indicated<br>for use in stimulating<br>neuromuscular tissue<br>for bulk muscle<br>excitation in the legs or<br>arms for rehabilitative<br>purposes.<br><br>Indications for Use for<br>Muscle Stimulators:<br>-Relaxation of Muscle<br>Spasms<br>-Prevention or<br>Retardation of Disuse<br>Atrophy<br>-Increasing Local<br>Blood Circulation<br>-Muscle Re-education<br>-Immediate Post- | Same |
| | blood circulation<br>- Muscle re-education<br>- Immediate post-<br>surgical stimulation of<br>calf muscles to<br>prevent venous<br>thrombosis<br>- Maintaining or<br>increasing range of<br>motion | Surgical Stimulation of<br>Calf Muscles to<br>Prevent Venous<br>Thrombosis<br>-Maintaining or<br>Increasing Range of<br>Motion | |
| Electrical<br>Protection | Class II, BF | Class II, BF | Same |
| Interface | Touch screen | Touch-Screen | Same |
| Type of Energy | Magnetic Field | Magnetic Field | Same |
| Type of<br>Applicator | Single Magnetic Coil | Single Magnetic Coil | Same |
| Number of<br>Magnetic Coils<br>in the Applicator | 1 | 1 | Same |
| Number of<br>Applicators | 2 | 2 | Same |
| Color Touch<br>Screen | 8 inch<br>20.32 cm<br>800 x 480 pixel | 8.4 inch<br>21.3 cm<br>800 x 400 pixel | Similar<br>A minute difference in the<br>touch screen size, and<br>enhanced resolution<br>quality with a higher level<br>of pixel do not raise new<br>questions of safety and<br>effectiveness. |
| Type of<br>Operation | Continuous | Continuous | Same |
| Pulse<br>Repetition<br>Range | 1-150 Hz | 1-150 Hz | Same |
| Magnetic Field<br>Intensity | Applicator A: 1.00-<br>2.5T<br>Applicator B: 1.00-<br>2.5T | 299-1 applicator:0.5-<br>1.8T<br>299-2 applicator:0.7-<br>2.5T | Similar |
| Pulse Duration<br>(±20%) | 280μs | 280μs | Same |
| Pulse<br>Amplitude | 0-100% | 0-100% | Same |
| Therapy Time | Up to 60 min | Up to 60 min | Same |
| Operating<br>Temperature | 10-30°C | 10-30°C | Same |
| Operating<br>Humidity | 30-85% | 30-75% | Similar |
| Shape of<br>Stimulation<br>Pulse | Sine, Biphasic | Sine, Biphasic | Same |
| Energy Source | 120V~, 60Hz, 1500A | 100-240 V AC, 50-60<br>Hz, Max 14A | Talent-Pro's energy source<br>is set up to be applicable to<br>the US customer's use. |
| External<br>Exchangeable<br>Fuse | Yes | Yes | Same |
| System<br>Dimensions<br>(Width x Height<br>x D/Length) | 541.6 x500.7<br>x1072.9 mm<br>(21.3 x 19.7 x 1072.9<br>in) | 500 x 580 x 230 mm<br>(20 x 23 x 9 in) | Similar, a minor difference<br>in the overall external<br>dimensions is not expected<br>to raise different questions<br>of the device profile in<br>safety and effectiveness<br>compared to the predicate. |
| System Weight | 60 kg (132.3 lb) | 33 kg (73lb) | Similar |
| Position | Vertical-On Castors | Vertical-On Castors | Same |
## Table 1: Substantial Equivalence Assessment
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
#### SAFETY & PERFORMANCE DATA VII.
The following safety and performance data were provided in support of this substantial equivalence determination.
#### A Biocompatibility Testing
In order to meet the FDA's consensus standards for biocompatibility requirements identified in the FDA Guidance titled, "Use of ISO 10993-1, Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing within a Risk Management Process (Issued June 16, 2016)", GLP biocompatibility testing of the proposed, Talent-Pro Electromagnetic Stimulation System has been completed with the favorable results of the biocompatibility testing.
Our biocompatibility risk assessments of the Talent-Pro Electromagnetic Stimulation System have concluded that it is warranted no additional biocompatibility testing of the proposed device is necessary.
# Biocompatibility Testing Evaluations of Talent-Pro Electromagnetic Stimulation System Per ISO 10993 and FDA Guidance (June 16, 2016):
- Cytotoxicity
- Repeated Patch Dermal Sensitization Test (GLP Buehler Method ● Modified for Medical Devices)
- ISO Primary Skin Irritation Test
#### A Design Verification and Validation Testing
Design Verification and Validation (DV&V) testing was performed to verify that the proposed Talent-Pro Electromagnetic Stimulator meets the pre-defined safety and performance requirements. Testing was also conducted to verify the effectiveness of the implemented risk control measures to mitigate the risks identified within the risk management process (per ISO 14971:2012, Medical Devices - Application of Risk Management to Medical Devices). The following design verification, or performance testing of Talent-Pro Electromagnetic Stimulator have been completed with the favorable test results, meeting the applicable ISO standards and FDA's recognized consensus standards related to evaluation of performance of Talent-Pro Electromagnetic Stimulator.
{9}------------------------------------------------
The favorable results of the design verification and validation tests have demonstrated that the design output of the proposed Talent-Pro Electromagnetic Stimulator matches its design input and is appropriate to be used for the intended use, which support its substantially equivalent profile to the predicate device.
The following design verification, validation, electrical safety and performance testing of the Talent-Pro Electromagnetic Stimulator have been comprehensively completed with the favorable test results:
# Design Verification Performed (Safety and Performance Testing)
- CH1 Magnetic Field Output Magnetic Field Strength Test ●
- CH2 Magnetic Field Output Magnetic Field Strength Test ●
- CH1 Magnetic Field Intensity Control Performance Test ●
- CH2 Magnetic Field Intensity Control Performance Test ●
- CH1 Magnetic Field Stimulation Pulse Width Test ●
- CH2 Magnetic Field Stimulation Pulse Width Test ●
- CH1 Magnetic Field Stimulation Pulse Frequency Test ●
- CH2 Magnetic Field Stimulation Pulse Frequency Test
- Stability Test of Magnetic Field Output ●
- Magnetic Field Safety Device Test ●
- Simultaneous Operation Prevention Test
- Medical Device Inspection Test ●
## IEC Electrical Safety and Performance Testing
- IEC 60601-1 .
- IEC 60601-1-6 ●
- IEC 62366 ●
- IEC 60601-2-10 ●
- IEC 60601-1-2 ●
## > Shipping Validation
- . Shipping Validation Test
The results of the shipping validation test demonstrate that the proposed, Talent-Pro Electromagnetic Stimulator has met all the pre-determined shipping validation test requirements.
{10}------------------------------------------------
#### VIII. CONCLUSIONS
In conclusion, the proposed Talent-Pro Electromagnetic Stimulator is assessed to be substantially equivalent to the predicate device. HPM-6000 (K160992). The proposed Talent-Pro Electromagnetic Stimulator shares the same intended use, indications for use and operation principle as the predicate device. HPM-6000 (K160992). The technological characteristics of the Talent-Pro Electromagnetic Stimulator are similar to the predicate and the minor differences are assessed not to raise different questions in terms of safety and effectiveness.
The comprehensive evaluations of the Talent-Pro Electromagnetic Stimulator along with the aforementioned design verification and validation testing assessments provide assurance that the Talent-Pro Electromagnetic Stimulator has met the predetermined design requirements per the FDA's recognized consensus standards which have been applied to the predicate device in the same manner.
The comprehensive assessments all point to the conclusion that the Talent-Pro Electromagnetic Stimulator is substantially equivalent to the predicate, HPM-6000 (K160992).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.