K201984 · Quest Medical, Inc. · DTR · Nov 20, 2020 · Cardiovascular
Device Facts
Record ID
K201984
Device Name
MPS 3 ND Myocardial Protection System
Applicant
Quest Medical, Inc.
Product Code
DTR · Cardiovascular
Decision Date
Nov 20, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4240
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Quest MPS 3 ND Myocardial Protection System, consisting of the Quest MPS 3 ND Console, the Quest MPS 3 Controller, and the Quest MPS 3 ND Disposables (Quest MPS 3 ND Delivery Set and optional Accessories) together is intended for use by perfusionists and physicians to deliver whole blood (from any arterial source) and/or cardioplegia solutions to the heart during open-heart surgery on either an arrested or beating heart for use up to six hours in duration.
Device Story
System aids perfusionists in cardioplegia delivery during cardiopulmonary bypass surgery. Inputs: whole blood from heart-lung machine, crystalloid from IV-bag, and arrest/additive agents. Console contains blood/crystalloid pumps, arrest/additive pumps, pressure/temperature monitors, and ultrasonic air detection sensors. Controller provides touchscreen interface for parameter selection, delivery control, and data logging. System mixes blood, crystalloid, and drugs in specified ratios; regulates flow, pressure, and delivery route. Temperature regulation relies on external heater-cooler connected to disposable heat exchanger. Used in OR by perfusionists/physicians. Output: cardioplegia solution delivered to patient heart. Benefits: precise myocardial protection during surgery.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench, nonclinical, and human factors testing, including software verification, electrical safety, EMC, battery compliance, package stability, leak testing, and simulated use.
Technological Characteristics
System consists of reusable console, touchscreen controller, and disposable delivery sets. Features blood/crystalloid pumps, drug pumps, and ultrasonic air sensors. Temperature regulation via external heater-cooler connected to disposable heat exchanger. Connectivity: sensor interface for monitoring. Software-controlled. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2; battery compliance per IEC 62133-2 and IEC 62281.
Indications for Use
Indicated for use by perfusionists and physicians to deliver whole blood and/or cardioplegia solutions to the heart during open-heart surgery on arrested or beating hearts for up to six hours.
Regulatory Classification
Identification
A cardiopulmonary bypass heat exchanger is a device, consisting of a heat exchange system used in extracorporeal circulation to warm or cool the blood or perfusion fluid flowing through the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 20, 2020
Quest Medical, Inc. Tosan Eweka Regulatory Affairs Supervisor One Allentown Parkway Allen, Texas 75002
Re: K201984
Trade/Device Name: MPS 3 ND Myocardial Protection System Regulation Number: 21 CFR 870.4240 Regulation Name: Cardiopulmonary bypass heat exchanger Regulatory Class: Class II Product Code: DTR Dated: October 21, 2020 Received: October 22, 2020
Dear Tosan Eweka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
#### K201984
Device Name
Quest MPS 3 ND Myocardial Protection System
#### Indications for Use (Describe)
The Quest MPS 3 ND Myocardial Protection System, consisting of the Quest MPS 3 ND Console, the Quest MPS 3 Controller, and the Quest MPS 3 ND Disposables (Quest MPS 3 ND Delivery Set and optional Accessories) together is intended for use by perfusionists and physicians to deliver whole blood (from any arterial source) and/or cardioplegia solutions to the heart during open-heart surgery on either an arrested or beating heart for use up to six hours in duration.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
-
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services -Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Notification K201984
# GENERAL INFORMATION [807.92(a)(1)]
#### Applicant:
Quest Medical, Inc. One Allentown Parkway Allen, TX 75002-4211 USA Phone: 972-332-6338 Fax: 972-390-2881
#### Contact Person:
Tosan Eweka Regulatory Affairs Supervisor Quest Medical, Inc. One Allentown Parkway Allen, TX 75002-4211 USA Phone: 972-332-6338 972-390-2881 Fax: Email: teweka@questmedical.com
#### Date Prepared: October 21, 2020
# DEVICE INFORMATION [807.92(a)(2)]
#### Trade Name:
Quest MPS 3 ND Myocardial Protection System (MPS 3 ND)
#### Generic/Common Name:
Cardiopulmonary Bypass Heat Exchanger
#### Classification:
Class II per 21CFR§870.4240
# Product Code: DTR
{4}------------------------------------------------
## PREDICATE DEVICE(S) [807.92(a)(3)]
MPS 3 Myocardial Protection System (K200438).
### DEVICE DESCRIPTION [807.92(a)(4)]
The Quest MPS 3 ND Myocardial Protection System (MPS 3 ND) is a software controlled system designed to aid the perfusionist in cardioplegia delivery to a patient during Cardiopulmonary Bypass Surgery. The Quest MPS 3 ND System combines blood from the heart-lung machine and crystalloid from the IV-bag in a specified ratio and then adds in the drug (arrest or additive agent) for delivery to the patient. The Quest MPS 3 ND System consists of a reusable MPS 3 ND Console, a reusable MPS 3 Controller, and single use MPS 3 ND Disposables.
The Quest MPS 3 ND Console incorporates two Blood/Crystalloid pumps (B/C pumps), pressure and temperature monitors, a sensor interface, an Arrest Agent pump, an Additive pump and ultra-sonic air detection sensors. The Quest MPS 3 ND Console monitors and controls the blood:crystalloid ratio, drug concentration, flow rate, pressure and delivery route of the cardioplegia solution delivered to the patient. The MPS 3 ND Console also monitors the temperature of the cardioplegia solution which is regulated by an external heater cooler device.
The Quest MPS 3 Controller is a touchscreen user interface utilized by the operator to select all parameters, initiate/stop cardioplegia delivery, monitor delivery parameters and view/save relevant case information and data.
The Quest MPS 3 ND Disposables includes single use delivery sets (also known as the "MPS 3 ND Delivery Sets"), delivery set accessories and blood bypass tubing. The MPS 3 ND Delivery Sets and accessories consist of a flexible cassette, a heat exchanger with water line, drug cartridges and relevant tubing and connectors used to complete the cardioplegia circuit for use with the Quest MPS 3 ND Console. The blood bypass tubing is used as a backup in the event the Quest MPS 3 ND Console becomes unusable.
#### INDICATIONS FOR USE [807.92(a)(5)]
The Quest MPS 3 ND Myocardial Protection System, consisting of the Quest MPS 3 ND Console, the Quest MPS 3 Controller, and the Quest MPS 3 ND Disposables (Quest MPS 3 ND Delivery Set and optional Accessories) together is intended for use by perfusionists and physicians to deliver whole blood (from any arterial source) and/or cardioplegia solutions to the heart during open-heart surgery on either an arrested or beating heart for use up to six hours in duration.
{5}------------------------------------------------
#### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE [807.92(a)(6)]
In demonstrating substantial equivalence of the Quest MPS 3 ND Myocardial Protection System to the predicate MPS 3 Myocardial Protection System, Quest Medical considered the following aspects:
- Intended Use/Indications for Use ●
- Anatomical Site ●
- Patient Population ●
- Principles of Operation ●
- Design Features/Technological Characteristics ●
- Performance Specifications ●
- Materials
- Labeling ●
The MPS 3 ND System has the same Indications for Use, principles of operation, duration of use, and is used in the same anatomical site in the same patient population as the predicate device. The MPS 3 ND System has similar design features, technological characteristics, performance specifications, patient contacting materials and labeling as the predicate device. The primary differences in technological characteristics between the predicate device and the subject device are outlined below.
- The predicate device console includes a water circulation system while the subject device console does not include a water circulation system. The water circulation system in the predicate device regulates and controls the temperature of cardioplegia solution delivered to the patient. With the predicate device, users can also use an external heater cooler device to achieve temperature control, thereby bypassing the internal water circulation system. With exclusion of a water circulation system in the subject device, the subject device console relies on an external heater cooler device to provide temperature regulated water which heats or cools the cardioplegia solution delivered through the device.
- The water tubing in the predicate device is reusable while the water tubing in the subject device is disposable. The subject device delivery set includes disposable water tubing with fittings attached to the water ports of the heat exchanger component of the Delivery Set which is not included in the predicate device Delivery Set. The disposable water tubing attached to the heat exchanger is routed through the rear panel of the console where it can be connected directly to the external heater cooler device. The disposable water tubing in the subject device serves the same function as the water tubing in the predicate device (delivery of temperature controlled water to the heat exchanger).
{6}------------------------------------------------
#### 510(k) SUMMARY
Results from performance testing conducted on the subject device demonstrates that the differences in technological characteristics between the subject device and the predicate do not raise different questions of safety or effectiveness. Thus, the MPS 3 ND Myocardial Protection System (MPS 3 ND) is substantially equivalent to the predicate device.
### PERFORMANCE DATA [807.92(b)]
All necessary bench, nonclinical, and human factors testing was conducted on the MPS 3 ND Myocardial Protection System (MPS 3 ND) to support a determination of substantial equivalence to the predicate device.
### Nonclinical Testing Summary [807.92(b)(1)]
The bench and nonclinical tests included:
- Software Verification Tests ●
- . IEC 60601-1 Electrical Safety Tests
- IEC 60601-1-2 Electromagnetic Compatibility Tests ●
- MPS 3 ND System Performance Tests ●
- IEC 62133-2 and IEC 62281Battery Compliance Tests ●
- Human Factors/Usability Test ●
- Package Stability/Transit Tests .
- Leak Test ●
- Tubing Bonding Strength ●
- Simulated Use ●
{7}------------------------------------------------
# Clinical Testing Summary [807.92(b)(2)]
Not applicable. Clinical testing was not performed to support this 510(k) submission.
# CONCLUSIONS [807.92(b)(3)]
The collective results of the performance testing demonstrate that the Quest MPS 3 ND Myocardial Protection System (MPS 3 ND) meets the established specifications necessary for consistent performance during its intended use. In addition, the collective performance testing demonstrate that the Quest MPS 3 ND Myocardial Protection System (MPS 3 ND) does not raise different questions of safety or effectiveness for Cardiopulmonary Bypass Surgery when compared to the predicate device. The results from the performance testing support the conclusion that the Quest MPS 3 ND Myocardial Protection System (MPS 3 ND) is substantially equivalent to the predicate device.
### SUMMARY
The Quest MPS 3 ND Myocardial Protection System (MPS 3 ND) is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.