Ritleng®+ and Ritleng®+ PVP

K201892 · Fci (France Chirurgie Instrumentation) Sas · OKS · Mar 11, 2021 · OP

Device Facts

Record IDK201892
Device NameRitleng®+ and Ritleng®+ PVP
ApplicantFci (France Chirurgie Instrumentation) Sas
Product CodeOKS · OP
Decision DateMar 11, 2021
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic, Pediatric

Indications for Use

The RITLENG®+ products are indicated in the treatment of epiphora in patients of 12 months and older in cases of: - Canalicular pathologies (stenoses, obstructions, lacerations). - - Congenital duct obstruction (Hasner valve stenosis), - - Dacryocystorhinostomy (conventional or laser). -

Device Story

Self-retaining nasal bicanalicular lacrimal stent; used for treatment of lacrimal duct obstructions. Device consists of silicone tube with two larger diameter portions for self-retention and PEEK thread guides at extremities. Silicone tube may be PVP-coated for wettability. Central body includes marking for placement between lacrimal puncta. Used by ophthalmologists in clinical settings. Device acts as mechanical stent to maintain patency of lacrimal drainage system; facilitates tear drainage; relieves epiphora symptoms. Benefits patient by restoring lacrimal function.

Clinical Evidence

Bench testing only. Biocompatibility supported by cytotoxicity testing (ISO 10993-5), chemical characterization (ISO 10993-18), and toxicological risk assessment. Sterilization, package integrity, and stability studies performed to applicable standards.

Technological Characteristics

Bicanalicular lacrimal stent. Materials: medical-grade silicone, PEEK thread guides. Optional PVP coating. Self-retaining design via larger diameter tube portions. Sterilization: Ethylene Oxide (SAL 10-6). Packaging: double Tyvek pouch.

Indications for Use

Indicated for patients 12 months and older with epiphora due to canalicular pathologies (stenoses, obstructions, lacerations), congenital duct obstruction (Hasner valve stenosis), or requiring dacryocystorhinostomy (conventional or laser).

Regulatory Classification

Identification

Lacrimal stents and intubations sets are intended to repair the lacrimal drainage system. Lacrimal stents and intubations sets are indicated for, but limited to, the treatment of epiphora in infants and adults; canalicular pathologies such as stenosis; obstruction or laceration; and conditions requiring dacryocystorhinostomy (conventional or laser); or imperforation of the nasolacrimal duct in an infant.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in a way that "FDA" is on top, followed by "U.S. FOOD & DRUG", and then "ADMINISTRATION" at the bottom. March 11, 2021 FCI (France Chirurgie Instrumentation) SAS % Barbara S. Fant, Pharm.D. President Clinical Research Consultants, Inc. 3308 Jefferson Avenue, Upper Level Cincinnati, OH 45220 Re: K201892 Trade/Device Name: Ritleng®+ and Ritleng®+ PVP Regulatory Class: Unclassified Product Code: OKS Dated: February 12, 2021 Received: February 16, 2021 Dear Barbara S. Fant: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K201892 Device Name RITLENG®+ RITLENG®+ PVP Indications for Use (Describe) The RITLENG®+ products are indicated in the treatment of 12 months and older in cases of: - Canalicular pathologies (stenoses, obstructions, lacerations), - Congenital duct obstruction (Hasner valve stenosis), - Dacryocystorhinostomy (conventional or laser). Type of Use (Select one or both, as applicable) | <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | |------------------------------------------------------------------------------------------------| | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | 510(k) Number: | K201892 | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Owner: | France Chirurgie Instrumentation SAS (FCI S.A.S.)<br>20-22 rue Louis Armand<br>75015 Paris, France<br>Telephone:<br>+33 1 53 98 98 98<br>Facsimile:<br>+33 1 53 98 98 99 | | Contact Person: | Barbara S. Fant, Pharm.D.<br>Clinical Research Consultants, Inc.<br>3308 Jefferson Avenue<br>Upper Level<br>Cincinnati, OH 45220<br>Phone:<br>(513) 961-8200<br>Facsimile:<br>(513) 961-2858 | | Date: | March 4, 2021 | | Trade Name: | RITLENG®+<br>RITLENG®+PVP | | Common name: | Bicanalicular Lacrimal Stent | | Classification Name: | Lacrimal Stents and Intubation Sets | | Product Code: | OKS | ## Identification of a Legally Marketed Predicate Device RITLENG®+ is substantially equivalent to the RITLENG® intubation marketed by FCI Ophthalmics, Inc., 510(k) Premarket Notification Number K955671, FDA Product Code OKS. ### General Description The RITLENG®+ is a self-retaining nasal bicanalicular intubation for the lacrimal ducts. It consists of a silicone tube with two larger diameter portions, connected at each extremity with a PEEK thread guide. Each thread guide has a wide diameter part and a narrow diameter part. The silicone tube may be coated with polyvinylpyrrolidone (PVP) to improve its wettability. The central body of the intubation displays a marking intended for the intubation placement between the lacrimal puncta. {4}------------------------------------------------ # Intended Use The RITLENG®+ products are indicated in the treatment of epiphora in patients of 12 months and older in cases of: - Canalicular pathologies (stenoses, obstructions, lacerations). - - Congenital duct obstruction (Hasner valve stenosis), - - Dacryocystorhinostomy (conventional or laser). - ### Comparison of Technological Characteristics RITLENG®+ is substantially equivalent to the RITLENG® Intubation as both devices have the identical intended use and intubation method, with similar indications for use statements. The implant part of both devices is made from the identical grade silicone: and the silicone tube of both devices may be coated with polyvinylpyrrolidone (PVP) to improve its wettability. Both devices are packaged in a double Tyvek pouch and sterilized by Ethylene Oxide to a sterility level of SAL 10-6. The difference between the two devices is the addition of the two larger diameter portions of the silicone tube to the RITLENG® that create a self-retaining (autostable) feature when placed in the lacrimal ducts. Based on this comparison, RITLENG®+ is substantially equivalent to the RITLENG® predicate device (K955671). ### Brief Summary of Non-Clinical Tests and Results RITLENG®+ has been designed and tested to the applicable standards. All nonclinical test results met the established specifications for the device. In-process controls and final product quality controls, including finished product release testing and inspection, assure that RITLENG®+ is manufactured within specifications. The biocompatibility of the device is supported by cytotoxicity testing per ISO 10993-5, three sets of chemical characterization testing using different patient contacting device components per ISO 10993-18, and toxicological risk assessment based on these chemical characterization results. The biocompatibility of the subject device was also supported by biocompatibility information of the materials used in the subject device to the materials used in the cleared predicate. Ethylene oxide sterilization specifications, package integrity studies, and stability studies were performed to the applicable standards; and the test results support the shelf-life and storage conditions for the device. {5}------------------------------------------------ The results from non-clinical testing demonstrate RITLENG®+ meets the established specifications for the device and that the test results and established specifications are substantially equivalent to those of the RITLENG® predicate device. ### Basis of Substantial Equivalence RITLENG®+ is substantially equivalent to the RITLENG® predicate device with respect to the intended use as a bicanalicular intubation device for the treatment of congenital lacrimal duct obstructions, and the biocompatibility of the medical grade silicone raw materials used to manufacture the device which are manufactured by FCI SAS and distributed in the USA by FCI Ophthalmics, Inc.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...