K201877 · Safkan, Inc. · KYZ · Nov 5, 2020 · General Hospital
Device Facts
Record ID
K201877
Device Name
OtoSet- Ear Cleaning System
Applicant
Safkan, Inc.
Product Code
KYZ · General Hospital
Decision Date
Nov 5, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6960
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The OtoSet - Ear Cleaning System (OtoSet) is indicated to remove earwax from the external ear canal with irrigation and suction in adults.
Device Story
OtoSet is an automated ear irrigation device for adult cerumen removal. System comprises adjustable head strap, two ear units, single-patient-use disposable ear tips, and waste containers. Each ear unit contains solution container, liquid pump, vacuum pump, PCB electronics, and rechargeable batteries. Device cleans left, right, or both ear canals simultaneously. Operated by clinician or user; provides automated irrigation and suction to clear ear canal. Output is physical removal of earwax. Benefits include efficient, automated cerumen management.
Clinical Evidence
Two clinical studies conducted. Study 1 (n=31 subjects, 62 ears) compared OtoSet to Welch Allyn Ear Wash System using a 0-3 Degree of Occlusion Scale. Study 2 (n=23 subjects, 31 ears) evaluated OtoSet effectiveness using the same 0-3 scale. Results demonstrated OtoSet is equivalent to the predicate in removing earwax from the external ear canal.
Technological Characteristics
Automated ear irrigation system. Components: silicone rubber ear tips/pads/straps, liquid/vacuum pumps, rechargeable batteries, PCB. Biocompatibility per ISO 10993-1. Software per IEC 62304. Electrical safety/EMC per ANSI/AAMI ES60601-1, IEC 60601-1-2, FCC 47 CFR 15, IEC 62133-2. Cleaning validation per ANSI/AAMI TIR 12 and ISO 10993-5.
Indications for Use
Indicated for removal of earwax from the external ear canal in adults via irrigation and suction.
Regulatory Classification
Identification
An irrigating syringe is a device intended for medical purposes that consists of a bulb or a piston syringe with an integral or a detachable tube. The device is used to irrigate, withdraw fluid from, or instill fluid into, a body cavity or wound.
Predicate Devices
Welch Allyn Ear Wash System (Class I Exempt)
Reference Devices
Bionix OtoClear (Class I Exempt)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square. The full name of the agency is "U.S. Food & Drug Administration".
November 5, 2020
SafKan, Inc. Allison Komiyama Principal Consultant Acknowledge Regulatory Strategies, LLC 2251 San Diego Avenue Ste B-257 San Diego, California 92110
Re: K201877
Trade/Device Name: OtoSet - Ear Cleaning System Regulation Number: 21 CFR 880.6960 Regulation Name: Irrigating syringe Regulatory Class: Class I Product Code: KYZ Dated: October 2, 2020 Received: October 5, 2020
Dear Allison Komiyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Malvina Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201877
Device Name OtoSet - Ear Cleaning System
Indications for Use (Describe)
The OtoSet - Ear Cleaning System (OtoSet) is indicated to remove earwax from the external ear canal with irrigation and suction in adults.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for SafKan Ear Care. The logo features a stylized "S" inside a circle on the left, followed by the text "SafKan" in a sans-serif font. Below "SafKan" is the text "EAR CARE" in smaller letters. The color scheme is a combination of gold and teal.
# 510(k) Summary K201877
## DATE PREPARED
November 3, 2020
## MANUFACTURER AND 510(k) OWNER
SafKan, Inc. 1665 E 18th Street #205, Tucson, AZ 85719, USA Telephone: Official Contact: Sahil Diwan, Co-Founder & CEO
## REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., R.A.C. Erin A. Gontang, Ph.D. AcKnowledge Regulatory Strategies, LLC Telephone: +1 (619) 458-9547 akomiyama@acknowledge-rs.com, egontang@acknowledge-rs.com Email: Website: www.AcKnowledge-RS.com
# DEVICE INFORMATION
Proprietary Name/Trade Name: OtoSet – Ear Cleaning System Common Name: Irrigating Syringe Regulation Number: 21 CFR 880.6960 Class: I Product Code: KYZ
#### PREDICATE DEVICE IDENTIFICATION
The OtoSet – Ear Cleaning System is substantially equivalent to the following devices:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|----------------|------------------------------------------------|-------------------|
| Class I Exempt | Welch Allyn Ear Wash System/ Welch Allyn, Inc. | ✓ |
| Class I Exempt | Bionix OtoClear/Bionix Development Corp. | Reference Device |
The predicate and reference devices have not been subject to a design related recall.
#### DEVICE DESCRIPTION
The OtoSet – Ear Cleaning System (OtoSet) is an automated ear irrigation device used to remove earwax, or cerumen, from the external ear canal of adults. The OtoSet consists of an adjustable head strap, two ear units, two single-patient-use disposable ear tips, and two singlepatient-use disposable waste containers.
In addition to the disposable ear tips and waste containers, each ear unit houses a solution
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for SafKan Ear Care. The logo features a stylized "S" inside a circle on the left, followed by the text "SafKan" in a clean, sans-serif font. Below "SafKan" is the text "EAR CARE" in a smaller font size.
# 510(k) Summary
container, liquid pump, vacuum pump, printed circuit board electronics, rechargeable batteries, and electrical and plumbing elements to connect these items together. The device can be used to clean the left, right, or both external ear canals at the same time.
The OtoSet includes components that come in contact with the patient during use. The singlepatient-use disposable ear tips, ear pads and head strap pad (all made of silicone rubber) are surface contacting components having limited contact duration (< 24 hours) with intact skin.
# INDICATIONS FOR USE
The OtoSet – Ear Cleaning System (OtoSet) is indicated to remove earwax from the external ear canal with irrigation and suction in adults.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
SafKan believes that the OtoSet is substantially equivalent to the predicate device based on the information summarized here:
The intended use and indications for use of the OtoSet are the same as the primary predicate device, the Welch Allyn Ear Wash System. Both devices are irrigation devices used to remove earwax from the external ear canal, and control irrigation pressure via a pump. The OtoSet and the Welch Allyn Ear Wash System also control suction pressure via vacuum.
The OtoSet and the predicate device use non-sterile single-patient-use disposable ear tips.
The technological characteristics of the OtoSet are very similar to the predicate device, in its function, principle of operation, performance, and biocompatibility. These technological characteristics have undergone testing to ensure the device is as safe and effective as the primary predicate.
#### SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for the OtoSet. The following tests were performed to demonstrate safety based on current industry standards:
Cleaning validation: Device cleaning was validated using ANSI/AAMI TIR 12 and ISO 10993-5.
Biocompatibility: Patient contacting material met the requirements outlined in ISO 10993-1.
Software Verification: The software development and testing were executed in compliance to IEC 62304.
{5}------------------------------------------------
# 510(k) Summary
Image /page/5/Picture/1 description: The image is a logo for SafKan Ear Care. The logo consists of a circular icon with the letter "S" inside, followed by the words "SafKan" and "EAR CARE" stacked on top of each other. The icon and the words are in a light brown color. The logo is simple and modern, and it is likely used to represent a business that provides ear care services.
Electromagnetic Compatibility and Electrical Safety: The subject device was tested in compliance to ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC 60601-1-2, FCC 47 CFR 15 Subpart B and ANSI C63.4, and IEC 62133-2.
Usability testing: The usability of the subject device was assessed by 15 individuals who adequately represented the intended user population. All participants successfully performed simulated scenarios and sub-tasks, and provided feedback on the use of the subject device.
# SUMMARY OF CLINICAL TESTING
Clinical testing of the OtoSet included two studies. The first study (n=31 subjects, 62 ears) evaluated the OtoSet against the device's primary predicate, the Class I Welch Allyn Ear Wash System. The objective of the study was to evaluate the effectiveness of earwax removal (using a 0-3 Degree of Occlusion Scale) of the OtoSet compared to the Welch Allyn Ear Wash System. The second study (n=23 subjects, 31 ears) evaluated the subject device's ability to effectively remove earwax (also using a 0-3 Degree of Occlusion Scale).
The data in these studies show that the OtoSet is equivalent to the predicate device in terms of removing earwax from the external ear canal. The results also support the proposed intended use of the OtoSet.
# CONCLUSION
Based on the testing performed, non-clinical (biocompatibility, software validation, electromagnetic compatibility and electrical safety, and usability testing), and clinical testing, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The similar indications for use, technological characteristics, and performance characteristics for the proposed OtoSet – Ear Cleaning System are assessed to be substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.