← Product Code [MXD](/productcode/MXD) · K201865

# Biomonitor III, Biomonitor IIIm (K201865)

_Biotronik, Inc. · MXD · Dec 8, 2020 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K201865

## Device Facts

- **Applicant:** Biotronik, Inc.
- **Product Code:** [MXD](/productcode/MXD.md)
- **Decision Date:** Dec 8, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The BIOMONITOR III/BIOMONITOR IIIm is indicated to detect the following cardiac arrhythmias: - atrial fibrillation - bradycardia - sudden rate drop - high ventricular rate (HVR) - asystole The BIOMONITOR 111/BIOMONITOR Illm is indicated for use in: - Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias - Patients who experience transient symptoms that may suggest a cardiac arrhythmia The device has not been tested for and it is not intended for pediatric use.

## Device Story

Programmable, subcutaneous insertable cardiac monitor; records subcutaneous ECGs (sECGs) and physiological parameters (including core body temperature). Automatically detects arrhythmias (AF, bradycardia, asystole, sudden rate drop, HVR); patient-activated recording via Remote Assistant III during symptomatic episodes. Wireless transmission of data to BIOTRONIK Home Monitoring® system for physician review. Enables remote management of patients at risk of arrhythmias. Temperature sensor data transmission and historical trend display added to monitor physiological status. Used in clinical settings; physician-managed.

## Clinical Evidence

No clinical performance data was submitted or relied upon. Bench testing and animal testing (swine model) were performed to validate temperature sensor accuracy (within 0.1°C).

## Technological Characteristics

Subcutaneous insertable cardiac monitor; 47.5 x 8.3 x 4.3 mm (can); 1.9 cc volume; 4.0 g weight. Senses sECG via two integrated electrodes. Includes temperature sensor. Wireless connectivity to BIOTRONIK Home Monitoring system. MR Conditional.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- BIOMONITOR III and BIOMONITOR IIIm ([K200444](/device/K200444.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.

December 08, 2020

BIOTRONIK, Inc. Jon Brumbaugh Vice President, Regulatory Affairs and Compliance 6024 Jean Road Lake Oswego, Oregon 97035

Re: K201865

Trade/Device Name: BIOMONITOR III, BIOMONITOR IIIm Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MXD Dated: October 21, 2020 Received: October 22, 2020

Dear Jon Brumbaugh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jennifer Shih Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201865

Device Name BIOMONITOR III BIOMONITOR IIIm

Indications for Use (Describe)

The BIOMONITOR III/BIOMONITOR IIIm is indicated to detect the following cardiac arrhythmias:

- · atrial fibrillation
- bradycardia
- sudden rate drop
- · high ventricular rate (HVR)
- asystole

The BIOMONITOR 111/BIOMONITOR Illm is indicated for use in:

- · Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias
- · Patients who experience transient symptoms that may suggest a cardiac arrhythmia

The device has not been tested for and it is not intended for pediatric use.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for BIOTRONIK. The logo consists of the letters "BIO" inside of a circle on the left, followed by the word "BIOTRONIK" in large, bold, dark blue letters. Below the word "BIOTRONIK" is the phrase "excellence for life" in a smaller, lighter blue font.

# 510(k) Summary

# BIOMONITOR III and BIOMONITOR IIIm, Implantable Cardiac Monitors

# Traditional 510(k) Premarket Notification

- 1. Submission Information

| Date prepared | September 2, 2020                                                                                                                                                              |
|---------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact       | Jon Brumbaugh<br>VP, Regulatory Affairs and<br>Compliance<br>BIOTRONIK, Inc.<br>6024 Jean Road<br>Lake Oswego, OR 97035<br>Phone (888) 345-0374<br>jon.brumbaugh@biotronik.com |

- Manufacturer BIOTRONIK SE & Co. KG Woermannkehre 1, 12359 Berlin, Germany Registration number 9610139
### 2. Subject Devices

| Trade Name | BIOMONITOR III and BIOMONITOR IIIm |
|------------|------------------------------------|
|------------|------------------------------------|

- Common Name Implantable Cardiac Monitor
- Classification Name Recorder, Event, Implantable Cardiac (With Arrhythmia Detection)
- Classification Class II (21 CFR 870.1025)
- Product Code MXD

### 3. Predicate Device

BIOTRONIK BIOMONITOR III and BIOMONITOR IIIm, K200444, cleared April 23, 2020

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for BIOTRONIK. The logo consists of a circular emblem with the letters "BIO" inside, followed by the word "BIOTRONIK" in bold, dark blue letters. Below the word "BIOTRONIK" is the tagline "excellence for life" in a smaller, lighter blue font.

### Device Descriptions 4.

BIOMONITOR III and BIOMONITOR IIIm are programmable, subcutaneous insertable cardiac monitors able to record subcutaneous ECGs (sECGs) and other physiological parameters.

The BIOMONITOR III and BIOMONITOR IIIm are designed to automatically record the occurrence of arrhythmias in a patient. Arrhythmia may be classified as atrial fibrillation (AF), bradyarrhythmia, asystole, sudden rate drop, or high ventricular rate. In addition, the BIOMONITOR III and BIOMONITOR IIIm can be activated by the patient using the Remote Assistant III to record cardiac rhythm during symptomatic episodes. BIOMONITOR III and BIOMONITOR IIIm may be used with the current legally marketed BIOTRONIK Home Monitoring® technology, which is an automatic, wireless, remote monitoring system for management of patients with implantable cardiac monitors.

### 5. Indications for Use

The indications for use for the BIOMONITOR III and BIOMONITOR IIIm are identical to the predicate device.

The BIOMONITOR III/BIOMONITOR IIIm are indicated to detect the following cardiac arrhythmias:

- atrial fibrillation
- bradycardia ●
- . sudden rate drop
- high ventricular rate (HVR)
- . asystole

The BIOMONITOR III/BIOMONITOR IIIm is indicated for:

- . Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias
- Patients who experience transient symptoms that may suggest a cardiac arrhythmia ●

The device has not been tested for and it is not intended for pediatric use.

### Technological Characteristics 6.

The substantial equivalence claim between the subject and the predicate devices is supported by the information included in the premarket notification. This includes the following information:

- Description of the subject and predicate devices including the addition of ● temperature change alerts and display through BIOTRONIK's Home Monitoring System
- . Intended use of the subject and predicate devices
- Performance of the subject and predicate devices ●
- Technological characteristics of the subject and predicate devices
- Validation testing ●

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for BIOTRONIK. The logo consists of a circular symbol with the letters "BIO" inside, followed by the word "BIOTRONIK" in bold, dark blue letters. Below the word "BIOTRONIK" is the tagline "excellence for life" in a smaller, lighter blue font.

K201865

| Table 1. Comparison of the Predicate and Subject Devices |                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                               |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technical Data                                           | Predicate BIOMONITOR III<br>and BIOMONITOR IIIm                                                                                                                                                                                                                                                                                  | BIOMONITOR III and<br>BIOMONITOR IIIm                                                                                                                                                                         |
| FDA Clearance                                            | K200444, Predicate                                                                                                                                                                                                                                                                                                               | Subjects of this 510(k)                                                                                                                                                                                       |
| Indications                                              | Patients with clinical syndromes or situations at increased risk<br>of cardiac arrhythmias<br>Patients who experience transient symptoms that may suggest<br>a cardiac arrhythmia<br>The device has not been tested for and it is not intended for<br>pediatric use                                                              |                                                                                                                                                                                                               |
| Principle of Operation                                   | The BIOMONITOR III/BIOMONITOR IIIm senses subcutaneous<br>electrocardiograms (SECG) using two integrated electrodes<br>and has the capability of detecting a number of arrhythmias.<br>Like the predicate device, BIOMONITOR III/BIOMONITOR IIIm<br>sends recorded SECG and statistics to the Home Monitoring<br>Service Center. |                                                                                                                                                                                                               |
| Temperature Sensor<br>Data                               | Transmission of temperature<br>sensor data from the subject<br>device to the Home<br>Monitoring System                                                                                                                                                                                                                           | Transmission of temperature<br>sensor data from the subject<br>device to the Home Monitoring<br>System, implementation of<br>associated alerts and display<br>of historical trends and<br>temperature changes |
| Dimensions (mm)<br>Length x Width x Height               | 47.5 x 8.3 x 4.3 (can)<br>77.5 x 8.6 x 4.6 w/ lead                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                               |
| Volume                                                   | 1.9 cc                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                               |
| Weight                                                   | 4.0 g                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                               |
| AT/AF                                                    | 40 s/episode<br>30 s prior auto activation<br>10 s post auto activation                                                                                                                                                                                                                                                          |                                                                                                                                                                                                               |
| MR Conditional                                           | Yes                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                               |

#### てったり o Drodic

The newly available feature included in this 510(k) is the ability to measure core body temperature and the effect on determination of substantial equivalence was minimal as there was no associated change to intended use, indications for use, or hardware and no additional questions regarding safety or effectiveness.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for BIOTRONIK. The logo consists of the letters "BIO" inside of a circle with a cross through it, followed by the word "BIOTRONIK" in large, bold, dark blue letters. Below the word "BIOTRONIK" is the phrase "excellence for life" in a smaller, lighter font.

#### 7. Non-Clinical Performance Data

The following performance data are provided in support of the substantial equivalence determination:

#### 7.1 Validation and Verification Testing:

The BIOMONITOR III and BIOMONITOR IIIm have undergone thorough validation and verification testing to ensure final device functionality. The following categories of tests were performed and passed:

- Verification Testing for BIOMONITOR III and BIOMONITOR IIIm Temperature Sensor .
#### 7.2 Animal Testing:

Performance of the temperature sensor was validated in swine for detection of fever to a measurement accuracy of within 0.1℃ when compared to a clinical reference

#### 8. Clinical Performance Data

No clinical performance data was submitted or relied upon in support of the substantial equivalence determination.

#### 9. Conclusion

The subject BIOMONITOR III/BIOMONITOR IIIm devices and predicate BIOMONITOR III/BIOMONITOR IIIm devices are identical. They have the same principle of operation and identical physical device characteristics as well as software features and functionality. The enabling of the temperature sensor for data transmission introduces no new issues of safety or effectiveness. Additionally, the hardware design remains unchanged. These aspects of equivalence are confirmed by testing provided within the application.

Therefore, the BIOMONITOR III/BIOMONITOR IIIm were determined to be substantially equivalent to the primary predicate devices.

Image /page/6/Picture/14 description: The image shows the text "Page 4 of 4". This indicates that it is the last page of a document. The text is in a simple, sans-serif font and is left-aligned. The numbers are bolded.

---

**Source:** [https://fda.innolitics.com/device/K201865](https://fda.innolitics.com/device/K201865)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
