← Product Code [PEW](/productcode/PEW) · K201837

# UroLift 2 System (formerly UL500), UroLift 2 Delivery Handle, UroLift 2 Implant Cartridge (K201837)

_Neotract, Inc. · PEW · Jul 31, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K201837

## Device Facts

- **Applicant:** Neotract, Inc.
- **Product Code:** [PEW](/productcode/PEW.md)
- **Decision Date:** Jul 31, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5530
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The UroLift System is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH), including lateral and median lobe hyperplasia, in men 45 years of age or older.

## Device Story

The UroLift 2 System is an implantable transprostatic tissue retractor system used to treat BPH-related urinary obstruction. The device is inserted through the penile orifice into the urethra. The clinician uses the delivery handle and implant cartridges to deploy UroLift implants transversely through the prostate. These implants retract the prostatic tissue, displacing the urethra toward the prostatic capsule to maintain an expanded urethral lumen. This reduces fluid obstruction and improves lower urinary tract symptoms. The procedure typically requires 2-6 implants. An optional scope seal allows for anatomical examination without removing the telescope. The device is intended for use by clinicians in a clinical setting. By mechanically holding the approximated inner and outer prostate tissue, the system provides a minimally invasive alternative to traditional surgical interventions, potentially benefiting patients by alleviating urinary symptoms.

## Clinical Evidence

Bench testing only. Non-clinical design verification included product performance, compatibility, and deployment testing on devices subjected to worst-case sterilization and transit conditions. Biocompatibility testing was conducted per ISO 10993-1, -5, -10, and -11. Sterilization was validated to a 10^-6 Sterility Assurance Level (SAL) via gamma irradiation. Functional testing confirmed device performance after a 12-month shelf life.

## Technological Characteristics

Implantable transprostatic tissue retractor system. Components: Delivery Handle (single-patient use), Implant Cartridges (single-use), UroLift Implants. Materials: Patient-contacting components (delivery handle/cartridge) are externally communicating, tissue/bone/dentin, limited contact (<24 hours). Sterilization: Gamma irradiation. Connectivity: None. Software: None.

## Regulatory Identification

An implantable transprostatic tissue retractor system is a prescription use device that consists of a delivery device and implant. The delivery device is inserted transurethrally and deploys the implant through the prostate. It is designed to increase prostatic urethral patency by providing prostate lobe tissue retraction while preserving the potential for future prostate procedures and is intended for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia in men.

## Special Controls

In combination with the general controls of the FD&C Act, the UroLift® System is subject to the following special controls:

- (1) The elements of the device that may contact the patient must be demonstrated to be biocompatible.
- (2) Performance data must demonstrate the sterility of the patient-contacting components of the device.
- (3) Performance data must support shelf life by demonstrating continued sterility of the device (of the patient-contacting components), package integrity and device functionality over the requested shelf life.
- (4) Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
	- (A) Deployment testing must be conducted
	- (B) Mechanical strength must be conducted
	- (C) Resistance-to-degradation testing must be conducted
- (5) Non-clinical testing must evaluate the compatibility of the device in a magnetic resonance (MR) environment
- (6) In vivo testing must demonstrate safe and effective use, assess the impact of the implants on the ability to perform subsequent treatments, document the adverse event profile associated with clinical use, and demonstrate that the device performs as intended under anticipated conditions of
	- use. The following performance characteristics must be tested:
	- (A) Deployment testing must be conducted
	- (B) Implant migration must be conducted
- (7) Labeling must bear all information required for safe and effective use of the device, and must include:
	- (A) Specific instructions, warnings, cautions, limitations, and the clinical training needed for the safe use of the device
	- (B) Information on the patient population for which the device has been demonstrated to be effective
	- (C) A detailed summary of the device technical parameters
	- (D) Information on how the device operates and the typical course of treatment
	- (E) An expiration date/shelf life
	- (F) A detailed summary of the device- and procedure-related complications or adverse events pertinent to use of the device.

*Classification.* Class II (special controls). The special controls for this device are:(1) The elements of the device that may contact the patient must be demonstrated to be biocompatible.
(2) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(3) Performance data must support shelf life by demonstrating continued sterility of the device (of the patient-contacting components), package integrity, and device functionality over the requested shelf life.
(4) Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Deployment testing must be conducted.
(ii) Mechanical strength must be conducted.
(iii) Resistance-to-degradation testing must be conducted.
(5) Non-clinical testing must evaluate the compatibility of the device in a magnetic resonance environment.
(6) In vivo testing must demonstrate safe and effective use, assess the impact of the implants on the ability to perform subsequent treatments, document the adverse event profile associated with clinical use, and demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Deployment testing must be conducted.
(ii) Implant migration must be conducted.
(7) Labeling must bear all information required for safe and effective use of the device, and must include:
(i) Specific instructions, warnings, cautions, limitations, and the clinical training needed for the safe use of the device.
(ii) Information on the patient population for which the device has been demonstrated to be effective.
(iii) A detailed summary of the device technical parameters.
(iv) Information on how the device operates and the typical course of treatment.
(v) An expiration date/shelf life.
(vi) A detailed summary of the device- and procedure-related complications or adverse events pertinent to use of the device.

## Predicate Devices

- NeoTract UroLift System (UL500) ([K173087](/device/K173087.md))

## Reference Devices

- NeoTract UroLift System (UL400) ([K193269](/device/K193269.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a bold, sans-serif font and the full name in a smaller, sans-serif font below it. The overall design is clean and professional, reflecting the agency's role in regulating food and drugs in the United States.

July 31, 2020

NeoTract, Inc. Brian Gall Regulatory Affairs Manager 4155 Hopyard Road Pleasanton, CA 94588

Re: K201837

Trade/Device Name: NeoTract® UroLift® 2 System (UL2) Regulation Number: 21 CFR§ 876.5530 Regulation Name: Implantable transprostatic tissue retractor system Regulatory Class: II Product Code: PEW Dated: June 30, 2020 Received: July 2, 2020

Dear Brian Gall:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm = identifies = combination = product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH and Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See DICE (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Martha W. Betz, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201837

Device Name NeoTract® UroLift® 2 System (UL2)

#### Indications for Use (Describe)

The UroLift 2 System is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH), including lateral and median lobe hyperplasia, in men 45 years of age or older.

| Type of Use (Select one or both, as applicable)                                      |                                               |
|--------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

### COMPANY INFORMATION

NeoTract, Inc. 4155 Hopyard Road Pleasanton, CA 94588 Registration Number: 3015181082

### SUBMISSION CORRESPONDENT

Brian Gall Requlatory Affairs Manager, Interventional Urology NeoTract, Inc. 4155 Hopvard Road Pleasanton, CA 94588

Telephone - 925.329.6547 E-mail - brian.gall@teleflex.com

### DATE PREPARED

30 June 2020

### DEVICE INFORMATION

| Trade Name:           | NeoTract® UroLift® 2 System (UL2)                                                                                        |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------|
| Common Name:          | Implantable transprostatic tissue retractor system                                                                       |
| Classification Name:  | Implantable transprostatic tissue retractor system                                                                       |
| Product Code:         | PEW                                                                                                                      |
| Regulation Number:    | 876.5530                                                                                                                 |
| Classification:       | II                                                                                                                       |
| Classification Panel: | Gastrorenal, ObGyn, General Hospital, and Urology Devices<br>(OHT3) Reproductive, Gynecology and Urology Devices (DHT3B) |

### DEVICE DESCRIPTION

The UroLift System is designed to access the prostatic urethra and deliver one UroLift Implant through a lobe of the prostate. The UroLift System is inserted into the urethra through the penile orifice and used to displace the urethra toward the prostatic capsule. The UroLift Implant is then deployed transversely through the prostatic tissue. Multiple implants are deployed in the UroLift System procedure. The implants secure the retracted position of the urethra, thereby maintaining an expanded urethral lumen, reducing fluid obstruction and improving lower urinary tract symptoms (LUTS). This is accomplished by holding the approximated position of the inner (urethral) tissue and the outer (capsular) tissue of the prostate with the UroLift Implant. The procedure typically requires 2-6 implants to retract the obstruction. The UroLift 2 System (UL2), most recently cleared in K173087 under the marketing model number of UL500, is comprised of the UroLift Delivery Handle (single patient use), the UroLift Implant Cartridges (singleuse) and the UroLift Implants (one implant per cartridge). Each Implant Cartridge is preloaded with one UroLift Implant. The Implant Cartridges fit into the Delivery Handle. Each patient procedure will use one dedicated Delivery Handle and the number of Implant Cartridges/implants necessary to perform a typical procedure (estimated 2-6

Image /page/3/Picture/13 description: The image shows the logos for neotract and Teleflex. The neotract logo is on the left, and the Teleflex logo is on the right. Underneath the logos, the text "INTERVENTIONAL UROLOGY" is written in a smaller font. The logos are both in blue and white.

{4}------------------------------------------------

implants). The UroLift 2 System is also provided with an optional Scope Seal which enables a clinician to examine the anatomy between implant deployments without removal of the telescope from the Delivery Handle.

# INTENDED USE

The UroLift System is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH), including lateral and median lobe hyperplasia, in men 45 years of age or older.

# CONTRAINDICATIONS

The UroLift System should not be used if the patient has:

- Prostate volume of >100 cc
- A urinary tract infection
- · Urethra conditions that may prevent insertion of delivery system into bladder
- · Urinary incontinence due to incompetent sphincter
- Current gross hematuria

# PREDICATE DEVICE

The predicate device is the UroLift UL500 System by NeoTract (K173087).

| Trade Name:           | NeoTract UroLift System (UL500)                                                                                             |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Common Name:          | Implantable transprostatic tissue retractor system                                                                          |
| Classification Name:  | Implantable transprostatic tissue retractor system                                                                          |
| Product Code:         | PEW                                                                                                                         |
| Regulation Number:    | 876.5530                                                                                                                    |
| Classification:       | II                                                                                                                          |
| Classification Panel: | Gastrorenal, ObGyn, General Hospital, and Urology<br>Devices (OHT3) Reproductive, Gynecology and Urology<br>Devices (DHT3B) |

A reference device is included in this s ion as well. The reference device UroLift UL400 System by NeoTract (K193269)

| Trade Name:           | NeoTract UroLift System (UL400)                                                                                             |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Common Name:          | Implantable transprostatic tissue retractor system                                                                          |
| Classification Name:  | Implantable transprostatic tissue retractor system                                                                          |
| Product Code:         | PEW                                                                                                                         |
| Regulation Number:    | 876.5530                                                                                                                    |
| Classification:       | II                                                                                                                          |
| Classification Panel: | Gastrorenal, ObGyn, General Hospital, and Urology<br>Devices (OHT3) Reproductive, Gynecology and Urology<br>Devices (DHT3B) |

# COMPARISON WITH THE PREDICATE DEVICE

The UroLift 2 System (formerly UL500) described in this submission is substantially equivalent to the previously cleared generations of the device. The UroLift 2 System was most recently cleared in K173087; previous clearances included K172359, K162345, and K153584. Minor device modifications have been made to the UL2 Delivery System (Delivery Handle and Implant Cartridge) that do not affect the overall safety and effectiveness of the UroLift procedure. No modifications have been made to the UroLift Implant.

Image /page/4/Picture/19 description: The image features the logos of "neoTract" and "Teleflex" side-by-side, with the text "INTERVENTIONAL UROLOGY" centered below. The "neoTract" logo is in black and white, while the "Teleflex" logo is in blue. The text "INTERVENTIONAL UROLOGY" is in a smaller, lighter font, providing additional context to the brands.

{5}------------------------------------------------

### COMPARISON WITH THE REFERENCE DEVICE

The UroLift 2 System described in this submission utilizes the cleared UroLift System UL400 for some of the changes to the design of the device including the redesigned distal tip of the proposed device to match the UL400 geometry, as well as the expanded contraindication to increase the prostate size limitation from 80cc to 100cc. The device parameters relevant to the placement of the UroLift implant are identical between the UroLift 2 System and the UroLift System UL400.

### PERFORMANCE TESTING

The design requirements for the UroLift 2 System were reviewed and non-clinical design verification testing was required to assure that the modifications of the proposed device did not impact the safe and effective use of the device. Non-clinical testing included product performance testing, compatibility testing, and deployment testing. The testing was performed on devices which had undergone worst case sterilization and transit testing. The test methods were equivalent to the previously cleared UroLift System, and all acceptance criteria were met.

# BIOCOMPATIBILITY TESTING

The UroLift 2 System has been tested for biocompatibility and passed the relevant tests according to ISO 10993-1: Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process and the FDA guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The modifications addressed in this Special 510(k) submission do not impact the UroLift Implant cleared previously. Parts of the Delivery Device (Delivery Handle and Implant Cartridge) are patient contacting. They are considered externally communicating: tissue, bone, dentin with limited (<24 hour) contact. These components were tested according to ISO 10993-1:2018. Biocompatibility testing was performed on worst case sterilized devices and included, as applicable:

- Cytotoxicity testing per ISO 10993-5:2009 Biological evaluation of medical . devices Part 5: Tests for in vitro cytotoxicity
- Sensitization testing per ISO 10993-10:2010 Biological evaluation of medical . devices Part 10: Tests for irritation and skin sensitization
- Intracutaneous Reactivity testing per ISO 10993-10:2010 - Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization
- ISO 10993-11:2017: Biological Evaluation of Medical Devices Part 11: Tests ● for systemic toxicity

# STERILIZATION AND SHELF-LIFE TESTING

The UroLift System has been validated to determine the minimum gamma irradiation dose to ensure a 10 ° Sterility Assurance Level (SAL). Due to the modifications of the device, a new sterilization was performed, which included recovery, total bioburden, bacteriostasis/fungistasis, and product sterility. In addition, new dose mapping studies were completed at the sterilization sites.

Device functional testing was performed in a manner equivalent to the predicate to ensure the device functioned as intended after the stated shelf life of 12 months. Additional shelf life studies may be performed to extend the shelf life in the future using equivalent test methods.

Image /page/5/Picture/14 description: The image shows the logos for "NeoTract" and "Teleflex" side by side. The NeoTract logo is on the left, and the Teleflex logo is on the right. Below the logos, the text "INTERVENTIONAL UROLOGY" is written in a smaller font. The logos and text are all in blue and black.

{6}------------------------------------------------

# CONCLUSION

The testing demonstrated the NeoTract UroLift 2 System is as safe and effective, has the same intended use, technological characteristics and principles of operation as the predicate device. Therefore, the NeoTract UroLift 2 System is substantially equivalent to the predicate device.

Image /page/6/Picture/4 description: The image shows the logos for "NeoTract" and "Teleflex" side by side. The "NeoTract" logo is on the left, and the "Teleflex" logo is on the right. Underneath the logos, the text "INTERVENTIONAL UROLOGY" is written in a smaller font. The logos and text are all in a dark blue color.

---

**Source:** [https://fda.innolitics.com/device/K201837](https://fda.innolitics.com/device/K201837)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
