CleaniCal is a Calcium Hydroxide paste that has a creamy consistency and is suitable for several indications including: Temporary disinfectant dressings in the obturation of root canals; Indirect pulp capping or management of deep caries lesions; or Direct pulp capping.
Device Story
CleaniCal is a pre-filled syringe containing calcium hydroxide paste (pH 12.3). Used by dental professionals in clinical settings for endodontic and restorative procedures. The paste is applied directly into root canals or onto exposed/deep pulp tissue to act as a disinfectant dressing or capping agent. The high pH environment provides antimicrobial properties. The device is supplied non-sterile and is intended for professional use to manage caries and facilitate root canal therapy.
Clinical Evidence
Bench testing only. Testing included visual inspection, capacity, pH, flow, and radio-opacity per ISO 6876. Biocompatibility testing included ISO 10993-5 (cytotoxicity), ISO 10993-10 (delayed hypersensitivity), ISO 10993-11 (acute/subacute systemic toxicity), ISO 10993-6 (implantation), and ISO 10993-3 (genotoxicity). All results met preset criteria.
Technological Characteristics
Calcium hydroxide paste (30% concentration, pH 12.3). Composition includes zirconium dioxide, n-Methyl-2-pyrrolidone, and hypromellose. Form factor is a pre-filled syringe. Non-sterile. Shelf life of 2 years. Testing performed per ISO 6876.
Indications for Use
Indicated for dental professionals performing temporary disinfectant dressings in root canal obturation, indirect pulp capping, management of deep caries lesions, or direct pulp capping.
Regulatory Classification
Identification
A calcium hydroxide cavity liner is a device material intended to be applied to the interior of a prepared cavity before insertion of restorative material, such as amalgam, to protect the pulp of a tooth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 12, 2021
Maruchi Jang Wook CEO 2-208, Medical Industry Complex Bldg., 42-10, Taejanggongdan Wonju-si, Gangwon-do 26311 REPUBLIC OF KOREA
Re: K201799
Trade/Device Name: CleaniCal Regulation Number: 21 CFR 872.3250 Regulation Name: Calcium Hydroxide Cavity Liner Regulatory Class: Class II Product Code: EJK, KIF Dated: December 28, 2020 Received: January 6, 2021
Dear Jang Wook:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### 4. INDICATION FOR USE STATEMENT
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
| 510(k) Number (if known) | K201799 |
|--------------------------|-----------|
| Device Name | CleaniCal |
Indications for Use (Describe)
CleaniCal is a Calcium Hydroxide paste that has a creamy consistency and is suitable for several indications including:
- Temporary disinfectant dressings in the obturation of root canals;
- Indirect pulp capping or management of deep caries lesions; or
- Direct pulp capping.
Type of Use (Select one or both, as applicable)
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{3}------------------------------------------------
## 5. 510(k) SUMMARY
#### 510(k) Summary
Date: June 15, 2020
# 1. SUBMITTER
- MARUCHI
2-208, Medical Industry Complex Bldg., 42-10, Taejanggongdan-gil, Wonju-si, Gangwon-do, 26311, Republic of Korea TEL : +82-33-734-0330 FAX : +82-33-746-2804 Contact Name: Sung Wook, Jang Email: ecometal@endocem.com
## 2. DEVICE
- ·Trade Name: CleaniCal
- ·Common Name: Calcium hydroxide paste
- · Classification Name: Liner, Cavity, Calcium hydroxide
- · Regulation Number 872.3250
·Class: 2
·Classification Product Code: EJK
- ·Subsequent Product Code: KIF
#### 3. Predicate Device
K060365, APEXCAL, Ivoclar Vivadent K170175, ENDOSEAL MTA, MARUCHI
## 4. DEVICE DESCRIPTION
CleaniCal is a type of pre-filled syringe to Calcium Hydroxide Paste with a pH 12.3.
## 5. INDICATIONS FOR USE
CleaniCal is a Calcium Hydroxide paste that has a creamy consistency and is suitable
{4}------------------------------------------------
for several indications including:
- * Temporary disinfectant dressings in the obturation of root canals;
- * Indirect pulp capping or management of deep caries lesions; or
- * Direct pulp capping.
## 6. PERFORMANCE TESTING (NON-CLINICAL)
The following test articles were tested based on the referenced standard. All the test results met the preset test criteria.
- Testing institution's method Visual, Capacity, pH .
- . ISO 6876 - Package, Extraneous matter, Flow and Radio-opacity
- ISO 10993-5 Cytotoxicity (Agar diffusion, MTT) •
- . ISO 10993-10 - Maximization test for delayed hypersensitivity (LLNA)
- ISO 10993-11 Acute systemic toxicity •
- ISO 10993-6 Implantation (Subcutaneous tissue) •
- ISO 10993-6 & ISO 10993-11 Subacute systemic toxicity, .
- ISO 10993-3 Genotoxicity (Bacterial Reverse Mutation, & Mammalian . Erythrocyte Micronucleus)
| | Proposed Device | Predicate Device | Reference Device | Discuss/Justify<br>the Differences |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | New | K060365 | K170175 | - |
| Trade Name | CleaniCal | APEXCAL | ENDOSEAL MTA | - |
| Manufacturer | MARUCHI | IVOCLAR<br>VIVADENT, INC. | MARUCHI | - |
| Common Name | Calcium<br>hydroxide paste | Calcium<br>hydroxide paste | Root<br>Material | Filling Equivalent |
| Classification<br>Name | Liner, Cavity,<br>Calcium<br>hydroxide | Liner, Cavity,<br>Calcium<br>hydroxide | Resin, Root canal<br>filling | Equivalent |
| Device Class | 2 | 2 | 2 | Equivalent |
| Product Code | EJK | EJK | KIF | Equivalent |
| Device<br>Description | CleaniCal is a type of<br>pre-filled<br>syringe to Calcium<br>Hydroxide Paste<br>with a pH 12.3. | ApexCal is a<br>creamy,<br>radiopaque, ready-<br>to-use calcium<br>hydroxide paste<br>with a pH above<br>12.5. | ENDOSEAL MTA<br>is an endodontic<br>sealer based on<br>MTA, providing a<br>root canal filling. It<br>is premixed and pre-<br>loaded in a syringe,<br>which allows<br>a complete filling of<br>the entire root canal<br>including accessory<br>and lateral canals. | Equivalent |
| Indications<br>for Use | CleaniCal is a<br>Calcium<br>Hydroxide paste<br>that has a creamy<br>consistency and is<br>suitable for several<br>indications<br>including:<br>* Temporary<br>disinfectant<br>dressings in the<br>obturation of root<br>canals;<br>* Indirect pulp<br>capping or<br>management of<br>deep caries | ApexCAL is a<br>Calcium<br>Hydroxide paste<br>that has a creamy<br>consistency and is<br>suitable for several<br>indications<br>including:<br>* Temporary<br>disinfectant<br>dressings in the<br>obturation of root<br>canals;<br>* Indirect pulp<br>capping or<br>management of<br>deep caries | * Permanent<br>obturation of the<br>root canal following<br>vital pulp-<br>extirpation<br>* Permanent<br>obturation of the<br>root canal following<br>removal of infected<br>or necrotic pulp and<br>the placement of<br>intracanal dressings. | Equivalent |
| | lesions; or<br>* Direct pulp capping. | lesions; or<br>* Direct pulp capping. | | |
| Intended user | Dental professional | Dental professional | Dental professional | Equivalent |
| Standards | ISO 6876 | ISO 6876 | ISO 6876 | Equivalent |
| Chemical<br>Composition | -Calcium<br>hydroxide<br>-Zirconium<br>dioxide<br>- Excipients (n-<br>Methyl-2-<br>pyrrolidone,<br>Hypromellose) | - Calcium<br>hydroxide<br>- Bismuth<br>carbonate<br>- Excipients<br>(polyethylene<br>glycol, glycerine,<br>water) | -Natural Pure<br>Cement<br>-Zirconium dioxide<br>-Bismuth trioxide<br>- Excipients<br>(Bentonite Clay, Methyl-2-<br>Pyrrolidone,<br>Hypromellose) | The main<br>component is<br>same, but some<br>other<br>components are<br>different.<br>However, the<br>biocompatibility<br>and the<br>performance<br>test results<br>supported that<br>the subject<br>device is<br>substantially<br>equivalent to the<br>predicate<br>devices. |
| Content of<br>calcium<br>hydroxide | 30 % | 29 % | | Equivalent |
| pH | 12.3 | 12.5 | | Equivalent |
| Liquid Formula | Paste type | Paste type | Paste type | Equivalent |
| Packaging | Pre-loaded syringe | Pre-loaded syringe | Pre-loaded syringe | Equivalent |
| Sterile | Non-sterile | Non-sterile | Non-sterile | Equivalent |
| Shelf Life | 2 years | 2 years | 2 years | Equivalent |
#### 7. SUBSTANITAL EQUIVALENCE
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
## 8. SUBSTANTIAL EQUIVALENCE DISCUSSION
CleaniCal has the same indications for use and the principle of operations as the predicate devices. It is intended purpose as they are placed into the root canal as a calcium hydroxide paste which met the requirement according to ISO 6876. It has similar physical and biocompatible properties and demonstrates comparable performance specifications to the predicate devices.
The chemical compositions might slightly different from the predicate devices, both are used calcium hydroxide main composition as medicament, and additional component is used to make its appropriate workability.
The bench and biocompatibility testing performed demonstrates that any differences in their technological characteristics do not raise any new questions as to safety and effectiveness. Therefore, it is concluded that CleaniCal is substantially equivalent to the predicate devices. Hence, its equivalent is acceptable because it shows no clinically significant difference in the performance and safety to the device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.