K201752 · Brainlab AG · HAW · Jan 29, 2021 · Neurology
Device Facts
Record ID
K201752
Device Name
Disposable Pre-calibrated Suction
Applicant
Brainlab AG
Product Code
HAW · Neurology
Decision Date
Jan 29, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
Disposable Pre-calibrated Suction is an accessory of the Cranial Image Guided Surgery System and intended to be used as a navigated suction device in any surgical procedure in which the use of the Cranial Image Guided Surgery System may be indicated. Surgical example procedures include: · Cranial resection of tumors and other lesions · Resection of skull base tumors or other lesions · AVM Resection
Device Story
Disposable navigated suction device; accessory to Brainlab optical Cranial Image Guided Surgery (IGS) systems. Input: optical infrared light signals reflected from flat markers on integrated tracking array. Operation: system tracks markers to localize suction tip position in real-time; pre-calibrated for immediate use. Output: visual representation of suction tip position within patient anatomy on IGS display. Used in OR by surgeons; provides guidance during cranial resections. Benefits: enables precise navigation during suctioning; reduces setup time via pre-calibration; single-use design ensures sterility.
Clinical Evidence
Bench testing only. Evaluated mechanical stability of handle-to-tube bond, instrument tracking accuracy (mean locational error 0.45-0.51 mm; mean angular error 0.19-0.23°), system integration/pre-calibration effectiveness, shelf-life, biocompatibility (ISO 10993-1, -4, -5, -7, -10, -11), sterility (10^-6 SAL), and sterile barrier integrity.
Indicated for patients undergoing cranial surgical procedures, including resection of tumors, lesions, or AVMs, where use of a Cranial Image Guided Surgery System is appropriate.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Stryker Navigation System with CranialMap software application (K162929)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
January 29, 2021
Brainlab AG Chiara Cunico Manager Regulatory Affairs Olof-Palme-Str. 9 Munich, Bavaria 81829 Germany
Re: K201752
Trade/Device Name: Disposable Pre-calibrated Suction Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: December 28, 2020 Received: December 31, 2020
Dear Chiara Cunico:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201752
Device Name Disposable Pre-Calibrated Suction
Indications for Use (Describe)
Disposable Pre-calibrated Suction is an accessory of the Cranial Image Guided Surgery System and intended to be used as a navigated suction device in any surgical procedure in which the use of the Cranial Image Guided Surgery System may be indicated.
Surgical example procedures include:
- · Cranial resection of tumors and other lesions
- · Resection of skull base tumors or other lesions
- · AVM Resection
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
January 29, 2021
| General Information | | | | | |
|----------------------------|--------------------------------------------------------|--|--|--|--|
| Manufacturer | Brainlab AG, Olof-Palme-Str. 9, 81829, Munich, Germany | | | | |
| Establishment Registration | 8043933 | | | | |
| Device Name | Neurological Stereotaxic Instrument | | | | |
| Trade Name | Disposable Pre-calibrated suction | | | | |
| Classification Name | Stereotaxic instrument | | | | |
| Product Codes | HAW | | | | |
| Regulation Number | 882.4560 | | | | |
| Regulatory Class | II | | | | |
| Panel | Neurology | | | | |
| Predicate Devices and K | 1. K082060 Cranial ENT IGS System | | | | |
| Numbers | 2. K092467 Disposable Stylet | | | | |
| Manufacturer | Brainlab AG | | | | |
| Contact Information | |
|-----------------------------------|----------------------------------------|
| Primary Contact | Alternate Contact |
| Chiara Cunico | Regulatory Affairs Brainlab |
| Manager RA | Phone: +49 89 99 15 68 0 |
| Phone: +49 89 99 15 68 0 | Fax: +49 89 99 15 68 5033 |
| Email: chiara.cunico@brainlab.com | Email: regulatory.affairs@brainlab.com |
#### 1. Indications for Use
Disposable Pre-calibrated Suction is an accessory of the Cranial Image Guided Surgery System and intended to be used as a navigated suction device in any surgical procedure in which the use of the Cranial Image Guided Surgery System may be indicated.
Surgical example procedures include:
- · Cranial resection of tumors and other lesions
- Resection of skull base tumors or other lesions
- AVM Resection
{5}------------------------------------------------
#### 2. Device Description
Disposable Pre-calibrated Suction is an accessory of the Cranial IGS system and intended to be used as a navigated suction device in any surgical procedure in which the use of the Cranial IGS system may be indicated.
The Brainlab Disposable Suction is an accessory for the currently released and developed Brainlab optical IGS systems for cranial procedures. The device will be pre-calibrated, i.e. it will be automatically recognized by the system and is immediately ready to use. By tracking the flat markers attached to the integrated tracking array, the instrument and thereby position of the tip can be located by the Brainlab optical IGS systems for cranial procedures.
The device is intended for single short term invasive use on an individual patient during a single procedure. This invasive device is used for a short-term limited contact (<24 hours) and can be in direct contact with the central nervous system (CNS) tissues and the cerebrospinal fluid (CSF).
### 3. Substantial Equivalence
The reusable ENT suction has the same intended use and indications for use as the Disposable Suction and hence, is considered to be a suitable primary predicate device. Furthermore, the secondary predicate - Disposable stylet, was cleared under K092467 and found to be substantially equivalent with the predicate device Cranial ENT IGS (K082060). It has comparable features in terms of: sterility, single-use and tracking technology as the Subject Device and is also intended for invasive patient contact, including CNS and CSF.
At a high level, the subject and primary predicate device are based on the following same technological elements:
- Optical infrared light based reflective marker tracking for instrument localization a)
- b) Suction and pointing functionality combined in one device
- Stainless steel suction tube that is in direct contact with patient c)
- Accessory to the Cranial ENT IGS system d)
The following technological differences exist between the subject and primary predicate device, but are all similar to the secondary predicate device:
- (1) Pre-calibrated versus manual calibrated instrument
- (2) Single use sterile versus reusable instrument
- (3) Flat Marker versus Spherical Marker Tracking
- (4) Integrated tracking array versus (de)attachable tracking array
- (5) Polycarbonate handle glued to suction tube versus total stainless steel instrument
The Subject Device has larger tube diameters of 2.7 mm and 4mm when compared with the predicate device (2 mm and 3.3 mm). This larger size does not introduce an additional risk to patient safety since these dimensions are identical to suction tubes of the cleared reference device K162929 "Stryker Navigation System with CranialMap software application". K162929 has similar indications for use as the subject device.
{6}------------------------------------------------
| | Disposable Pre-<br>calibrated Suction<br>(Brainlab) | Posterior-Fossa<br>Suction Tube (Stryker) | Comparison |
|---------------|-----------------------------------------------------|-------------------------------------------|------------|
| Tube diameter | 2.7mm and 4.0 mm | 2.7mm and 4.0 mm | Identical |
These differences do not raise any new questions of safety or effectiveness compared to the predicate device.
#### 4. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
The substantial equivalence of the Disposable Suction is furthermore extensively verified. Technical performance has been tested on the following main aspects:
- Mechanical stability testing: mechanical stability of polycarbonate handle glued to a) suction tube has proven to be sufficient for intended use.
- Instrument tracking accuracy testing: instrument accuracy with flat marker technology b) has proven to meet accuracy requirements.
- Instrument system integration testing: pre-calibration of the Disposable Suction has C) proven to be effective within the Cranial IGS system.
- Shelf-life testing: technical performance (mechanical stability and accuracy) have proven d) to still meet requirement after shelf-life.
- Biocompatibility testing: biocompatibility has been stated for all materials of the e) Disposable Suction, also considering the manufacturing process.
- Sterility testing: sterilization has effectively shown that the sterility level of 106 f) has been reached.
- Sterile barrier system integrity testing: packaging validation has proven the integrity of g) the sterile barriers system of the double blister packaging.
| | REF 52184-01 | | REF 52185-01 | |
|-------------------------|------------------|---------------|------------------|---------------|
| | Locational error | Angular error | Locational error | Angular error |
| Mean | 0.45 mm | 0.19 ° | 0.51 mm | 0.23 ° |
| Standard deviation | 0.11 mm | 0.05 ° | 0.09 mm | 0.06 ° |
| 95th percentile | 0.93 mm | 0.27 ° | 1.01 mm | 0.31 ° |
| 99% confidence interval | 0.53 mm | 0.22 ° | 0.59 mm | 0.28 ° |
Table: Locational and angular accuracy Disposable Suction (REF 52184-01 and REF 52185-01)
#### Biocompatibility testing
The biological safety as a part of Disposable Suction basic safety has been assessed during biological safety evaluation. For a description of the used materials characterization, biological endpoint evaluation as well as for details about the performed verification tests refer to the Biological Risk Assessment by NAMSA performed according to ISO 10993-1 FDA Guidance "Use of International Standard ISO10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk
management process". The relevant "Biological Effect" endpoints have been assessed to identify
{7}------------------------------------------------
existing physical and/or chemical information and assess necessity of additional testing in order to demonstrate biological safety for the device acc. to its intended use..
Following biocompatibility tests have been performed (by ISO 17025 accredited laboratories) on the release candidate products manufactured, packaged and sterilized according to final manufacturing process:
- Cytotoxicity on tube test acc. to ISO 10993-5 ●
- Cytotoxicity on handle test acc. to ISO 10993-5
- Acute Systemic Toxicity test acc. to ISO 10993-11 ●
- Irritation test acc. to ISO 10993-10
- Pyrogen test acc. to European Pharmacopeia, 9th edition, 2016 .
- . Pyrogen test acc. to USP 42 - NF 37
- . Sensitization test acc. to ISO 10993-10
- Particle test acc. to USP [788] ●
- Bioburden test before and after packaging acc. to ISO 11737-1 .
- EO residual gas testing acc. to ISO 10993-7 ●
- Hemolysis tests acc. to ISO 10993-4 ●
# 5. Conclusion
The comparison of the Subject Device with the predicate devices shows that the Subject Device has similar functionality, intended use, technological characteristics, and typical users as the predicate devices. Verification and validation activities ensured that the design specifications are met and that the Device does not introduce new issues concerning safety and effectiveness. Hence, the Subject Device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.