The Pentacam AXL Wave is intended to image the anterior segment of the eye which includes the cornea, pupil, anterior chamber and lens. It is indicated for the evaluation of corneal shape, condition of the lens (opaque crystalline lens), the anterior chamber angle, anterior chamber depth, the volume of the anterior chamber, anterior or posterior cortical opacity, the location of cataracts using cross slit imaging with densitometry, corneal thickness, axial length, white-to-white distance, optical aberrations of the eye, and retroillumination imaging. The Pentacam AXL Wave also performs calculations to assist physicians in determining the power of the intraocular lens for implantation.
Device Story
Device images anterior eye segment (cornea, pupil, anterior chamber, lens) using rotating Scheimpflug camera; captures 100 images in <2 seconds; generates 3D model of anterior segment. Integrates Michelson interferometer for axial length measurement and Hartmann-Shack aberrometer for optical aberration assessment. Retroillumination imaging uses slit-lamp illumination method. Operated by clinicians in clinical settings. Data transmitted to connected PC for analysis and display. Physician uses 3D models and measurements to determine IOL power for cataract surgery. Benefits include comprehensive diagnostic data for surgical planning and post-op assessment.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, ISO 15004-1, ISO 15004-2, ISO 22665, and IEC 60825-1. Software verification and validation performed.
Technological Characteristics
Scheimpflug camera (CMOS), Michelson interferometer (880 nm SLD), Hartmann-Shack aberrometer, and slit-lamp retroillumination. IR LEDs (810 nm) for illumination. Connectivity via PC. Class II device. Software-based 3D modeling and densitometry analysis.
Indications for Use
Indicated for evaluation of corneal shape, lens condition (opaque crystalline lens), anterior chamber angle/depth/volume, cortical opacity, cataract location, corneal thickness, axial length, white-to-white distance, optical aberrations, and retroillumination imaging in patients requiring anterior segment assessment and IOL power calculation.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
October 21, 2020
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
OCULUS Optikgerate GmbH % Randy Prebula Partner Hogan Lovells, US LLP 555 Thirteenth Street, NW Washington, District of Columbia 20004
Re: K201724
Trade/Device Name: Pentacam AXL Wave Regulation Number: 21 CFR 886.1850 Regulation Name: AC-Powered Slitlamp Biomicroscope Regulatory Class: Class II Product Code: MXK, HJO Dated: September 18, 2020 Received: September 18, 2020
Dear Mr. Prebula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201724
Device Name Pentacam AXL Wave
#### Indications for Use (Describe)
The Pentacam AXL Wave is intended to image the anterior segment of the eye which includes the cornea, pupil, anterior chamber and lens. It is indicated for the evaluation of
- corneal shape,
- condition of the lens (opaque crystalline lens),
- the anterior chamber angle,
- anterior chamber depth,
- the volume of the anterior chamber,
- anterior or posterior cortical opacity,
- the location of cataracts using cross slit imaging with densitometry,
- corneal thickness,
- axial length,
- white-to-white distance.
- optical aberrations of the eye,
- and retroillumination imaging.
The Pentacam AXL Wave also performs calculations to assist physicians in determining the power of the intraocular lens for implantation.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> |
|---------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
### OCULUS Pentacam AXL Wave
#### General Information
## Applicant:
OCULUS Optikgeräte GmbH Müncholzhäuserstr. 29 35582 Wetzlar Germany Phone: +49(0)641 2005-0 +49(0)641 2005-255 Fax:
#### Contact Person:
Mr. Eckhard Loh Head of Quality and Regulatory Affairs OCULUS Optikgeräte GmbH Müncholzhäuserstr. 29 35582 Wetzlar Germany Phone: +49 (0) 641 2005-0 Fax: +49 (0) 641 2005-255
Summary Prepared: October 20, 2020
# Device Information
| Classification Name: | Device, analysis, anterior segment<br>(AC-powered slitlamp biomicroscope,) |
|-----------------------|----------------------------------------------------------------------------|
| Trade/Propriety Name: | Pentacam AXL Wave |
| Common Name: | Ophthalmic biomicroscope |
| Regulation Number: | 21 C.F.R. § 886.1850 |
| Device class: | II |
| Product Code: | MXK, HJO |
#### Predicate Devices
| Primary Predicate: | OCULUS Optikgeräte GmbH, Pentacam AXL (K152311 |
|----------------------|------------------------------------------------|
| Secondary Predicate: | LUNEAU SAS, VX120 (K143086) |
{4}------------------------------------------------
# Intended Use / Indications for Use
The Pentacam AXL Wave is intended to image the anterior segment of the eye which includes the cornea, pupil, anterior chamber and lens. It is indicated for the evaluation of
- corneal shape,
- condition of the lens (opaque crystalline lens),
- the anterior chamber angle,
- anterior chamber depth,
- the volume of the anterior chamber,
- anterior or posterior cortical opacity,
- the location of cataracts using cross slit imaging with densitometry,
- corneal thickness,
- axial length,
- white-to-white distance,
- optical aberrations of the eve.
- and retroillumination imaging. ●
The Pentacam AXL Wave also performs calculations to assist physicians in determining the power of the intraocular lens for implantation.
# Product Description/Technological Characteristics
The Pentacam AXL Wave is intended to image the anterior segment of the eye to measure eye components, such as corneal thickness, anterior chamber depth, corneal cylinder, corneal cylinder axis and white-to-white-distance. The axial length of the eye can be measured by a built-in interferometer. An also integrated aberrometer can determine the optical aberrations of the eye. By using retroillumination imaging, the back-lit eye can be observed.
The measured parameters can be used by physicians to calculate the power of the intraocular lens (IOL) implanted during a cataract surgery.
While rotating around the eve. the Pentacam AXL Wave captures Scheimpflug images of the anterior eye segment through varying axes. The Scheimpflug images created during an examination are transmitted to a connected PC.
Scheimpflug images can be captured within maximum of two seconds. Up to 138,000 genuine height values are measured and analyzed from the Scheimpflug images.
The Scheimpflug images are the basis for the height data which are used to calculate a mathematical 3D model of the anterior eye segment.
The mathematical 3D model, corrected for eye movements, provides the basis for all subsequent analysis.
The axial length of the eye is measured by interferometry measurements are done by a common Hartmann-Shack-Aberrometer. The retroillumination works similar to the illumination method of slit-lamps.
{5}------------------------------------------------
# New features in the device
The Pentacam AXL Wave combines the functions of the cleared Oculus Pentacam AXL device (K152311) with aberrometry measurements done by a common Hartmann-Shack-Aberrometer and retroillumination which works like the respective illumination method of slit-lamps. The new capabilities (aberrometry measurement and retroillumination imaging) are "add-ons" to the functionality of the Pentacam AXL that do not modify or impact the other capabilities in any way.
The Pentacam AXL Wave has functions subject to FDA premarket review as well as exempted functions that are not subject to FDA premarket review ("optical aberration of the eye" and "retroillumination imaging"). For these applications. FDA assessed those functions only to the extent that they could adversely impact the safety and effectiveness of the functions subject to FDA premarket review.
# Performance Testing
EMC and electrical safety of the subject devices were evaluated using the following consensus standards: IEC 60601-1; IEC 60601-1-2.
Testing was also performed in accordance with ISO 15004-1, ISO 15004-2, ISO 22665 and IEC 60825-1 to demonstrate compliance with these standards.
Software verification and validation testing has also been performed.
# Substantial Equivalence Discussion
As explained in detail below, the Pentacam AXL Wave is substantially equivalent to other legally marketed Scheimpfluq imaging devices and optical aberrometers. Specifically, the Pentacam AXL Wave is substantially equivalent to the Pentacam AXL (K152311) and the Visionix VX120 (K143086).The Pentacam AXL Wave has the same general intended use and similar indications, technological characteristics, and principles of operation as the previously cleared predicates. A substantial equivalence chart comparing the similarities and differences between the Pentacam AXL Wave and its predicate devices is provided at the end of this section, below. As explained in more detail below, minor differences in the technological characteristics do not raise new questions of safety or efficacy. Bench and clinical testing demonstrate that the Pentacam AXL Wave is as safe and effective as its predicate devices.
# Intended Use/ Indications for Use
The Pentacam AXL Wave has the same intended use as the predicate Pentacam AXL device, with the only difference being the addition of the ability to evaluate optical aberration of the eye and conduct retroillumination imaging. These additional functionalities are similar to the indications identified for the other predicate device. VX120.
The cleared Pentacam AXL and VX120 are also indicated to perform measurements and calculations. The Pentacam AXL Wave is further indicated for performing all these measurements in one device. The minor differences in the way that the indications are described do not alter the intended use of the device by creating a new intended therapeutic, diagnostic, prosthetic, or surgical
{6}------------------------------------------------
use of the device compared to the predicates. These measurements are substantially equivalent to the indications for the cleared Oculus Pentacam AXL device and LUNEAU VX120.
As described below, each of the measurements for the Pentacam AXL Wave device are performed individually and are substantially equivalent in terms of technology, safety and performance to two identified predicates. The subject Pentacam AXL Wave thus meets the first requirement for establishing substantial equivalence.
# Technological Characteristics
The Pentacam AXL Wave combines the measurement capabilities of the cleared Oculus Pentacam AXL and the cleared VX120.
The combination of the two predicate devices' cleared technologies in one device does not impact the ability of either technology to operate per its separate intended use. As explained in more detail below, the Scheimpflug camera and optical interferometer of the Pentacam AXL Wave still operates in fundamentally the same manner as the Pentacam AXL and are not interfered with as a result of the additional functionalities.
Further, the Hartman-Shack aberrometer and retro-illuminator are fundamentally the same as the same features that are incorporated into the VX120. Because each technological component has already been used for the same purposes as used in the subject device, and the technological characteristics operate independently without interference from the other technological features, no new or different questions of safety or effectiveness are raised by the subject device.
The Pentacam AXL Wave and the VX120 both serve the function of wavefront aberrometer by using a Hartman-Shack aberrometer.
Both systems measure an optical wavefront using a wavefront sensor (Hartman-Shack sensor).
The Pentacam AXL Wave and the VX120 also serve the function of retroillumination by using illumination method of a slit-lamp biomicroscope.
The Pentacam AXL Wave also performs measurements of the corneal shape and cataracts, similarly to the cleared Pentacam AXL. Both the Pentacam AXL Wave and the Pentacam AXL use a digital camera with comparable resolution and the Scheimpflug principle to photograph the anterior segment of the eye and use the photographs to generate data and measurements about the eye.
Both devices also use IR LEDs for eye illumination, blue slit lamps with a slit slight of 14 mm as the light source for slit illumination, and a digital CMOS camera as the overview camera.
Minor differences, such as the spherical measurement range, which is reduces from -20D to +20D in the predicate device to -10D to +6D for the subject device, do not raise new questions of safety and effectiveness because the measurement range continues to be appropriate for the relevant patient group that is to be measured using the device.
The Pentacam AXL Wave and the Pentacam AXL use similar Pentacam software to display the measurements and three-dimensional displays. The software has been modified since the
{7}------------------------------------------------
clearance of the Pentacam AXL to display the additional measurements gained from the added wavefront aberrometer measurements and retro illumination.
Overall, the Pentacam AXL Wave has substantially equivalent technological characteristics to the predicate devices. The differences between the subject device and the cleared predicates are minor, and bench and clinical tests have confirmed the safety and effectiveness of the new device in achieving the indications for use.
| Device Type | Subject device | Predicate device<br>(Scheimpflug imaging, axial length, etc.) |
|-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | Pentacam AXL Wave | Pentacam AXL |
| Manufacturer name | OCULUS Optikgeraete GmbH | OCULUS Optikgeraete GmbH |
| 510(k) Number | TBD | K152311 |
| | The Pentacam AXL Wave is intended to image the anterior segment of the eye which includes the cornea, pupil, anterior chamber and lens. It is indicated for the evaluation of corneal shape, condition of the lens (opaque crystalline lens), the anterior chamber angle, anterior chamber depth, the volume of the anterior chamber, anterior or posterior cortical opacity, the location of cataracts using cross slit imaging with densitometry, corneal thickness, axial length, white-to-white distance, optical aberrations of the eye, and retroillumination imaging. | The Pentacam AXL is designed to take photos of the anterior segment of the eye which includes the cornea, pupil, anterior chamber and lens of the eye. To evaluate: corneal shape, analyze condition of the lens (opaque crystalline lens), analyze the anterior chamber angle, analyze anterior chamber depth, analyze the volume of the anterior chamber, analyze anterior or posterior cortical opacity, analyze the location of cataracts (nuclear, sub capsular and or cortical), using cross slit imaging with densitometry, corneal thickness, axial length (by optical biometry), white-to-white distance. |
| | The Pentacam AXL Wave also performs calculations to assist physicians in determining the power of the intraocular lens for implantation. | The Pentacam AXL also performs calculations to assist physicians in determining the power of the intraocular lens for implantation |
| Scheimpflug Camera | Digital CMOS camera | Digital CCD camera |
| Resolution of<br>Scheimpflug camera | 1392x1040 pixel | 1392x1040 pixel |
| Slit length | 14 mm | 14 mm |
| Slit width | 35 μm | 35 μm |
| Light source for slit<br>illumination | Blue slit lamp (LED 470 nm) | Blue slit lamp (LED 470 nm) |
| Light source for eye<br>illumination | 4 IR LEDs (810 nm) | 4 IR LEDs (810 nm) |
| Device Type | Subject device | Predicate device<br>(Scheimpflug imaging, axial length, etc.) |
| Images | 100 in 2 seconds | 100 in 2 seconds |
| Rotation | 360° | 360° |
| Measuring points | up to 138,000 | up to 138,000 |
| Working distance | 80 mm | 80 mm |
| Physical principle | Scheimpflug principle / Optical<br>interferometer | Scheimpflug principle / Optical<br>interferometer |
| Calculation principle | Edge detection and 3D regression<br>polynomial functions; detection of<br>interference pattern and consideration of<br>mirror position | Edge detection and 3D regression<br>polynomial functions; detection of<br>interference pattern and consideration of<br>mirror position |
| Auto release function | Yes | Yes |
| Auto movement system | No | No |
| Fixation light | Yes, red LED (640 nm) | Yes, red LED (640 nm) |
| Overview camera | Digital CMOS camera | Digital CMOS camera |
| Resolution of overview<br>camera | 1280 x 960 | 640 x 480 |
| Measurement range -<br>Axial length | 14 – 40 mm | 14 – 40 mm |
| Light source for<br>interferometer | IR Super luminescence diode (SLD) | IR Super luminescence diode (SLD) |
| Wavelength | 880 nm | 880 nm |
| SLD-Power for<br>measurement | 0.7 mW | 0.84 mW |
| SLD-Power for<br>alignment | None | None |
| Pulse width | 520 ms | 400 ms |
| IEC 60825-1<br>classification | Class 1 laser product | Class 1 laser product |
| Embedded laser class<br>(not accessible) | 3R | 3R |
| White-to-white<br>measurement principle | Analysis of overview image | Analysis of overview image |
| Power supply | External PSU: | External PSU: |
| | HMEG49 Medical power adapter<br>Input: 90- 264 V AC: 47 – 63 Hz<br>Output: 24 V DC, 2.1 A | HMEG49 Medical power adapter<br>Input: 90- 264 V AC: 47 - 63 Hz<br>Output: 24 V DC, 2.1 A |
| Power consumption | max 35 W | max 35 W |
| Protection class | 2 | 2 |
| Protection type | IP20 | IP20 |
| Device type | B | B |
| Dimensions | 278 x 320-400 x 502-532 mm (WxDxH) | 274 mm x 316 mm x 500-530 (WxDxH) |
| Weight | 9 kg (only measuring head) | 13.2 kg |
| Technological Characteristics with the Relevant Predicate Device) | | |
| Device Type | Applicant device | Predicate device<br>(Aberrometry, retroillumination imaging) |
| Model | Pentacam AXL Wave | VX120…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.