← Product Code [QHG](/productcode/QHG) · K201676

# Barricaid Instrument Tray (K201676)

_Intrinsic Therapeutics, Inc. · QHG · Jul 16, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K201676

## Device Facts

- **Applicant:** Intrinsic Therapeutics, Inc.
- **Product Code:** [QHG](/productcode/QHG.md)
- **Decision Date:** Jul 16, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.4510
- **Device Class:** Class 2
- **Review Panel:** Orthopedic

## Indications for Use

The Intrinsic Therapeutics Defect Measurement Tool for the Barricaid® Annular Closure Device is intended to aid in determining if a patient meets the indications for the Barricaid® Annular Closure Device by assessing the defect size in the annulus fibrosus following limited discectory. This tool is indicated to only be used with the Intrinsic Therapeutics Barricaid® Annular Closure Device.

## Device Story

The Barricaid® Instrument Kit consists of manual surgical instruments, including a Defect Measurement Tool, alignment trials, a slotted mallet, and a retraction wedge, housed in a sterilization tray. The Defect Measurement Tool is used by surgeons during limited discectomy procedures to physically measure the size of the defect in the annulus fibrosus. This measurement determines if the patient meets the specific anatomical criteria required for the implantation of the Barricaid® Annular Closure Device. The kit is used in a clinical/surgical setting. The modification in this submission involves a reduction in the size of the sterilization tray. The device provides a physical measurement to assist the surgeon in clinical decision-making regarding the suitability of the Barricaid® implant for a specific patient, potentially improving surgical outcomes by ensuring proper patient selection.

## Clinical Evidence

No clinical testing was performed. Substantial equivalence is supported by bench testing, including sterilization validation, cleaning validation, and shipping condition evaluation based on worst-case scenarios.

## Technological Characteristics

Manual surgical instruments made of stainless steel. Sterilization tray composed of stainless steel with stainless steel or silicone brackets. The device is a mechanical measurement tool; no energy source, software, or connectivity features.

## Regulatory Identification

Manual surgical instrument for appropriate patient select for orthopedic implant is a device used to measure an anatomical feature(s) to determine appropriate patient selection for an orthopedic implant. The characteristics of the instrument are defined by the specifications set for use of the orthopedic implant in terms of geometry, surgical technique, and use of the device.

## Predicate Devices

- Defect Measurement Tool (Product Code QHG)

## Reference Devices

- Barricaid® Annular Closure Device (ACD) ([P160050](/device/P160050.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 16, 2020

Intrinsic Therapeutics, Inc. Regina Shih VP, Regulatory Affairs 30 Commerce Way Woburn, Massachusetts 01801

Re: K201676

Trade/Device Name: Barricaid® Instrument Tray Regulation Number: 21 CFR 888.4510 Regulation Name: Manual Surgical Instrument for Appropriate Patient Selection for Orthopedic Implant Regulatory Class: Class II Product Code: OHG Dated: June 19, 2020 Received: June 19, 2020

Dear Regina Shih:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent Showalter, Ph.D. Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201676

Device Name Barricaid® Instrument Kit

#### Indications for Use (Describe)

The Intrinsic Therapeutics Defect Measurement Tool for the Barricaid® Annular Closure Device is intended to aid in determining if a patient meets the indications for the Barricaid® Annular Closure Device by assessing the defect size in the annulus fibrosus following limited discectory. This tool is indicated to only be used with the Intrinsic Therapeutics Barricaid® Annular Closure Device.

| Type of Use (Select one or both, as applicable)                                             |                                                                                            |
|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">    Prescription Use (Part 21 CFR 801 Subpart D)  </span> | <span style="font-family: Arial;">    Over-The-Counter Use (21 CFR 801 Subpart C)  </span> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Intrinsic Therapeutics. The logo consists of a blue circle with a stylized "I" inside, followed by the words "INTRINSIC" in blue and "THERAPEUTICS" in teal. The "I" in the circle is designed to look like a spine.

## 510(k) Summary (21 CFR 807.92)

K201676 Page 1 of 3

#### 1. Submitter

| Submitter Name: | Intrinsic Therapeutics            |
|-----------------|-----------------------------------|
| Address:        | 30 Commerce Way, Woburn, MA 01801 |
| Telephone:      | 781-932-0222                      |
| Contact Person: | Regina Shih                       |
| E-mail:         | RShih@in-Thera.com                |
| Date:           | July 14, 2020                     |

#### II. Device Information

Device Name: Barricaid® Instrument Kit Common Name: Defect measurement Tool Classification Panel: Orthopedic Regulation Number: 21 CFR 888.4510 Regulatory Class: 2 Product Code: QHG

## III. Predicate Device

Primary Predicate: Defect Measurement Tool (Product Code QHG, Regulation 21 CFR 888.4510)

Reference Device: Barricaid® Annular Closure Device (ACD) – P160050

{4}------------------------------------------------

## IV. Device Description

The Barricaid® Instrument kit contains a number of manual surgical instruments as well as the Defect Measurement Tool. This kit is provided with all of the instruments with a sterilization tray. The Defect Measurement Tool is intended to measure the annular defect to determine if a patient meets the indications for use for the Barricaid® Annular Closure Device. All instruments are made from stainless steel. This kit is intended to be used only with the Intrinsic Therapeutics Barricaid® Annular Closure Device (ACD).

The purpose of this submission is to modify and reduce the size of the sterilization tray used to contain the reusable instruments necessary to surgically implant or remove the Barricaid® Annular closure device.

### V. Indications for Use

The Intrinsic Therapeutics Defect Measurement Tool for the Barricaid® Annular Closure Device is intended to aid in determining if a patient meets the indications for use defined for the Barricaid® Annular Closure Device by assessing the defect size in the annulus fibrosus following limited discectomy. This tool is indicated to only be used with the Intrinsic Therapeutics Barricaid® Annular Closure Device.

## VI. Comparison of Technological Characteristics with the Predicate Device

The instruments included in the subject Barricaid® Instrument Kit are identical to the kit in the previously approved kit (P160050) and include the Defect Measurement Tools, Alignment Trials, Slotted Mallet, and the Retraction Wedge. The dimension of the sterilization tray is reduced from the predicate sterilization tray. The sterilization tray is comprised of stainlesssteel, and the brackets composed of stainless-steel or silicone material, which is identical to the predicate.

### VII. Performance Data

- . Bench Testing
Scientific evaluation of sterilization validation and cleaning validation based on worst case. Scientific evaluation of shipping condition based on worst case and market history.

- Clinical Testing .
No new clinical testing was performed.

- Biocompatibility .
No new biocompatibility testing was performed.

{5}------------------------------------------------

## VIII. Conclusion

Based on the changes to the tray, the risk analysis, rationale and previous testing demonstrated that there are no additional risks that have been detected beyond acceptance level. The bench testing conducted supports the proposed device as substantially equivalent to the predicate device (Defect Measurement Tool (Product Code QHG, Regulation 21 CFR 888.4510)).

---

**Source:** [https://fda.innolitics.com/device/K201676](https://fda.innolitics.com/device/K201676)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
