← Product Code [LZO](/productcode/LZO) · K201673

# SMS Cementless Stem (K201673)

_Medacta International S.A. · LZO · Sep 18, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K201673

## Device Facts

- **Applicant:** Medacta International S.A.
- **Product Code:** [LZO](/productcode/LZO.md)
- **Decision Date:** Sep 18, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The hip prosthesis SMS is designed for cementless use in total or partial hip arthroplasty, for primary or revision surgery. Hip replacement is indicated in the following cases: · Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia · Avascular necrosis of the femoral head · Acute traumatic fracture of the femoral head or neck · Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip arthroplasty, hip resurfacing replacement or total hip arthroplasty.

## Device Story

SMS Cementless Stem is a bone-preserving short femoral stem for total or partial hip arthroplasty. Device functions as a structural implant for proximal femoral fixation. Used in surgical settings by orthopedic surgeons. Implanted into the femur to replace diseased or damaged hip joint components; provides mechanical stability for joint articulation. Benefits include restoration of joint function and pain relief. Device is a line extension adding sizes 1V2, 2V2, 14, and 15 to existing SMS product line.

## Clinical Evidence

No clinical studies were conducted. Substantial equivalence is supported by non-clinical bench testing, including range of motion (EN ISO 21535), fatigue testing (ISO 7206-4, ISO 7206-6), pull-off force testing (ASTM F2009-00), and coating integrity testing (ISO 13779-1, ASTM F1147-99).

## Technological Characteristics

Cementless femoral stem; materials and coatings consistent with predicate devices. Mechanical testing performed per ISO 7206-4/6 and ASTM F2009-00. Coating testing per ISO 13779-1 and ASTM F1147-99. Device is a modular component of a total hip system.

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- Zimmer Porolock MIS Stem ([K071723](/device/K071723.md))
- SMS ([K181693](/device/K181693.md))

## Reference Devices

- AMIStem-H ([K093944](/device/K093944.md))
- Quadra-S ([K093944](/device/K093944.md))
- Quadra-H ([K093944](/device/K093944.md))
- AMIStem ([K121011](/device/K121011.md))
- Quadra ([K121011](/device/K121011.md))
- AMIStem-P ([K173794](/device/K173794.md))
- AMIStem-P Collared ([K173794](/device/K173794.md))
- AMIStem-H Proximal Coating ([K173794](/device/K173794.md))
- Quadra-H ([K082792](/device/K082792.md))
- Quadra-R ([K082792](/device/K082792.md))
- AMIStem-H Proximal Coating ([K161635](/device/K161635.md))
- MiniMAX ([K170845](/device/K170845.md))
- Mpact DM ([K143453](/device/K143453.md))
- VersafitCup ([K083116](/device/K083116.md))
- VersafitCup ([K092265](/device/K092265.md))
- VersafitCup CC Trio ([K103352](/device/K103352.md))
- Mpact ([K103721](/device/K103721.md))
- Mpact ([K132879](/device/K132879.md))
- Mpact 3D Metal ([K171966](/device/K171966.md))
- Medacta Bipolar Head ([K091967](/device/K091967.md))
- CoCr ball heads ([K072857](/device/K072857.md))
- CoCr ball heads ([K080885](/device/K080885.md))
- CoCr ball heads ([K103721](/device/K103721.md))
- Endo Head ([K111145](/device/K111145.md))
- MectaCer BIOLOX Forte ([K073337](/device/K073337.md))
- MectaCer BIOLOX Delta Femoral Heads ([K112115](/device/K112115.md))
- MectaCer BIOLOX Option Heads ([K131518](/device/K131518.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 18, 2020

Medacta International SA % Chris Lussier Director, Quality and Regulatory Medacta USA 3973 Delp Street Memphis, Tennessee 38118

Re: K201673

Trade/Device Name: SMS Cementless Stem Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, MEH, KWY, LZY Dated: August 27, 2020 Received: August 28, 2020

Dear Chris Lussier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Vesa Vuniqi, M.S. Assistant Director DHTA: Division of Joints Arthroplasty OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

### K201673

Device Name SMS Cementless Stem

Indications for Use (Describe)

The hip prosthesis SMS is designed for cementless use in total or partial hip arthroplasty, for primary or revision surgery. Hip replacement is indicated in the following cases:

· Severely painful and/or disabled joint as a result of arthritis, rheumatoid polyarthritis or congenital hip dysplasia

· Avascular necrosis of the femoral head

· Acute traumatic fracture of the femoral head or neck

· Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip arthroplasty, hip resurfacing replacement or total hip arthroplasty.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

#### I. Submitter

Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66

Contact Person: Stefano Baj, Regulatory Affairs Director, Medacta International SA Applicant Correspondent: Chris Lussier, Senior Director of Quality and Regulatory, Medacta USA Date Prepared: June 18, 2020

#### II. Device

| Device Proprietary Name: | SMS Cementless Stem                                                                              |
|--------------------------|--------------------------------------------------------------------------------------------------|
| Common or Usual Name:    | Femoral Stems                                                                                    |
| Classification Name:     | Hip joint, metal/ceramic/polymer, semi-constrained, cemented or nonporous, uncemented prosthesis |
| Primary Product Code:    | LZO, MEH, KWY, LZY                                                                               |
| Regulation Number:       | 21 CFR 888.3353, 21 CFR 888.3390, 21 CFR 888.3360                                                |
| Device Classification    | II                                                                                               |

#### III. Predicate Device

Substantial equivalence is claimed to the following primary predicate device:

- Primary Predicate: Zimmer Porolock MIS Stem, K071723, Zimmer GmbH; and
- Additional Predicate: SMS, K181693, Medacta International SA. ●

#### IV. Device Description

The SMS femoral stem is a cementless bone preserving short stem designed for proximal fixation in total or partial hip arthroplasty for primary or revision surgery.

The SMS implants are comprised of the following products:

- SMS Cementless Solid Standard Stem (available in 11 sizes from size 3 to 13); and ●
- . SMS Cementless Solid Lateralized Stem (available in 11 sizes - from size 3 to 13).

Both are available on the US market via the clearance - K181693.

{4}------------------------------------------------

Concerning the new sizes of the solid version: 1V2, 2V2, size 14, and size 15; both STD and LAT versions have been introduced as a product range extension to the stems currently marketed as the SMS product line (K181693).

The SMS implants are line extensions to Medacta's Total Hip Prosthesis - AMIStem-H, Quadra-S and Quadra-H Femoral Stems (K093944), AMIStem and Quadra - Line Extension (K121011), AMIStem-P, AMIStem-P Collared and AMIStem-H Proximal Coating Femoral Stems (K173794), Quadra-H and Quadra-R Femoral Stems (K082792), AMIStem-H Proximal Coating (K161635), MiniMAX (K170845), and SMS (K181693).

The SMS implants are part of the Medacta Total Hip Prosthesis system. The Medacta Total Hip Prosthesis system consists of femoral stems, modular femoral heads and acetabular components. The acetabular components consist of metal cups and liners made of ultra-high molecular weight polyethylene (UHMWPE) or Highcross highly crosslinked ultra-high molecular weight polyethylene (HXUHMWPE). Acetabular components include the Mpact DM (K143453), VersafitCup (K083116 and K092265), VersafitCup CC Trio (K103352), Mpact (K103721 and K132879), Mpact 3D Metal (K171966) and Medacta Bipolar Head (K091967).

The SMS stems can be combined with the CoCr ball heads (K072857, K080885 and K103721), Endo Head (K111145) or with the MectaCer BIOLOX® Forte (K073337), MectaCer BIOLOX® Delta Femoral Heads (K112115) or MectaCer BIOLOX® Option Heads (K131518).

#### V. Indications for Use

The hip prosthesis SMS is designed for cementless use in total or partial hip arthroplasty, for primary or revision surgery. Hip replacement is indicated in the following cases:

- . Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, rheumatoid polyarthritis or congenital hip dysplasia
- . Avascular necrosis of the femoral head
- Acute traumatic fracture of the femoral head or neck
- Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip ● arthroplasty, hip resurfacing replacement or total hip arthroplasty.

#### VI. Comparison of Technological Characteristics

The SMS range extension implants and the predicate devices share the following characteristics:

- CCD angle;
- cementless; ●
- material of construction; ●
- coatings and coatings composition;
- biocompatibility;
- device usage;
- sterility;
- shelf life; and ●
- packaging. ●

{5}------------------------------------------------

SMS is technologically different from the predicate devices as follows:

- . sizes:
- stem lengths; ●

#### VII. Performance Data

Testing was conducted according to written protocols with acceptance criteria that were based on standards. The following mechanical studies were performed in substantial equivalence determination:

Non-Clinical Studies:

- Performance Tests ●
	- o range of motion (ROM): EN ISO 21535:2009 Non-Active Surgical Implants -Joint Replacement Implants - Specific Requirements for Hip-Joint Replacement Implants;
	- o fatigue testing: ISO 7206-4 Third Edition 2010-06-15 Implants for Surgery -Partial and Total Hip Joint Prostheses - Part 4: Determination of Endurance Properties and Performance of Stemmed Femoral Components [Including AMENDMENT 1 (2016)];
	- fatigue testing: ISO 7206-6 Second Edition 2013-11-15 Implants for Surgery o Partial and Total Hip Joint Prostheses - Part 6: Determination of Endurance Properties of Head and Neck Region of Stemmed Femoral Components;
	- pull off force testing: ASTM F2009-00 (Reapproved 2011) Standard Test Method o for Determining the Axial Disassembly Force of Taper Connections of Modular Prostheses.
- Coating Tests ●
	- o ISO 13779-1 Second Edition 2008-10-01 Implants for Surgery Hydroxyapatite -Part 1: Ceramic Hydroxyapatite: and
	- ASTM F1147-99 Standard Test Method for Tension Testing of Calcium Phosphate o and Metal Coatings.
- Pyrogenicity ●
	- Bacterial Endotoxin Test (LAL test) was conducted according to European o Pharmacopoeia §2.6.14 (which is equivalent to USP chapter <85>) and pyrogen test according to USP chapter <151> for pyrogenicity determination; and
	- the subject devices are not labeled as non-pyrogenic or pyrogen free. o

Clinical Studies:

- No clinical studies were conducted. .
### VIII. Conclusion

Based on the above information, the SMS implants new sizes 1V2, 2V2, size 14, and size 15 both STD and LAT version are substantially equivalent to the identified predicate devices.

Substantial equivalence has been demonstrated through a comparison of intended use, design, and technological characteristics, as well as performance evaluations. The SMS range extension implants

{6}------------------------------------------------

are as safe and effective as the predicate devices, Zimmer Porolock MIS Stem (K071723) and SMS solid stem, (K181693).

---

**Source:** [https://fda.innolitics.com/device/K201673](https://fda.innolitics.com/device/K201673)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com