K201670 · Siemens Medical Solutions USA, Inc. · KPR · Oct 21, 2020 · Radiology
Device Facts
Record ID
K201670
Device Name
YSIO X.pree
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
KPR · Radiology
Decision Date
Oct 21, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
AI/ML, Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Auto Cropping
AI-based algorithm
—
—
—
—
—
—
Indications for Use
The device is a digital X-ray system to generate X-ray images from the whole body including the skull, chest, abdomen, and extremities. The acquired images support medical professionals to make diagnostic and/or therapeutic decisions. Generic clinical benefits of radiographic examinations within the intended use are applicable for this device. YSIO X.pree is not for mammography examinations.
Device Story
Modular digital radiography X-ray system; components include ceiling-suspended X-ray tube, Bucky wall stand, Bucky table, generator, and wireless/fixed detectors. Inputs: X-ray radiation patterns converted to digital images. Features: collimator-mounted camera for patient positioning/collimation; touch-based user interface on tube suspension. Operated by healthcare professionals in clinical settings. Output: digital X-ray images for diagnostic/therapeutic decision-making. Benefits: improved workflow via touch interface and camera-assisted positioning; supports broad patient population including bariatric/pediatric. System monitors performance; blocks functions/displays errors if faults occur.
Clinical Evidence
Bench testing only. Performance verified through DQE and MTF measurements for detectors, confirming comparability to predicate. Compliance with IEC 60601 series (safety, EMC, radiation protection), ISO 14971 (risk management), and DICOM standards demonstrated.
Technological Characteristics
Stationary digital X-ray system. Components: Polydoros R80 generator, OPTITOP X-ray tube, Digital Multileaf Collimator N, Trixell Pixium detectors. Connectivity: DICOM compliant. Software: syngo XR on Windows 10. Sterilization: N/A (non-invasive). Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-28, 60601-2-43, 60601-2-54, ISO 10993-1, IEC 62304.
Indications for Use
Indicated for radiographic imaging of the whole body (skull, chest, abdomen, extremities) in pediatric, adult, and bariatric patients; includes emergency applications. Not for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
October 21, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Siemens Medical Solutions USA, Inc. % Mr. Andrew Turner Regulatory Affairs Specialist 40 Liberty Boulevard, 65-1A MALVERN PA 19355
Re: K201670
Trade/Device Name: YSIO X.pree Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR Dated: September 17, 2020 Received: September 21, 2020
Dear Mr. Turner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201670
Device Name YSIO X.Pree
Indications for Use (Describe)
The device is a digital X-ray system to generate X-ray images from the whole body including the skull, chest, abdomen, and extremities. The acquired images support medical professionals to make diagnostic and/or therapeutic decisions. Generic clinical benefits of radiographic examinations within the intended use are applicable for this device.
YSIO X.pree is not for mammography examinations.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------|---------------------------------------------------------|
| <span></span> | <span></span> |
| <div></div> Prescription Use (Part 21 CFR 801 Subpart D) | <div></div> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots. Below the words is the text "K201670".
## 510(k) Summary: YSIO X.pree
Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355
#### Date Prepared: August 5, 2020
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
#### 1. General Information:
Importer / Distributor: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Establishment Registration Number: 2240869
## Location of Manufacturing Site
Siemens Healthcare GmbH Siemensstr. 1 91301 Forchheim, Germany Establishment Registration Number: 3004977335
#### 2. Contact Person:
Andrew Turner Requlatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, Mail Stop 65-1A Malvern, PA 19355, USA Phone: 610-850-5627 Email: andrew.turner@siemens-healthineers.com
#### 3. Device Name and Classification:
Trade Name: YSIO X.pree System, X-Ray, Stationary Classification Name: Classification Panel: Radiology Classification Regulation: 21 CFR § 892.1680 Device Class: Class II Product Code: KPR
#### 4. Legally Marketed Predicate Device
Trade Name: Ysio Max 510(k) #: K181270 Classification Name: System, X-Ray, Stationary Classification Panel: Radiology
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
Classification Regulation: 21 CFR §892.1680 Device Class: Class II Product Code: KPR
#### 5. Device Description:
The YSIO X.pree is a radiography X-ray system. It is designed as a modular system with components such as a ceiling suspension with X-ray tube, Bucky wall stand, Bucky table, X-ray generator, portable wireless and fixed integrated detectors that may be combined into different configurations to meet specific customer needs. The following modifications have been made to the cleared predicate device:
- New image system with touch user interface o
- Added a camera at collimator to support clinical workflow o
#### Indications for Use: 6.
The device is a digital X-ray system to generate X-ray images from the whole body including the skull, chest, abdomen, and extremities. The acquired images support medical professionals to make diagnostic and/or therapeutic decisions. Generic clinical benefits of radiographic examinations within the intended use are applicable for this device.
YSIO X.pree is not for mammography examinations.
#### 7. Substantial Equivalence:
The YSIO X.pree with VA10 is substantially equivalent to the commercially available Ysio Max VF10 ( K181270 ).
### 8. Summary of Technological Characteristics of the Subiect Device as Compared with the Predicate Devices:
The YSIO X.pree is comparable in indications for use, design, material, functionality, technology, energy source and is substantially equivalent to the commercially available Ysio Max. It uses the same or similar components cleared with Ysio Max (e.g. tube, collimator, table, wallstand, detector or generator).
Many components of the subject device have the same technological characteristics as those from the predicate device.
Testing and validation have been successfully completed and test results show that the subject device YSIO X.pree with all of its components is comparable and therefore substantially equivalent to the predicate device.
The modifications made to the subject device YSIO X.pree are within the scope of intended use of 510(k) cleared predicate device Ysio Max (tomographic exposure is no longer available) and they also don't alter the fundamental scientific technology compared to predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern.
## Table 1: Comparison of the Subject to the Predicate
| Attribute | Subject Device<br>YSIO X.pree | Predicate Device<br>Ysio Max K181279 | Comparison /<br>Remarks |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | YSIO X.pree is a device<br>intended to visualize<br>anatomical structures by<br>converting an X-ray<br>pattern into a visible<br>image. YSIO X.pree<br>enables radiographic<br>exposures of the whole<br>body and may be used on<br>pediatric, adult and<br>bariatric patients. It can<br>also be used for<br>emergency applications.<br>YSIO X.pree is not for<br>mammography<br>examinations. | Ysio Max is a device<br>intended to visualize<br>anatomical structures by<br>converting an X-ray pattern<br>into a visible image. Ysio<br>Max enables radiographic<br>and tomographic exposures<br>of the whole body and may<br>be used on pediatric, adult<br>and bariatric patients. It can<br>also be used for emergency<br>applications.<br>Ysio Max is not for<br>mammography<br>examinations. | The option for<br>tomographic<br>exposures is not<br>available for the<br>subject device<br>due to low<br>customer use in<br>previous devices.<br>Ysio X.pree was<br>initially developed<br>without this<br>feature and<br>therefore this<br>does not alter the<br>safety and<br>effectiveness. |
| Product Code | KPR | KPR | same |
| X-Ray | | | |
| Generator | Polydoros R80<br>65/80 kW | Polydoros R80<br>65/80 kW | same |
| X-Ray tube | OPTITOP<br>150/40/80/HC-100 | OPTITOP<br>150/40/80/HC-100 | same |
| X-ray<br>techniques | Radiography | Radiography | same |
| Collimator | Digital Multileaf<br>Collimator N | Digital Multileaf<br>Collimator N | same |
| Air kerma | Kerma X | Kerma X | same |
| CARE | Combined<br>Applications to<br>Reduce Exposure | Combined<br>Applications to<br>Reduce Exposure | same |
| Touch user<br>interface on<br>tube<br>suspension | New touchscreen in<br>landscape format | Touchscreen in portrait<br>format. | New touchscreen |
| Digital Imaging | | | |
| Fixed detector<br>for table and<br>wall stand | Trixell Pixium 4343RC<br>"Max Static" | Trixell Pixium 4343RC<br>"Max Static" | same |
| Large mobile<br>detector | Trixell Pixium<br>3543EZh "MAX wi-D" | Trixell Pixium<br>3543EZh "MAX wi-D" | same |
| Small mobile<br>detector | Pixium 2430EZ "MAX<br>mini" | Pixium 2430EZ "MAX mini" | small changes in<br>DQE and MTF,<br>see table below,<br>does not affect<br>safety and<br>effectiveness. |
| | | | effectiveness |
| Digital imaging<br>system | syngo XR | Fluorospot Compact | Different, new<br>imaging system |
| | Operating system<br>Windows 10 | Operating system Windows<br>10 | same |
| | Operated via touch screen | Operated with mouse and<br>keyboard | Different, does<br>not affect safety<br>or effectiveness |
| | Image processing with<br>MyExam IQ | Image processing with<br>Diamond View Plus | Slightly different,<br>based on same<br>processing<br>software. Does<br>not affect safety<br>or effectiveness |
| | Al-based Auto Cropping | Auto Cropping | New Algorithm,<br>does not affect<br>safety or<br>effectiveness |
| | Acquisition and Image<br>processing parameters<br>selected via clinical<br>protocols | Acquisition and Image<br>processing parameters<br>selected via Organ<br>Programs | Improved, does<br>not affect safety<br>or effectiveness |
| Other Features and Components | | | |
| | Table with fixed detector<br>and table with bucky | Table with fixed detector<br>and table with bucky | same |
| Patient table | Standard tabletop and flat<br>tabletop | Standard tabletop | Additional flat<br>tabletop for easier<br>patient<br>positioning, does<br>not affect safety<br>or effectiveness |
| Wall stand | Wall stand with fixed<br>detector and wall stand<br>with bucky | Wall stand with fixed<br>detector and wall stand with<br>bucky | same |
| Camera | Live camera for patient<br>positioning and collimation | N/A | New, does not<br>affect safety or<br>effectiveness |
| Wireless<br>Remote<br>Control | Yes, same type | Yes, same type | same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots.
# Table 2: Comparison of detector parameters for detector Pixium 2430EZ
| Mobile Rad detector | YSIO X.pree | Ysio Max / K181279 |
|--------------------------|-------------------|--------------------|
| Siemens Name | MAX mini | MAX mini |
| Trixell name | Pixium 2430EZ | Pixium 2430EZ |
| Dimensions (active area) | 28.4 cm x 22.5 cm | 28.4 cm x 22.5 cm |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" are a series of orange dots arranged in a circular pattern.
| Matrix size | 1920 x 1520 | 1920 x 1520 |
|-----------------|--------------------|--------------------|
| DQE in %; 2 µGy | 70 % at 0.05 lp/mm | 66 % at 0.05 lp/mm |
| | 51 % at 1 lp/mm | 50 % at 1 lp/mm |
| | 42 % at 2 lp/mm | 40 % at 2 lp/mm |
| | 29 % at 3 lp/mm | 24 % at 3 lp/mm |
| | 19 % at Nyquist | 17 % at Nyquist |
| MTF in % | 63 % at 1 lp/mm | 61 % at 1 lp/mm |
| | 35 % at 2 lp/mm | 31 % at 2 lp/mm |
| | 19 % at 3 lp/mm | 15 % at 3 lp/mm |
| | 15 % at Nyquist | 12 % at Nyquist |
The MAX mini has now the same image performance as the MAX Wi-D.
#### 9. Summary of Non-Clinical Tests:
The YSIO X.pree was tested and complies with the voluntary standards listed in the table below:
| Reference Number and Date | Title of Standard |
|----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1<br>Edition 3.1 (IEC 60601-1:2005<br>+ Cor.:2006 + Cor.:2007 +<br>A1:2012) | Medical Electrical Equipment - Part 1: General Requirements<br>for Safety |
| IEC 60601-1-2<br>Edition 4.0<br>(IEC 60601-1-2:2014) | Medical Electrical Equipment - Part 1-2: General requirements<br>for basic safety and essential performance - Collateral<br>Standard: Electromagnetic Compatibility -- Requirements and<br>Tests |
| IEC 60601-1-3<br>Edition 2.1<br>(IEC 60601-1-3:2008 +<br>A1:2013) | Medical electrical equipment - Part 1-3: General requirements<br>for basic safety and essential performance - Collateral<br>Standard: Radiation protection in diagnostic X-ray equipment |
| IEC 60601-2-28 Edition 3.0<br>(IEC 60601-2-28: 2017) | Medical electrical equipment - Part 2-28: Particular<br>requirements for the basic safety and essential performance of<br>X-ray tube assemblies for medical diagnosis |
| IEC 60601-2-43<br>Edition 2.1<br>(IEC 60601-2-43:2010 +<br>A1:2017) | Medical electrical equipment - Part 2-43: Particular<br>requirements for the safety of X-ray equipment for<br>interventional procedures |
| IEC 60601-2-54<br>Edition 1.2<br>(IEC 60601-2-54:2009 +<br>Cor.:2010 + Cor.:2011 +<br>A1:2015 + A2:2018) | Medical electrical equipment - Part 2-54: Particular<br>requirements for the basic safety and essential performance of<br>X-ray equipment for radiography and radioscopy |
## Table 3: Non-clinical performance testing
{8}------------------------------------------------
| IEC 62366-1<br>Edition 1.0<br>(IEC 62366-1:2015) | Medical devices - Application of usability engineering to<br>medical devices |
|------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 14971:2019 | medical devices - application of risk management to medical<br>devices |
| IEC 62304<br>Edition 1.1<br>(IEC 62304:<br>2006 + A1:2015) | Medical device software - Software life cycle processes |
| IEC 61910-1<br>Edition 1.0<br>(IEC 61910:<br>2014) | Medical electrical equipment - Radiation dose documentation -<br>Part 1: Radiation dose structured reports for radiography and<br>radioscopy (IEC 61910-1:2014) |
| PS 3.1 - 3.20 2016 | Digital Imaging and Communications in Medicine (DICOM) Set |
| ISO 10993-1: 2013 | Biological evaluation of medical devices -- Part 1: Evaluation<br>and testing within a risk management process |
The following quality assurance measures were applied to the development of the system:
- . Risk Analysis
- Requirement Specification Reviews ●
- . Design Reviews
- Integration testing (System verification and validation) ●
#### 10. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features, including visual and audible warnings, are incorporated into the system design. In addition, the YSIO X.pree is continuously monitored and if an error occurs the system functions will be blocked and an error message will be displaved.
Furthermore, the operators are health care professionals familiar with and responsible for the x-ray examinations to be performed. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and all equipment is subject to final performance testing.
#### 11. Conclusion as to Substantial Equivalence:
The YSIO X.pree has the same fundamental scientific technology and performance characteristics as the predicate, Ysio Max (K181270). The YSIO X.pree has the same intended use and a simplified indications for use. Therefore, the YSIO X.pree is substantially equivalent to the predicate Ysio Max
#### 12. Guidance documents
The following FDA guidance documents were utilized in this Premarket Notification:
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
- Content of Premarket Submissions for Management of Cyberse ● Devices Guidance for Industry and Food and Drug Administration Staff Document lssued on: October 2, 2014
- . Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices Guidance for Industry and Food and Drug Administration Staff Document issued on July 11, 2016. The draft of this document was issued on November 2, 2015.
- . Pediatric Information for X-ray Imaging Device Premarket Notifications Guidance for Industry and Food and Drug Administration Staff Document issued on November 28, 2017.
- Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices . Guidance for Industry and Food and Drug Administration Staff Document issued on: September 1, 2016
- Guidance for Industry and FDA Staff Guidance for the Content of Premarket ● Submissions for Software contained in Medical Devices, Document issued on: May 11, 2005
- Guidance for Industry and FDA Staff Recognition and Use of Consensus . Standards Document issued on: September 17, 2007
- . The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] Guidance for Industry and Food and Drug Administration Staff Document issued on: July 28, 2014
- . Radio Frequency Wireless Technology in Medical Devices Guidance for Industry and Food and Drug Administration Staff Document issued on: August 14, 2013
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.