K201669 · Nanowear, Inc. · DXH · Nov 6, 2020 · Cardiovascular
Device Facts
Record ID
K201669
Device Name
SimpleSENSE
Applicant
Nanowear, Inc.
Product Code
DXH · Cardiovascular
Decision Date
Nov 6, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2920
Device Class
Class 2
Indications for Use
The SimpleSENSE System is intended for use at home, or at a healthcare facility, under the direction of a licensed medical professional, to record, display and store the following physiological data: a) 2 leads of Electrocardiogram; b) Respiration rate measured through thoracic impedance; c) Heart Sounds; and d) Activity including posture. The device is intended for use when the clinician decides to evaluate the physiologic signals as an aid to diagnosis and treatment. The SimpleSENSE System is intended to be used by patients at rest and not performing any activities or movements. ECG recordings are indicated for the manual assessment of cardiac rhythm disturbances. The device does not produce alarms and is not intended for active patient monitoring (real-time). The device is not intended for use as life supporting equipment on high-risk patients such as critical care patients. The device is not intended for use in the presence of a pacemaker.
Device Story
SimpleSENSE is a wearable textile garment with integrated nanosensor electrodes and a MEMS microphone; captures 2-lead ECG, respiration rate (via thoracic impedance), heart sounds, and activity/posture. System includes a Signal Acquisition Unit (SAU) for data storage/transmission and a mobile application for iPhone (iOS 13.4) to initiate recordings and forward reports. Used at home or healthcare facilities by patients at rest; operated under clinician direction. Data is wirelessly transmitted to a smartphone and stored on a removable microSD card. Clinicians review the recorded data to aid in diagnosis and treatment of cardiac rhythm disturbances. Device does not provide real-time monitoring or alarms.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included verification of multiparametric data capture, Bluetooth/iPhone connectivity, data encryption, battery safety, SAU durability, microSD card performance, biocompatibility of the garment, sensor performance (ECG, thoracic impedance, MEMS microphone), garment mechanical/electrical performance, and shelf life.
Technological Characteristics
Wearable textile garment with integrated nanosensor electrodes and MEMS microphone. ECG: Bipolar lead instrumentation amplifier, Sigma-Delta ADC. Respiration: 4-point probe thoracic impedance. Power: Internal Li-Ion rechargeable battery. Connectivity: Bluetooth wireless to iPhone. Storage: Removable microSD card. Software: Mobile application for recording/transmission.
Indications for Use
Indicated for adult patients at rest, under direction of a licensed medical professional, for recording/storing ECG, respiration rate (thoracic impedance), heart sounds, and activity/posture as an aid to diagnosis and treatment. Contraindicated for high-risk/critical care patients, patients with pacemakers, and active real-time monitoring.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a human figure, and the text on the right is in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line.
November 6, 2020
Nanowear Inc. % Melissa Walker President & CTO Graematter, Inc. 1324 Clarkson Clayton Ctr, #332 St Louis, Missouri 63011
Re: K201669
Trade/Device Name: SimpleSENSE Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter And Receiver Regulatory Class: Class II Product Code: DXH, DSB, BZQ, DPS, DQD Dated: October 4, 2020 Received: October 6, 2020
Dear Melissa Walker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K201669
Device Name Simple SENSE
#### Indications for Use (Describe)
The SimpleSENSE System is intended for use at home, or at a healthcare facility, under the direction of a licensed medical professional, to record, display and store the following physiological data: a) 2 leads of Electrocardiogram; b) Respiration rate measured through thoracic impedance; c) Heart Sounds; and d) Activity including posture. The device is intended for use when the clinician decides to evaluate the physiologic signals as an aid to diagnosis and treatment. The SimpleSENSE System is intended to be used by patients at rest and not performing any activities or movements. ECG recordings are indicated for the manual assessment of cardiac rhythm disturbances. The device does not produce alarms and is not intended for active patient monitoring (real-time). The device is not intended for use as life supporting equipment on high-risk patients such as critical care patients. The device is not intended for use in the presence of a pacemaker.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K201669 510(k) Summary
Submitter's information
Nanowear, Inc. 53 Boerum Pl, Suite 3F Brooklyn, NY 11201 United States 718-637-4815
Contact: Venk Varadan 53 Boerum Pl, Suite 3F Brooklyn, NY 11201 United States Phone: 718-637-4815 Date: June 15 2020
Classification The classification for the new device is shown in the table below.
| 21 CFR<br>Reference | Product<br>Code | Class | Trade Name | Classification Name |
|---------------------|-----------------|-------|-------------|--------------------------------------------------------------|
| §870.2920 | DXH | 2 | | Transmitters and Receivers,<br>Electrocardiograph, Telephone |
| §870.2770 | DSB | 2 | SimpleSENSE | Plethysmograph, Impedance |
| §868.2375 | BZQ | 2 | SimpleSENSE | Monitor, Breathing Frequency |
| §870.2340 | DPS | 2 | SimpleSENSE | Electrocardiograph |
| §870.1875 | DQD | 2 | SimpleSENSE | Stethoscope, Electronic |
Predicate The predicate device is: devices · K161431 SimplECG cleared on 11/30/2016 from Nanowear, Inc. The following devices are considered as reference devices: • K160981 Patient Monitor, models elite V5, elite V6 and elite V8 cleared on 12/22/2016 from Edan Instruments, Inc. • K151319 Eko Electronic Stethoscope System cleared on 5/18/2015 from Eko Devices, Inc. The SimpleSENSE System is intended for use at home, or at a healthcare facility, under Indications for use the direction of a licensed medical professional, to record, display and store the following physiological data: a) 2 leads of Electrocardiogram; b) Respiration rate measured through thoracic impedance; c) Heart Sounds; and d) Activity including posture. The device is intended for use when the clinician decides to evaluate the physiologic signals of adult patients as an aid to diagnosis and treatment. The SimpleSENSE System is intended to be used by patients at rest and not performing any activities or movements. ECG recordings are indicated for the manual assessment of cardiac rhythm disturbances. The device does not produce alarms and is not intended for active patient monitoring (real-time). The device is not intended for use as life supporting equipment on high-risk patients such as critical care patients. The device is not intended for use in the presence of a pacemaker.
{4}------------------------------------------------
Device The Nanowear SimpleSENSE device is the next generation diagnostic description monitoring technology that captures electrocardiographic (ECG) signals, respiration rate though thoracic impedance, heart sounds, activity including posture with sensors embedded on a wearable textile garment. The signals are stored and wirelessly transmitted to a smartphone, and forwarded to a medical professional for review.
> The garment is designed to be unobtrusive to everyday activity and provide an easy and efficient means of capturing ECG, respiration rate, heart sounds and activity data from patients.
> The garment is designed to be unobtrusive to everyday activity and provide an easy and efficient means of capturing ECG data from patients. The device consists of three (3) components:
- The SimpleSENSE Garment: an integrated network of nanosensor electrodes for measuring ECG and respiratory rate from thoracic impedance, and incorporating a MEMS microphone for measuring heart sounds.
- The SimpleSENSE Signal Acquisition Unit (SAU): data acquisition, storage, and transmission to an iPhone 7 using iOS 13.4. Incorporates an accelerometer to measure activity.
- The SimpleSENSE Mobile Application: mobile application for to start/stop a recording and to forward the test report to the medical professional.
{5}------------------------------------------------
| Characteristics | The table below provides a comparison of the New Device with the predicate and reference devices. | | | | |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Nanowear Inc.<br>SimpleSENSE - New<br>Device<br>Current Submission | Nanowear Inc. SimplECG<br>K161431 | Edan Instruments, Patient<br>Monitor, models elite V5,<br>elite V6 and elite V8<br>K160981 | Eko Devices Inc. Eko<br>Electronic<br>Stethoscope System<br>K151319 | Comparison |
| Indications for<br>Use Statement | The SimpleSENSE system is<br>intended for use at home,<br>or healthcare facility, under<br>the direction of a licensed<br>medical professional, to<br>record, display and store<br>the following physiological<br>data: a) 2 leads of<br>Electrocardiogram b)<br>Respiration rate measured<br>through thoracic<br>impedance; c) Heart<br>Sounds; and d) Activity<br>including posture. The<br>device is intended for use<br>when the clinician decides<br>to evaluate the physiologic<br>signals of adult patients as<br>an aid to diagnosis and<br>treatment. The<br>SimpleSENSE System is<br>intended to be used by<br>patients at rest and not<br>performing any activities or<br>movements. ECG | The SimplECG is intended<br>to aid in the diagnostic<br>evaluation of patients, 21<br>years of age and above, on<br>the order of a physician,<br>who experience transient<br>symptoms which may<br>suggest the need for<br>monitoring to manually<br>assess their cardiac rhythm<br>disturbance. ECG data is<br>recorded, stored,<br>transferred and displayed<br>wirelessly for review by a<br>physician who is skilled in<br>rhythm interpretation. | The monitors are intended<br>to be used for monitoring,<br>storing, and reviewing of,<br>and to generate alarms for,<br>multiple physiological<br>parameters of adults,<br>pediatrics and neonates.<br>The monitors are intended<br>for use by trained<br>healthcare professionals in<br>hospital environments. The<br>monitored physiological<br>parameters include: ECG,<br>respiration (RESP),<br>temperature (TEMP),<br>oxygen saturation of<br>arterial blood (SpO2), pulse<br>rate (PR), non-invasive<br>blood pressure (NIBP),<br>invasive blood pressure<br>(IBP), carbon dioxide<br>(CO2), cardiac output<br>(C.O.), anesthetic gas (AG),<br>bispectral index (BIS),<br>respiration mechanics (RM) | The Eko Electronic<br>Stethoscope System is<br>intended to be used as<br>a part of a physical<br>assessment of a<br>patient by healthcare<br>professionals for<br>diagnostic decision<br>support in clinical<br>settings. Eko is<br>intended for use on<br>pediatric and adult<br>patients. It can<br>electronically amplify,<br>filter and transfer<br>sounds to the<br>accompanying mobile<br>application for storage<br>and sharing. It can<br>used to record heart<br>sounds and cardiac<br>murmurs, bruits,<br>respiratory sounds and<br>abdominal sounds<br>during physical | The SimpleSENSE offers<br>the same ECG measures<br>as the SimplECG; heart<br>sounds as in the Eko<br>device, and a subset of<br>the measures included in<br>the Edan Patient Monitor<br>device.<br>The absence of the<br>measures does not affect<br>the substantial<br>equivalence of the New<br>Device. |
| Characteristic | Nanowear Inc.<br>SimpleSENSE - New<br>Device<br>Current Submission | Nanowear Inc. SimplECG<br>K161431 | Edan Instruments, Patient<br>Monitor, models elite V5,<br>elite V6 and elite V8<br>K160981 | Eko Devices Inc. Eko<br>Electronic<br>Stethoscope System<br>K151319 | Comparison |
| | recordings are indicated for<br>the manual assessment of<br>cardiac rhythm<br>disturbances. The device<br>does not produce alarms<br>and is not intended for<br>active patient<br>monitoring(real-time). The<br>device is not intended for<br>use as life supporting<br>equipment on high-risk<br>patients such as critical care<br>patients. The device is not<br>intended for use in the<br>presence of a pacemaker. | | and impedance<br>cardiography (ICG). BIS is<br>intended for use on adult<br>and pediatric patients.<br>ICG monitoring is intended<br>for use on adults only. The<br>arrhythmia detection and<br>ST Segment analysis are<br>intended for adult patients.<br>The monitors are<br>additionally intended for<br>use during patient<br>transport inside hospitals.<br>The monitors are not<br>intended for MRI<br>environments. | examination in normal<br>patients or those with<br>suspected diseases of<br>the cardiac, vascular,<br>pulmonary or<br>abdominal organ<br>systems. | |
| Product Code(s) | DXH, DPS, DQD, DSB, BZQ | DXH | MHX, DSI, DRT, CBQ, DXN,<br>DSK, CBR, DQA, NHO, CBS,<br>NHQ, NHP, CCK, DSB, CCL,<br>BZQ, BZK, DPS, FLL, DRG,<br>MLD | DQD | Similar |
| Characteristic | Nanowear Inc.<br>SimpleSENSE - New<br>Device<br>Current Submission | Nanowear Inc. SimplECG<br>K161431 | Edan Instruments, Patient<br>Monitor, models elite V5,<br>elite V6 and elite V8<br>K160981 | Eko Devices Inc. Eko<br>Electronic<br>Stethoscope System<br>K151319 | Comparison |
| Acquired Data | Electrocardiogram<br>(EKG/ECG)<br>Respiration Rate derived<br>from thoracic impedance<br>Heart Sounds | Electrocardiogram<br>(EKG/ECG) | ECG, Respiration Rate,<br>Temperature SpO2, pulse<br>rate non-invasive blood<br>pressure (NIBP), invasive<br>blood pressure (IBP),<br>carbon dioxide (CO2),<br>cardiac output (C.O.),<br>anesthetic gas (AG),<br>Bispectral Index (BIS),<br>respiration mechanics (RM)<br>Impedance Cardiography<br>(ICG). | Heart Sounds | Same measures as those<br>found in the predicate &<br>reference devices. The<br>New Device captures the<br>ECG signal in the same<br>way as the SimplECG<br>predicate. The New<br>Device captures some of<br>the measures in the Edan<br>reference and the heart<br>sounds as in the Eko<br>reference. The absence of<br>some measures does not<br>affect the substantial<br>equivalence of the<br>SimpleSENSE device. |
| | Electrocardiogram<br>(EKG/ECG): Textile-based<br>Nanosensors | Electrocardiogram<br>(EKG/ECG): Textile-based<br>Nanosensors | Not Applicable | Not Applicable | Same sensors as in the<br>SimplECG device are used<br>to capture signals |
| Sensor<br>Technology | Respiration Rate: Textile-<br>based Nanosensors<br>measure thoracic<br>impedance and<br>respiration is derived<br>from thoracic impedance | Not Applicable | Thoracic<br>Impedance/Impedance<br>Cardiogram (ICG), ECG:<br>Disposable Silver or<br>Silver/Silver Chloride<br>electrode.<br>Respiration Rate: derived<br>from thoracic impedance | Not Applicable | Both the New Device and<br>the Edan reference<br>measure and derive<br>respiration rate using<br>thoracic impedance<br>captured via a sensor.<br>The sensors in the two<br>devices are different,<br>however both types of<br>sensors are used for |
| Characteristic | Nanowear Inc.<br>SimpleSENSE - New<br>Device<br>Current Submission | Nanowear Inc. SimplECG<br>K161431 | Edan Instruments, Patient<br>Monitor, models elite V5,<br>elite V6 and elite V8<br>K160981 | Eko Devices Inc. Eko<br>Electronic<br>Stethoscope System<br>K151319 | Comparison |
| | | | | | physiologic signal capture<br>in other cleared devices. |
| | Heart Sound:<br>Microelectromechanical<br>(MEMS) microphone | Not Applicable | Not Applicable | Heart Sound: Uses a<br>standard stethoscope<br>diaphragm. Method of<br>transduction and<br>digitization of sound<br>waves is not Available | The SimpleSENSE device<br>uses a different sensor<br>than the Eko reference.<br>However, the signal<br>acquisition and location<br>used in both the<br>SimpleSENSE device and<br>the Eko reference are<br>similar. |
| | Electrocardiogram<br>(EKG/ECG): Standard<br>Bipolar lead<br>instrumentation amplifier<br>and Sigma-Delta Analog<br>to Digital Converter. | Electrocardiogram<br>(EKG/ECG): Standard<br>Bipolar lead<br>instrumentation amplifier<br>and Sigma-Delta Analog to<br>Digital Converter | Not Applicable | Not Applicable | The signal acquisition<br>method in the New Device<br>for capturing ECG signal is<br>the same as the SimplECG<br>predicate. |
| Signal<br>acquisition<br>method | Respiration rate: Thoracic<br>Impedance is measured<br>using four-point probe<br>using low amplitude<br>current applied to the<br>body and impedance<br>measured from voltage<br>drop derived from<br>thoracic impedance.<br>(Respiration Rate<br>detection range 6 - 22 | Not Applicable | Thoracic<br>Impedance/Impedance<br>Cardiograph: four-point<br>probe using low amplitude<br>current applied to the body<br>and impedance measured<br>from voltage drop.<br>Respiration Rate: derived<br>from thoracic impedance | Not Applicable | The signal acquisition<br>method for capturing<br>thoracic impedance is the<br>same as the Edan<br>reference. |
| Characteristic | Nanowear Inc.<br>SimpleSENSE - New<br>Device<br>Current Submission | Nanowear Inc. SimplECG<br>K161431 | Edan Instruments, Patient<br>Monitor, models elite V5,<br>elite V6 and elite V8<br>K160981 | Eko Devices Inc. Eko<br>Electronic<br>Stethoscope System<br>K151319 | Comparison |
| | breaths per minute (BPM)<br>with accuracy ± 2 BPM) | | | | |
| | Heart Sound: solid state<br>Microelectromechanical<br>(MEMS) microphone<br>embedded in garment<br>and located near Apex of<br>heart. | Not Applicable | Not Applicable | Heart Sound: sensor<br>embedded in a<br>standard stethoscope<br>attachment that can<br>amplify, digitize and<br>transmit data. | The signal acquisition<br>method for detecting<br>heart sounds is similar to<br>the Eko reference. Both<br>devices are placed in<br>similar locations to detect<br>and capture heart<br>sounds. |
| Display Type | No on-device display. | No on-device display. | On Device Display of<br>waveforms acquired and<br>derived parameters is<br>available. Data can be<br>downloaded from the<br>device through an Ethernet<br>connection or a USB<br>connection. | No on-device display.<br>Display of data is<br>available on a<br>smartphone through a<br>smartphone app or on<br>the web using web<br>services. | Data display for the New<br>Device is the same as the<br>SimplECG predicate and<br>the Eko reference that<br>rely on wireless<br>transmission and display<br>on a smartphone or web-<br>based display. |
| Display<br>Requirement | User provided display<br>hardware for a<br>healthcare professional<br>to view the recorded<br>data. A general-purpose<br>PC/Laptop/Desktop or<br>mobile device may be<br>used. | User provided display<br>hardware for a healthcare<br>professional to view the<br>recorded data. A general-<br>purpose<br>PC/Laptop/Desktop may be<br>used. | No requirement for<br>additional user provided<br>hardware for display | Display available<br>through User provided<br>smartphone or tablet.<br>Display also available<br>through Web browser<br>on any<br>Laptop/PC/Desktop. | The Edan reference<br>includes a display<br>(hardware) while the New<br>Device, the SimplECG, and<br>the Eko reference all<br>display the data on the<br>user's hardware and via<br>web based display. |
| Power Source | Internally powered using<br>Li-Ion rechargeable<br>battery | Single Use Lithium AA type<br>batteries | Rechargeable Lithium-Ion<br>Battery and AC Mains<br>supported. | Internally powered<br>using Li-Ion<br>rechargeable battery | All of the devices are able<br>to operate on battery<br>power. The Edan<br>reference has an option |
| Characteristic | Nanowear Inc.<br>SimpleSENSE - New<br>Device<br>Current Submission | Nanowear Inc. SimplECG<br>K161431 | Edan Instruments, Patient<br>Monitor, models elite V5,<br>elite V6 and elite V8<br>K160981 | Eko Devices Inc. Eko<br>Electronic<br>Stethoscope System<br>K151319 | Comparison |
| Internal<br>Memory/data | Removeable MicroSD<br>card | Removeable MicroSD card | Solid state memory non-<br>removeable. | Solid state memory<br>non-removeable | All devices use memory<br>devices to store the data<br>collected. The New<br>Device and the SimplECG<br>predicate use removeable<br>cards while the Edan and<br>Eko references use non-<br>removable memory<br>devices. |
| Communication<br>Interface | Wireless transceiver<br>using Bluetooth | Wireless transceiver using<br>Bluetooth…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.