K201650 · Uro-1, Inc. · KNW · Jan 15, 2021 · Gastroenterology, Urology
Device Facts
Record ID
K201650
Device Name
VMCore Biopsy Needle
Applicant
Uro-1, Inc.
Product Code
KNW · Gastroenterology, Urology
Decision Date
Jan 15, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The VMCore biopsy needle is intended for use with the Bard Magnum reusable biopsy instrument for biopsies of soft tissues, such as the prostate, lung, kidney, or liver. The VMCore biopsy needle is not intended for use in bone
Device Story
VMCore Biopsy Needle is a disposable, two-part stainless steel needle set (outer cannula and core collector) with plastic hubs. Used with Bard Magnum reusable biopsy instrument to collect soft tissue samples (prostate, lung, kidney, liver). Operated by physicians in hospitals, clinics, or offices under image guidance. Device functions via single puncture to retrieve tissue core. Output is a physical tissue sample for diagnostic evaluation. Benefits include minimally invasive tissue acquisition for pathology.
Clinical Evidence
No clinical testing performed. Bench testing included real-time aging, in vitro tissue collection volume/weight analysis at various depths, hub-to-needle attachment strength, and cannula bend strength. Biocompatibility (cytotoxicity) was established via comparison to the reference device (K180214).
Indicated for use with the Bard Magnum reusable biopsy instrument for soft tissue biopsies (prostate, lung, kidney, liver). Not for use in bone.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
January 15, 2021
URO-1, Inc. Thomas Lawson, Ph.D. Director, Regulatory Affairs 111 North Chestnut Street. Suite 106 Winston Salem, NC 27101
Re: K201650 Trade/Device Name: VMCore Biopsy Needle Regulation Number: 21 CFR§ 876.1075 Regulation Name: Gastroenterology-Urology-Biopsy Instrument Regulatory Class: II Product Code: KNW Dated: December 14, 2020 Received: December 15, 2020
Dear Thomas Lawson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thelma Valdes, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201650
Device Name VMCore Biopsy Needle
#### Indications for Use (Describe)
The VMCore biopsy needle is intended for use with the Bard Magnum reusable biopsy instrument for biopsies of soft tissues, such as the prostate, lung, kidney, or liver. The VMCore biopsy needle is not intended for use in bone
| Type of Use (Select one or both, as applicable) | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| <table><tr><td><span> <b> ☑ </b> </span> Prescription Use (Part 21 CFR 801 Subpart D) </td><td><span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </td></tr></table> | <span> <b> ☑ </b> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
| <span> <b> ☑ </b> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{4}------------------------------------------------
# SECTION 5.
# 510(k) SUMMARY
### General Information
| Submitter | URO-1, Inc. |
|-------------------------|----------------------------------------------------------------------------------|
| Address | URO-1, Inc.<br>111 North Chestnut Street<br>Suite 106<br>Winston Salem, NC 27101 |
| FDA Registration Number | No yet assigned |
| Correspondence Person | Thomas Lawson, PhD |
| Contact Information | Email: drthomlawson@gmail.com<br>Phone: 510-206-1794 |
| Date Prepared | 15 January 2021 |
# Proposed Device
| Trade Name | VMCore Biopsy Needle |
|-----------------------|--------------------------------------------------|
| Common Name | VMCore Biopsy Needle |
| Regulation Number and | 21 CFR§876.1075, Gastroenterology-Urology Biopsy |
| Classification Name | Instrument |
| Product Code | KNW |
| Regulatory Class | II |
### Predicate Device
| Trade Name | RP Cutting Needle |
|------------------------------------------------------------------------------|----------------------------------------------------------------|
| Common Name | RP Cutting Needle |
| Premarket Notification | K092059 |
| Regulation Number and<br>Classification Name | 21 CFR§876.1075, Gastroenterology-Urology Biopsy<br>Instrument |
| Product Code | KNW |
| Regulatory Class | II |
| Note: This predicate device has not been subject to a design-related recall. | |
{5}------------------------------------------------
| Trade Name | Reprise Bladder Injection System |
|------------------------------------------------------------------------------|--------------------------------------------|
| Common Name | Reprise Injector |
| Premarket Notification | K180214 |
| Regulation Number and<br>Classification Name | 21 CFR§876.1500, Endoscope and Accessories |
| Product Code | FBK |
| Regulatory Class | II |
| Note: This reference device has not been subject to a design-related recall. | |
### Device Description
The VMCore Biopsy Needle Set is a set of two disposable needles used for collection of tissue samples. The set consists of an Outer Cannula and the Core Collector. The needles are made of stainless steel and have a plastic hub attached at their proximal ends.
### Indications for Use
The indications for use for the VMCore Biopsy Needle is:
The VMCore Biopsy Needle is intended for use with the Bard Magnum reusable biopsy instrument for biopsies of soft tissue, such as prostate, lung, kidney, or liver. The VMCore Biopsy Needle is not intended for use in bone.
Both the subject device and the predicate device have intended use of obtaining tissue samples during biopsy of soft tissue.
#### Comparison of Technological Characteristics with the Predicate Device
URO-1, Inc. has identified the RP Cutting Needle (K092059) as the predicate device. The VMCore Biopsy Needle is substantially equivalent to the predicate device based upon the following similarities:
- The intended use of the predicate device and the subject device is for obtaining 1. biopsies from soft tissue;
- 2. The two devices are introduced into the body under imaging control;
{6}------------------------------------------------
- 3. Both needles are advanced into tissue via a biopsy gun, specifically the Bard Magnum Biopsy Reusable Biopsy Instrument;
- 4. The subject device is similar to the predicate device in terms of dimensions of the needles: and
- న్. Both devices are made from biocompatible materials.
The Repris Bladder Injection System (K180214) is also manufactured for Uro-1, Inc. and is a reference device in this submission since the VMCore Biopsy Needle will use the material for its needles as did the Repris device, will be packaged in the exact same pouch as is the Repris device for a single unit, they both are sterilized by radiation.
| | Subject Device | Predicate Device |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | VMCore Biopsy<br>Needle | RP Cutting Needle<br>(Riverpoint Medical) |
| | (URO-1, Inc.)<br>(This Submission) | K092059 |
| Device Class | II | II |
| FDA Product Code | KNW | KNW |
| Product<br>Classification | 876.1075 | 876.1075 |
| Indication for Use | The VMCore Biopsy<br>Needle is indicated for<br>use with the Bard<br>Magnum reusable<br>biopsy instrument for<br>biopsies of soft tissue,<br>such as prostate, lung,<br>kidney, or liver. | The RP Cutting Needle is<br>indicated for use by<br>medical professionals<br>with the Bard Magnum<br>reusable biopsy<br>instrument for biopsies of<br>soft tissue, such as liver,<br>lung, kidney, or prostate. |
| Intended use | To obtain biopsies from<br>soft tissue, such as | To obtain biopsies from<br>soft tissue, such as liver, |
## Comparison of the VMCore Biopsy Needle to the predicate device, the RP Cutting Needle.
{7}------------------------------------------------
VMCore Biopsy Needle
| | prostate, lung, kidney, or liver. It is not intended for use in bone. | lung, kidney, or prostate. It is not intended for use in bone. |
|----------------------------------|-----------------------------------------------------------------------|----------------------------------------------------------------|
| Technical<br>Characteristics | | |
| Components of the<br>Set | Outer Cannula<br>Core Collector Needle | Outer Cannula<br>Biopsy Needle |
| Outer Diameter<br>(OD) of needle | 16 & 18 gauge | 14 to 20 gauge |
| Length of needle<br>assembly | 20 cm and 25 cm | 10 and 25 cm |
| Needle material | Stainless Steel | Same |
| Needle hub<br>material | Polycarbonate<br>Acrylonitrile butadiene<br>styrene (ABS) | Plastic not identified in<br>labeling |
| Operational<br>Characteristics | | |
| Mechanism of<br>action | Single puncture and<br>sample | Same |
| Accessory Device | Bard Magnum Biopsy<br>Instrument | Same |
| Method of<br>advancement | Advanced under image<br>guidance | Same |
| Location of<br>Procedure | Hospitals, clinics,<br>physician offices | Same |
| Duration of use | < 24 hours | < 24 hours |
| Single-use catheter | Yes | Yes |
| Provided Sterile | Yes | Yes |
### Performance Data
The performance testing conducted establishes that the VMCore Biopsy Needle does not raise new questions of the safety and effectiveness.
#### Biocompatibility testing
The VMCore Biopsy Needle is made from Stainless Steel 304 and passed cytotoxicity testing following sterilization. This material is the same as was used in the construction of another device marketed by Uro-1, Inc., the Reprise Bladder Injection System, and that
{8}------------------------------------------------
K201650
device was reviewed and cleared under K180214. The results of testing the Repris device for the following tests:
- o Cytotoxicity;
- Sensitization; and O
- Irritation. o
are comparable to expected results with the VMCore Biopsy Needle so they were not necessary to repeat.
#### Mechanical Testing
The mechanical testing of the subject device included:
- Dimensions of the device at one-month real-time aging; ●
- Volume of tissue collection in an in vitro model .
- Volume of tissue collected from 3 different depths of insertion in an in vitro . model
- Measurement of the tissue weight collected in the first sample taken and the last ● sample taken by the device from an in vitro model;
- Measure of the strength of the attachment of the hub to the needle; and ●
- Bend strength of the needle cannula.
#### Animal Testing
No preclinical testing of the subject device was necessary.
#### Clinical Studies
No clinical testing of the subject device was necessary.
#### Conclusion
The information submitted in this premarket notification confirms that the VMCore Biopsy Needle raises no new questions of safety and effectiveness and that the VMCore Biopsy Needle is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.