← Product Code [KWQ](/productcode/KWQ) · K201587

# icotec Anterior Cervical Plate System (K201587)

_Icotec AG · KWQ · Oct 15, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K201587

## Device Facts

- **Applicant:** Icotec AG
- **Product Code:** [KWQ](/productcode/KWQ.md)
- **Decision Date:** Oct 15, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The icotec Anterior Cervical Plate System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine (C2-T1) in whom life expectancy is of insufficient duration to permit achievement of fusion.

## Device Story

Anterior cervical plate system consisting of plates and self-tapping screws; used for anterior cervical fixation (C2-T1). Plates are precontoured, 18mm wide, 1-4 segments (21-94mm length). Screws (4.0/4.25mm diameter) feature fully threaded conical heads that lock into plate holes to prevent pullout; screw angulation prevents penetration of lower vertebral end plate. System facilitates conventional ventral surgical approaches. Used by surgeons in clinical settings to restore spinal column integrity in tumor patients. Benefits include temporary stabilization of cervical spine in patients unable to undergo fusion.

## Clinical Evidence

No clinical data. Evidence based on bench testing including static compression bending, static torsion, and dynamic compression bending per ASTM F1717, and biocompatibility assessment.

## Technological Characteristics

Spinal intervertebral body fixation orthosis. System includes precontoured plates (18mm width) and self-tapping bone screws with threaded conical heads for locking. Mechanical testing performed per ASTM F1717. Biocompatible materials.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Aesculap, Inc. (AIS) ABC2 Cervical Plating System ([K050813](/device/K050813.md))
- EBI VueLock™ Anterior Cervical Plate System ([K023133](/device/K023133.md))
- Synthes Spine Anterior CSLP System ([K030866](/device/K030866.md))

## Reference Devices

- VADER®one Pedicle System MIS ([K190545](/device/K190545.md))
- Lightmore® Pedicle System 6.0 ([K193423](/device/K193423.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 15, 2020

icotec ag % Margeaux Rogers Associate Director, Regulatory Affairs MCRA, LLC 1050 K Street NW Suite 1000 Washington, District of Columbia 20001

Re: K201587

Trade/Device Name: icotec Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWO Dated: September 18, 2020 Received: September 18, 2020

Dear Margeaux Rogers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-

combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201587

Device Name icotec Anterior Cervical Plate System

#### Indications for Use (Describe)

The icotec Anterior Cervical Plate System is integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage turnors involving the cervical spine (C2-T1) in whom life expectancy is of insufficient duration to permit achievement of fusion.

| Type of Use (Select one or both, as applicable)                                                          |                                                                                                         |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Device Trade Name: | icotec Anterior Cervical Plate System                                                                                                                  |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:      | icotec ag<br>Industriestrasse 12<br>9450 Altstätten<br>Switzerland<br>www.icotec-medical.com<br>Phone: +41 71 757.0000                                 |
| Contact:           | Ms. Marina Hess<br>CQO/Management Representative<br>icotec ag                                                                                          |
| Prepared by:       | Ms. Margeaux Rogers<br>Associate Director, Regulatory Affairs<br>MCRA, LLC<br>1050 K Street NW, Suite 1000<br>Washington, DC 20001<br>mrogers@mcra.com |
| Date Prepared:     | September 18th, 2020                                                                                                                                   |
| Classifications:   | 21 CFR §888.3060, Spinal intervertebral body fixation orthosis                                                                                         |
| Class:             | II                                                                                                                                                     |
| Product Codes:     | KWQ                                                                                                                                                    |

### Indications for Use:

The icotec Anterior Cervical Plate System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine (C2-T1) in whom life expectancy is of insufficient duration to permit achievement of fusion.

### Device Description:

The icotec Anterior Cervical Plate System consists of plates and screws intended for use in anterior cervical fixation from C2 to T1. The system is developed to allow for conventional ventral approaches to the cervical spine.

The icotec Anterior Cervical Plate System is available in various plate lengths and with selftapping screws for the specific adaption to the patient's anatomy. The icotec Anterior Cervical Plates are 18mm in width and come as 1- to 4-segmental implants with lengths ranging from 21

{4}------------------------------------------------

up to 94mm. The plates are precontoured to better fit patient anatomy. The screw holes in the plates are conical and threaded.

The icotec ACP self-tapping screws are available in diameters of 4.0 and 4.25mm with lengths of 13 and 15mm. The fully threaded bone screws have threaded conical heads to firmly lock into the plate. The conical threaded screw heads are designed to block pullout while screw angulation in the cranial direction prevents screws from penetrating through the lower end plate of the vertebral body.

# Primary Predicate Devices:

The icotec Anterior Cervical Plate System is substantially equivalent in device indications, design, and performance to the following primary predicate device.

- Aesculap, Inc. (AIS) ABC2 Cervical Plating System (K050813) ●
# Additional Predicate Devices:

The following additional predicate devices are presented to further demonstrate substantial equivalence since these devices contribute incremental benefit to the substantial equivalence profile with the device indications, design, mechanical performance, and clinical performance:

- EBI VueLock™ Anterior Cervical Plate System (K023133)
- Synthes Spine Anterior CSLP System (K030866) ●

The following reference device is substantially equivalent to the subject device with respect to materials and manufacturing technology and methods:

- VADER®one Pedicle System MIS and Lightmore® Pedicle System 6.0 (K190545, ● K193423)
### Performance Testing Summary:

The testing of the icotec Anterior Cervical Plate System includes:

- Static compression bending, static torsion and dynamic compression bending per ASTM F1717
- -Biocompatibility Assessment
- Clinical Data -

# Substantial Equivalence:

The subject devices were demonstrated to be substantially equivalent to predicates cited above with respect to indications, design, materials, function, manufacturing, and performance. The nonclinical tests performed by the company, including static compression, static torsion and dynamic compression per ASTM F1717, met the pre-defined acceptance criteria. The results of the performed tests demonstrate that the icotec Anterior Cervical Plate System is substantially equivalent to the legally marketed predicate devices.

{5}------------------------------------------------

# Conclusion:

The purpose of the traditional 510(k) is to receive regulatory clearance to introduce the icotec Anterior Cervical Plate System to interstate commerce. Substantial equivalence has been demonstrated to the cited predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K201587](https://fda.innolitics.com/device/K201587)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
