← Product Code [MNR](/productcode/MNR) · K201495

# Maxxi Flow Sensor (K201495)

_Neurovirtual USA, Inc. · MNR · Dec 14, 2020 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K201495

## Device Facts

- **Applicant:** Neurovirtual USA, Inc.
- **Product Code:** [MNR](/productcode/MNR.md)
- **Decision Date:** Dec 14, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.2375
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology

## Indications for Use

The Maxxi Flow sensor is a respiratory sensor. It is placed under the patient's and produce a signal that is directly proportional to the temperature changes of air inhaled and exhaled during respiration. Maxxi Flow sensors generate a small analog electrical signal that provides a clear, reliable indication of respiration airflow. This sensor is intended to be used with polysomnography devices and adult patients. Intended to be use in a sleep laboratory, clinics or hospitals.

## Device Story

Respiratory thermocouple sensor; placed under patient nostrils; detects temperature changes between ambient (inhalation) and lung (exhalation) air. Thermocouple elements generate small analog electrical signal proportional to temperature change; signal transmitted via insulated Teflon wire to polysomnography (PSG) amplifier. Used in sleep labs, clinics, or hospitals by clinicians. Provides clear, reliable indication of respiration airflow; aids in monitoring breathing frequency. Benefits patient by enabling accurate respiratory data acquisition during sleep studies.

## Clinical Evidence

Bench testing only. Comparative performance testing against predicate (Easy Flow) evaluated dimensions, cable length, visual aspects, sensor resistance, and output signal frequency/amplitude (0-30 Hz range). Results confirmed equivalence within ±15% for signal metrics and ±10% for dimensions. Environmental temperature testing showed consistent frequency and minimal amplitude/resistance variance compared to predicate.

## Technological Characteristics

Thermocouple-based respiratory flow sensor. Materials: Thermoplastic elastomers (TPC-ETs) for sensor body, insulated Teflon wire. Connectivity: DIN 42-802 touch-proof connector. Signal: Analog AC, 0-30 Hz frequency range. Resistance: <10Ω. Dimensions: 0.032 x 1.36 x 0.23 inches. Standards: ISO 13485:2003, ISO 10993-1:2009, IEC 60601-1 (3rd ed), IEC 60601-1-2:2014. Non-powered, passive device.

## Regulatory Identification

A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.

## Predicate Devices

- Thermocouple Flow Sensor (Easy Flow) ([K922112](/device/K922112.md))

## Reference Devices

- ULTIMA AIRFLOW SENSOR ([K981445](/device/K981445.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 14, 2020

Neurovirtual USA, Inc. Eduardo Faria CEO 3303 W Commercial Blvd #100 Fort Lauderdale, Florida 33309

Re: K201495

Trade/Device Name: Maxxi Flow Sensor Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: MNR Dated: November 10, 2020 Received: November 12, 2020

Dear Eduardo Faria:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Malvina B. Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201495

Device Name Maxxi Flow Sensor

Indications for Use (Describe)

The Maxxi Flow sensor is a respiratory sensor. It is placed under the patient's and produce a signal that is directly proportional to the temperature changes of air inhaled and exhaled during respiration. Maxxi Flow sensors generate a small analog electrical signal that provides a clear, reliable indication of respiration airflow. This sensor is intended to be used with polysomnography devices and adult patients.

Intended to be use in a sleep laboratory, clinics or hospitals.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D)  | Over-The-Counter Use (21 CFR 801 Subpart C) |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Neurovirtual. The logo consists of a stylized waveform in shades of orange on the left, followed by the word "NEUROVIRTUAL" in blue. The "O" in "NEUROVIRTUAL" is replaced with an orange circle.

## 510(k) SUMMARY

Submitter's Name: Neurovirtual USA, Inc.

Owner / Operator Registration Number: 3006125362 FEI Number: 3006125362 Manufacture Registration Number: 3006125362

Address: 3303 W Commercial Blvd #100 Fort Lauderdale, FL 33309 USA

#### Phone and Fax Numbers

Phone: (786) 693-8200 Fax: (305) 393-8429

Contact Person: Eduardo J. Faria

Preparation Date: December 11, 2020

#### Classification Name:

Common / Usual Name: Breathing frequency monitor Proprietary Name: Maxxi Flow Sensor Product Code: MNR Class: Class II Regulation: 21 CFR 868.2375

## Device Description:

MAXXI FLOW SENSOR Thermocouple based sensors detect change of breath temperature between ambient temperature (inhalation) and lung temperature (exhalation). A thermocouple placed in front of a nostril detects breathing as a temperature change.

The product is offered in 2 different length sizes, 3ft and 7ft.

Maxxi Flow sensor is compatible with Neurovirtual PSG devices.

## Substantial Equivalence:

The Maxxi Flow Sensor is substantial equivalent with the following products:

Primary Predicate

| 510(k) Number | Model                                | Company                                             |
|---------------|--------------------------------------|-----------------------------------------------------|
| K922112       | Thermocouple Flow Sensor (Easy Flow) | S.L.P. Ltd.<br>Former EPM INFORMATION SYSTEMS, INC. |

Reference device:

| 510(k) Number | Model                 | Company               |
|---------------|-----------------------|-----------------------|
| K981445       | ULTIMA AIRFLOW SENSOR | BRAEBON MEDICAL CORP. |

The reference predicate was added in order to support to demonstrate the substantial equivalence with the subject device because it has similar anatomical and technological characteristics, as well it is not powered with battery which is the same as the subject device.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Neurovirtual. The logo consists of a stylized waveform graphic in shades of orange, followed by the word "NEUROVIRTUAL" in blue, block letters. The "O" in Neurovirtual is replaced with an orange circle.

- 1. Indications for Use:

| Indications for Use Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                  |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Neurovirtual                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | S.L.P. Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Braebon                                                                                          |
| Maxxi Flow Sensor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Thermocouple Flow Sensor (Easy flow)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | ULTIMA AIRFLOW SENSOR                                                                            |
| The Maxxi Flow sensor is a respiratory<br>sensor. It is placed under the patient's<br>nostrils and produce a signal that is<br>directly proportional to the temperature<br>changes of air inhaled and exhaled during<br>respiration. Maxxi Flow sensors generate<br>a small analog electrical signal that<br>provides a clear, reliable indication of<br>respiration airflow. This sensor is intended<br>to be used with polysomnography devices<br>and adult patients.<br><br>Intended to be use in a sleep laboratory,<br>clinics or hospitals. | The Easy flow sensor is a respiratory<br>sensor. It is placed under the patient's<br>nostrils and produce a signal that is<br>directly proportional to the temperature<br>changes of air inhaled and exhaled during<br>respiration. Easy flow sensors generate a<br>small analog electrical signal that<br>provides a clear, reliable indication of<br>respiration airflow. This sensor is<br>intended to be used in polysomnography<br>devices<br><br>Intended to be use in a sleep laboratory,<br>clinics or hospitals | A Qualitative measure of respiratory<br>airflow for recording onto a data<br>acquisition system. |

## 2. Technological Characteristics Comparison:

The predicate devices used to establish substantial equivalence for the Maxxi Flow Sensor are outlined below. This section of this submission will provide a comparison of design, materials, and technical specifications of the Maxxi Flow Sensor to each of the predicate devices stratified by functional modality.

| Specifications Comparison |                                                                                                              |                                                                                                           |                                                                                                              |
|---------------------------|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| Device                    | Neurovirtual<br>Maxxi Flow Sensor                                                                            | S.L.P. Ltd.<br>Thermocouple Flow Sensor (Easy<br>flow)                                                    | Braebon<br>ULTIMA AIRFLOW SENSOR                                                                             |
| 510(k) Number             | K201495                                                                                                      | K922112                                                                                                   | K981445                                                                                                      |
| Regulation                | 21 CFR 868.2375                                                                                              | 21 CFR 868.2375                                                                                           | 21 CFR 868.2375                                                                                              |
| Classification            | MNR                                                                                                          | BZQ                                                                                                       | BZQ                                                                                                          |
| Target population         | Adults                                                                                                       | All Ages                                                                                                  | All Ages                                                                                                     |
| Environment               | sleep laboratory, clinics or<br>hospitals                                                                    | sleep laboratory, clinics or<br>hospitals                                                                 | sleep laboratory, clinics or<br>hospitals                                                                    |
| Application               | Respiratory thermocouple flow<br>sensor                                                                      | Respiratory thermocouple flow<br>sensor                                                                   | Respiratory thermocouple flow<br>sensor                                                                      |
| Mechanical application    | Sensor is placed under the<br>patient's nostrils                                                             | Sensor is placed under the<br>patient's nostrils                                                          | Sensor is placed under the<br>patient's nostrils                                                             |
| Structure                 | Thermocouple elements<br>covered by TPC plastic<br>injection with a wire to<br>connect to the PSG amplifier. | Thermocouple elements covered<br>by TPC plastic injection with a wire<br>to connect to the PSG amplifier. | Thermocouple elements<br>covered by TPC plastic injection<br>with a wire to connect to the<br>PSG amplifier. |
| Contact Material          | Thermoplastic elastomers<br>(TPC-ETs)                                                                        | Thermoplastic elastomers (TPC-<br>ETs)                                                                    | Thermoplastic elastomers (TPC-<br>ETs)                                                                       |
| Wire Material             | Insulated Teflon wire                                                                                        | Insulated Teflon wire                                                                                     | Insulated Teflon wire                                                                                        |
| Wire Colors               | White                                                                                                        | White                                                                                                     | White                                                                                                        |
| Connector                 | DIN 42-802<br>touch proof                                                                                    | DIN 42-802<br>touch proof                                                                                 | DIN 42-802<br>touch proof                                                                                    |
| Signal output             | AC signal                                                                                                    | AC signal                                                                                                 | AC signal                                                                                                    |
| Signal Frequency range    | 0-30 HZ                                                                                                      | 0-30 HZ                                                                                                   | 0-30 HZ                                                                                                      |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Neurovirtual. The logo features a stylized waveform graphic in shades of orange, with the darker orange portion resembling an EKG. To the right of the waveform is the company name, "NEUROVIRTUAL," in bold, dark blue letters. The "O" in "NEUROVIRTUAL" is a lighter orange circle.

| Sensor Resistance | <10Ω                     | <10Ω                     | Not disclosed by manufacturer |
|-------------------|--------------------------|--------------------------|-------------------------------|
| Sensor Dimensions | .032 x 1.36 x 0.23 (in)  | 0.30 x 1.3 x 0.22 (in)   | 0.30 x 1.3 x 0.22 (in)        |
| Single-Use        | No                       | No                       | No                            |
| Cable length      | 3ft and 7ft              | 3ft and 7ft              | 3ft and 7ft                   |
| Package           | Plastic bag              | Plastic bag              | Plastic bag                   |
| Pictures          | Image: sensor with wires | Image: sensor with wires | Image: sensor with wires      |

Discussion: The Maxxi Flow and the predicate devices, are substantially equivalent in technology, function, and intended use, all sensors are indicated to provide respiratory airflow signal for a polysomnography device; all devices use the same method of action; all devices use equivalent material and all devices provide equivalent output signals.

#### I) Applied Standards:

In order to reach high guality and effectiveness the Maxxi Flow Sensor is produced in compliance with the following standards:

- ISO 13485:2003, "Medical Devices, Quality Management Systems: Requirements for Regulatory -Purposes" and FDA GMP "Good Manufacturing Practices".
- -ISO10993-1:2009, "Biological Evaluation of Medical Devices"
- -60601-1 3a edition - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- -60601-1-2:2014 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

#### J) Non-clinical Testing:

The test performance comparing the Maxxi Flow and the predicate device Easy Flow manufactured by S.L.P Ltd was performed and the summary results are show in the summary below.

| Test              | Description                                                                                                                                                              | Criteria                                                                    | Results                                                                                                                    |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Parts Dimensions  | Verify if the dimensions of the<br>thermocouple unit are within the<br>acceptable range and equivalent to the<br>predicate                                               | Allowance: +-10%                                                            | The dimensions of the thermocouple unit are<br>within the acceptable criteria as when<br>compared to the predicate device. |
| Cable Length      | Verify if the wire lead length is within the<br>acceptable range and equivalent to the<br>predicate                                                                      | Expected: 3 ft and 7 ft<br>Allowance: +-5%                                  | The cable length of the sensor is within the<br>acceptable criteria as when compared to the<br>predicate device.           |
| Visual Conditions | Verify the visual aspects of the product and<br>equivalency with the predicate. Cable<br>aspects, labeling, flexibility, plastic finishing,<br>and connector conditions. | All aspects must be substantially<br>equivalent to the predicate<br>device. | Maxxi Flow sensor was inspected, and the result<br>is equivalent to the predicate device.                                  |
| Sensor Resistance | Verify the resistance of the sensor using the<br>ohmmeter in a controlled test environment<br>where the temperature is stable.                                           | Expected: 0-10 ohms                                                         | The Maxxi Flow was inspected, and the<br>resistance is between the acceptable criteria<br>range. (2.0 ohms)                |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Neurovirtual. The logo features a stylized waveform graphic in shades of orange on the left, followed by the word "NEUROVIRTUAL" in blue. The "O" in Neurovirtual is a solid orange circle.

| Output signals<br>Frequency and Amplitude<br>Tests | The sensor Maxxi Flow and the predicate<br>Easy Flow were connected to the same PSG<br>recorder and the acquired data was<br>compared in frequency and amplitude. | Allowance: +-15% | Both sensors acquired equivalent signals within<br>the acceptable range. |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|--------------------------------------------------------------------------|
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|--------------------------------------------------------------------------|

Additionally, the test performance with different environment temperatures was performed. The results showed exact same frequency as the predicate, with minimal difference on voltage amplitude and resistance measurement values.

Conclusion: The Maxxi Flow and the predicate device Easy Flow manufactured by S.L.P Ltd., are substantially equivalent in technology, function, and intended use.

---

**Source:** [https://fda.innolitics.com/device/K201495](https://fda.innolitics.com/device/K201495)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
