← Product Code [NXC](/productcode/NXC) · K201450

# Straight T Clear Dental Aligner (K201450)

_Straight T, Inc. · NXC · Sep 10, 2020 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K201450

## Device Facts

- **Applicant:** Straight T, Inc.
- **Product Code:** [NXC](/productcode/NXC.md)
- **Decision Date:** Sep 10, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.5470
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Straight T Clear Dental Aligners indicated for use in the alignment of permanent teeth (i.e. all second molars) through orthodontic treatment of misalignment and malocclusion.

## Device Story

Straight T Clear Dental Aligner is a series of thermoformed plastic trays used for orthodontic tooth movement. Process: clinician obtains intraoral scans or physical impressions; data imported into 3Shape software for treatment planning; 3D printer creates molds; plastic sheets thermoformed over molds. Trays worn by patient 20-21 hours/day, removed for eating/cleaning. Clinician monitors treatment progress. Device exerts continuous gentle force to move teeth to desired positions. Used in dental clinics under professional supervision.

## Clinical Evidence

Bench testing only. No clinical or animal testing performed. Biocompatibility testing conducted per ISO 10993 (Ames Assay, Cytotoxicity, Intracutaneous/Intradermal Reactivity, Oral Mucosa Irritation, Maximization for Delayed-Type Hypersensitivity, Subacute Systemic Toxicity). Physical properties tested per ASTM standards (D638, D790, D792, D570, D5420).

## Technological Characteristics

Material: thermoformed polyurethane. Form factor: sequential intraoral plastic trays. Energy source: none (passive). Connectivity: uses 3Shape software for CAD/CAM treatment planning. Manufacturing: 3D printing of molds followed by thermoforming. Biocompatibility: ISO 10993 compliant. Physical testing: ASTM D638 (tensile/elongation), ASTM D790 (flexural), ASTM D792 (specific gravity), ASTM D570 (water absorption), ASTM D5420 (impact strength).

## Regulatory Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

## Predicate Devices

- Smylio Invisible Clear Aligners ([K173784](/device/K173784.md))

## Reference Devices

- 3Shape Software ([K180491](/device/K180491.md))
- 3Shape Software ([K152086](/device/K152086.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 10, 2020

Straight T, Inc. % Robert Dean President Compliance Systems International, LLC. 1083 Delaware Ave. Buffalo, New York 14209

Re: K201450

Trade/Device Name: Straight T Clear Dental Aligner Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: July 30, 2020 Received: August 4, 2020

Dear Robert Dean:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120                        |
|-----------------------------------------|---------------------------------------------------------|
| Food and Drug Administration            | Expiration Date: 06/30/2020<br>See PRA Statement below. |

**Indications for Use**

| 510(k) Number (if known)                        | K201450                                                                                                                                                                                                                                                                                                           |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                                     | Straight T Clear Dental Aligner                                                                                                                                                                                                                                                                                   |
| Indications for Use (Describe)                  | The Straight T Clear Dental Aligners indicated for use in the alignment of permanent teeth (i.e. all second molars) through orthodontic treatment of misalignment and malocclusion.                                                                                                                               |
| Type of Use (Select one or both, as applicable) | <div> <span> <svg height="12" width="12"> <path d="M1 5 L4 8 L11 1" fill="none" stroke="black" stroke-width="2"></path> </svg> </span>             Prescription Use (Part 21 CFR 801 Subpart D)             <span> <input type="checkbox"/>Over-The-Counter Use (21 CFR 801 Subpart C)             </span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to.

Department of Health and Human Services  

Food and Drug Administration  

Office of Chief Information Officer  

Paperwork Reduction Act (PRA) Staff  

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

## K201450

## 510K Summary

| Submitter Name:                                                          | Straight T, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Trade Name:                                       | Submission Device                                                                                                                                                                               | Predicate Device                                                                                                                                                            |
|--------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address:                                                       | 16952 W. Bell Rd #301, Surprise AZ 85374                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                   | Straight T, Inc.<br>Clear Dental Aligners                                                                                                                                                       | K173784<br>Smylio Invisible Clear Aligners<br>K173785                                                                                                                       |
| Phone Number:                                                            | (623)474-384                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 510(k) Number                                     |                                                                                                                                                                                                 |                                                                                                                                                                             |
| Contact Person:                                                          | Minh Trinh                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Manufacturer                                      | Straight T                                                                                                                                                                                      | Smylio                                                                                                                                                                      |
| Date Prepared:                                                           | May 20, 2020                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Classification #,<br>Product Code<br>Device Class | 21 CFR 852.5470<br>NXC<br>2                                                                                                                                                                     | 21 CFR 852.5470<br>NXC<br>2                                                                                                                                                 |
| Device Trade Name:                                                       | Straight T Clear Dental Aligner                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Intended Use                                      | The Straight T Clear Dental Aligners<br>indicated for use in the alignment of<br>permanent teeth (i.e. all second molars)<br>through orthodontic treatment of<br>misalignment and malocclusion. | Smylio Invisible Clear Aligners is indicated<br>for use in the alignment of permanent<br>teeth through orthodontic treatment of<br>misalignment and malocclusion.           |
| Common Name                                                              | Aligners, Sequential                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Mode of Action                                    | Alignment of teeth by application of<br>continuous gentle force, by sequential use<br>of preformed plastic trays.                                                                               | Alignment of teeth by application of<br>continuous gentle force, by sequential use<br>of preformed plastic trays.                                                           |
| Classification Name<br>Number<br>Product Code<br>Regulatory Class        | Orthodontic Plastic Bracket<br>21 CFR 872.5470<br>NXC<br>2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Method of Use                                     | Each preformed plastic tray is worn by<br>the patient as prescribed by the dental<br>practitioner, usually a few weeks prior to<br>using the next sequential Aligners tray.                     | Each preformed plastic tray is worn by<br>the patient as prescribed by the dental<br>practitioner, usually a few weeks prior to<br>using the next sequential Aligners tray. |
| Primary Predicate<br>Device:                                             | K173784, Smylio Invisible Clear Aligner                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Material                                          | Thin thermoformed<br>polyurethane                                                                                                                                                               | Thin thermoformed<br>polyurethane                                                                                                                                           |
| Statement of<br>Indications<br>for Use                                   | The Straight T Clear Dental Aligners indicated for use in the alignment of<br>permanent teeth (i.e. all second molars) through orthodontic treatment of<br>misalignment and malocclusion.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Biocompatible                                     | Yes                                                                                                                                                                                             | Yes                                                                                                                                                                         |
| Device Description<br>and Summary of<br>Technological<br>Characteristics | A dental health professional (e.g. orthodontist or dentist) prescribes the Straight T<br>Dental Aligners based on an assessment of the patient's teeth. The dental health<br>professional (dentist/orthodontist) takes intraoral scans or physical impressions of<br>the patient's teeth, determines a course of treatment with the system, and<br>completes a prescription form using a standard dental software used for tooth<br>alignment. Straight T Clear Dental Aligner are intraoral thermoformed plastic<br>aligners that are worn 20 to 21 hours per day and are designed to be used in a<br>sequence, each aligner providing a gentle continuous force, to allow for the<br>movement of teeth to the final desired position. The aligners are to be removed for<br>eating and for cleaning. Straight T Clear Dental Aligners are fabricated using a<br>three-step process.<br>The first step is to obtain the dimensions and details of the patient's baseline<br>dentition. This is generally done using an oral scan data or a physical impression.<br>This scanned data (digital CAD/CAM models or patient models) are imported into<br>specialized dental software for treatment planning. The second step is the printing<br>of 3D models of the treatment plan for use in step 3 (thermoforming). In the second<br>step, Straight T Dental utilizes a software application to plan the treatment by<br>creating a series of sequential models that gradually position the teeth into their<br>final desired position. The treatment plan is sent to the doctor for approval. Upon<br>approval, a 3D printer is used to create the molds needed for each treatment step<br>to provide the surface around which the aligner is thermoformed. The final step is<br>the thermoforming of a plastic sheet material to each of the sequential treatment<br>steps. This process is done using a standard thermoforming equipment and the<br>appropriate material as outlined in this submission. The trays are provided to the<br>dental health care professional who provides them to the patient in sequential<br>stages, confirming fit and design. The dental health professional monitors treatment<br>from the moment the first aligner is delivered to when the final aligner is finished<br>and treatment is complete. | OTC or Rx                                         | Rx                                                                                                                                                                                              | Rx                                                                                                                                                                          |
| Software Use                                                             | Yes, 3Shape K180491                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes, 3Shape K 152086                              |                                                                                                                                                                                                 |                                                                                                                                                                             |
| Sterile                                                                  | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | No                                                |                                                                                                                                                                                                 |                                                                                                                                                                             |

{4}------------------------------------------------

Differences between Straight T Clear Dental Aligner compared to predicate device

| Straight T Clear Dental Aligner                                                                                                                                                               | S & E Effect                                                                                                                    | Smylio K173784                                                                                                       |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Straight T prepares the treatment plan<br>in Step 2 of the manufacturing process<br>for subsequent approval by a doctor.                                                                      | No effect, both treatment plans<br>are doctor approved.                                                                         | Smylio K173784<br>doctor prepares<br>the treatment plan                                                              |
| Straight T uses 3Shape Software<br>K180491                                                                                                                                                    | No effect, 3Shape Software<br>K180491 is FDA 510K cleared, the<br>use/manufacturing process has<br>been validated by Straight T | Smylio uses 3Shape<br>Software K152086                                                                               |
| Straight T biocompatibility summary<br>applied ISO 10993<br>-3, Biological evaluation of medical<br>devices — Part 3: Tests for<br>genotoxicity, carcinogenicity and<br>reproductive toxicity | No effect on biocompatibility. ISO<br>7405 directly references the same<br>test as conducted using ISO 10993<br>et.al.          | Smylio<br>biocompatibility<br>summary<br>references ISO<br>7405, Dentistry –<br>Evaluation of<br>biocompatibility of |

{5}------------------------------------------------

| Mechanism of Action                                  | In the same manner as the predicate device, each aligner exerts gentle force to achieve progressive realignment of the teeth until the final correction has been attained. This is based on the treatment plan and proceeds over time.                                                                                                                                                                 |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Testing                                       | Biocompatibility                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                      | Contact of the device to the patient's oral tissue requires the Aligners material to be biocompatible. The thermoplastic polyurethane has been tested according to Good Laboratory Practices for its biocompatibility according to ISO 10993, as follows:                                                                                                                                              |
|                                                      | Part 3 (Bacterial Mutagenicity - Ames Assay)                                                                                                                                                                                                                                                                                                                                                           |
|                                                      | Part 5 (Cytotoxicity Elution - MEM),                                                                                                                                                                                                                                                                                                                                                                   |
|                                                      | Part 10 (Intracutaneous/Intradermal)                                                                                                                                                                                                                                                                                                                                                                   |
|                                                      | Reactivity), Part 10 (Oral Mucosa Irritation),                                                                                                                                                                                                                                                                                                                                                         |
|                                                      | Part 10 (Maximization for Delayed-Type Hypersensitivity),                                                                                                                                                                                                                                                                                                                                              |
|                                                      | Part 11 (Subacute Systemic Toxicity)                                                                                                                                                                                                                                                                                                                                                                   |
|                                                      | Animal   Human Testing                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                      | No animal or human testing were required for this product because it is composed of the same materials and has a similar design and method of manufacture/fabrication in comparison to the predicate device.                                                                                                                                                                                           |
| Non-Clinical Physical acceptance Properties Testing: | Device material tested to the following standards and meet the criteria                                                                                                                                                                                                                                                                                                                                |
|                                                      | <ul><li>Elongation @ Yield (%) ASTM D638</li><li>Elongation @ Break (%) ASTM D638</li><li>Tensile @ Yield (PSI) ASTM D638</li><li>Tensile Strength (PSI) ASTM D638</li><li>Tensile Modulus (PSI) ASTM D638</li><li>Flexural Modulus (PSI) ASTM D790</li><li>Flexural Strength (PSI) ASTM D790</li><li>Specific Gravity g.cm3 ASTM D792</li><li>Water Absorption (%)24 hours @ 23°C ASTM D570</li></ul> |

- Gardner Impact Strength 23°C J/mm ASTM D5420

{6}------------------------------------------------

| • -5, Biological evaluation of medical<br>devices -- Part 5: Tests for in vitro<br>cytotoxicity               | medical devices<br>used in dentistry |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------|
| • -10, Biological evaluation of medical<br>devices -- Part 10: Tests for irritation<br>and skin sensitization |                                      |
| • -11, Biological evaluation of medical<br>devices -- Part 11: Tests for systemic<br>toxicity                 |                                      |

The intended use of the Straight T, Inc. Clear Dental Aligner is the same to that of the primary predicate device as they are both intended for correcting dental malocclusion in patients with permanent dentition.

It has a similar technological principle, and the device characteristics are similar to the predicate device. The mode of operation and the material used to fabricate the aligner trays is the same as the predicate device. There are minor differences comparing Straight T Clear Dental Aligner to the predicate Smylio Invisible Clear Aligners which do not affect substantial equivalence or safety and effectiveness.

## Substantial Equivalence Conclusion

Thus, based on the above it can be concluded that Straight T Clear Dental Aligners is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K201450](https://fda.innolitics.com/device/K201450)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
