Molecular Flow Simulations is designed as a planning system for pre- and intra-operative planning of stereotactic or image-guided surgeries. It is specially designed to display anatomical images of a patient acquired with magnetic resonance (MR) or computed tomography (CT) scanners as well as images derived from diffusion tensor imaging (DTI) data acquired with MR. Molecular Flow Simulations is a dedicated tool for planning trajectories of intracranial catheters. Guidelines for the catheter placement, such as from catheter suppliers, can be visualized and displayed to support the surgeon in improving catheter placement planning. The guidelines, in combination with anatomical information, can be used to suggest areas that are compliant with the guidelines. Molecular Flow Simulations does not generate or create rules for the placement of intracranial catheters by any means. Molecular Flow Simulations uses MR-DTI and T2- weighted MR images to suggest likely volumes of fluid distribution.
Device Story
Software-only planning system for stereotactic/image-guided intracranial catheter surgery. Inputs: MR/CT anatomical images, MR-DTI data. Operation: displays patient anatomy; visualizes catheter placement guidelines; calculates/predicts 3D infusate distribution based on patient-specific data and mathematical modeling (including flux boundary conditions, capillary loss maps, and poroelastic expansion). Output: surgical plan with suggested catheter trajectories and fluid distribution volumes. Used in clinical settings by surgeons. Benefits: improves catheter placement accuracy; optimizes infusate distribution; identifies potential leakage pathways (e.g., sulci).
Clinical Evidence
Bench testing only. Validation included verification of functional specifications, data integrity, and user performance testing of the manual. Requirements Traceability Matrix (RTM) confirmed all requirements and risks were tested. No clinical data.
Technological Characteristics
Software-only application for standard PC/laptop or intranet access. Uses MR-DTI and T2-weighted MR imaging. Features 3D distribution simulation, flux boundary modeling, capillary loss maps, and poroelastic expansion modeling. Risk management per ISO 14971.
Indications for Use
Indicated for pre- and intra-operative planning of stereotactic or image-guided intracranial catheter placement surgeries. Uses MR and CT anatomical images and MR-DTI data to visualize catheter trajectories and suggest fluid distribution volumes.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 3, 2021
Therataxis, LLC % Al Memmolo President, Convergent Clinical, Inc. 6648 Surf Crest St. Carlsbad, California 92011
Re: K201435
Trade/Device Name: Molecular Flow Simulations Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW, QRI Dated: November 11, 2021 Received: November 15, 2021
Dear Al Memmolo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201435
Device Name Molecular Flow Simulations
#### Indications for Use (Describe)
Molecular Flow Simulations is designed as a planning system for pre- and intra-operative planning of stereotactic or image-guided surgeries. It is specially designed to display anatomical images of a patient acquired with magnetic resonance (MR) or computed tomography (CT) scanners as well as images derived from diffusion tensor imaging (DTI) data acquired with MR. Molecular Flow Simulations is a dedicated tool for planning trajectories of intracranial catheters. Guidelines for the catheter placement, such as from catheter suppliers, can be visualized and displayed to support the surgeon in improving catheter placement planning. The guidelines, in combination with anatomical information, can be used to suggest areas that are compliant with the guidelines. Molecular Flow Simulations does not generate or create rules for the placement of intracranial catheters by any means. Molecular Flow Simulations uses MR-DTI and T2- weighted MR images to suggest likely volumes of fluid distribution.
The primary mode of action for Molecular Flow Simulations is a device for creating stereotactic or image-guided surgical plans, especially for the creation of plans for the placement of intracranial catheters.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> |
|---------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| A. Device Information: | 510(k) Summary |
|--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Category | Comments |
| Sponsor: | Therataxis, LLC<br>Johns Hopkins Eastern Building Suite B305<br>1101 East 33rd Street Baltimore, MD 21208<br>(443) 451-7154 |
| Correspondent Contact<br>Information: | Al Memmolo<br>Convergent Clinical, Inc.<br>Carlsbad, CA 92011<br>Email: almemmolo@gmail.com |
| Device Common Name: | Planning System, Stereotaxic Instrument |
| Device Regulation & Name: | CFR 882.4560 Stereotaxic Instrument |
| Classification & Product Code:<br>510(k) Number: | Classification: Class II Product Code: HAW, QRI<br>K201435 |
| Device Proprietary Name: | Molecular Flow Simulations |
# Molecular Flow Simulations
## Predicate Device Information:
| Predicate Device: | iPlan Flow |
|----------------------------------------------------|----------------------------------------------------------|
| Predicate Device Manufacturer: | Brainlab AG |
| Predicate Device Common Name: | iPlan Flow |
| Predicate Device Premarket Notification # | K053164 |
| Predicate Device Classification & Name | Classification: Class II<br>Name: Stereotaxic Instrument |
| Predicate Device Classification<br>& Product Code: | Classification: Class II<br>Product Code: HAW |
## B. Date Summary Prepared
December 3, 2021
## C. Description of Device
Molecular Flow Simulations is a software tool running on a standard, standalone computer (PC or Laptop), or being accessible via the intranet connection, that can be used by surgeons for pre- or intraoperative planning of treatments based on stereotactic systems or image guided surgery systems. The system is a software-only medical device to be installed on common IT hardware.
{5}------------------------------------------------
#### D. Indications for Use
Molecular Flow Simulations is designed as a planning system for pre- and intra-operative planning of stereotactic or image-guided surgeries. It is specially designed to display anatomical images of a patient acquired with magnetic resonance (MR) or computed tomography (CT) scanners as well as images derived from diffusion tensor imaging (DTI)-data acquired with MR. Molecular Flow Simulations is a dedicated tool for planning trajectories of intracranial catheters. Guidelines for the catheter placement, such as from catheter suppliers, can be visualized and displayed to support the surgeon in improving catheter placement planning. The guidelines, in combination with anatomical information, can be used to suggest areas that are compliant with the guidelines. Molecular Flow Simulations does not generate or create rules for the placement of intracranial catheters by any means. Molecular Flow Simulations uses MR-DTI and T2weighted MR images to suggest likely volumes of fluid distribution.
The primary mode of action for Molecular Flow Simulations is a device for creating stereotactic or imageguided surgical plans, especially for the creation of plans for the placement of intracranial catheters.
| Characteristic | Application Device:<br>Molecular Flow<br>Simulations<br>(K201435) | Predicate Device:<br>iPlan Flow<br>(K053164) | Impact on<br>Substantial<br>Equivalence |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|
| Company | Therataxis, LLC | Brainlab | - |
| Regulation<br>Number | 882.4560 | 882.4560 | Identical |
| Product Code | HAW | HAW | Identical |
| Intended Use | Molecular Flow Simulations<br>is designed as a planning<br>system for pre- and<br>intraoperative planning of<br>stereotactic or image guided<br>surgery treatments. It is<br>specially designed to display<br>anatomical images of a<br>patient acquired with MR<br>and/or CT as well as images<br>derived from DTI-data<br>acquired with Magnetic<br>Resonance Imaging (MRI). | iPlan Flow is designed as a<br>planning system for pre-<br>and intraoperative planning<br>of stereotactic or image<br>guided surgery treatments.<br>It is specially designed to<br>display anatomical images<br>of a patient acquired with<br>MR and/or CT as well as<br>images derived from DTI-data<br>acquired with<br>Magnetic Resonance<br>Imaging (MRI). | Equivalent |
| Indications for Use | Molecular Flow | iPlan Flow | K201435 |
| | Molecular Flow<br>Simulations is designed as<br>a planning system for pre-<br>and intra-operative<br>planning of stereotactic or<br>image-guided surgeries. It<br>is specially designed to<br>display anatomical images<br>of a patient acquired with<br>magnetic resonance (MR)<br>or computed tomography<br>(CT) scanners as well as<br>images derived from<br>diffusion tensor imaging<br>(DTI)-data acquired with<br>MR. Molecular Flow<br>Simulations is a dedicated<br>tool for planning<br>trajectories of intracranial<br>catheters. Guidelines for the<br>catheter placement, such as<br>from catheter suppliers, can<br>be visualized and displayed<br>to support the surgeon in<br>improving catheter<br>placement planning. The<br>guidelines, in combination<br>with anatomical<br>information, can be used to<br>suggest areas that are<br>compliant with the<br>guidelines. Molecular Flow<br>Simulations does not<br>generate or create rules for<br>the placement of<br>intracranial catheters by<br>any means. Molecular Flow<br>Simulations uses MR-DTI<br>and T2- weighted MR<br>images to suggest likely<br>volumes of fluid<br>distribution. | iPlan Flow is designed as a<br>planning system for pre-<br>and intraoperative planning<br>of stereotactic or image<br>guided surgery treatments.<br>It is specially designed to<br>display anatomical images<br>of a patient acquired with<br>MR and/or CT as well as<br>images derived from DTI-<br>data acquired with<br>Magnetic Resonance<br>Imaging (MRI). iPlan Flow<br>is a dedicated tool for<br>planning trajectories of<br>intra-cranial catheters.<br>Guidelines for the catheter<br>placement e.g. from<br>catheter suppliers can be<br>visualized and displayed to<br>support the surgeon in<br>improving catheter<br>placement planning. The<br>guidelines, in combination<br>with anatomical<br>information, can be used to<br>suggest areas that are<br>compliant with the<br>guidelines. iPlan Flow does<br>not generate or create rules<br>for the placement of<br>intracranial catheters by<br>any means. iPlan Flow uses<br>MR-DTI and T2- weighted<br>MR images to suggest<br>likely volumes of fluid<br>distribution. | Equivalent |
| | The primary mode of action<br>for Molecular Flow<br>Simulations is a device for<br>creating stereotactic or<br>image-guided surgical plans,<br>especially for the creation of<br>plans for the placement of<br>intracranial catheters. | The Primary mode of<br>action for iPlan Flow is a<br>device for creating<br>treatment plans for<br>stereotactic or image<br>guided surgical treatment,<br>especially for the creation<br>of plans for the placement<br>of intra-cranial catheters.<br>The treatment plans can be<br>used in conjunction with<br>other BrainLAB medical<br>devices such as<br>VectorVision for image<br>guided surgical treatment. | |
| | | | K201435, Pag |
| Technology | Molecular Flow<br>Simulations can be used<br>for the planning of<br>intracranial catheters, with<br>image guided surgery.<br>Guidelines for the exact<br>placement of intracranial<br>catheters can be<br>visualized. These<br>guidelines comprise the<br>minimal depth of the<br>catheter tip in the brain<br>tissue, the minimal<br>distance of the catheter tip<br>from intra-cranial surfaces<br>and the minimal distance<br>between different catheter<br>tips. The depth guideline<br>can be calculated from the<br>flow rate and the catheter<br>diameter and warnings<br>will be displayed if the<br>trajectory of a planned<br>catheter is likely to cross<br>an intra-cranial surface.<br>Molecular Flow<br>Simulations is able to<br>calculate a likely fluid<br>distribution from the<br>planned catheter positions<br>to support the physician in<br>his decision about<br>appropriate catheter<br>positions. These features<br>enable the surgeon to<br>better plan and place intra-<br>cranial catheters. | iPlan Flow can be used for<br>the planning of<br>intracranial catheters, with<br>image guided surgery.<br>Guidelines for the exact<br>placement of intracranial<br>catheters can be<br>visualized. These<br>guidelines comprise the<br>minimal depth of the<br>catheter tip in the brain<br>tissue, the minimal<br>distance of the catheter tip<br>from intra-cranial surfaces<br>and the minimal distance<br>between different catheter<br>tips. The depth guideline<br>can be calculated from the<br>flow rate and the catheter<br>diameter and warnings<br>will be displayed if the<br>trajectory of a planned<br>catheter is likely to cross<br>an intra-cranial surface.<br>iPlan Flow is able to<br>calculate a likely fluid<br>distribution from the<br>planned catheter positions<br>to support the physician<br>in his decision about<br>appropriate catheter<br>positions. These features<br>enable the surgeon to<br>better plan and place intra-<br>cranial catheters. | Technology is<br>identical. |
| Characteristic | Application Device:<br>Molecular Flow<br>Simulations<br>(K201435) | Predicate Device:<br>iPlan Flow<br>(K053164) | Impact on<br>Substantial<br>Equivalence |
| Features | Comprehensive Target<br>Planning: Automatically<br>defines margins around<br>the tumor. | Comprehensive Target<br>Planning: Automatically<br>defines margins around<br>the tumor. | |
| Features | Guided Entry Point<br>Planning: The sulcus<br>detection feature<br>automatically segments<br>cerebrospinal fluid spaces<br>as they are potential<br>leakage pathways. | Guided Entry Point<br>Planning: The sulcus<br>detection feature<br>automatically segments<br>cerebrospinal fluid spaces<br>as they are potential<br>leakage pathways. | Equivalent |
| | Guided Catheter<br>Planning: Catheter<br>planning guidelines<br>indicate potential<br>backflow and other drug<br>leakage pathways around<br>the catheter tip in order to<br>maximize infusate<br>distribution. | Guided Catheter<br>Planning: Catheter<br>planning guidelines<br>indicate potential<br>backflow and other drug<br>leakage pathways around<br>the catheter tip in order to<br>maximize infusate<br>distribution. | |
| | 3D Distribution<br>Simulation: The likely<br>distribution of infusate for<br>a given catheter position<br>is predicted in 3D based<br>on patient-specific<br>information and<br>mathematical modeling.<br>Flux boundary conditions<br>can be utilized. This more<br>accurately models CED<br>infusions, in which the<br>infusate is delivered at a<br>constant flow rate rather<br>than a constant pressure. | 3D Distribution<br>Simulation: The likely<br>distribution of infusate for<br>a given catheter position<br>is predicted in 3D based<br>on patient-specific<br>information and<br>mathematical modeling. | |
| | 3D capillary loss maps<br>from DCE, poroelastic<br>expansion of the<br>extracellular space, and<br>flux boundary modeling of<br>the infusion sources. | | |
#### E. Comparison of the Technological Characteristics
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
#### F. Summary of Supporting Data
Several tests were performed, in concordance with Duke University, using synthetic, animal and human imaging. Therefore, Molecular Flow Simulations does not raise any new questions regarding safety and effectiveness.
### G. Discussion of Performance Testing
All necessary testing has been performed with Molecular Flow Simulations to assure substantial equivalence to the predicate device.
Summary of non-clinical tests:
Molecular Flow Simulations is a software application; therefore no electrical safety or electromagnetic testing was required.
Testing conducted to demonstrate software validation and substantial equivalence included:
- . Verification testing that product meets product performance and functional specifications.
- Verification that data submitted is stored properly to maintain data integrity (e.g. no loss of data or corruption).
- User performance testing to demonstrate adequate instructional utility of the User Manual. ●
The Requirements Traceability Matrix (RTM) provides a mapping between requirements, risks, test cases, and shows related test results. The RTM confirms that there was a test case authored and executed for all requirements and any applicable risks.
After extensive bench testing to performance requirements and criteria established in accordance with application of ISO 14971, risk analysis standard, no new issues of safety, performance, technology or intended use were identified.
## H. Conclusion
Upon reviewing the technical information provided in this submission and comparing intended use, principle of operation, performance data, and overall technological characteristics, Molecular Flow Simulations is determined to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.