← Product Code [GCJ](/productcode/GCJ) · K201434

# Connected OR Hub with Voice Control, Connected OR Hub with Device Control (K201434)

_Stryker · GCJ · Jul 28, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K201434

## Device Facts

- **Applicant:** Stryker
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jul 28, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The use of the Connected OR Hub with Device and Voice Control system is to allow for voice control and remote control of medical device settings by surgeons or operating room personnel, the need to manually operate those devices compatible with the Connected OR Hub with Device and Voice Control or to rely on verbal communication between the surgeon and other opersonnel in order to adjust the surgical equipment. It also has additional digital documentation functionality to electronically capture, transfer, store and display medical device data (non-medical device function), which is independent of the functions or parameters of any attached Stryker device.

## Device Story

Connected OR Hub acts as a centralized control console for operating room equipment. Inputs include surgical video feeds, voice commands via wireless headset, and IR remote signals. The system transforms these inputs to adjust settings on connected devices (cameras, light sources, insufflators, pumps, RF/shaver systems) and automates smoke detection and evacuation. It also captures and stores non-diagnostic medical device data. Used in the OR by surgeons and staff; the hub connects to peripheral 'Spokes' via Ethernet to facilitate wireless control data transfer. Output is displayed on monitors and stored digitally. Benefits include reduced manual interaction with equipment and improved OR workflow efficiency.

## Clinical Evidence

Bench testing only. No clinical data required. Testing included software validation (IEC 62304), usability (IEC 62366-1), electrical safety (ANSI/AAMI ES60601-1), EMC (IEC 60601-1-2), and performance bench testing against specifications and predicate devices.

## Technological Characteristics

Console-based system with capacitive touchscreen, IR remote, and wireless headset. Connectivity via wired device control ports and Ethernet-linked Spokes. Embedded Microsoft Windows 10. Custom chipset, HDD/SSD storage. Video I/O: DVI, RGBHV, HDMI. Electrical safety: IEC 60601-1; EMC: IEC 60601-1-2.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Connected OR Hub with Voice and Device Control ([K181258](/device/K181258.md), [K192172](/device/K192172.md))
- ZMED VP4000 Video Processor ([K130929](/device/K130929.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with a stylized image of a human figure. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" in a larger font and "ADMINISTRATION" in a smaller font below it. The logo is clean and professional, reflecting the organization's role in regulating food and drugs.

July 28, 2020

Stryker Endoscopy Meagan Jones Principal Regulatory Affairs Specialist 5900 Optical Court San Jose, California 95131

Re: K201434

Trade/Device Name: Connected OR Hub with Device and Voice Control Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: May 28, 2020 Received: June 1, 2020

Dear Meagan Jones:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Digitally signed by Long H. Chen
Long H. Chen -S-S
Date: 2020.07.28 14:57:20 -04'00'

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201434

#### Device Name

Connected OR Hub with Device and Voice Control

Indications for Use (Describe)

Intended use/Indications for Use:

The use of the Connected OR Hub with Device and Voice Control system is to allow for voice control and remote control of medical device settings by surgeons or operating room personnel, the need to manually operate those devices compatible with the Connected OR Hub with Device and Voice Control or to rely on verbal communication between the surgeon and other opersonnel in order to adjust the surgical equipment. It also has additional digital documentation functionality to electronically capture, transfer, store and display medical device data (non-medical device function), which is independent of the functions or parameters of any attached Stryker device.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R Part 807.92(c).

## Submitter:

| Applicant:      | Stryker Endoscopy<br>5900 Optical Court<br>San Jose, CA 95138                                                       |
|-----------------|---------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Meagan Jones<br>Principal Regulatory Affairs Specialist<br>Phone: (214) 701-2186<br>Email: meagan.jones@stryker.com |
| Date Prepared:  | May 28, 2020                                                                                                        |

### Subject Device:

| Name of Device:         | Connected OR Hub with Device and Voice Control |
|-------------------------|------------------------------------------------|
| Common or Usual Name    | SDC4K Information Management System            |
| Classification Name:    | Laparoscope, General & Plastic Surgery         |
| Regulation Number       | 21 C.F.R. §876.1500                            |
| Regulatory Class:       | II                                             |
| Product Code:           | GCJ                                            |
| Subsequent Product Code | HRX                                            |
| 510(k) Review Panel:    | General and Plastic Surgery                    |

### Predicate Device(s):

| Stryker's Connected OR Hub with Device and Voice Control | K181258, K192172 |
|----------------------------------------------------------|------------------|
| ZMED VP4000 Video Processor                              | K130929          |

NOTE: The predicate devices have not been subject to a design-related recall.

### Device Description:

The Connected OR Hub with Device and Voice Control is a medical device that allows the surgeon to control the state, selection, and settings of any compatible device attached to it. It also has operating room documentation functionalities (Class I device function) to electronically capture, transfer, store and display medical device data independently of the functions or parameters of the connected medical device.

The Connected OR Hub with Device and Voice Control consists of the following:

{4}------------------------------------------------

- A Connected OR Hub Console a.
- A Device control package b.
- A Voice control package C.

The Connected OR Hub with Device and Voice Control with the Video Image Processing (VIP) software feature incorporates and automates an enhanced image algorithm which detects and digitally removes smoke from surgical images; and, automates the detection and initiates smoke evacuation through compatible laparoscopic insufflator(s) during endoscopic procedures.

## Indications for Use:

The use of the Connected OR Hub with Voice and Device Control system is to allow for remote and voice control of medical device settings by surgeons or operating room personnel, thereby eliminating the need to manually operate those devices compatible with the Connected OR Hub with Voice and Device Control or to rely on verbal communication between the surgeon and other operating room personnel in order to adjust the surgical equipment. It also has additional digital documentation functionality to electronically capture, transfer, store and display medical device data (Class I device function), which is independent of the functions or parameters of any attached Stryker device.

| Item                                                                                | Subject Device<br>Connected OR Hub with<br>Device and Voice Control                                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Devices                                                                                |                                                                                                                                                                                                                                                                                                   |                                                           |
|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
|                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Connected OR Hub with Voice and Device<br>Control (primary predicate)                            | ZMED VP4000 Video<br>Processor<br>(secondary predicate)                                                                                                                                                                                                                                           |                                                           |
| Submission<br>Reference                                                             | Current submission                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | K192172, K181258                                                                                 | K130929                                                                                                                                                                                                                                                                                           |                                                           |
| Product Code                                                                        | GCJ, HRX                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same as subject<br>device                                                                        | LLZ                                                                                                                                                                                                                                                                                               |                                                           |
| Classification                                                                      | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Class II                                                                                         | Class II                                                                                                                                                                                                                                                                                          |                                                           |
| Intended<br>Use /Indications for<br>Use                                             | The use of the Connected OR<br>Hub with Device and Voice<br>Control system is to allow for<br>voice control and remote<br>control of medical device<br>settings by surgeons or<br>operating room personnel,<br>thereby eliminating the need to<br>manually operate those<br>devices compatible with the<br>Connected OR Hub with<br>Device and Voice Control or<br>to rely on verbal<br>communication between the                                                                   | Same as subject<br>device.                                                                       | The ZMED VP4000<br>Video Processor is<br>intended for use in any<br>application where a<br>viewing device<br>(fluoroscope, endoscope,<br>laparoscope, etc.) and<br>monitor are incorporated<br>to aid in diagnosis and<br>treatment of a disease<br>such as an arthroscopy<br>or cholecystectomy. |                                                           |
| Item                                                                                | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Predicate Devices                                                                                |                                                                                                                                                                                                                                                                                                   |                                                           |
|                                                                                     | Connected OR Hub with<br>Device and Voice Control                                                                                                                                                                                                                                                                                                                                                                                                                                   | Connected OR Hub with Voice and Device Control (primary predicate)                               | ZMED VP4000 Video Processor (secondary predicate)                                                                                                                                                                                                                                                 |                                                           |
|                                                                                     | surgeon and other operating<br>room personnel in order to<br>adjust the surgical equipment.<br>It also has additional digital<br>documentation functionality to<br>electronically capture, transfer,<br>store and display medical<br>device data (non-medical<br>device function), which is<br>independent of the functions<br>or parameters of any attached<br>Stryker device.                                                                                                     |                                                                                                  |                                                                                                                                                                                                                                                                                                   |                                                           |
| Manufacturer                                                                        | Stryker                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same as subject<br>device                                                                        | ZMED, Inc.                                                                                                                                                                                                                                                                                        |                                                           |
| Principles of<br>Operation                                                          | Use of IR remote control for<br>device control and RF<br>communication for voice<br>control of connected devices.                                                                                                                                                                                                                                                                                                                                                                   | Same as subject<br>device.                                                                       | Provides enhanced video<br>contrast mapping which<br>facilitates better<br>visualization of subtle<br>(e.g. low contrast) image<br>features                                                                                                                                                       |                                                           |
| Device Components                                                                   | Connected OR Hub console Device Control Software Voice Control Software                                                                                                                                                                                                                                                                                                                                                                                                             | Same as subject<br>device                                                                        | VP4000 Video Processor<br>(with enhanced imaging<br>algorithm)                                                                                                                                                                                                                                    |                                                           |
| Feature(s)                                                                          | Smoke<br>Detection<br>                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Manual and automated                                                                             | N/A – No smoke<br>detection functionality                                                                                                                                                                                                                                                         | Not applicable - No<br>smoke detection<br>functionality   |
|                                                                                     | Enhanced<br>Imaging                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Manual and automated                                                                             | Manual (through<br>ZMED VP4000 Video<br>processor)                                                                                                                                                                                                                                                | Automated enhanced<br>imaging                             |
|                                                                                     | Smoke<br>Evacuation                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Manual and automated                                                                             | Manual (through<br>Device and Voice<br>control)                                                                                                                                                                                                                                                   | Not applicable - No<br>smoke evacuation<br>functionality. |
| Documentation<br>Functionalities (Class<br>I/Non-Medical Device<br>functionalities) | Capture, transfer and display<br>image/video of various<br>formats for recording purposes<br>only (not for diagnosis or<br>treatment evaluation).                                                                                                                                                                                                                                                                                                                                   | Same as subject<br>device                                                                        | Not applicable - no<br>Class I/Non-Medical<br>Device functionality<br>included in Clarity.                                                                                                                                                                                                        |                                                           |
| Device Control User<br>Interface                                                    | Capacitive touch Graphical User Interface on LCD touchscreen Voice Recognition and Control via wireless headset Device Control via IR Remote Control                                                                                                                                                                                                                                                                                                                                | Same as subject<br>device                                                                        | Operation is required<br>from outside sterile field.<br>Front panel provides<br>On/Off and one-of-six<br>preset selection controls,<br>with additional buttons                                                                                                                                    |                                                           |
| Item                                                                                | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Predicate Devices                                                                                |                                                                                                                                                                                                                                                                                                   |                                                           |
|                                                                                     | Connected OR Hub with<br>Device and Voice Control                                                                                                                                                                                                                                                                                                                                                                                                                                   | Connected OR Hub<br>with Voice and Device<br>Control (primary<br>predicate)                      | ZMED VP4000 Video<br>Processor<br>(secondary predicate)                                                                                                                                                                                                                                           |                                                           |
|                                                                                     | • Device Control via Camera<br>Head directional keypad                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                  | for custom adjustment of<br>each preset.                                                                                                                                                                                                                                                          |                                                           |
| Connection to<br>Controllable Devices                                               | Wired connection to<br>Connected OR Hub's device<br>control ports via device<br>control cables.<br><br>Connected OR Hub is<br>connected to master<br>Connected OR Spoke via an<br>Ethernet cable, while devices<br>at remote locations within the<br>same OR are connected to the<br>slave Connected OR Spoke via<br>device control cables. The<br>master and slave Spokes act as<br>the wireless transfer medium<br>to transfer device control data<br>to / from Connected OR Hub. | Same as subject<br>device                                                                        | Not applicable – no<br>devices are controlled<br>with the Clarity Console.<br>It is meant to be<br>connected into a video<br>stream between a camera<br>and surgical display.                                                                                                                     |                                                           |
| Controllable Devices                                                                | Surgical Cameras, Light<br>Sources, Insufflators, Pumps,<br>RF and Shaver System,<br>Wireless Monitor                                                                                                                                                                                                                                                                                                                                                                               | Same as subject<br>device.                                                                       | Not applicable - No<br>controllable devices.                                                                                                                                                                                                                                                      |                                                           |
| Embedded Software<br>Design                                                         | Embedded Microsoft<br>Windows 10                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same as subject<br>device                                                                        | Embedded software.                                                                                                                                                                                                                                                                                |                                                           |
| Electronic Circuit<br>Design                                                        | Custom designed chipset,<br>storage solution and Capture<br>Card.<br><br>CD/DVD drive: Not included<br>in chassis<br><br>On-board storage: Hard Disk<br>Drive (HDD) and Solid-State<br>Drive (SSD)                                                                                                                                                                                                                                                                                  | Same as subject<br>device                                                                        | Custom design.                                                                                                                                                                                                                                                                                    |                                                           |
| Video Input and<br>Output                                                           | Input: DVI, RGBHV and<br>HDMI<br>Output: DVI, HDMI                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same as subject<br>device.                                                                       | Input: DVI,<br>Output: DVI                                                                                                                                                                                                                                                                        |                                                           |
| Power rating                                                                        | 100-240VAC ~50/60 Hz,<br>4A/2A maximum                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same as subject<br>device.                                                                       | Same as subject device.                                                                                                                                                                                                                                                                           |                                                           |
| Electrical Safety                                                                   | IEC 60601-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same as subject<br>device.                                                                       | Same as subject device.                                                                                                                                                                                                                                                                           |                                                           |
| Item                                                                                | Subject Device<br>Connected OR Hub with<br>Device and Voice Control                                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Devices<br>Connected OR Hub<br>with Voice and Device<br>Control (primary<br>predicate) | ZMED VP4000 Video<br>Processor<br>(secondary predicate)                                                                                                                                                                                                                                           |                                                           |
| EMC                                                                                 | IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same as subject device.                                                                          | Same as subject device.                                                                                                                                                                                                                                                                           |                                                           |

#### Comparison of Technological Characteristics with the Predicate Device:

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

#### Performance Data:

Testing was completed in accordance with the following:

| Test                                  | Method                                                                                                             | Conclusion |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------|------------|
| Software Validation &<br>Verification | IEC 62304:2015                                                                                                     | PASS       |
| Usability                             | IEC 62366-1:2015                                                                                                   | PASS       |
| Electrical Safety                     | ANSI/AAMI ES60601-1:2005/(R)2012 and<br>A1:2012                                                                    | PASS       |
| EMC                                   | IEC 60601-1-2:2014                                                                                                 | PASS       |
| Performance - Bench                   | In accordance with device input specifications<br>and comparative testing to currently legally<br>marketed device. | PASS       |

NOTE: The Connected OR Hub with Device and Voice Control with the VIP feature does not require clinical studies to support the determination of substantial equivalence.

### Conclusions:

The Connected OR Hub with Device and Voice Control is substantially equivalent in design, intended use, principles of operation, technological characteristics and safety features to the predicate devices. There are no new issues of safety and/or effectiveness introduced by the Connected OR Hub with Device and Voice Control.

---

**Source:** [https://fda.innolitics.com/device/K201434](https://fda.innolitics.com/device/K201434)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
