Visage Breast Density is a software application intended for use with compatible full field digital mammography and digital breast tomosynthesis systems. Visage Breast Density from a mammography study and provides an ACR BI-RADS Atlas 5th Edition breast density category to aid radiologists in the assessment of breast tissue composition. Visage Breast Density produces adjunctive information. It is not a diagnostic aid. Visage Breast Density is based on the Visage 7 product for distributing, viewing, processing, and archiving medical images.
Device Story
Software application; add-on module to Visage 7 PACS. Inputs: full field digital mammography and digital breast tomosynthesis images. Processing: convolutional neural network (CNN) computes likelihoods for four ACR BI-RADS breast density categories; individual image classifications merged for study-level classification. Output: breast density category displayed on Visage 7 client monitor; stored in Visage 7 server database. Used in clinical settings by radiologists. Benefits: provides adjunctive information to assist in breast tissue composition assessment.
Clinical Evidence
Bench testing using two independent datasets (500 and 700 studies). Ground truth established by consensus of three board-certified, MQSA-qualified radiologists per site. Performance metrics included accuracy per category and total accuracy for four-category classification and binary 'dense' vs 'non-dense' classification. Results showed similar accuracy to the predicate device.
Technological Characteristics
Software-only application; add-on module for Visage 7 PACS. Employs convolutional neural network (CNN) for image classification. Operates on digital mammography and digital breast tomosynthesis images. Integrated into existing PACS server/client architecture.
Indications for Use
Indicated for use by radiologists to aid in the assessment of breast tissue composition from mammography studies, providing an ACR BI-RADS Atlas 5th Edition breast density category (A, B, C, or D). Not a diagnostic aid.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.
Visage Imaging GmbH % Juliane Dinter Lead Project Manager QiP GmbH Struveweg 40 LUDWIGSFELDE, BRANDENBURG 14974 GERMANY
January 29, 3021
Re: K201411
Trade/Device Name: Visage Breast Density Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: QIH Dated: December 21, 2020 Received: December 23, 2020
Dear Juliane Dinter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills Division Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201411
Device Name Visage Breast Density
Indications for Use (Describe)
Visage Breast Density is a software application intended for use with compatible full field digital mammography and digital breast tomosynthesis systems. Visage Breast Density from a mammography study and provides an ACR BI-RADS Atlas 5th Edition breast density category to aid radiologists in the assessment of breast tissue composition. Visage Breast Density produces adjunctive information. It is not a diagnostic aid.
Visage Breast Density is based on the Visage 7 product for distributing, viewing, processing, and archiving medical images.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.6 2.4L5.6 10.4L2.4 7.2L3.2 6.4L5.6 8.8L12.8 1.6L13.6 2.4Z" fill="black"></path> <rect height="15" rx="3.5" stroke="black" width="15" x="0.5" y="0.5"></rect> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect height="15" rx="3.5" stroke="black" width="15" x="0.5" y="0.5"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Traditional 510(k) Summary Visage Breast Density
Image /page/3/Picture/2 description: The image shows the logo for Visage Imaging. The logo consists of a geometric shape in the upper left corner, with the words "Visage Imaging" to the right of it. The geometric shape is made up of three overlapping squares in green, gray, and black. The word "Imaging" has a registered trademark symbol next to it.
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807.92.
## Submitter
| Name | Visage Imaging GmbH |
|----------------|----------------------------------------------------------------------|
| Address | Lepsiusstrasse 70<br>12163 Berlin<br>Germany |
| Contact Person | Stefan Strowich<br>Manager of Quality Systems and Regulatory Affairs |
| E-mail | sstrowich@visageimaging.com |
| Phone Number | +49 30 700968-41 |
| Date Prepared | January 28, 2021 |
#### Device
| Trade Name | Visage Breast Density |
|---------------------|--------------------------------------------------|
| Common Name | Medical Imaging Software |
| Classification Name | Automated Radiological Image Processing Software |
| Regulatory Number | 21 CFR 892.2050 |
| Product Code | QIH |
| Device Class | Class II |
| Review Panel | Radiology |
#### Predicate Device
| Trade Name | PowerLook Density Assessment |
|---------------------|-------------------------------------|
| Common Name | Medical Imaging Software |
| Classification Name | System, Image Processing, Radiology |
| Regulatory Number | 21 CFR 892.2050 |
| Product Code | LLZ |
| Device Class | Class II |
| Review Panel | Radiology |
| Submitter | iCAD, Inc. |
| 510(k) Number | K180125 |
| Clearance | April 5, 2018 |
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Visage Imaging. The logo consists of a geometric shape in the upper left corner, with the words "Visage Imaging" to the right of it. The geometric shape is made up of three overlapping squares in green and gray. The words "Visage Imaging" are in a serif font, with the word "Imaging" in a slightly smaller font than the word "Visage".
## Device Description
Visage Breast Density is a software application that assesses breast density from a mammography study and provides a density category A, B, C, or D according to the ACR BI-RADS Atlas 5th Edition to aid radiologists in the assessment of breast tissue composition.
Visage Breast Density employs a convolutional network (CNN) for the automatic classification of breast density. The CNN has been trained on a large database of mammography exams. When applied to a mammography image, the CNN computes four likelihoods corresponding to the four breast density categories. The classifications of the individual images are merged into a general classification of the mammography study.
Visage Breast Density is designed as an add-on module to the Visage 71 product for distributing, viewing, processing, and archiving medical images. The assessment of breast density is performed from mammography studies stored on the Visage 7 server. The resulting breast density classification is displayed by the Visage 7 client on a computer monitor and stored in the database on the Visage 7 server.
#### Indications for Use
Visage Breast Density is a software application intended for use with compatible full field digital mammography and digital breast tomosynthesis systems. Visage Breast Density assesses breast density from a mammography study and provides an ACR BI-RADS Atlas 5th Edition breast density category to aid radiologists in the assessment of breast tissue composition. Visage Breast Density produces adjunctive information. It is not a diagnostic aid.
Visage Breast Density is based on the Visage 7 product for distributing, viewing, processing, and archiving medical images.
#### Comparison with Predicate Device
The technical characteristics of Visage Breast Density and the predicate device PowerLook Density Assessment are comparable.
The devices under comparison are software applications that assess breast density and provide a density category according to the ACR BI-RADS Atlas 5th Edition. Both devices are intended as an interpretative aid in the assessment of breast tissue composition. They have the same product classification, same level of concern, same patient population and similar intended users.
Visage Breast Density uses a trained convolutional network with digital mammography images. The predicate device determines the density category of tomosynthesis images based on a calculation of area covered by dense tissue. Despite the different analysis techniques, non-clinical testing showed that Visage Breast Density achieves similar accuracies compared to the predicate device.
Both applications are designed as a software extension meant to operate within a larger software framework intended for managing medical images.
² Visage 7 is a Class II medical device. It has been cleared with the former trade name Visage PACS/CS under K082269.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Visage Imaging. The logo consists of a geometric shape in the upper left corner, with the words "Visage Imaging" to the right. The geometric shape is made up of three overlapping squares in green and gray. The text is in a bold, sans-serif font.
## Performance Data
Visage Breast Density was tested against two independent data sets of 500 and 700 studies from two different sites. Three board certified radiologists with MQSA qualification per site performed a breast density classification and the consensus of the three reviewers was determined for each study.
The predicted breast density category of Visage Breast Density was related to the ground truth from the clinical reports and the consensus of the three reviewers. The accuracies per category and the total accuracy were computed for the classification into the four categories of the ACR BI-RADS Atlas 5ª Edition as well as for the binary classification 'dense' versus 'non-dense'.
The resulting accuracies were compared to the corresponding values of the predicate device. Visage Breast Density achieved similar accuracies per category and similar total accuracies compared to the predicate device.
The validation showed that the results of Visage Breast Density regarding the classification of breast density categories are in sufficient accordance with the results of the predicate device.
#### Conclusion
The differences in technological characteristics between Visage Breast Density and the predicate device do not raise new questions of safety or effectiveness. The differences in indications for use do not change the intended use nor raise new questions of safety and effectiveness.
Based on the information presented in this premarket notification, and in accordance with the indications for use, technological characteristics and validation testing, it is concluded that Visage Breast Density is substantially equivalent to the predicate device PowerLook Density Assessment.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.